(Stroke. 1995;26:2044-2047.)
© 1995 American Heart Association, Inc.
Articles |
From the Department of Neurology, King's College Hospital School of Medicine and Dentistry, and the Institute of Psychiatry, London, UK.
Correspondence to Dr Hugh Markus, Department of Neurology, Institute of Psychiatry, De Crespigny Park, London, SE5 8AF, UK.
| Abstract |
|---|
|
|
|---|
Methods To test the importance of this potential problem, the same 25 embolic signals recorded as the audio signal on digital audiotape were each played repeatedly through a transcranial Doppler ultrasound (TCD) system using fast Fourier transform analysis. An older system with no time-window overlap was used, and a more modern system was also used in which three different degrees of overlap were used: -9%, 27%, and 57%. The number of signals audible but not appearing on the spectral display was recorded. The variability in the relative intensity increase for the same embolic signal played repeatedly was estimated by calculating the coefficient of variation of the relative intensity increase.
Results With the older system, 39/500 (7.8%) of embolic signals were missed. With the newer system, the number of embolic signals missed was fewer and decreased with increasing degrees of overlap (10/500 for -9% overlap, 1/500 for 27% overlap, and 0/500 for 57% overlap). For those setups in which embolic signals were missed, there was a highly significant relationship between duration of embolic signal and number of signals missed. In parallel with these results, the coefficient of variation of the relative intensity increase became progressively less with increasing degrees of time-window overlap. For all processing setups, the coefficient of variation was greater for the less intense and shorter duration signals, but this dependence, as estimated by the slope of the regression line, became less strong with higher degrees of overlap.
Conclusions Inadequate degrees of fast Fourier transform time-window overlap will result in the failure of current TCD machines to detect embolic signals. Furthermore, this and the time windowing currently usually used may result in variability in the relative intensity increase of identical embolic signals. These factors need to be taken into account when comparing data on the frequencies of embolic signals recorded by different researchers and in the design of future TCD equipment.
Key Words: cerebral embolism ultrasonics
| Introduction |
|---|
|
|
|---|
A specific theoretical problem is the speed of processing and time
windowing of current commercially available transcranial
Doppler (TCD) machines.12 This problem is suggested by
the observation that a typical clicking or chirping sound
characteristic of an embolic signal may be heard in the audio
Doppler signal, without a high-intensity signal appearing on
the spectral display, and that this is particularly common for
low-intensity and short-duration signals, such as those
detected in patients with carotid artery disease. Most TCD machines use
a fast Fourier transform (FFT) during signal processing; each FFT is
then displayed as one column on the time/velocity (or frequency)
display. The execution of each FFT requires a specific amount of time,
and depending on the speed of the processor, a variable amount of
the FFT time window can be overlapped with the next FFT. If no overlap
occurs, there is the possibility that an embolic signal may pass
through the sample volume but arrive between the sampling periods of
the two time windows and not be displayed on the screen (Fig 1
, top). Many currently used TCD
machines were designed for flow-velocity measurement, not with
embolus detection specifically in mind, and are equipped with computers
unable to process the Doppler signal rapidly enough to ensure a
high degree of FFT overlap; this could theoretically account for
embolic signals being missed. A related problem that may arise from an
inadequate degree of FFT overlap is variability in the measurement of
the relative intensity of an embolic signal. To improve the spectral
appearance, a time window (such as a Hanning time window) is usually
used during signal analysis.13 This increases the
relative amplitude of signals occurring during the center of the time
window, but signals occurring at the temporal edges of the time window
will have their relative amplitude reduced (Fig 1
). If
there is sufficient overlap of the time windows, this will not create a
problem, but if the overlap is insufficient, an embolic signal being
sampled in the center of an FFT sampling time is likely to result in a
more intense embolic signal than one that is sampled at the temporal
edges of a sampling period (Fig 1
, bottom). Embolic
signal intensity depends on both embolus size and
material1 14 and may allow limited characterization of
embolus type and size15 ; however, estimates of intensity
increase for short-duration embolic signals could be inaccurate if
such processing difficulties occur.
|
In this study, the importance of processing speed and FFT time-window overlap was examined using embolic signals recorded from patients, which were subsequently analyzed in standard TCD systems.
| Materials and Methods |
|---|
|
|
|---|
Twenty-five embolic signals were analyzed. Each embolic signal was played back 20 times using each of four different processing setups. (1) The first was a TC2000 equipped with a 286 processor. This is an older system with no time-window overlap, although determination of the exact degree of overlap is not possible. For the other setups, the same TC2000 was used equipped with a 386 processor in place of the 286 processor. By altering the sweep speed, three degrees of FFT overlap were used: (2) 57% (sweep time, 3.5 seconds); (3) 27% (sweep time, 6.0 seconds); and (4) -9% (sweep time, 8.9 seconds). Overlap was calculated as overlap(%)=PRF/FFT points ([FFT points/PRF]-[sweep time/columns]), where FFT points is the number of points in the FFT (128 in this study), sweep time is time(s) taken for the signal to sweep the screen once, and columns is the number of columns in the spectral display (512 in this study).
In practice, because the processor performs both FFT analysis and other functions simultaneously, the above calculations are an approximation; the actual degree of overlap is likely to be less than stated, but the relative differences between the different settings will be similar.
For each embolic signal and processing setup, the number of embolic signals missed (heard as a typical sound but not displayed) was recorded. Where a visible embolic signal was present, the maximum relative intensity increase of each embolic signal during each of the 20 playbacks was measured. To calculate the variability of the relative intensity increase associated with a single embolic signal, the coefficient of variation (%) of relative intensity increase across the 20 playbacks was then calculated for each embolic signal and setup combination.
The embolic signals were selected to represent a range of durations and relative intensity increases. Relative intensity increase (decibels) was calculated from the equation 10 log (maximum intensity of embolic signal/mean background intensity), and duration was calculated from the duration (number of time columns) of the high-intensity (>4 dB) embolic signal. Fourteen signals were from patients with symptomatic carotid artery disease, and 11 were from patients with mechanical heart valves. Mean±SD relative intensity increase was 8.3±1.42 dB (range, 5.0 to 9.9 dB), and mean±SD duration was 32.4±18.6 milliseconds (range, 5 to 80 milliseconds); values were higher for valve than carotid emboli (intensity increase: 9.2 versus 7.9 dB, t test P=.0028; duration: 47.7 versus 20.4 milliseconds, t test P=.0001). When calculating correlations, the values of relative intensity increase and duration of embolic signal for each embolic signal were taken from the replay (of the 20 replays) in which the embolic signal had the maximum relative intensity increase; it was assumed that during this replay the embolic signal was processed nearest to the center of the time window.
| Results |
|---|
|
|
|---|
corrected for ties=-0.815,
P=.0001). No embolic signals with a duration of more than 40
milliseconds were missed. More carotid emboli than valve embolic
signals were missed (mean±SD, 2.64±2.06 versus 0.18±0.41;
t test P=.0007), which was consistent
with the carotid embolic signals being of shorter duration and lower
intensity.
|
386 Processor
The number of embolic signals missed was fewer and decreased with
increasing degrees of FFT overlap (10/500 for -9% overlap, 1/500
for 27% overlap, and 0/500 for 57% overlap). For analysis
using the -9% overlap, there was a highly significant
relationship between duration and number missed (Spearman's
corrected for ties=-0.567, P=.01); no embolic signals
with a duration greater than 15 milliseconds were missed.
Paralleling these results, the coefficient of variation of the relative
intensity increase became progressively less with increasing degrees of
FFT overlap (Fig 3
). For all processing
setups, the coefficient of variation was greater for the shorter
duration signals, but this dependency, as estimated by the regression
coefficient and the slope of the regression line between duration of
embolic signal and coefficient of variation in relative intensity
increase, became less strong with higher degrees of FFT overlap and was
no longer significant for the highest degree of overlap. Values for
Pearson correlation coefficient (r) and the slope of the
regression line were r=-.566, P=.0032, and
slope=- 0.125 dB/ms for -9% overlap; r=-.404,
P=.045, and slope=-0.09 dB/ms for 27% overlap; and
r=.239, P=.25, and slope=-0.034 dB/ms for
57% overlap.
|
| Discussion |
|---|
|
|
|---|
The number of "missed" signals is greater for short-duration low-intensity signals. Therefore, it is unlikely that a significant number of embolic signals recorded from patients with mechanical heart valves will be missed, as these signals have a higher intensity compared with those in patients with carotid artery disease.15 However, this is likely to be a problem in studies in carotid stenosis, and there has been a particularly wide variation in the frequency of embolic signals in patients with carotid artery stenosis, in whom the embolic signals are often of low intensity and short duration. Furthermore, many current studies use a threshold relative intensity increase as one of the criteria for identifying an embolic signal; the variation in relative intensity increase for smaller signals may result in only a proportion of identical embolic signals being detected in the middle of the time frame, therefore exceeding the threshold intensity used in identifying embolic signals. This may account for variation in the number of embolic signals reported in different studies even when embolic signals are not completely "missed" on the spectral display.
The variability in the intensity of the same embolic signals will also be important if information on embolus size and material is to be derived from analysis of the spectral display. Experimental studies have demonstrated a relationship between embolus size and both relative intensity increase and duration of high-intensity signal.14 16 These studies were performed using larger experimental emboli, for which the effect of inadequate FFT overlap is likely to be unimportant. The situation may be different for the much lower amplitude and shorter duration signals recorded in conditions such as carotid artery disease and atrial fibrillation.
A number of alternative strategies will reduce this problem. A higher degree of overlap is possible with the newest TCD machines, which use more powerful processors, but this will be reduced in these machines if recordings are made from multiple channels simultaneously. Alternative time windows that are not so heavily weighted toward the signal in the center of the time frame may reduce the variability of the relative intensity increase. Different methods of signal analysis, such as Wigner analysis with its higher temporal resolution, offer an alternative approach.17
The detection of embolic signals may offer a useful tool in the investigation and management of patients with or at risk for cerebrovascular disease. However, before it can be used in routine clinical practice and before multicenter prospective outcome studies can be performed, the reason for the differences in the frequency of embolic signals in different centers needs to be determined. While this may represent differences in patient groups, this study demonstrates that machine characteristics are important and that the degree of overlap can have dramatic effects on the number of embolic signals visualized on the spectral display. Future studies should take the degree of time frame overlap into account; a simple assessment of its importance for different systems can be performed by recording the Doppler audio signal before it has undergone its FFT analysis and repeatedly replaying the portion containing an embolic signal. The degrees of overlap will vary within individual machines according to recording parameters such as pulse repetition frequency and sweep speed, and these need to be accounted for in any such analysis.
| Acknowledgments |
|---|
Received July 25, 1995; revision received July 25, 1995; accepted July 27, 1995.
| References |
|---|
|
|
|---|
2.
Siebler M, Kleinschmidt A, Sitzer M, Steinmetz H,
Freund HJ. Cerebral microembolism in symptomatic and
asymptomatic high-grade internal carotid artery
stenosis. Neurology.. 1994;44:615-618.
3. Georgiadis D, Grosset DG, Quin RO, Nichol JA, Bone I, Lees KR. Detection of intracranial emboli in patients with carotid disease. Eur J Vasc Surg.. 1994;8:309-314. [Medline] [Order article via Infotrieve]
4. Georgiadis D, Grosset DG, Kelman A, Faichney A, Lees KR. Prevalence and characteristics of intracranial microemboli signals in patients with different types of prosthetic cardiac valves. Stroke.. 1994;25:587-592. [Abstract]
5. Markus HS, Droste DW, Brown MM. Detection of asymptomatic cerebral embolic signals with Doppler ultrasound. Lancet.. 1994;343:1011-1012. [Medline] [Order article via Infotrieve]
6. Rams JJ, Davis DA, Lolley DM, Berger MP, Spencer M. Detection of microemboli in patients with artificial heart valves using transcranial Doppler: preliminary observations. J Heart Valve Dis.. 1993;2:37-41. [Medline] [Order article via Infotrieve]
7. Tong DC, Bolger A, Albers GW. Incidence of transcranial Doppler-detected cerebral microemboli in patients referred for echocardiography. Stroke.. 1994;25:2138-2141. [Abstract]
8. Pugsley W, Klinger L, Paschalis C, Treasure T, Harrison M, Newman S. The impact of microemboli during cardiopulmonary bypass on neuropsychological functioning. Stroke.. 1994;25:1393-1399. [Abstract]
9.
Spencer MP, Thomas GI, Nicholls SC, Sauvage
LR. Detection of middle cerebral artery emboli during carotid
endarterectomy using transcranial
Doppler ultrasonography. Stroke.. 1990;21:415-423.
10. Babikian VL, Hyde C, Pochay V, Winter MR. Clinical correlates of high-intensity transient signals detected on transcranial Doppler sonography in patients with cerebrovascular disease. Stroke.. 1994;25:1570-1573. [Abstract]
11. Ries S, Schminke U, Daffertshofer M, Hennerici M. Emboli detection by TCD in patients with cerebral ischaemia. Cerebrovasc Dis. 1994;4-3S:22. Abstract.
12.
Siebler M, Rose G, Sitzer M, Bender A, Steinmetz
H. Real-time identification of cerebral microemboli with US
feature detection by a neural network. Radiology. 1994;192:739-742.
13. Oppenheim AV, Schafer RW. Discrete-Time Signal Processing. Englewood Cliffs, NJ: Prentice-Hall Inc; 1989.
14.
Markus HS, Brown MM. Differentiation between
different pathological cerebral embolic materials using
transcranial Doppler in an in vitro model.
Stroke.. 1993;24:1-5.
15.
Grosset DG, Georgiadis D, Kelman AW, Lees KR.
Quantification of ultrasound emboli signals in patients with cardiac
and carotid disease. Stroke.. 1993;24:1922-1924.
16. Russell D, Brucher R, Madden K, Clark WM, Sandset PM, Zivin J. The intensity of the Doppler signal caused by arterial emboli depends on embolus size. In: Oka M, von Reutern GM, Furuhata H, Kodaira K, eds. Recent Advances in Neurosonology. Amsterdam, Netherlands: Elsevier Science Publishing Co; 1992.
17. Smith JL, Evans DH, Fan L, Thrush AJ, Naylor AR. Processing Doppler ultrasound signals from blood borne emboli. Ultrasound Med Biol. 1994;20:455-462.
This article has been cited by other articles:
![]() |
S. Gao, K. S. Wong, T. Hansberg, W. W. M. Lam, D. W. Droste, and E. B. Ringelstein Microembolic Signal Predicts Recurrent Cerebral Ischemic Events in Acute Stroke Patients With Middle Cerebral Artery Stenosis Stroke, December 1, 2004; 35(12): 2832 - 2836. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Bernd Ringelstein, D. W. Droste, V. L. Babikian, D. H. Evans, D. G. Grosset, M. Kaps, H. S. Markus, D. Russell, and M. Siebler Consensus on Microembolus Detection by TCD Stroke, March 1, 1998; 29(3): 725 - 729. [Abstract] [Full Text] [PDF] |
||||
![]() |
L Valton, V Larrue, A P. Le Traon, and G Geraud Cerebral microembolism in patients with stroke or transient ischaemic attack as a risk factor for early recurrence J. Neurol. Neurosurg. Psychiatry, December 1, 1997; 63(6): 784 - 787. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. S. Markus and J. Molloy Use of a Decibel Threshold in Detecting Doppler Embolic Signals Stroke, April 1, 1997; 28(4): 692 - 695. [Abstract] [Full Text] |
||||
![]() |
H.-C. Koennecke, H. Mast, S. S. Trocio Jr, R. L. Sacco, J. L. P. Thompson, and J. P. Mohr Microemboli in Patients With Vertebrobasilar Ischemia : Association With Vertebrobasilar and Cardiac Lesions Stroke, March 1, 1997; 28(3): 593 - 596. [Abstract] [Full Text] |
||||
![]() |
E. V. Van Zuilen, W. H. Mess, C. Jansen, I. Van Der Tweel, J. Van Gijn, and R. G.A. Ackerstaff Automatic Embolus Detection Compared With Human Experts: A Doppler Ultrasound Study Stroke, October 1, 1996; 27(10): 1840 - 1843. [Abstract] [Full Text] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|