(Stroke. 2000;31:1335.)
© 2000 American Heart Association, Inc.
Original Contributions |
From the Department of Clinical Neurosciences (M.C., R.D., Z.K., H.S.M.), St Georges Hospital Medical School, London, UK; Nicolet-EME GmbH (G.R.), Kleinostheim, Germany; the Department of Neurology (R.D., D.W.D.), Munster, Germany; the Department of Clinical Neurophysiology (R.A.), St Antonius Hospital, Utrecht, Netherlands; the Department of Neurology (V.B.), Boston University of Medicine, Boston, Mass; the Department of Neurology (D.G.), Southern General Hospital, Glasgow, UK; the Department of Neurology (M.S.), University of Dusseldorf, Dusseldorf, Germany; and the Department of Neurology (L.V.), University of Toulouse, Toulouse, France.
Correspondence to Prof Hugh Markus, Clinical Neuroscience, St Georges Hospital Medical School, Cranmer Terrace, London, SW17 ORE, England. E-mail h.markus{at}sghms.ac.uk
Background and PurposeThe clinical application of Doppler detection of circulating cerebral emboli will depend on a reliable automated system of embolic signal detection; such a system is not currently available. Previous studies have shown that frequency filtering increases the ratio of embolic signal to background signal intensity and that the incorporation of such an approach into an offline automated detection system markedly improved performance. In this study, we evaluated an online version of the system. In a single-center study, we compared its performance with that of a human expert on data from 2 clinical situations, carotid stenosis and the period immediately after carotid endarterectomy. Because the human expert is currently the "gold standard" for embolic signal detection, we also compared the performance of the system with an international panel of human experts in a multicenter study.
MethodsIn the single-center evaluation, the performance of the software was tested against that of a human expert on 20 hours of data from 21 patients with carotid stenosis and 18 hours of data from 9 patients that was recorded after carotid endarterectomy. For the multicenter evaluation, a separate 2-hour data set, recorded from 5 patients after carotid endarterectomy, was analyzed by 6 different human experts using the same equipment and by the software. Agreement was assessed by determining the probability of agreement.
ResultsIn the 20 hours of carotid stenosis data, there
were 140 embolic signals with an intensity of
7 dB. With the software
set at a confidence threshold of 60%, a sensitivity of 85.7% and a
specificity of 88.9% for detection of embolic signals were obtained.
At higher confidence thresholds, a specificity >95% could be
obtained, but this was at the expense of a lower sensitivity. In the 18
hours of postcarotid endarterectomy data, there
were 411 embolic signals of
7-dB intensity. When the same confidence
threshold was used, a sensitivity of 95.4% and a specificity of 97.5%
were obtained. In the multicenter evaluation, a total of 127 events
were recorded as embolic signals by at least 1 center. The total
number of embolic signals detected by the 6 different centers was 84,
93, 108, 92, 63, and 78. The software set at a confidence threshold of
60% detected 90 events as embolic signals. The mean probability of
agreement, including all human experts and the software, was 0.83, and
this was higher than that for 2 human experts and lower than that for 4
human experts. The mean values for the 6 human observers were averaged
to give P=0.84, which was similar to that of the
software.
ConclusionsBy using the frequency specificity of the intensity increase occurring with embolic signals, we have developed an automated detection system with a much improved sensitivity. Its performance was equal to that of some human experts and only slightly below the mean performance of a panel of human experts
Key Words: carotid artery diseases cerebral embolism observer variation signal processing, computer-assisted ultrasonography
This article has been cited by other articles:
![]() |
R. M. Kwee, R. J. van Oostenbrugge, L. Hofstra, G. J. Teule, J.M.A. van Engelshoven, W. H. Mess, and M. E. Kooi Identifying vulnerable carotid plaques by noninvasive imaging Neurology, June 10, 2008; 70(24_Part_2): 2401 - 2409. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Dittrich, M. A. Ritter, M. Kaps, M. Siebler, K. Lees, V. Larrue, D. G. Nabavi, E. B. Ringelstein, H. S. Markus, and D. W. Droste The Use of Embolic Signal Detection in Multicenter Trials to Evaluate Antiplatelet Efficacy: Signal Analysis and Quality Control Mechanisms in the CARESS (Clopidogrel and Aspirin for Reduction of Emboli in Symptomatic carotid Stenosis) Trial Stroke, April 1, 2006; 37(4): 1065 - 1069. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Kalairajah, A. J. Cossey, G. M. Verrall, G. Ludbrook, and A. J. Spriggins Are systemic emboli reduced in computer-assisted knee surgery?: A PROSPECTIVE, RANDOMISED, CLINICAL TRIAL J Bone Joint Surg Br, February 1, 2006; 88-B(2): 198 - 202. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. L. Abbott, B. R. Chambers, J. L. Stork, C. R. Levi, C. F. Bladin, and G. A. Donnan Embolic Signals And Prediction of Ipsilateral Stroke or Transient Ischemic Attack in Asymptomatic Carotid Stenosis: A Multicenter Prospective Cohort Study Stroke, June 1, 2005; 36(6): 1128 - 1133. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. A Rodriguez, K. A Williams, A. Babaev, F. Rubens, and H. J Nathan Effect of perfusionist technique on cerebral embolization during cardiopulmonary bypass Perfusion, January 1, 2005; 20(1): 3 - 10. [Abstract] [PDF] |
||||
![]() |
I. K. Moppett and R. P. Mahajan Transcranial Doppler ultrasonography in anaesthesia and intensive care Br. J. Anaesth., November 1, 2004; 93(5): 710 - 724. [Full Text] [PDF] |
||||
![]() |
A. D. Mackinnon, R. Aaslid, and H. S. Markus Long-Term Ambulatory Monitoring for Cerebral Emboli Using Transcranial Doppler Ultrasound Stroke, January 1, 2004; 35(1): 73 - 78. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Laas, S. Kseibi, M. Perthel, A. Klingbeil, L'E. El-Ayoubi, and A. Alken Impact of high intensity transient signals on the choice of mechanical aortic valve substitutes Eur. J. Cardiothorac. Surg., January 1, 2003; 23(1): 93 - 96. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Brucher and D. Russell Automatic Online Embolus Detection and Artifact Rejection With the First Multifrequency Transcranial Doppler Stroke, August 1, 2002; 33(8): 1969 - 1974. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Devuyst, G.A. Darbellay, J.-M. Vesin, V. Kemeny, M. Ritter, D.W. Droste, C. Molina, J. Serena, R. Sztajzel, P. Ruchat, et al. Automatic Classification of HITS Into Artifacts or Solid or Gaseous Emboli by a Wavelet Representation Combined With Dual-Gate TCD Stroke, December 1, 2001; 32(12): 2803 - 2809. [Abstract] [Full Text] [PDF] |
||||
|
Stroke Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2000 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |