(Stroke. 1996;27:683-686.)
© 1996 American Heart Association, Inc.
Articles |
From the Department of Neurology, University of Münster, and the Technical University of Cologne (J.G.) (Germany).
Correspondence to D. Georgiadis, MD, Department of Neurology, Martin-Luther-University, Ernst-Grube Str 40, 06097 Halle, Germany.
| Abstract |
|---|
|
|
|---|
Methods According to the coincidence method, high-intensity transients should only be classified as microembolic signals if they appear sequentially in the two sample volumes monitored and within a defined time window calculated from the blood velocity and the spatial distance between the insonation depths. Part A: microbubbles were introduced in a continuous-flow bench model of the middle cerebral artery to evaluate the accuracy of the multigate probe in embolus detection. Part B: in the subjects and patients, the minimal and maximum time delays in the appearance of microembolic signals in the two middle cerebral artery sample volumes were calculated as 0.01 second and set at 0.1 second, respectively. The multigate probe was used to monitor (1) 5 normal volunteers in whom 1008 artifact signals were produced, (2) 2 patients undergoing aortic valve replacement surgery, and (3) 12 patients with potential cardiac or carotid embolic sources.
Results In the bench model, 95.5% of microembolic signals produced by microbubbles appeared in the two sample volumes with a time delay between 0.02 and 0.05 second, while in the remaining 4.5% a shorter passage time of 0.01 second was measured. A total of 1968 high-intensity signals were recorded in subjects and patients. All but 20 of these (99%) appeared in both monitoring channels within the above time frame. To summarize, 996 (98.8%) of the 1004 artifact signals and 943 (98.1%) of the 961 microembolic signals were correctly classified.
Conclusions Application of the coincidence theory to distinguish microembolic signals from artifacts provides a promising new technique with high sensitivity and specificity that could decisively improve the validity of embolus detection.
Key Words: diagnostic imaging Doppler embolism
| Introduction |
|---|
|
|
|---|
The purpose of this study was to evaluate the feasibility of the differentiation between microembolic and artifact signals with the use of the coincidence method, originally described by Bothe and Kohlhörster.7 They used this approach to investigate cosmic radiation, ie, to distinguish real energetic particles from random artifacts. The occurrence of two or more events caused by the same physical quality (for example, an excited physical state, a beam of laser light, or a current fluid) strongly correlates with time. The occurrence of "true" events should therefore take place within a predefined time window.
To apply this approach in embolus detection, simultaneous monitoring of spatially separate segments of a vessel is required. "Multigate" monitoring of blood flow velocity within two or more spatially separated segments of the same vessel and its simultaneous analysis through separate Doppler channels was originally described as a method to assess blood velocity profiles.8 The use of a multigate device for embolus detection was first proposed by Rune Aaslid during the 8th International Symposium on Cerebral Hemodynamics in Münster, Germany, in September 1994. Passage of emboli through the artery under study would theoretically result in the detection of high-intensity signals in all channels in a certain sequence and with a certain time delay between them defined by the direction and flow velocity of the blood column and the distance between the separate sample volumes. In contrast, artifacts should appear in all channels simultaneously or in only one channel.
| Subjects and Methods |
|---|
|
|
|---|
t, which depends on its velocity and the distance between the
two points.
t is also the minimal time required for a transported
particle to cover the same distance. The maximum time required (
t')
depends on a number of parameters, including the flow
profile of the fluid and the surface characteristics of the particle.
Assuming that a particle is registered at point A at the time
tA and another particle is registered at point B at the
time tB, we regard these events as
"coincidental" if the condition
![]() |
is satisfied. In this case it is clear that the same particle
was registered consecutively at both points. The restriction of the
elapsing time after time tA (
t') is essential to
distinguish the particle under observation from random or subsequent
particles.
The same principles apply for monitoring of the middle cerebral artery
(MCA) for the detection of MES with TCD if two spatially separate
segments A and B of the artery are insonated simultaneously
and the data recorded on two separate Doppler channels (Fig 1
). The same current blood passes the
sample volume A at a time tA and the sample volume B at a
time tB. The distance between the sample volumes and the
blood flow velocity determine the length of
t.
Simultaneous recording of echoes on both channels
allows the identification of artifact signals, since they are
recorded at either time tA or tB on both
channels, or on one channel only. Consequently, coincidental signals
can be interpreted as being caused by true particles transported by the
streaming blood.
|
The passage time
t was calculated from both the peak
systolic blood flow velocity
(Vs1-2) in the arterial
segment under study and the spatial distance
(s1-2) between the insonation
depths 1 and 2 of the sample volumes according to the formula
![]() |
where
Vs1-2=Vs1+Vs2/2
and Vs1 and Vs2 are the peak systolic
flow velocities at insonation depths 1 and 2, respectively.
t also
reflects the minimal time required for an embolus to travel from depth
1 to depth 2. A longer time interval than
t may be necessary for
this passage, since the route of the embolus through the artery is not
predictable, eg, it is unclear whether it travels the shortest way in
the middle of the blood column or occasionally bumps on the wall of the
vessel, thus traveling a longer distance. In our paradigm, a strict
limitation of the appropriate length of the time window (t+
t')
within which the emboli should have passed the arterial
segment of interest is less important, since microemboli very rarely
appear only milliseconds after each other. For this reason, the
parameter t+
t' was set at 0.1 second.
A prototype multigate probe was used (Eden Medical Electronics). This was a 2-MHz pulsed-wave probe consisting of one transmitting and two receiving channels. The burst duration of this probe was 13 microseconds (when the sample volume was set at a length of 10 mm), and the pulse repetition frequency was 6.25 kHz. Measurement of the blood flow velocities at both insonation depths was performed by splitting the reflected echoes and feeding them into separate receiving channels. The probe was used with a standard TCD device (Pioneer TC 4040, Eden Medical Electronics). Signal analysis was based on a 128-point fast Fourier transform. Data from both channels were recorded on digital audiotapes using a two-channel digital audiotape recorder (Sony 59ES), whereby only one of the quadrature audio signals was recorded per channel. The time of the appearance of MES and artifacts was documented based on the time display of the digital audiotape equipment. An experienced observer was present during all monitoring sessions.
In Vitro Studies
A bench model of the MCA was devised. This consisted of a
continuous-flow diaphragm pump (PAR-MAX 3, ITT Jabsco), plastic
tubing 3.5 mm in diameter, and a solution of cellulose with an average
particle size of 20 µm (Sigmacell type 20, Sigma Chemical Co)
mimicking human blood. The plastic tubing was inserted in a specially
constructed chamber filled with ultrasonic gel. Rigid tubing was used,
since the flexibility of the tube does not influence flow dynamics in a
constant flow situation. The probe was mounted in the chamber in a
position providing a 10° insonation angle of the simulated MCA. Since
the power of the ultrasonic signal was not of interest in this setting,
no shell imitating the temporal bone was intersected between the probe
and the ultrasonic gel. Microbubbles were produced by agitating water
in a syringe. Subsequently, the solution containing microbubbles was
injected into the system through an 80-µm filter via a syringe
through a side port. Before all experiments, the model was degassed for
30 minutes with the use of a vacuum pump (R Halm Elektromotorenfabrik).
Fifteen minutes of tentative TCD monitoring was performed after
degassing to ensure the absence of microbubbles in the circuit. All TCD
settings were kept constant throughout the study (sample volume, 8 mm;
gain, 14).
Two hundred separate microbubbles were introduced into the circulation of the model. MES were recorded from paired sampling depths of 50 and 68 mm and 50 and 62 mm, respectively, with a constant flow velocity of 60 cm/s. The calculated passage time for emboli was 0.03 for the first and 0.02 second for the second combination.
Patient Studies
The MCA was identified according to generally accepted
criteria9 and the probe subsequently fixed on the temporal
skull with an elasticized band. Monitoring was performed at two
different insonation depths with a constant distance of 10 mm between
them, usually at 56 and 46 mm. The whole length of the main stem of the
MCA was investigated in 2-mm steps to exclude the presence of
significant velocity changes between the two sample volumes, eg, due to
MCA stenosis or hyperperfusion. After obtaining a satisfactory
signal, the sample volume was gradually reduced in both channels as far
as possible provided that a sufficient background MCA signal could be
maintained. The final sample volumes used varied between 6 and 10 mm in
length.
Five healthy volunteers, 2 patients undergoing cardiac surgery for aortic valve replacement, and a total of 12 patients with potential cardiac or carotid embolic sources were used to investigate the feasibility of the new TCD technique. Each control subject underwent a 30-minute monitoring procedure. A total of 1008 artifact signals were evaluated; they occurred either naturally by proband movements, slight probe dislocation, accidental hitting of the probe by the proband, chewing, or speaking, or they were actively produced by the observer by lightly touching the probe or its fixation.
Two patients undergoing cardiac surgery for aortic valve replacement (Tecna, Baxter) were continuously monitored during the surgical procedure. A total of 862 MES were detected according to conventional criteria, ie, (1) short duration (<0.1 second for signals appearing in systole, <0.3 second for signals appearing in diastole), (2) characteristic sound, (3) random appearance in the cardiac cycle, and (4) a minimum intensity of 4 dB above the background MCA signal.2
Additionally, 12 patients expected to reveal MES in their MCAs
(mean±SE age, 51±4 years; 9 men, 3 women) were recruited from the
departments of cardiothoracic surgery (n=7) and neurology (n=5). Five
of them had prosthetic cardiac valves and two a left
ventricular assist device (Novacor, Baxter). Neurological
patients had high-grade (>70%) symptomatic carotid
artery stenosis diagnosed with extracranial duplex (n=5) and
additional cerebral angiography (n=2). Clinical details and monitoring
data of these patients are listed in the Table
. The
overall monitoring time was 8 hours and 15 minutes. A total of 99 MES
were identified using the four conventional criteria listed above.
|
Off-line evaluation of both artifacts and MES consisted of
measurement of the exact time of occurrence of each signal within each
channel. This was done with an accuracy of 0.01 second. The maximum
blood velocity (vb) in centimeters per second was also
calculated. Both parameters were calculated using standard
software incorporated in the Doppler device. Subsequently, the time
between the appearance of the signals in the two channels was
evaluated. The smallest distinguishable time difference was 0.01
second. For the purposes of this study, the channel with the greater
insonation depth was always coded as "channel 1" as opposed to
the channel with the more shallow insonation depth, which was coded
"channel 2." Every true MCA embolus traveling orthogradely first
had to pass channel 1 and then channel 2. The distance between the two
sampling depths was 1 cm, and peak systolic flow velocities
ranged from 47 to 85 cm/s. The longest time delay was calculated
according to
t=1 cm/47
cm · s-1=0.021 s, and the shortest
one according to
t=1 cm/85
cm · s-1=0.012 s.
t was therefore
set at 0.01 second and the time window for the identification of MES
between 0.01 and 0.1 second.
| Results |
|---|
|
|
|---|
Patient Studies
Four (0.4%) of the 1008 artifact signals recorded in the
normal volunteers appeared in only one channel (one in channel 1 and
three in channel 2). Of the remaining 1004 signals, 992 appeared
simultaneously in both channels and 12 with a time delay of
0.01 second. Thus, the specificity of this technique was quite high
(98.7%).
Of the 862 intraoperatively detected MES, 15 (1.7%) were viewed in
only one channel (13 appeared only in channel 1 and two only in channel
2). This could be explained by escape of microemboli via MCA branches
originating from the MCA stem between the two sample volumes, a longer
time delay due to "sticking" of the emboli on the
arterial wall, or a failure to detect them in the second
channel as a result of insufficient time overlap of the Doppler
device. Of the remaining 847 signals, 845 (99.7%) fulfilled the
characteristic criteria of emboli according to the coincidence method,
ie, with respect to the time delay of their appearance in the two
channels (Table
).
Similar results were obtained in the 12 patients with potential embolic sources: of 99 MES, 1 was viewed in only one channel (1%), while the remaining 98 MES (99%) were correctly characterized as such by the multigate approach.
To summarize the results of the three groups, the described method
correctly identified 996 of 1008 artifact signals and 943 (845+98) of
961 MES (Fig 2
). Thus, the sensitivity of the new device
(true positive/all positive) was 98.1% and the specificity (true
negative/all negative) 98.8%.
|
| Discussion |
|---|
|
|
|---|
The failure rate of 3.1% is possibly due in part to measurement error caused by imprecise placement of the cursors. This could be avoided with the application of a more sophisticated software and a better time resolution of the Doppler device. Presumably, a small failure rate due to signals viewed in only one channel cannot be avoided, but this was minimal in our setting and accounted for only 0.4% of the artifact and 1.4% of the MES.
The main limitation of this technique is that its applicability to individual patients has not yet been evaluated. Sampling of the MCA at two different depths spatially separated by a distance of approximately 10 mm was possible in all patients in our study. Larger studies with the multigate probe are needed to determine its applicability in other settings and to develop guidelines for further refinement. The current time resolution of the Doppler device allows application of this new technique only in patients with an MCA velocity lower than 100 cm/s. While this limitation is fulfilled by most individuals,9 it prohibits the use of this method in patients with MCA stenosis or cerebral hyperperfusion. Again, improvement of the temporal resolution of the Doppler equipment is required to extend its applicability to practically every clinical setting.
The distribution of the time delay in our model was sharper than that during the in vivo studies. This could be due to the homogeneity of size and composition of the embolic material in the in vitro study. Additionally, the interactions between microbubbles and in vitro circulating fluid are thought to be quantitatively less pronounced than those between blood and formed particles.
Most MES recorded in vivo in our study (585 of 961, 60.9%) appeared with a time delay of 0.02 to 0.03 second, which suggests that the bulk of microemboli traveled with a speed corresponding to the systolic blood velocity. The speed of embolic particles is mainly influenced by their size and their surface characteristics. We hypothesize that the distribution of different flowing velocities within the laminar layers of the blood column and the heterogeneous size of the particles considerably contribute to the broad scatter of passage times between the two sample volumes.
In conclusion, our results suggest that the application of the coincidence theory to distinguish MES from artifact signals provides a promising new technique that could decisively improve the validity of embolus detection.
| Acknowledgments |
|---|
Received September 25, 1995; revision received December 14, 1995; accepted December 15, 1995.
| References |
|---|
|
|
|---|
2. Georgiadis D, Grosset DG, Kelman AW, Faichney A, Lees KR. Incidence and characteristics of intracranial microemboli signals in patients with different types of prosthetic cardiac valves. Stroke. 1994;25:587-592. [Abstract]
3. Brucher R, Russell D. Spectral characteristics of emboli with artefact suppression. Stroke. 1993;24:510. Abstract.
4.
Markus HS, Loh A, Brown MM. Computerized
detection of cerebral emboli and discrimination from artifact using
Doppler ultrasound. Stroke. 1993;24:1667-1672.
5.
Siebler M, Rose G, Sitzer M, Bender A, Steinmetz
H. Real-time identification of cerebral microemboli with
ultrasonic feature detection by a neuronal network.
Radiology. 1994;192:739-742.
6. Smith JL, Evans DH, Fan L, Thrush AJ, Naylor AR. Processing of Doppler ultrasound signals from blood borne emboli. Ultrasound Med Biol. 1994;20:455-462.
7. Bothe W, Kohlhörster W. Das Wesen der Höhenstrahlung. Zeitschrift für Physik. 1929;29:751-777.
8. Hocks APG, Reneman RS, Peronnau PA. A multi-gate pulsed Doppler system with serial processing. IEEE Trans Sonics Ultrason SU. 1981;28:242-247.
9. Ringelstein EB, Otis SM, Spaar-Kahlscheuer B, Niggermeier E. Transcranial Doppler sonography: anatomical landmarks and normal velocity values. Ultrasound Med Biol. 1990;8:745-761.
This article has been cited by other articles:
![]() |
D. G. Nabavi, J. Stockmann, C. Schmid, M. Schneider, D. Hammel, H. H. Scheld, and E. B. Ringelstein Doppler microembolic load predicts risk of thromboembolic complications in Novacor patients J. Thorac. Cardiovasc. Surg., July 1, 2003; 126(1): 160 - 167. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. L. Taylor, M. A. Borger, R. D. Weisel, L. Fedorko, and C. M. Feindel Cerebral microemboli during cardiopulmonary bypass: increased emboli during perfusionist interventions Ann. Thorac. Surg., July 1, 1999; 68(1): 89 - 93. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Lagos, F. M. Cabezas, and G. Deklunder Microemboli in Cerebral Circulation and Alteration of Cognitive Abilities in Patients With Mechanical Prosthetic Heart Valves • Response Stroke, May 1, 1999; 30(5): 1150 - 1150. [Full Text] [PDF] |
||||
![]() |
V. Kemeny, D. W. Droste, S. Hermes, D. G. Nabavi, G. Schulte-Altedorneburg, M. Siebler, and E. B. Ringelstein Automatic Embolus Detection by a Neural Network Stroke, April 1, 1999; 30(4): 807 - 810. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Furui, K. Hanzawa, H. Ohzeki, T. Nakajima, N. Fukuhara, M. Takamori, J.P. Mohr, and C. Stapf "Tail Sign" Associated With Microembolic Signals • Editorial Comment Stroke, April 1, 1999; 30(4): 863 - 866. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Ries, K. Tiemann, C. Pohl, C. Bauer, M. Mundo, and H. Becher High-Resolution Emboli Detection and Differentiation by Characteristic Postembolic Spectral Patterns Stroke, March 1, 1998; 29(3): 668 - 672. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Bernd Ringelstein, D. W. Droste, V. L. Babikian, D. H. Evans, D. G. Grosset, M. Kaps, H. S. Markus, D. Russell, and M. Siebler Consensus on Microembolus Detection by TCD Stroke, March 1, 1998; 29(3): 725 - 729. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Georgiadis, A. Wenzel, H. R. Zerkowski, S. Zierz, and A. Lindner Automated Intraoperative Detection of Doppler Microembolic Signals Using the Bigate Approach Stroke, January 1, 1998; 29(1): 137 - 139. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Sliwka, C. Klotzsch, O. Popescu, K. Brandt, P. Schmidt, P. Berlit, and J. Noth Do Chronic Middle Cerebral Artery Stenoses Represent an Embolic Focus? : A Multirange Transcranial Doppler Study Stroke, July 1, 1997; 28(7): 1324 - 1327. [Abstract] [Full Text] |
||||
![]() |
H. S. Markus and J. Molloy Use of a Decibel Threshold in Detecting Doppler Embolic Signals Stroke, April 1, 1997; 28(4): 692 - 695. [Abstract] [Full Text] |
||||
![]() |
D. W. Droste, G. Hagedorn, A. Notzold, H.-J. Siemens, H. H. Sievers, and M. Kaps Bigated Transcranial Doppler for the Detection of Clinically Silent Circulating Emboli in Normal Persons and Patients With Prosthetic Cardiac Valves Stroke, March 1, 1997; 28(3): 588 - 592. [Abstract] [Full Text] |
||||
![]() |
U. Sliwka, A. Lingnau, W.-D. Stohlmann, P. Schmidt, M. Mull, R. R. Diehl, and J. Noth Prevalence and Time Course of Microembolic Signals in Patients With Acute Stroke: A Prospective Study Stroke, February 1, 1997; 28(2): 358 - 363. [Abstract] [Full Text] |
||||
![]() |
J. Molloy and H. S. Markus Multigated Doppler Ultrasound in the Detection of Emboli in a Flow Model and Embolic Signals in Patients Stroke, September 1, 1996; 27(9): 1548 - 1552. [Abstract] [Full Text] |
||||
![]() |
D. G. Nabavi, D. Georgiadis, T. Mumme, P. Zunker, and E. B. Ringelstein Detection of Microembolic Signals in Patients With Middle Cerebral Artery Stenosis by Means of a Bigate Probe: A Pilot Study Stroke, August 1, 1996; 27(8): 1347 - 1349. [Abstract] [Full Text] |
||||
![]() |
D.G. Nabavi, D. Georgiadis, T. Mumme, C. Schmid, T.G. Mackay, H.H. Scheld, and E.B. Ringelstein Clinical Relevance of Intracranial Microembolic Signals in Patients With Left Ventricular Assist Devices : A Prospective Study Stroke, May 1, 1996; 27(5): 891 - 896. [Abstract] [Full Text] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Stroke Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1996 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |