(Stroke. 1995;26:2048-2052.)
© 1995 American Heart Association, Inc.
Articles |
From the Department of Neurology, University at Lübeck (M.K., G.S.), and Byk Gulden Pharmaceuticals, Clinical Pharmacology, Konstanz (P.S., K.-D.B., H.B., W.W.), Germany.
Correspondence to Prof Dr M. Kaps, Klinik für Neurologie, Universität zu Lübeck, Ratzeburger Allee 160, D-23538 Lübeck, Germany.
| Abstract |
|---|
|
|
|---|
Methods In two phase I studies, 8 healthy volunteers received a spherosome suspension containing a phospholipid as the active ingredient. The intravenous injection was performed in three doses (2.5, 5, and 10 mL) at four different injection rates (0.25, 0.5, and 1 mL/s and bolus). The duration and degree of the signal enhancement were measured by two transcranial ultrasonic procedures presently used in clinical practice: transcranial Doppler sonography (TCD) and transcranial color-coded sonography (TCCS). The assessment of tolerability was based on chemical laboratory parameters and hemodynamic data (heart rate, blood pressure, electrocardiogram) and on questionnaires relating to general well-being.
Results BY963 was tolerated without complications. All 38 administrations of the echo contrast medium produced a marked increase in the TCD signal (>30 dB) in the intracranial basal cerebral arteries. To obtain the optimum time window for diagnostic use, higher doses with slower injection rates are advantageous. The duration of optimal contrasting was 42 to 68 seconds (TCD) and 12 to 132 seconds (TCCS), depending on the method and mode of administration. Bolus injections gave rise to an increased incidence of color artifacts.
Conclusions BY963 significantly improves intracranial Doppler imaging while being well tolerated. The signal enhancement lasts long enough for TCCS to display all basal cerebral arteries after just one injection.
Key Words: cerebrovascular disorders diagnostic imaging transcranial Doppler ultrasonics
| Introduction |
|---|
|
|
|---|
The most serious problem of transcranial ultrasound diagnostics, however, is still the acoustic "window failure."2 Depending on age, sex, and race, approximately 18% of stroke patients cannot be examined in this way because of insufficient ultrasound transmission through the temporal region of the skull, and the percentage of failures can be as high as 50% among nonwhites.3
High hopes are therefore pinned on the use of echo contrast media, which allow a significant signal enhancement.4 5 6 7 8 An ideal echo contrast medium must be well tolerated, capable of intravenous administration, and sufficiently stable to pass through the pulmonary circulation and the heart into cerebral circulation. There should also be a standardized microbubble size, and the course of image degradation should be calculable as a function of time.
Studies with a galactose microbubble suspension (SHU 508
A) have shown that by appropriate enhancement of the signal, the distal
brain arteries and the veins can also be imaged, even though as a rule
they escape noncontrast examination.6 8 According to
recent experience, the time course, duration, and extent of the
Doppler signal enhancement depend to some extent on the echo
contrast medium used. In the present study we have investigated for
the first time the cerebral circulation using a new standardized
substance (BY963) that contains a phospholipid derived from soybean
lecithins as its active ingredient (Table 1
).
|
The aim of the study was to test the clinical tolerability of BY963 and to clarify how its diagnostic usefulness is influenced by the dose and mode of administration. For this purpose the signal enhancement was measured with two different transcranial ultrasonic techniques (TCD and TCCS).
| Subjects and Methods |
|---|
|
|
|---|
The spherosome suspension of the echo contrast medium was prepared
immediately before the injection by agitating 5 mL of the lyophilizate
with 0.18 mL of air in a special mixing chamber. The system comprises
two syringes connected by an adapter, and the carrier solution is
prepared for use by repeated agitation of first one side and then the
other to obtain a homogeneous suspension of air-filled
microbubbles with a standardized bubble size of approximately 3.8 mm
(Table 1
).
Tolerability and Drug Safety
The tolerability of the echo contrast medium injections was
assessed on the basis of 12-channel electrocardiographic
recordings, blood pressure and pulse measurements,
questionnaires relating to changes in general well-being, clinical
laboratory parameters, and urinalyses. The tolerability
data were collected 30 minutes before and 3, 5, and 24 hours after the
injections and 14 days after the end of the study. During the injection
an electrocardiogram was recorded continuously in
one lead.
TCD Studies
The TCD recordings were made with the use of a pulsed
ultrasound system (Multi-Dop X, DWL) following the usual
guidelines.9 The sound signals were recorded
continuously with the use of a 2-MHz head fixed over the temporal
acoustic window with a spectacle-frame holder. The signal
enhancement was quantified with the use of an on-line fast Fourier
transformation and displayed on the screen in flash color in 3-dB
steps.
At the start of the study the Doppler sample volume was centered on the main stem of the right MCA, and the signal amplification was then reduced until the frequency spectrum was no longer discernible on the screen. The contrast medium was administered with the instrument kept at this constant setting.
BY963 (Table 1
) was administered into the right cubital
vein in three healthy volunteers (age, 23 to 27 years; height, 1.79 to
1.89 m; weight, 70 to 76 kg) in two doses (2.5, 5 mL) and at three
different speeds (0.25, 0.5, 1.0 mL/s). After each of the six
injections, 0.9% saline was used to rinse any residues of the contrast
medium out of the arm veins. The data were recorded on diskettes
and on videotape.
The evaluation was performed off-line by two experienced
investigators who determined the time to the appearance of the echo
contrast medium in the MCA main stem after the start of injection (Fig 1
). This time to contrast appearance is
designated TA. In addition, the maximum signal enhancement was
recorded on the basis of the color-coded image, as was the
duration of homogeneous representation of the
Doppler frequency spectrum (as a parameter of
diagnostic value). The decline of the effect of the echo
contrast medium was recognizable acoustically by a characteristic
"splashing" noise and optically by interruptions in the envelope
curve of the Doppler frequency spectrum (Fig 1
). The
following parameters were measured after each contrast
medium injection: (1) the latency period (TA) from the start of
injection to the detection of echo contrast medium in the MCA (in
seconds) and (2) maximum signal enhancement (in decibels) and the
duration of the optimum/homogeneous enhancement without
overcontrasting or fragmentation of the Doppler envelope curve (in
seconds).
|
TCCS Studies
For the color imaging studies we used a standard 64-channel
phased-array system with a 2.7-MHz ultrasonic head (HP Sonos 1500,
Hewlett-Packard Co). The imaging of the basal cerebral arteries was
performed transtemporally with an overall view in the axial
plane.10 11 Before administration of the echo contrast
medium, the color amplification was reduced so that parenchymatous
structures were still visible but not the color coding of the cerebral
arteries.
Five more healthy volunteers (age, 26 to 36 years; height, 1.66 to 1.89 m; weight, 63 to 99 kg) were injected through a cubital vein with three different doses (2.5, 5.0, 10 mL) of BY963 at an injection rate of 0.25 or 1.00 mL/s or as a bolus (3 mL/s). An interval of 1 hour was allowed between injections to ensure that the previous dose of echo contrast medium had been completely washed out.
The echo contrast effect was recorded on videotape and evaluated
off-line. Four different phases were defined for the purposes of
this evaluation (Fig 2
): phase I, latency
period from the beginning of injection of BY963 to the first color
image of the cerebral arteries (Fig 2
, top left panel);
phase II, excessive enhancement of the signal, with color artifacts
("blooming") and limited diagnostic usefulness; phase
III, optimum imaging of the basal cerebral arteries (Fig 2
, top right panel); and phase IV,
"fragmentation" of the color signal, again with limited
diagnostic usefulness (Fig 2
, bottom left
panel). Recording was stopped after 3 minutes even if some
contrast effects were still discernible in individual artery sections
(Fig 2
, bottom right panel).
|
| Results |
|---|
|
|
|---|
TCD Studies
Table 2
shows that the TA varied from 12 to 28
seconds. The TA depends on the heart rate and on the speed of
injection. Individually (ie, at a constant heart rate) the TA varied
only slightly (1.7±2.4 seconds; maximum, 4 seconds), independent of
the infusion velocity. Higher injection rates decreased the TA. The
signal amplitude was in all cases increased by more than 30 dB.
|
The duration of optimum contrasting was dose dependent and lasted from
42 to 68 seconds (Table 2
). A higher dose of BY963
increased the optimum diagnostic time window. Slow speed of
injection is preferable because it results in a shorter phase of
overcontrasting.
TCCS Studies
After all 20 injections of BY963, we noted a distinct enhancement
of the color signal, which permitted differentiation of the
large-caliber basal cerebral arteries on both ipsilateral and
contralateral sides; moreover, on the ipsilateral side the distal MCA
branches, which otherwise escape detection, were imaged (Fig 2
).
Evaluations of the TCCS images showed that the TA depended on the mode
of injection and not on the dose (Fig 3
; Table 3
). Bolus injections resulted in a
prolonged blooming phase at the expense of the period of optimum
diagnostic usefulness. The length of phase III was dose
dependent (minimum, 12 seconds; maximum, 132 seconds). In all tests,
phase IV lasted more than 3 minutes after the start of injection.
|
|
| Discussion |
|---|
|
|
|---|
No relevant adverse events were observed in this study, as in all previous studies with BY963.12 The tolerability even in high doses of up to 30 mL as a bolus injection appears to be good (M.K. et al, unpublished data). In particular, there were no clinical central nervous symptoms. Further studies of drug safety using more differentiated cerebral function tests and electroencephalographic monitoring are currently under way.
Since all of the healthy volunteers provided good transcranial ultrasound images even without echo contrast media, after adjusting the TCD measurement volume in the main stem of the MCA (in the TCCS projection plane), we reduced the Doppler gain until no Doppler signal was detectable. This adjustment was not altered during the course of the study and served as the reference for evaluating the echo contrast effect. As the next step we envisage studies under clinical conditions in patients with the problem of poor ultrasound penetration.
The results of TCD and TCCS are in agreement regarding the typical course of the echo contrast effect after BY963. After intravenous injection into a cubital vein (TCD studies), 12 to 28 seconds pass (varying between individuals) before an enhancement of the Doppler signals from the MCA becomes apparent. A phase of overcontrasting follows, which eventually (depending on the mode of administration) passes into a phase of diagnostic interest, during which optimum imaging of the cerebral arteries is obtained, free from unwanted signal artifacts. Both in TCD and in TCCS the injection rate has a particular significance for diagnostic usefulness. Bolus injections shorten TA and prolong overenhancement (blooming), but they do not prolong the diagnostically useful time. For the same injection rate the TAs measured in TCD studies agree well with the values from TCCS (20 to 22 seconds for an injection rate of 0.25 mL/s). Comparable measurements have been reported by other working groups using galactose particles as the contrast substance.5 6 8
In subjects with healthy hearts the TA, which we have expressed in seconds, depends on the heart rate: the echo contrast medium enters the cerebral circulation sooner when the heart rate is faster. In individuals (ie, for the same heart rate) the TAs remained constant regardless of the amount of substance injected.
TCD Studies
The time for which the envelope curve of the Doppler frequency
spectrum remained uninterrupted and free from artifacts was used to
quantify the duration of the diagnostic effect. During this
time quantitative analysis of the spectrum delivers relevant
and useful data.13 In the phase that follows
quantification is more problematic, but the positions of
the brain vessels can still be identified. Thus, the meaning of
"diagnostically useful" is to some extent a question
of definition.
The signal enhancement was quantified with the use of a color scale for
relative decibel values. Since in the later phases the Doppler
frequency spectrum becomes inhomogeneous (Fig 1C
) but individual particles of contrast medium
continue to amplify the signal by more than 30 dB even after several
minutes have elapsed, we have not constructed a dose-effect curve
in this study. BY963 can also be detected in the jugular vein in
healthy volunteers, ie, it passes through the capillaries and into the
venous system (M.K. et al, unpublished data, 1994) and undergoes
recirculation.
TCCS Studies
Because of the different assessment criteria, the duration of the
diagnostically relevant window cannot be compared with the
values from TCD. Experience shows that after a slow injection of 10 mL
BY963 (0.25 mL/s), all basal cerebral arteries may be imaged
simultaneously for a period of approximately 1.5 minutes,
even when unfavorable technical parameters have been
selected (eg, axial view plane of the entire brain, 2.7-MHz sector
probe, insensitive color parameters, and gain reduction).
In the case of acute stroke, it would therefore be possible to identify
patients with a main trunk occlusion in the middle or posterior
cerebral arteries on arrival at the hospital in a reliable manner and
without difficulty or any great expense. Diagnosis would thus be
considerably hastened, and better planning of the next steps in
treatment would be possible. Recent clinical echo contrast studies in
patients with cerebrovascular disease have demonstrated that the change
in diagnosis from uncertain to firmly established was possible in the
majority of the cases.8
Distal arterial sections were also imaged, but they were not analyzed in greater detail because the study protocol specified a fixed projection plane. More detailed studies on this topic have been performed with the use of galactose microparticles, which revealed otherwise invisible intracranial arterial and venous segments by color duplex sonography.6 8
In TCCS studies superimposed artifacts have a special impact; individual arterial sections run together or become round and swollen, and clouds of color and intense reflections can complicate an assessment considerably. TCCS seems to be more sensitive to overcontrasting than TCD. Even after phase III (optimum investigation conditions), the cerebral arteries can still be localized by TCCS for some time, although the images of arterial segments are no longer connected.
While all transpulmonary echo contrast media used thus far in neuroangiology depend for their action on encapsulated microbubbles of air, there are significant differences between carriers. The diagnostic properties will also be different. Because of methodological differences, a comparison of our results with other echo contrast media is difficult, since we used healthy volunteers rather than patients with insufficient temporal acoustic transparency. Moreover, the intervals measured depend on how the echo contrast effect is defined, and the technical parameters of the study (eg, signal amplification) play a special part. In routine examinations the initial overcontrasting phase would be reduced simply by turning down the amplification on the instrument. Similarly, the diagnostic window could be considerably prolonged in the later phase simply by increasing amplification.
It must be emphasized that the data on administration methods are applicable only to the imaging of basal cerebral arteries. Other regions of the vascular system (eg, coronary or peripheral arteries) may require a different mode of administration.
| Selected Abbreviations and Acronyms |
|---|
|
| Acknowledgments |
|---|
Received May 2, 1995; revision received July 5, 1995; accepted July 7, 1995.
| References |
|---|
|
|
|---|
2.
Bornstein NM, Norris JW.
Transcranial Doppler sonography is at present of
limited value. Arch Neurol. 1994;51:1057-1059.
3.
Halsey JH. Effect of emitted power on waveform
intensity in transcranial Doppler.
Stroke. 1990;21:1573-1578.
4. Ries F, Kaal K, Schultheiss R, Solymosi L, Schlief R. Air bubbles as a contrast medium in transcranial Doppler sonography. J Neuroimaging.. 1991;1:173-178. [Medline] [Order article via Infotrieve]
5.
Ries F, Honisch C, Lambertz M, Schlief R. A
transpulmonary contrast medium enhances the
transcranial Doppler signal in humans.
Stroke. 1993;24:1903-1909.
6.
Bogdahn U, Becker G, Schlief R, Reddig J, Hassel
W. Contrast-enhanced transcranial
color-coded real-time sonography: results of a phase-two
study. Stroke. 1993;24:676-684.
7. Haggag KJ, Russel D, Brucher R, Dahl A, Jakobsen J, Muan B. Colour duplex studies of the cranial vasculature after intravenous contrast. Cerebrovasc Dis. 1994;4(suppl 3):4. Abstract.
8.
Otis S, Rush M, Boyajian R.
Contrast-enhanced transcranial imaging: results of an
American phase-two study. Stroke. 1995;26:203-209.
9. Aaslid R, Markwalder TM, Nornes H. Noninvasive transcranial Doppler ultrasound recording of flow velocity in basal cerebral artery. J Neurosurg. 1992;57:769-774.
10.
Bogdahn U, Becker G, Winkler J, Greiner K, Perez J,
Meurers J. Transcranial color coded real-time
sonography in adults. Stroke. 1990;21:1680-1688.
11. Kaps M. Extra- und intrakranielle Farbduplexsonographie. Berlin, Germany: Springer-Verlag; 1994.
12. Bubenheimer P, Schaffer P, Beller KD, Bliesath H. Erste Anwendung des kapillargängigen Ultraschallkontrastmittels BY963 zur 2D- und Farbdopplerechokardiographie am Menschen. Ultraschall Klin Prax. 1993;8:185.
13. Petrick J, Schlief R, Affeld K, Zomack M, Urbank A. Der Effekt von Ultraschallkontrastmitteln auf die Dopplerfrequenz: Eine in vitro Studie. In: Program and abstracts of the Ultraschalldiagnostik 94 Dreiländertreffen; October 26-29, 1994; Basel, Switzerland.
This article has been cited by other articles:
![]() |
G. Deklunder Role of ultrasound and contrast-enhanced ultrasound in patients with cerebrovascular disease Eur. Heart J. Suppl., March 1, 2002; 4(suppl_C): C51 - C55. [Abstract] [PDF] |
||||
![]() |
T. Postert, B. Braun, S. Meves, O. Koster, H. Przuntek, S. Weber, and T. Buttner Contrast-Enhanced Transcranial Color-Coded Sonography in Acute Hemispheric Brain Infarction Stroke, September 1, 1999; 30 (9): e1819 - 1826. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Postert, A. Muhs, S. Meves, J. Federlein, H. Przuntek, and T. Buttner Transient Response Harmonic Imaging : An Ultrasound Technique Related to Brain Perfusion Stroke, September 1, 1998; 29(9): 1901 - 1907. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. G. Nabavi, D. W. Droste, V. Kemeny, G. Schulte-Altedorneburg, S. Weber, and E. B. Ringelstein Potential and Limitations of Echocontrast-Enhanced Ultrasonography in Acute Stroke Patients : A Pilot Study Stroke, May 1, 1998; 29(5): 949 - 954. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Postert, J. Federlein, B. Braun, O. Koster, C. Bornke, H. Przuntek, T. Buttner, R. W. Baumgartner, M. Arnold, F. Gonner, et al. Contrast-Enhanced Transcranial Color-Coded Real-Time Sonography: A Reliable Tool for the Diagnosis of Middle Cerebral Artery Trunk Occlusion in Patients With Insufficient Temporal Bone Window • Response Stroke, May 1, 1998; 29(5): 1070 - 1073. [Full Text] |
||||
![]() |
S. Knecht, M. Deppe, A. Ebner, H. Henningsen, T. Huber, H. Jokeit, and E.-B. Ringelstein Noninvasive Determination of Language Lateralization by Functional Transcranial Doppler Sonography : A Comparison With the Wada Test Stroke, January 1, 1998; 29(1): 82 - 86. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. W. Baumgartner, M. Arnold, F. Gonner, I. Staikow, C. Herrmann, A. Rivoir, and R. M. Muri Contrast-Enhanced Transcranial Color-Coded Duplex Sonography in Ischemic Cerebrovascular Disease Stroke, December 1, 1997; 28(12): 2473 - 2478. [Abstract] [Full Text] |
||||
![]() |
G. Seidel and M. Kaps Harmonic Imaging of the Vertebrobasilar System Stroke, August 1, 1997; 28(8): 1610 - 1613. [Abstract] [Full Text] |
||||
![]() |
M. Kaps, P. Schaffer, K.-D. Beller, K.-D. Eng, G. Seidel, H. Bliesath, E. Diletti, and E. Eng Characteristics of Transcranial Doppler Signal Enhancement Using a Phospholipid-Containing Echocontrast Agent Stroke, May 1, 1997; 28(5): 1006 - 1008. [Abstract] [Full Text] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Stroke Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1995 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |