(Stroke. 1997;28:2453-2456.)
© 1997 American Heart Association, Inc.
Articles |
From the Departments of Neurology (D.W.D., T.H., V.K., G.S.-A., D.G.N., E.B.R.) and Cardiothoracic and Vascular Surgery (D.H., H.H.S.), University of Münster, and Department of Neurology, Medical University of Lübeck (M.K.) (Germany).
Correspondence to Dr Dirk W. Droste, Klinik und Poliklinik für Neurologie der WWU Münster, Albert-Schweitzer-Str 33, D-48129 Münster, Germany.
| Abstract |
|---|
|
|
|---|
Methods We performed 1-hour transcranial Doppler sonographic recordings from the middle cerebral or posterior cerebral artery in 20 patients with mechanical prosthetic heart valves, in 78 patients with an arterial embolic source, and in 20 control subjects. During 30 minutes of this recording, the patients inspired room air and 6 L of oxygen per minute via a loosely fitting facial mask; during the remaining 30 minutes, they breathed room air only.
Results There was a significant decline of embolic signals (ES) under oxygen in the patients with mechanical prosthetic cardiac valves (144 ES without oxygen versus 63 ES with oxygen; P=.002) but not in the patients with arterial embolic sources (145 ES without oxygen versus 135 ES with oxygen; P=NS). In the control subjects, no ES were found.
Conclusions ES in patients with mechanical prosthetic cardiac valves correspond mainly to gas bubbles. Oxygen inhibits the cavitation process of mechanical prosthetic heart valves or speeds up redissolution of gas bubbles generated by cavitation. In contrast, solid microemboli originating from thrombus or atheroma cannot be suppressed by oxygen inhalation. This simple method of oxygen inhalation should help to clarify the composition of microemboli in various clinical and experimental settings.
Key Words: cerebral embolism cerebrovascular disorders heart valve prosthesis oxygen ultrasonics
| Introduction |
|---|
|
|
|---|
In a previous communication, we described a prompt decrease of the number of ES under the inhalation of oxygen in six patients with mechanical prosthetic cardiac valves, and we suggested that microemboli in these patients are predominantly gaseous.10 In this report we want to validate these findings in an additional larger cohort and to demonstrate that microemboli from arterial sources are predominantly nongaseous.
| Subjects and Methods |
|---|
|
|
|---|
The MCA or the PCA (P1or P2 segment, PCA) was continuously insonated through the temporal window for 1 hour. A 2-MHz probe was mounted on the temporal plane and secured in a head ribbon. A small sample volume of 4 mm in length and a low gain provided a setting guaranteeing optimal ES discrimination from the background spectrum.4 This setting was maintained throughout the recordings. The investigations were well tolerated by the subjects without side effects. The sametranscranial pulsed Doppler ultrasound device (TC4040, Nicolet-EME) was used for all studies and for the blinded off-line evaluation of the data. The machine employed a 128-point FFT analysis and used a graded color scale to display the intensity of the received Doppler signal. FFT time frame overlap was 67%. The Doppler signal was recorded onto DAT with normal speed for documentation and off-line control. The tapes were given numbers. No patient details could be deduced from these tapes. An experienced observer's analysis of ES comprised (1) listening to each signal and (2) watching each signal on the screen at highest speed. This analysis was done blinded off-line by reintroducing the DAT signal into the FFT processor. The following definition for ES was used: typical visible and audible (click, chirp, whistle) short-duration, high-intensity signal within the Doppler flow spectrum with its occurrence at random in the cardiac cycle, and an intensity increase of 3 dB or more above the background signal.1 2 11 Interobserver and intercenter validations were performed to guarantee reproducibility and stability of assessment. In a randomized way, oxygen was applied at a dose of 6 L/min by a loosely fitting facial mask during either the first 30 minutes or the last 30 minutes of the recording time. We also assessed off-line from the DAT tapes in a period of 20 seconds sampled at random between the 10th and the 20th and between the 40th and 50th minute of the recording the mean peak systolic and end-diastolic velocity from each heart beat as well as the heart rate from each subject over 20 seconds. We then calculated the mean maximal peak systolic and the mean maximal end-diastolic blood flow velocity from all the heart beats in these 20 seconds. The heart rate per 20 seconds was tripled to obtain the heart rate per minute.
Patients With Mechanical Prosthetic Cardiac Valves
Three women and 17 men aged 25 to 81 years (mean age, 56 years)
were investigated. All of them had received aortic valve prostheses,
and 7 of them had an additional mitral valve prosthesis
(Edwards Tekna, models 3200 A and 9200 mol/L; 1 patient had a St
Jude medical aortic valve). The time between the operation and the
ultrasound investigation varied from 37 to 1358 days (mean, 702 days).
In 13 patients the right MCA was insonated; in the remaining 7 the left
MCA was the artery of interest. Insonation depths varied from 44 to
56 mm. Seven patients were in atrial fibrillation during the
embolus detection. Two patients had minor plaques, and 1 patient had a
60% stenosis of the extracranial ICA proximal to the artery
under investigation. Two patients had cerebral or retinal
ischemic events 881 and 902 days, respectively, before the
recording. Three patients were smokers, 2 were diabetic, 5 had
arterial hypertension, and 5 had
hyperlipidemia. All patients were on oral
anticoagulation.
Patients With Arterial Embolic Sources
Twenty-one women and 57 men, aged 39 to 84 years (mean age, 57
years) were investigated in a vessel downstream to a proven
arterial embolic source. None of these patients had an
artificial cardiac valve. In 7 patients we recorded from the right
PCA, in 3 patients from the left PCA, in 38 patients from the right
MCA, and in 30 patients from the left MCA. Insonation depths varied
from 44 to 56 mm for the MCA insonation (in 1 patient 58 mm)
and from 60 to 66 mm for investigating the PCA. In the MCA
recorded group, patients 1 to 61 had ipsilateral extracranial
carotid artery disease (47 ICA or CCA stenoses
60%, 6 severe
plaques, 5 ICA occlusions, 3 ICA dissections). Patients 62 to 68 had a
corresponding MCA or carotid siphon stenosis, and 1 had a
proximal MCA occlusion. In the PCA group, patients 69 to 71 had
high-grade basilar artery stenosis, patients 72 and 73 had a
corresponding high-grade PCA stenosis, patient 74 had a
proximal subclavian artery stenosis, patient 75 had an
intracranial vertebral artery occlusion on one side and an extracranial
vertebral artery stenosis on the other side, patients 76 and 77
had vertebral artery dissection, and patient 78 had an extracranial
vertebral artery stenosis. Two patients were in atrial
fibrillation during the recording. Of all 78 patients, 35 had
experienced cerebral or retinal ischemic events in the
corresponding arterial territory in the past. Twenty-one
patients were smokers, 13 were diabetic, 44 had arterial
hypertension, and 38 suffered from hyperlipidemia.
Thirteen patients were on oral anticoagulation, 45 on antiaggregants,
and 24 on intravenous heparin.
Control Subjects
Twenty subjects without any ischemic retinal or
cerebrovascular events and a normal ultrasound examination served as a
control group (9 men and 11 women; age range, 19 to 77 years; mean age,
26 years). Four subjects were smokers, 1 was diabetic, none were
hypertensive, and 1 suffered from hyperlipidemia. No
subject was in atrial fibrillation during the recording. In 11
subjects we recorded from the left MCA and in 9 subjects from the
right MCA. Insonation depths varied from 46 to 54 mm.
Statistical Analysis
For the statistical analysis, the numbers of ES as well
as the systolic and end-diastolic blood flow
velocity and the heart rate with or without oxygen inhalation were
compared with the nonparametric Wilcoxon test.
Three groups were considered separately: (1) 20 patients with
mechanical prosthetic cardiac valves, (2) 78 patients with
arterial embolic sources, and (3) the control group (n=20).
We also compared the heart rate and the maximal systolic and
end-diastolic blood flow velocity using the
Wilcoxon test for the each group. Statistical significance was
declared at the .05 level.
| Results |
|---|
|
|
|---|
|
Sixteen patients with ES had more ES without oxygen inhalation than with oxygen inhalation, and only 1 patient had more ES with oxygen inhalation (P=.002). Mean±SD peak systolic and end-diastolic blood flow velocity and heart rate were 72.8±9.2 cm/s, 31.4±6.7 cm/s, and 72.3±11.6/min, respectively, without oxygen inhalation and 71.6±6.8 cm/s, 29.9±5.8 cm/s, and 72.3±11.4/min, respectively, with oxygen inhalation. None of these differences were significant.
Patients With Arterial Embolic Sources
ES were detected in 22 of 78 patients varying from
1 to 71/h in the individual patients. There were a total of 145 ES
without oxygen (mean±SD, 1.9±5.3; range, 1 to 33/30 min in the
individual patients) and 135 ES with oxygen (mean±SD, 1.7±5.3; range,
1 to 38/30 min in the individual patients). Twelve patients had more ES
without oxygen, and 9 had more ES with oxygen. One patient had the same
amount of ES under the two conditions. When the Wilcoxon test
was performed, the comparison showed no significant difference
(P=.34) between the number of ES under baseline conditions
and under oxygen inhalation. Mean±SD peak systolic and
end-diastolic blood flow velocity and heart rate were
65.6±20.9 cm/s, 30.0±10.9 cm/s, and 70.9±13.5/min, respectively,
without oxygen inhalation and 65.9±21.6 cm/s, 29.7±10.8 cm/s, and
68.4±12.4/min, respectively, with oxygen inhalation. None of these
differences were significant.
Control Subjects
None of these subjects showed ES. Mean±SD peak systolic
and end-diastolic blood flow velocity and heart rate were
90.0±20.3 cm/s, 44.1±12.7 cm/s, and 72.7±10.7/min, respectively,
without oxygen inhalation and 87.5±20.9 cm/s, 41.9±13.3 cm/s, and
70.0±10.6/min, respectively, with oxygen inhalation. None of these
differences were significant.
| Discussion |
|---|
|
|
|---|
Moreover, our data strongly support the view that clinically silent circulating microemboli in patients with arterial thromboembolic sources are predominantly nongaseous. Whether these signals correspond to reversible platelet aggregates, thrombus, or plaque debris remains unclear.
Differences in blood flow velocity or heart rate due to oxygen inhalation did not exist and therefore cannot account for the above observations.19
The simple application of oxygen described in this report may be used not only in patients with different cardiac valves but also in other thromboembolic sources such as cardiac assist devices. Oxygen may also be helpful in tube models to better define the nature and potential concomitant risk of microemboli.20
| Selected Abbreviations and Acronyms |
|---|
|
| Acknowledgments |
|---|
Received March 20, 1007; revision received July 11, 1997; accepted August 28, 1997.
| References |
|---|
|
|
|---|
2. Spencer MP. Detection of cerebral arterial emboli. In: Newell DW, Aaslid R, eds. Transcranial Doppler. New York, NY: Raven Press Ltd; 1992:216230.
3.
Markus HS, Brown MM. Differentiation between different
pathological cerebral embolic materials using transcranial
Doppler in an in vitro model. Stroke. 1993;24:15.
4. Droste DW, Markus HS, Brown MM. The effect of different settings of ultrasound pulse amplitude, gain and sample volume on the appearance of emboli studied in a transcranial Doppler model. Cerebrovasc Dis. 1994;4:152156.
5. Droste DW, Markus HS, Nassiri D, Brown MM. The effect of velocity on the appearance of embolic signals studied in transcranial Doppler models. Stroke. 1994;25:986991.[Abstract]
6. Georgiadis D, Mackay TG, Kelman AW, Grosset DG, Wheatley DJ, Lees KR. Differentiation between gaseous and formed embolic materials in vivo: application in prosthetic heart valve patients. Stroke. 1994;25:15591563.[Abstract]
7.
Sturzenegger M, Beer JH, Rihs F. Monitoring combined
antithrombotic treatments in patients with prosthetic heart
valves using transcranial Doppler and coagulation
markers. Stroke. 1995;26:6369.
8.
Braekken SK, Russell D, Brucher R, Svennevig J.
Incidence and frequency of cerebral embolic signals in patients with a
similar bileaflet mechanical heart valve. Stroke. 1995;26:12251230.
9. Tong DC, Bolger A, Albers GW. Incidence of transcranial Dopplerdetected cerebral microemboli in patients referred for echocardiography. Stroke. 1994;25:21382141.[Abstract]
10.
Kaps M, Hansen J, Weiher M, Tiffert K, Kayser I, Droste
DW. Clinically silent microemboli in patients with artificial
prosthetic aortic valves are predominantly gaseous and not
solid. Stroke. 1997;28:322325.
11.
Consensus Committee of the 9th International Cerebral
Hemodynamic Symposium. Basic identification criteria of
Doppler microembolic signals. Stroke. 1995;26:1123.
12. Lee CS, Chandran KB, Chen LD. Cavitation dynamics of mechanical heart valve prostheses. Artif Organs. 1994;18:758767.[Medline] [Order article via Infotrieve]
13. Richard G, Beavan A, Strzepa P. Cavitation threshold ranking and erosion characteristics of bileaflet heart valve prostheses. J Heart Valve Dis. 1994;3(suppl 1):S94S101.
14. Shu MC, Leuer LH, Armitage TL, Schneider TE, Christiansen DR. In vitro observations of mechanical heart valve cavitation. J Heart Valve Dis. 1994;3(suppl 1):S85S92.
15. Graf T, Reul H, Detlefs C, Wilmes R, Rau G. Causes and formation of cavitation in mechanical heart valves. J Heart Valve Dis. 1994;3(suppl 1):S49S64.
16. Herman BA, Carey RF. A protocol for the evaluation of the cavi-tation potential of mechanical heart valves. J Heart Valve Dis. 1994;3(suppl 1):S128S130.
17. Burchfiel CM, Hammermeister KE, Krause-Steinrauf H, Sethi GK, Henderson WG, Crawford MH, Wong M. Left atrial dimension and risk of systemic embolism in patients with a prosthetic heart valve. J Am Coll Cardiol. 1990;15:3241.[Abstract]
18. Kuntze CE, Ebels T, Eijgelaar A, Van der Heide JNH. Rates of thromboembolism with three different mechanical heart valve prostheses: randomised study. Lancet. 1989;3:514517.
19. Bew SA, Field LM, Droste DW, Razis P. The effect of high concentrations of inspired oxygen on middle cerebral artery blood velocity measured by transcranial Doppler. Exp Physiol. 1994;79:593596.[Abstract]
20.
Nabavi DG, Georgiadis D, Mumme T, Schmid C, Mackay TG,
Scheld HH, Ringelstein EB. Clinical relevance of intracranial
microembolic signals in patients with left
ventricular assist devices: a prospective study.
Stroke. 1996;27:891896.
This article has been cited by other articles:
![]() |
A. King and H. S. Markus Doppler Embolic Signals in Cerebrovascular Disease and Prediction of Stroke Risk: A Systematic Review and Meta-Analysis * Supplemental Online References Stroke, December 1, 2009; 40(12): 3711 - 3717. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Guerrieri Wolf, B. P. Choudhary, Y. Abu-Omar, and D. P. Taggart Solid and gaseous cerebral microembolization after biologic and mechanical aortic valve replacement: Investigation with multirange and multifrequency transcranial Doppler ultrasound J. Thorac. Cardiovasc. Surg., March 1, 2008; 135(3): 512 - 520. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Dittrich and E. B. Ringelstein Occurrence and Clinical Impact of Microembolic Signals During or After Cardiosurgical Procedures Stroke, February 1, 2008; 39(2): 503 - 511. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. H. Thoennissen, M. Schneider, A. Allroggen, M. Ritter, R. Dittrich, C. Schmid, H. H. Scheld, E. B. Ringelstein, and D. G. Nabavi High level of cerebral microembolization in patients supported with the DeBakey left ventricular assist device J. Thorac. Cardiovasc. Surg., October 1, 2005; 130(4): 1159 - 1166. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Perthel, S. Kseibi, A. Bendisch, and J. Laas The dynamic bubble trap reduces microbubbles in extracorporeal circulation and high intensity transient signals in the middle cerebral artery: a case report Perfusion, September 1, 2003; 18(5): 325 - 329. [Abstract] [PDF] |
||||
![]() |
J. H. Rapp, X. M. Pan, B. Yu, R. A. Swanson, R. T. Higashida, P. Simpson, and D. Saloner Cerebral Ischemia and Infarction From Atheroemboli <100 {micro}m in Size Stroke, August 1, 2003; 34(8): 1976 - 1980. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. G. Nabavi, J. Stockmann, C. Schmid, M. Schneider, D. Hammel, H. H. Scheld, and E. B. Ringelstein Doppler microembolic load predicts risk of thromboembolic complications in Novacor patients J. Thorac. Cardiovasc. Surg., July 1, 2003; 126(1): 160 - 167. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Laas, S. Kseibi, M. Perthel, A. Klingbeil, L'E. El-Ayoubi, and A. Alken Impact of high intensity transient signals on the choice of mechanical aortic valve substitutes Eur. J. Cardiothorac. Surg., January 1, 2003; 23(1): 93 - 96. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Russell and R. Brucher Online Automatic Discrimination Between Solid and Gaseous Cerebral Microemboli With the First Multifrequency Transcranial Doppler Stroke, August 1, 2002; 33(8): 1975 - 1980. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. W. Droste, K. Kuhne, R. M. Schaefer, and E. B. Ringelstein Detection of microemboli in the subclavian vein of patients undergoing haemodialysis and haemodiafiltration using pulsed Doppler ultrasound Nephrol. Dial. Transplant., March 1, 2002; 17(3): 462 - 466. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. J. Feiwell, L. Besmertis, R. Sarkar, D. A. Saloner, and J. H. Rapp Detection of Clinically Silent Infarcts after Carotid Endarterectomy by Use of Diffusion-weighted Imaging AJNR Am. J. Neuroradiol., April 1, 2001; 22(4): 646 - 649. [Abstract] [Full Text] |
||||
![]() |
V. Kemeny, D. W. Droste, S. Hermes, D. G. Nabavi, G. Schulte-Altedorneburg, M. Siebler, and E. B. Ringelstein Automatic Embolus Detection by a Neural Network Stroke, April 1, 1999; 30(4): 807 - 810. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Stroke Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1997 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |