(Stroke. 1997;28:1328-1329.)
© 1997 American Heart Association, Inc.
Articles |
From the Departments of Neurology (D.G., A.L., S.Z.) and Cardiothoracic Surgery, Martin-Luther University Halle-Wittenberg, Halle/Saale, Germany (M.P., Y.G., H.R.Z.).
Correspondence to D. Georgiadis, MD, Department of Neurology, Martin-Luther University Halle-Wittenberg, Ernst-Grube-Str 40, 06097 Halle/Saale, Germany. E-mail dimitrios.georgiadis{at}medizin.uni-halle.de
Background and Purpose The aim of this study was the evaluation of the prevalence and counts of Doppler microembolic signals (MES) in children with prosthetic cardiac valves and their comparison to those obtained in corresponding adult patients.
Patients and Methods Nine children and 43 adults with ATS valves implanted in the aortic position were monitored over both middle cerebral arteries with transcranial Doppler ultrasound. MES were identified on-line according to standard criteria. Heart rate and rhythm, valve type, size and duration, patients' height, International Normalized Ratio, and prevalence of neurological complications were obtained from all study participants.
Results MES prevalence and counts were significantly higher in children compared with adult patients (100% versus 25.5% and 58 [18.5 to 115.5] versus 5.5 [2 to 10.5], median, 95% CI, respectively). No corresponding differences in valve size or duration of valve implant were evident, but children had faster heart rates and were significantly smaller compared with adults. A positive correlation between patients' size, heart rate, and MES counts was noted.
Conclusions MES counts in children with mechanical prosthetic valves are significantly higher compared with those in corresponding adults. We hypothesize that this is due to (1) the shorter distance between aortic valve and middle cerebral artery, since cavitation bubbles have a short life span and are bound to dissolve with time, and (2) the faster heart rate in children, resulting in a higher number of valve closures per minute.
Key Words: embolism ultrasonics heart valve prosthesis
This article has been cited by other articles:
![]() |
R. Brey and M. Carolin Detection of cerebral microembolic signals by transcranial Doppler may be a useful part of the equation in determining stroke risk in patients with antiphospholipid antibody syndrome Lupus, January 1, 1997; 6(8): 621 - 624. [PDF] |
||||
|
Stroke Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1997 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |