(Stroke. 1997;28:1307-1310.)
© 1997 American Heart Association, Inc.
Articles |
From the Department of Clinical Neuroscience (H.S.M.), King's College School of Medicine and Dentistry and the Institute of Psychiatry, London, UK; the Department of Neurophysiology (R.A.), St Antonius Ziekenhuis, Niuewegein, Netherlands; the Department of Neurology (V.B.), Boston (Mass) University School of Medicine; the Department of Neurology (C.B., C.L.), Austin and Repatriation Medical Centre, Melbourne, Australia; the Department of Neurology (D.D.), University of Munster (Germany); the Department of Neurology (D.G.), Southern General Hospital, Glasgow, UK; the Department of Neurology (D.R.), University of Oslo (Norway); the Department of Neurology (M.S.), University of Dusseldorf (Germany); and the Department of Neurology (C.T.), Bowman Gray School of Medicine, Winston-Salem, NC.
Correspondence to Dr Hugh Markus, Department of Clinical Neurosciences, Institute of Psychiatry, De Crespigny Park, London, SE5 8AF, UK. E-mail h.markus{at}iop.bpmf.ac.uk
| Abstract |
|---|
|
|
|---|
Methods Each center performed blinded analysis of a taped audio Doppler signal composed of transcranial Doppler middle cerebral artery recordings from 6 patients with symptomatic carotid artery stenosis. The exact time of any embolic signal was recorded. Six centers also measured the intensity increase of any embolic signals detected. Interobserver agreement was determined by a method based on the proportion of specific agreement.
Results Seven centers reported between 39 and 55 signals, but one center reported 142 embolic signals. The probability of agreement between observers was .678, which rose to .791 when the data from the highest reporting center were excluded. Introducing a decibel threshold resulted in a significant increase in the probability of agreement; a decibel threshold of >7 dB resulted in a probability of agreement of .902. Intensity measurements made by different centers were usually highly correlated, but this was not always the case, and 3 of the 15 correlations were not significant. The absolute values of the intensities measured varied between centers by as much as 40%.
Conclusions Although most centers report similar numbers of embolic signals, some use less specific criteria and report more events. The use of a decibel threshold improves reproducibility. However, intensity thresholds developed by one center cannot be directly transferred without validation to another center; differing methods of measurement are being used, and this results in different intensity values for the same embolic signals, even when the same equipment is used.
Key Words: carotid artery diseases cerebral embolism observer variation ultrasonics
| Introduction |
|---|
|
|
|---|
| Subjects and Methods |
|---|
|
|
|---|
No intensity threshold was specified, but centers were asked to record the relative intensity increase of any embolic signals. Three centers did not measure relative intensity increase. The remaining six centers used two general types of methods of measurement of relative intensity increase. First, the intensity increase of the embolic signal was calculated from the color-coded intensity scale on the screen. This method was used by centers 2, 5, and 9. The intensity of the embolic signal relative to a reference region in the spectra not including an embolic signal was determined. However, the reference regions used by each center were not necessarily identical. Second, the intensity of each embolic signal was calculated using automatic embolus software supplied with the machine. The algorithm determines the power of the embolic signal over the whole spectral line; similarly, the background power is calculated over the whole spectral line using a running average of background intensity over the preceding spectral lines. This method was used by centers 3, 6, and 8.
There were no moments when all observers recorded no abnormality, because the method of data collection did not allow for this. Therefore, we used a method of analysis that is independent of the number of observations, in which both observers would not detect an abnormality. This is an extension of the proportion of specific agreement.7 However, in this analysis we have taken into account that there are more than two observers.7 We estimated the probability that a second observer would record an abnormality if the first observer (or observers) recorded such an abnormality, as has previously been used to examine interrater agreement in embolic signal detection.7 This allows calculation of the probability that a specified observer will identify an embolic signal compared with the performance of one or more of the other observers. We determined the effect of introducing a signal intensity threshold (decibel threshold) on the agreement between observers by repeating the probability analysis only on signals above specific decibel thresholds. For this analysis, decibel thresholds determined by the coordinating center that had prepared the data were used, along with the first of the two measurement methods mentioned above. The peak intensity increase of the embolic signal was determined from the color-coded spectral display and compared with that of background spectra at the same frequency and the point of the preceding or subsequent cardiac cycle. During analysis, if the coordinating center had not identified an embolic signal at a point at which another center had detected one, that time point on the tape was reexamined. If an embolic signal could be then identified, its intensity was recorded. If no obvious embolic signal could be identified at that time, the maximum intensity increase in the spectral display at that time point was measured.
| Results |
|---|
|
|
|---|
|
There was a highly significant relationship between the number of
observers identifying any specific embolic signal and the intensity of
that embolic signal (Spearman's
=.837, P<.0001) (see
Figure
). Consistent with this, introducing a
decibel threshold resulted in a significant increase in the probability
of agreement (see Table 2
). Using a decibel threshold of
>6 dB resulted in a probability of agreement of .902, similar to that
reported in previous studies.6
|
|
Six centers were able to perform intensity measurements on embolic
signals. Correlation between measurements made by different centers is
shown in Table 3
. The correlation between most centers
was good, but this was not always the case, and 3 of the 15
correlations were not significant. In addition, the absolute values of
the intensities measured varied between centers, with some centers
recording significantly higher intensities. Analysis
was performed of the measured intensities of the 23 embolic signals
detected by all centers (Table 4
). As can be seen, there
was a 40% difference in mean measured intensities between highest and
lowest centers.
|
|
| Discussion |
|---|
|
|
|---|
Our results showed marked differences in the measured intensity of the
same embolic signals between different centers. Although there was a
good correlation between values in many centers, there was little or no
correlation in a minority of cases. Furthermore, even where there was
good correlation, the absolute values of intensity measurements were
quite different. This has important implications if an intensity
threshold is to be used. It implies that a threshold validated in one
center cannot be used by another center, even using the same equipment,
unless cross validation is performed. These differences are not
surprising, because intensity measurements are crucially dependent on
the method of measurement.9 Intensity is calculated from
the logarithm of the ratio of the power increase associated with the
embolic signal to that of the Doppler spectrum in the absence of
any embolic signal. The power increase associated with the embolic
signal may have been measured in a number of ways. These may include
the peak increase at one velocity, the area under the power increase
measured both across frequencies and across time, and the power
increase along the whole spectral line, taking into account the power
increase of the embolic signal and also of the background Doppler
spectrum at other velocities. Similarly, the background power may be
measured in a number of ways. The background power can be measured at
the same velocity or at all velocities, at the same point of the
cardiac cycle or averaging across the whole cardiac cycle, and only
within the Doppler spectrum or along the whole spectral line. For
example, the background intensity is higher in diastole
than in systole; therefore, the position of the cardiac cycle will
alter measurements. Similarly, if the whole spectral line is used, for
technically poor recordings with artifactual noise, the
background power will appear higher, resulting in a lower intensity
increase of the embolic signal. In a recent analysis, the use
of a 7-dB threshold developed using one standard method of measurement
resulted in 95% detection of embolic signals, whereas using another
method on the same data on the same equipment resulted in only
50%
detection.9 In clinical practice, differences may be even
greater as the intensity increase of an embolic signal will depend on
recording parameters such as sample volume.
Our results have a number of important implications. First, differences in reported frequencies of embolic signals in different studies may at least partly reflect different criteria for their identification. Second, the previously suggested consensus criteria for the identification of embolic signals7 are not sufficiently precise to result in reproducible identification of embolic signals in different centers. Third, the use of a decibel threshold will result in a marked improvement in interobserver agreement, and using a threshold set at the upper limit of random episodes of Doppler speckle in normal subjects will result in good probability of agreements on the order of .9. This seems a reasonable approach to improving reproducibility. However, our results demonstrate that different centers measure intensity in very different ways and that standardization of methods is required if a single-decibel threshold is to be used in consensus criteria. Until this occurs, each center needs to select an appropriate decibel threshold that is likely to be individual to their equipment, recording settings, and method of measurement. The use of validated automated detection systems may improve this area. Fourth, until validated automated detection systems are available, multicenter studies examining the predictive value of embolic signals should include blinded analysis of data by a single central reading center.
| Acknowledgments |
|---|
Received March 13, 1997; revision received April 24, 1997; accepted April 28, 1997.
| References |
|---|
|
|
|---|
2.
Markus HS, Thomson N, Brown MM.
Asymptomatic cerebral embolic signals in
symptomatic and asymptomatic carotid artery
disease. Brain. 1995;118:1005-1011.
3.
Sitzer M, Muller W, Siebler M, Hort W, Kniemeyer HW,
Jancke L, Steinmetz H. Plaque ulceration and lumen thrombus are
the main sources of cerebral microemboli in high-grade internal carotid
artery stenosis. Stroke. 1995;26:1231-1233.
4.
Valton L, Larrue V, Arrue P, Geraud G, Bes A.
Asymptomatic cerebral embolic signals in patients with
carotid stenosis: correlation with the appearance of plaque
ulceration on angiography. Stroke. 1995;26:813-815.
5.
Siebler M, Nachtmann A, Sitzer M, Rose G, Kleinschmidt
A, Rademacher J, Steinmetz H. Cerebral microembolism and the
risk of ischemia in asymptomatic high-grade
internal carotid artery stenosis. Stroke. 1995;26:2184-2186.
6.
Markus HS, Bland JM, Rose G, Sitzer M, Siebler
M. How good is intercenter agreement in the identification of
embolic signals in carotid artery disease? Stroke. 1996;27:1249-1252.
7.
Consensus Committee of the Ninth International
Cerebral Hemodynamic Symposium. Basic identification
criteria of Doppler microembolic signals.
Stroke. 1995;26:1123.
8.
Siebler M, Sitzer M, Rose G, Bendfeldt D, Steinmetz
H. Silent cerebral embolism caused by neurologically
symptomatic high-grade carotid stenosis: event
rates before and after carotid
endarterectomy. Brain. 1993;116:1005-1015.
9.
Markus HS, Molloy J. The use of a decibel
threshold in detecting Doppler embolic signals.
Stroke. 1997;28:692-697.
This article has been cited by other articles:
![]() |
H. Poppert, S. Sadikovic, K. Sander, O. Wolf, and D. Sander Embolic Signals in Unselected Stroke Patients: Prevalence and Diagnostic Benefit Stroke, August 1, 2006; 37(8): 2039 - 2043. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Dittrich, M. A. Ritter, M. Kaps, M. Siebler, K. Lees, V. Larrue, D. G. Nabavi, E. B. Ringelstein, H. S. Markus, and D. W. Droste The Use of Embolic Signal Detection in Multicenter Trials to Evaluate Antiplatelet Efficacy: Signal Analysis and Quality Control Mechanisms in the CARESS (Clopidogrel and Aspirin for Reduction of Emboli in Symptomatic carotid Stenosis) Trial Stroke, April 1, 2006; 37(4): 1065 - 1069. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. A Rodriguez, K. A Williams, A. Babaev, F. Rubens, and H. J Nathan Effect of perfusionist technique on cerebral embolization during cardiopulmonary bypass Perfusion, January 1, 2005; 20(1): 3 - 10. [Abstract] [PDF] |
||||
![]() |
M. A. Sloan, A. V. Alexandrov, C. H. Tegeler, M. P. Spencer, L. R. Caplan, E. Feldmann, L. R. Wechsler, D. W. Newell, C. R. Gomez, V. L. Babikian, et al. Assessment: Transcranial Doppler ultrasonography: Report of the Therapeutics and Technology Assessment Subcommittee of the American Academy of Neurology Neurology, May 11, 2004; 62(9): 1468 - 1481. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Orlandi, S. Fanucchi, F. Sartucci, L. Murri, C. D. Liapis, J. D. Kakisis, and A. G. Kostakis Can Microembolic Signals Identify Unstable Plaques Affecting Symptomatology in Carotid Stenosis? * Response Stroke, July 1, 2002; 33 (7): 1744 - 1746. [Full Text] [PDF] |
||||
![]() |
G. Devuyst, G.A. Darbellay, J.-M. Vesin, V. Kemeny, M. Ritter, D.W. Droste, C. Molina, J. Serena, R. Sztajzel, P. Ruchat, et al. Automatic Classification of HITS Into Artifacts or Solid or Gaseous Emboli by a Wavelet Representation Combined With Dual-Gate TCD Stroke, December 1, 2001; 32(12): 2803 - 2809. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Orlandi, S. Fanucchi, C. Fioretti, G. Acerbi, M. Puglioli, R. Padolecchia, F. Sartucci, and L. Murri Characteristics of Cerebral Microembolism During Carotid Stenting and Angioplasty Alone Arch Neurol, September 1, 2001; 58(9): 1410 - 1413. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Markus Monitoring Embolism in Real Time Circulation, August 22, 2000; 102(8): 826 - 828. [Full Text] [PDF] |
||||
![]() |
M. Cullinane, G. Reid, R. Dittrich, Z. Kaposzta, R. Ackerstaff, V. Babikian, D. W. Droste, D. Grossett, M. Siebler, L. Valton, et al. Evaluation of New Online Automated Embolic Signal Detection Algorithm, Including Comparison With Panel of International Experts Stroke, June 1, 2000; 31(6): 1335 - 1341. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Markus, M. Cullinane, and G. Reid Improved Automated Detection of Embolic Signals Using a Novel Frequency Filtering Approach Stroke, August 1, 1999; 30(8): 1610 - 1615. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. L. Taylor, M. A. Borger, R. D. Weisel, L. Fedorko, and C. M. Feindel Cerebral microemboli during cardiopulmonary bypass: increased emboli during perfusionist interventions Ann. Thorac. Surg., July 1, 1999; 68(1): 89 - 93. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Kemeny, D. W. Droste, S. Hermes, D. G. Nabavi, G. Schulte-Altedorneburg, M. Siebler, and E. B. Ringelstein Automatic Embolus Detection by a Neural Network Stroke, April 1, 1999; 30(4): 807 - 810. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Valton, V. Larrue, A. P. le Traon, P. Massabuau, and G. Geraud Microembolic Signals and Risk of Early Recurrence in Patients With Stroke or Transient Ischemic Attack Stroke, October 1, 1998; 29(10): 2125 - 2128. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Jander, M. Sitzer, R. Schumann, M. Schroeter, M. Siebler, H. Steinmetz, and G. Stoll Inflammation in High-Grade Carotid Stenosis : A Possible Role for Macrophages and T Cells in Plaque Destabilization Stroke, August 1, 1998; 29(8): 1625 - 1630. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. W. M. Keunen, C. J. Stam, D. L. J. Tavy, W. H. Mess, B. M. Titulaer, and R. G. A. Ackerstaff Preliminary Report of Detecting Microembolic Signals in Transcranial Doppler Time Series With Nonlinear Forecasting Stroke, August 1, 1998; 29(8): 1638 - 1643. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Brey and M. Carolin Detection of cerebral microembolic signals by transcranial Doppler may be a useful part of the equation in determining stroke risk in patients with antiphospholipid antibody syndrome Lupus, January 1, 1997; 6(8): 621 - 624. [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Stroke Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1997 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |