From the Departments of Neurology (F.R., C.P., C.B., M.M.) and Cardiology
(K.T., H.B.), University of Bonn (Germany).
Correspondence to Fernand Ries, MD, Department of Neurology, Sigmund-Freud Str 25, D-53105 Bonn, Germany.
MethodsTwo hundred seventy-six formed emboli, consisting of
different biological and nonbiological materials and as air bubbles,
were injected into a flow phantom with artificial blood vessels and
perfused in a steady or a pulsatile way. Embolic passage was assessed
with a modified 2.5-MHz pulsed Duplex machine and a commercial 2-MHz
Doppler system. Embolic HITS were analyzed using
internationally accepted criteria for the audiovisual characteristics
of HITS. Doppler spectra changes associated with HITS were
evaluated by means of a specially developed high-resolution
analysis of Doppler raw data.
ResultsSeventy-seven percent of all embolic events could be
identified using conventional audiovisual criteria for embolic HITS.
Analysis of Doppler spectra showed that all injected emboli
generated high-amplitude signals with a minimum of at least 3 dB above
background level. In addition, using high-resolution processing,
specific changes in Doppler spectral patterns could be identified
after all embolic HITS caused by solid particles. These postembolic
spectral patterns were always characterized by a Doppler frequency
shift decreasing in time and resembling the letter lambda (
ConclusionsIn this study, highly specific changes in Doppler
spectral patterns associated with microembolic HITS
could be characterized, resulting in further criteria for the
differentiation between microembolic signals and
artifact in Doppler emboli detection. The sensitivity of the
detection of these signals can be increased by high-resolution
analysis of raw Doppler data.
The purpose of this study was to evaluate the specific influence
of embolic material on Doppler power spectra to establish further
criteria for embolus identification. This approach in emboli detection
requires the development of a more fundamental analysis of
Doppler raw data. Thus, a Doppler system with a very high time
resolution and dynamic range was used, and Doppler raw data were
analyzed with software tools that could reduce the influence of
Doppler speckle interfering with embolic signals.
Doppler measurements were performed in a straight polyethylene tube
with a diameter of 6 mm. Laminar flow in the tube was guaranteed
by a tube segment 40 cm long without any mechanical devices, curves, or
edges. The insonation area was located in a glass tank filled with
castor oil for adaptation of acoustic properties and acoustic coupling.
The system was degassed with a vacuum pump and an interposed air
trap.
Microemboli consisted of polyethylene (n=198), cellulose (n=57),
clotted blood (n=11), and air bubbles (n=10). Polyethylene balls had
diameters of 0.8 mm to 1.0 mm (n=83), 1.35 mm (n=58),
and 2.65 mm (n=57), with a SD of ±5%, clotted blood particles of
2 to 4 mm in size, and cellulose particles of about 2 mm. The
size of the emboli was measured using an electronic caliper gauge. Air
bubbles with volumes up to 1 mm3 were
produced by manual injection with a micrometer syringe. One
hundred ten emboli were injected under steady flow conditions, and 166
under pulsatile flow conditions. Solid emboli were injected through an
injection port that was connected via a three-way tap to the circuit
120 cm upstream from the insonation area. Air bubble injection site was
located only 40 cm upstream from the insonation area to avoid
fractionating of the bubbles. Artificial HITS were actively produced by
the observer by lightly tapping on the probe or its fixation.
Doppler measurements were performed with a commercially available
Hewlett-Packard SONOS 1000 Duplex machine, equipped with a 2.5-MHz
linear array transducer. Doppler probes were fixed to a specially
developed tripod. The insonation depth was 60 mm with an
insonation angle of 40°. Continuous correction of the insonation
angle was done to provide corresponding velocity measurements. Sample
volume was set to include the whole cross-sectional area of the tube.
The dynamic range of the Doppler data was about 55 dB at lowest
gain setting. Peak power of flow signal was set to 30% of maximum
resolution on the ultrasound device (HP SONOS). In a dual display mode,
B-mode information and Doppler spectra could be investigated
instantaneously to correlate the passage of the embolus through the
sample volume with the spectral Doppler data.
Emboli were identified on-line according to the conventional
audiovisual criteria for embolic HITS with a minimum difference of at
least 3 dB between the HITS and the background flow
signal.7 Off-line evaluation of raw Doppler
data was done using a specific software that allows high-resolution
Doppler frequency spectrum
analysis.11 Unprocessed data of 256
short-time Fast Fourier Transformations were stored on a data
acquisition unit (486 DX2/66, 8 MB RAM, HD 2 Gbyte). An automated power
spectral analysis was triggered by the trigger signal of the
pulsatile pump. An averaging algorithm allowed the calculation of a
representative spectral pattern based on corresponding
segments of different pulse cycles, thus eliminating Doppler
speckle (ensemble averaging, so-called "Bartlett
procedure"12 ). This averaged pulse
calculated from 10 to 20 different pulse cycles could be subtracted
from single pulses that contained HITS. Thus, only the difference
between the averaged flow signal and the flow pattern of the
HITS-related pulses was displayed, so that HITS could be
analyzed off-line with respect to subtle changes in Doppler
spectrum. Using this procedure, all detected postembolic spectral
changes were described in terms of intensity, duration, and Doppler
frequency shift. For further statistical processing, data of
polyethylene emboli of different sizes under pulsatile flow conditions
were compared using an ANOVA (one-way) statistical
analysis.
Under pulsatile flow conditions, Doppler recordings were
performed simultaneously to registrations with the SONOS
1000 machine with two different MHz probes of a multigate
transcranial Doppler device (DWL-Multidop X, maximum
time resolution 64 short-time Fast Fourier Transformations).
Doppler angle was 15°, two sample volumes of 8 mm were set
into the center of the tube with a distance of 10 mm from sample
volume to sample volume. For off-line evaluation, Doppler spectral
patterns and unprocessed raw data were analyzed for both
channels.
Off-line evaluation of postembolic spectral patterns using the
high-resolution Doppler raw data recordings (HP SONOS 1000)
showed characteristic postembolic spectral patterns following solid
emboli in all registrations for all experimental conditions (steady
versus pulsatile flow; biological versus nonbiological material). These
spectral patterns were characterized by a Doppler frequency shift
decreasing in time and resembling the letter lambda, they will
therefore be described as
In spectral Doppler, postembolic patterns occurred after appearance
of a short-lasting very weak signal or no signal at all, 10 to 20 ms
after the last spectrum of a HITS. Initially, the intensity of these
signals was clearly above the level of the background flow pattern,
then rapidly decreased with time. In 60% of all
Registrations with the dual display mode showed exact coincidence of
the embolus passage through the sample volume in B-mode and HITS in
spectral Doppler. Postembolic spectral signals occurred clearly
after the passage of the embolus and sometimes were even detected at a
time period when the embolic particle had left the display on the
screen.
The DWL machine was less sensitive for the detection of these specific
spectral patterns associated with microembolic HITS.
Nevertheless, postembolic spectral changes resembling the
These results indicate that the
Until now, no specific postembolic spectral patterns have been
described. However, a review of the current literature concerning
emboli detection in patients revealed examples of spectral changes
resembling the
We assume that postembolic changes in Doppler spectra early after
the passage of emboli can be explained by Doppler reflection
phenomena caused by postembolic flow disturbances. Ultrasound
studies on different flow conditions demonstrate that Doppler power
increases significantly with the onset of flow
distortion,16 17 a finding that could explain the
higher intensity in the first third of the
Other technical methods to increase the specificity of embolus
identification are based on automated detection software that includes
neuronal network expert systems or a bigate comparison of signals
supposed to be of embolic origin. Neuronal network systems compare
actual HITS to formerly identified "real" embolic
signals.8 9 Embolus detection with these systems
is limited by the possibility of systematic errors caused by an
insufficient identification of reference signals. The validity of
reference signal identification could be improved by the
analysis of postembolic spectral patterns. Bigated Doppler
devices evaluate the time delay between the occurrence of HITS in two
spatially separate Doppler sample volumes using the coincidence
method.10 If this method is not applicable for
methodological reasons (eg, in case of a short middle cerebral artery)
analysis of postembolic spectral patterns for embolus
identification could still be done using only one sample volume. An in
vivo study comparing the specificity and sensitivity of embolus
detection using high-resolution spectral analysis versus the
bigate procedure and conventional criteria is currently in
progress.
In conclusion, the present study has shown that the
analysis of Doppler power spectra can provide specific
information for the detection of cerebral microemboli, which can be
used as additional criteria for embolus identification with ultrasound.
For clinical application of these results, transcranial
Doppler systems have to include a high-resolution raw data
analysis in clinical settings to evaluate specific postembolic
changes of Doppler spectra in vivo.
Received June 26, 1997;
revision received December 22, 1997;
accepted January 6, 1998.
2.
Ries F, Eicke M. Auswirkung der extrakorporalen
Zirkulation auf die intrazerebrale Hämodynamik. In: Widder B, ed.
Transkranielle Dopplersonographie bei cerebrovaskulären
Erkrankungen. Heidelberg: Springer Verlag; 1987:100103.
3.
Spencer MP, Thomas GI, Nicholls SC, Sauvage LR.
Detection of middle cerebral artery emboli during carotid
endarterectomy using transcranial
Doppler sonography. Stroke. 1990;21:415423.
4.
Markus HS, Tegeler CH. Experimental aspects of
high-intensity transient signals in the detection of emboli.
J Clin Ultrasound. 1995;23:8187.[Medline]
[Order article via Infotrieve]
5.
Russel D, Madden KP, Clark, WM, Sandset PM, Zivin JA.
Detection of arterial emboli using Doppler ultrasound
in rabbits. Stroke. 1991;22:253258.
6.
Markus HS, Harrison MJ. Microembolic
signal detection using ultrasound. Stroke. 1995;26:15171519. Editorial.
7.
Consensus Committee of the Ninth International
Cerebral Hemodynamics Symposium. Basic identification
criteria of Doppler microembolic signals.
Stroke. 1995;26:1123.
8.
Markus HS, Loh A, Brown MM. Computerized
detection of cerebral emboli and discrimination from artifact using
Doppler ultrasound. Stroke. 1993;24:16671672.
9.
Siebler M, Sitzer M, Bender A, Steinmetz H. Real-time
identification of cerebral microemboli with US feature detection by a
neural network. Radiology. 1994;192:739742.
10.
Georgiadis D, Goeke J, Hill M, König M, Nabavi
DG, Stögbauer F, Zunker P, Ringelstein EB. A novel technique for
identification of Doppler microembolic signals
based on the coincidence method. In vitro and in vivo evaluation.
Stroke. 1996;27:683686.
11.
Becher H, Tiemann K, Steck J. DOPPDOC-A new system for
evaluation of ultrasound backscatter by Doppler power measurement.
Eur Heart J. 1996;17:140. Abstract.
12.
Evans DH. Doppler Ultrasound-Physics,
Instrumentation and Clinical Applications. New York, NY: John
Wiley & Sons; 1989.
13.
Ries F, Tiemann K, Bauer C, Mundo M, Becher H. High
resolution emboli detection and differentiation by characteristic
spectral flow disturbances. Thrombosis. 1996;6:78.
Abstract.
14.
Ackerstaff RGA. Carotid
endarterectomy and intraoperative emboli detection.
Echocardiography. 1996;13:543550.[Medline]
[Order article via Infotrieve]
15.
Spencer MP, Thomas GI, Nicholls SC, Sauvage LR.
Detection of middle cerebral artery emboli during carotid
endarterectomy using transcranial
Doppler ultrasonography. Stroke. 1990;21:415423.
16.
Shung KK, Yuan Y, Fei DY, Tarbell JM. Effect of flow
disturbance on ultrasonic backscatter from blood. J
Acoust Soc Am. 1984;75:12651272.[Medline]
[Order article via Infotrieve]
17.
Bascom PAJ, Cobbold RSC. On a fractal packing approach
for understanding ultrasonic backscattering from blood. J Acoust
Soc Am. 1995;98:30403049.[Medline]
[Order article via Infotrieve]
© 1998 American Heart Association, Inc.
Original Contributions
High-Resolution Emboli Detection and Differentiation by Characteristic Postembolic Spectral Patterns
![]()
Abstract
Top
Abstract
Introduction
Subjects and Methods
Results
Discussion
References
Background and
PurposeHigh-intensity transient signals (HITS) detected by
transcranial Doppler ultrasonography correspond to
microemboli in intracranial arteries. The purpose of this study was to
develop new diagnostic criteria for the differentiation of
these microembolic signals from artifact, based on a
high-resolution analysis of Doppler power spectra in an in
vitro model.
).
Duration and appearance of the postembolic spectral patterns were
mainly influenced by the size and velocity of the embolus. Similar
phenomena could not be found in case of embolism by either small air
bubbles or in case of provoked artifact registration. Using a
commercial Doppler system specific, we documented postembolic
spectral patterns in 47% of injected emboli.
Key Words: diagnostic imaging embolism spectrum analysis ultrasonography, Doppler
![]()
Introduction
Top
Abstract
Introduction
Subjects and Methods
Results
Discussion
References
Cerebral microemboli
presenting as high-intensity transient signals (HITS) were first
described during transcranial Doppler monitoring of
cardiac1 2 and carotid
surgeries.3 Since these reports, the phenomenon
of embolic HITS has been described in experimental
settings,4 5 whereas the clinical relevance of
cerebral microemboli is still controversial.6 One
major problem of the current clinical application of Doppler
embolus detection is the differentiation between true embolic signals
and artifact. Moreover, it is difficult to differentiate true embolic
signals from spontaneous speckling in the background signal. In this
context, most common algorithms for embolus identification are based on
the analysis of the internationally accepted basic audiovisual
characteristics of HITS.7 Additionally, some
centers use automated detection software including neuronal network
expert systems8 9 or a multigate comparison of
signals supposed to be of embolic origin.10
Although the differentiation between embolic HITS and artifact is
principally feasible with these procedures, they are all subject to
different methodological restrictions, and sometimes the underlying
techniques still have to be verified.
![]()
Subjects and Methods
Top
Abstract
Introduction
Subjects and Methods
Results
Discussion
References
Experiments were performed using an in vitro flow phantom. This
phantom consisted of an electronically controlled pump system
generating a steady or a pulsatile flow pattern. The pump was served by
a 5-L reservoir filled with washed bovine erythrocytes or a blood
analog, consisting of a mix of 60/40% tyrode/glycerine solution and
cellulose particles. Volumetric flow of steady flow patterns was 150
mL/min. Pulsatile flow was simulated with an arterial flow
pattern, stroke volume was 8.5 mL, frequency was 60 pulses/min. A
variable Windkessel was interposed in the circuit to control
volumetric flow during diastole.
![]()
Results
Top
Abstract
Introduction
Subjects and Methods
Results
Discussion
References
Seventy-seven percent of all embolic events could be detected
on-line according to the conventional audiovisual characteristics of
HITS, and 23% did not create a sufficient audible
signal.7 Referring to off-line analysis,
all injected emboli generated a high-amplitude signal with a minimum of
at least 3 dB above background level that could always be detected by
both Doppler devices. In 29% of all events, the passage of emboli
resulted in an excessive increase of signal intensity above the dynamic
range of the Doppler system (HP SONOS 1000). Intensity of embolic
HITS was positively correlated to embolus size and strongly influenced
by the acoustical properties of the detected material. Depending on the
acoustical impedance, small air bubbles generated high-amplitude
signals compared with the relatively low intensity of large
polyethylene particles. HITS had a mean duration of 40 ms with a SD of
6 ms (steady flow conditions). Mean velocity of the emboli passing the
Doppler beam, defined as Doppler frequency shift at maximum
amplitude of HITS, was 32.4 cm/s (±3.7 cm/s) under steady flow
conditions and 38.9 cm/s (±11.2 cm/s) under pulsatile flow conditions.
Embolus velocity was significantly influenced by embolus size (ANOVA:
F ratio=8.9, P<.001; see Table
). Post hoc group
analysis revealed significant differences in velocity between
faster small emboli (0.8 to 1.0 mm) and slower large emboli
(2.65 mm).
View this table:
[in a new window]
Table 1. Embolus Velocity, Duration, and Slope of Postembolic Spectral
Patterns Following Polyethylene Emboli Under Pulsatile Flow Conditions
-signs13 in the
following part of this manuscript (see
-sign in Figure 1
for the HP-SONOS machine, in Figure 2
for the DWL-machine). All injected air
bubbles produced strong HITS but were never followed by
-signs even
in those cases when signal intensity was much higher than in solid
particles followed by strong postembolic spectral changes. Large
clusters of air bubbles (only caused by degassing of the system)
resulted in detection of postembolic spectral changes resembling a
-sign.
-Signs were never detected after artificial HITS that were
not of embolic origin, produced by mechanical irritation of the
Doppler probe.

View larger version (126K):
[in a new window]
Figure 1. Power spectrum display of 2.0-mmsized blood clot
embolus under pulsatile flow conditions. After passage of the embolus,
postembolic spectral patterns defined as
-sign (see text for
explanation) was observed for about 180 ms. Computation of
power/spectrum shows HITS and postembolic changes as well
(bottom).

View larger version (59K):
[in a new window]
Figure 2. Power spectra (left) and raw data (right) were
analyzed for duration and intensity of HITS and for
differentiation of specific postembolic spectral patterns with the DWL
machine. The second part of the signal begins
simultaneously with
-sign of Doppler spectra.
-signs, the
velocity of the signal finally passed zero to a reverse flow direction.
In the remaining 40%, the spectral patterns did not pass the baseline.
Duration of
-signs following the initiation of a HITS period was
always longer than the preceding HITS. The mean duration of all
detected
-signs was 230 ms (±54 ms) under steady flow conditions
and 169 ms (±53 ms) under pulsatile flow conditions. For polyethylene
emboli under pulsatile flow conditions, duration of
-signs was
significantly influenced by embolus size (ANOVA: F
ratio=5.3, P<.01; see Table
). Post hoc group
analysis revealed significant differences between short-lasting
small emboli (0.8 to 1.0 mm in size) and long-lasting large emboli
(2.65 mm in size). Further specifications of
-signs include the
slope of the signals, defined as the differential velocity to time
(dv/dt) of the decrease of the Doppler frequency shift. The mean
slope of all detected
-signs was 248 cm/s2
(±67 cm/s2) under steady flow conditions and 356
cm/s2 (±154 cm/s2) under
pulsatile flow conditions. For polyethylene emboli under pulsatile flow
conditions, this slope was again significantly influenced by embolus
size (ANOVA: F ratio=7.3, P<.001; see Table
).
Post hoc group analysis revealed significant differences
between steeper
-signs of small emboli (0.8 to 1.0 mm in size)
and flatter
-signs of large emboli (2.65 mm in size).
Furthermore, the slope of the
-signs showed a strong and highly
significant positive linear correlation to the initial velocity of the
embolus (for polyethylene emboli with a size between 0.8 and 1 mm
under pulsatile flow conditions r=.88, P<.001,
scatter plot shown in Fig 3
).

View larger version (19K):
[in a new window]
Figure 3. Scatterplot of the highly significant correlation
between the initial velocity of the embolus and the slope of the
postembolic spectral patterns. Example for polyethylene particles with
a size between 0.8 and 1 mm under pulsatile flow conditions.
-sign
could be detected in 46% of all injections under pulsatile flow
conditions. These signals showed again a clear distinction to the
preceding HITS due to reflection by the embolus particle and could
always be seen in both channels of the multigate probes and in the raw
data sets as well. The duration of postembolic signals in spectral
Doppler and Doppler raw data display was identical (Figure 2
).
![]()
Discussion
Top
Abstract
Introduction
Subjects and Methods
Results
Discussion
References
The purpose of this study was to evaluate the influence of embolic
material on Doppler power spectra to develop new
diagnostic criteria for the differentiation of HITS due to
the passage of microemboli from artifact. Our results demonstrate that
embolic HITS are followed by characteristic postembolic spectral
patterns with a Doppler frequency shift decreasing in time
and resembling the letter lambda. These postembolic changes of
Doppler spectra, which will be called the
-sign,13 could be detected after the passage
of all injected solid particles consisting of different biological and
nonbiological materials. In the case of air embolism,
-signs could
only be detected following the larger bubbles caused by degassing the
system, whereas
-signs were never seen following HITS due to
artifact registration.
-sign is highly specific for
embolic HITS, especially for those caused by solid particles. Thus, the
detection of specific postembolic spectral patterns may
represent an important addition to the widely accepted
consensus criteria for the identification of
microembolic signals caused by different thrombotic
material. These criteria include the evaluation of intensity, duration,
appearance, and acoustic characteristics of
HITS.7 When they were used in the present
study, embolic events could only be identified in 77%, while specific
postembolic spectral patterns were documented following all embolic
HITS. These results suggest that the analysis of postembolic
spectral patterns could also increase the sensitivity of embolus
detection, especially for those HITS that only create a weak audible
signal. As the sensitivity for the detection of
-signs dropped to
47% in a commercial Doppler system designed for emboli detection,
a high-resolution analysis of background subtracted Doppler
raw data is necessary for application of spectral analysis in
embolus detection.
-sign, randomly shown in the presented
figures.14 15 The reproducible appearance of this
phenomenon, which had not been considered in these studies, could be
evaluated first in our laboratory under the optimized conditions of an
in vitro flow phantom. Moreover, the development of a specific
algorithm for emboli detection even allowed the detection of
characteristic spectral changes due to the embolic passage of material,
usually causing HITS of lower intensity. Therefore, the implementation
of this technique to commercially available Doppler devices should
allow one to use this phenomenon in clinical studies to differentiate
embolic HITS from high-intensity artifact as well as from background
Doppler speckle.
-sign compared with the
background flow signal. However, technical factors concerning beam
geometry may play a major role in the generation of this phenomenon.
The linear decline of spectral patterns of the postembolic signal
frequently passing zero to a reverse flow direction (as well as the
strong correlation between the initial embolus velocity and the slope
of the signal) could not be explained by mere turbulence effects. For
further explanation of the
-sign, the physical interactions
between postembolic Doppler spectral patterns, and technical
settings of the Doppler device, flow conditions and properties of
the emboli have to be investigated using mathematical models and flow
simulation studies.
![]()
Acknowledgments
We thank Thomas Schlosser, Dorothee Stähler, Vera
Lambrecht, Rustam Mundegar, and the Fraunhofer Institute for Biomedical
Engineering, St Ingbert, Germany for their contribution to this
study.
![]()
References
Top
Abstract
Introduction
Subjects and Methods
Results
Discussion
References
1.
Padayachee TS, Parsons S, Theobald R, Linley
J, Gosling RG, Deverall PB.The detection of microemboli in the
middle cerebral artery during cardiopulmonary bypass: a
transcranial Doppler ultrasound investigation using
membrane and bubble oxygenators. Ann Thorac Surg. 1987;44:298302.[Abstract]
This article has been cited by other articles:
![]() |
T. Kofidis, S. Fischer, R. Leyh, H. Mair, M. Deckert, R. Haberl, A. Haverich, and B. Reichart Clinical relevance of intracranial high intensity transient signals in patients following prosthetic aortic valve replacement Eur. J. Cardiothorac. Surg., January 1, 2002; 21(1): 22 - 26. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. A. Rodriguez, G. Cornel, N. A. Weerasena, and W. M. Splinter Effect of Trendelenburg head position during cardiac deairing on cerebral microemboli in children: A randomized controlled trial J. Thorac. Cardiovasc. Surg., January 1, 2001; 121(1): 0003 - 9. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. W Droste, R. Dittrich, V. Kemeny, G. Schulte-Altedorneburg, and E B. Ringelstein Prevalence and frequency of microembolic signals in 105 patients with extracranial carotid artery occlusive disease J. Neurol. Neurosurg. Psychiatry, October 1, 1999; 67(4): 525 - 528. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Furui, K. Hanzawa, H. Ohzeki, T. Nakajima, N. Fukuhara, M. Takamori, J.P. Mohr, and C. Stapf "Tail Sign" Associated With Microembolic Signals • Editorial Comment Stroke, April 1, 1999; 30(4): 863 - 866. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Stroke Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1998 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |