MethodsThirty patients with 30 intracranial aneurysms were investigated (8 men, 22 women; mean±SD age 54±17 years). The TCCS examinations were performed with a 2-MHz probe using the power mode. The 3D system (3D-Echotech, Germany) consisted of an electromagnet, which induced a low-intensity magnetic field near the head of the patient. A magnetic position sensor was attached to the ultrasound probe and transmitted the spatial orientation of the probe to a workstation, which also received the corresponding 2D-images from the video-port of the duplex machine. The echo contrast enhancer D-galactose (Levovist, Schering, Germany) was used in all patients to improve the signal-to-noise ratio. All patients underwent presurgical digital subtraction angiography (DSA) to demonstrate the aneurysm.
ResultsTwenty-nine of 30 angiographically proven intracranial aneurysms (97%) were detected by 3D-TCCS. The aneurysmal diameter estimated by DSA ranged from 3 to 16 mm (mean 7.2±3.6 mm). A comparison of the 3 main diameters of each aneurysm revealed a correlation coefficient of 0.95 between DSA and 3D-TCCS. The 3D determination of the aneurysmal size by 2 experienced sonographers correlated with 0.96.
Conclusions3D-TCCS is a new, noninvasive method to investigate intracranial aneurysms. The differentiation between artifacts and true changes of the vessel anatomy is much easier in 3D-TCCS than in conventional 2D-TCCS. The new method yields an excellent correlation with the gold standard, DSA. Because the same 3D-TCCS data can be postprocessed by different investigators, it may be possible to improve reproducibility and increase the objectivity of transcranial color-coded duplex sonography.
Key Words: aneurysm angiography cerebral circulation ultrasonography, Doppler, transcranial
| Introduction |
|---|
|
|
|---|
| Subjects and Methods |
|---|
|
|
|---|
The free-hand system (3D-Echotech) used in this study is adaptable to every commercially available color-coded duplex system. The 3D system consists of an electromagnet, which induces a low-intensity magnetic field near the head of the patient. The composite magnetic field is generated from an array of 3 coils at 90° angles to each other to produce 3 magnetic fields, yielding a 3D orientation. A magnetic position sensor is attached to the ultrasound probe and transmits the spatial orientation (x, y, and z axes) of the probe to a workstation (2x Pentium 400-MHz processors, 256 MB ram), which also receives the corresponding 2D images from the video port of the duplex machine. This approach allows the operator to move the transducer without constraint. To acquire a 3D data block of the circle of Willis, the transducer must be tilted around an axis that is perpendicular to the central axis of the transducer (fanlike rotation) and moved slowly across the temporal bone window or the occipital approach. Over an interval of 30 to 40 seconds, 150 2D images, together with the spatial information, are stored on the hard disk of the workstation. During offline analysis a reconstruction algorithm is able to extract the color-coded information from the 3D data set. The Windows NT 4.0based software provides a photorealistic surface rendering of the investigated vessels. With use of different threshold values and cutting tools, it is possible to reduce artifact. An experienced ultrasonographer familiar with the equipment can produce an animated 3D image of the circle of Willis in 5 minutes. All patients were additionally investigated with 4-vessel digital subtraction angiography (DSA).
Thirty patients with subarachnoidal hemorrhage due to 30 angiographically proven intracranial aneurysms (8 men, 22 women; mean±SD age 54±17years) were involved. For the TCCS examination the investigator was aware of the blood distribution in the CT scans but was blinded to the precise location and dimensions of the aneurysm. In addition, 3D-TCCS data concerning the approximate size (craniocaudal, ventrodorsal, and mediolateral diameter), location, and orientation of the aneurysm were compared with angiographic, intraoperative, and 2D-TCCS data, respectively.
All 2D-TCCS studies with and without contrast were performed by one of the authors (G.L.). The recording of the 3D data and following 3D-TCCS offline analysis was performed by another author (A.B.). With a second offline-analysis by another operator (C.K.), we were able to calculate the interobserver correlation. For this second evaluation the data sets of all 30 patients were analyzed in a random order, without the knowledge of the previously performed 2D-TCCS examinations.
Statistical Analysis
DSA and 3D-TCCS findings regarding the aneurysmal
diameter were compared by linear regression analysis. The
correlation coefficient was calculated for the estimation of
aneurysmal size by 2 different investigators through use of
3D-TCCS. The overall accuracy was applied for the angiographic and
sonographic estimation of the aneurysmal location.
| Results |
|---|
|
|
|---|
|
|
|
|
|
The offline 3D reconstruction was performed by 2 experienced sonographers who were familiar with the method. Both investigators were only aware of the infratentorial/supratentorial location of the aneurysm or the affected hemisphere, respectively. The overall agreement was 100% for the location of the aneurysm. The determination of the aneurysmal size correlated with 0.96 between both investigators.
Correlation of NonContrast-Enhanced 3D-TCCS With
DSA
Only 9 aneurysms (30%) could be detected with
noncontrast-enhanced 3D-TCCS. The mean±SD angiographic size of the
detected aneurysms was 8.4±3.5 mm. The failure to detect
the remaining aneurysms must be attributed to an unfavorable
location (n=8), an insufficient bone window (n=4), or a diameter of
<6 mm (n=9). Three tubular aneurysms of the anterior
communicating artery were not diagnosed, because they could not be
differentiated from both distal segments (A2) of the anterior cerebral
artery. The noncontrast-enhanced 3D-TCCS examinations frequently
revealed vessel discontinuity, and a differentiation from background
signals was more difficult.
Correlation of Native 2D-TCCS, Contrast-Enhanced 2D-TCCS,
and DSA
Only 9 aneurysms (30%) could be detected with native
2D-TCCS. The mean±SD angiographic size of the detected
aneurysms was 8.4±3.5 mm. The failure to detect the
remaining aneurysms must be attributed to the same factors that
were mentioned for the noncontrast-enhanced 3D-TCCS examination.
In 20 of 30 aneurysms (67%), contrast-enhanced 2D-TCCS was
able to demonstrate the intracranial aneurysm. The remaining 10
aneurysms had a diameter of
6 mm (n=8) and/or showed an
unfavourable location (n=4) (Table
). The main aneurysmal
diameter estimated with 2D-TCCS revealed a lower correlation
coefficient (0.615) for DSA than that obtained with 3D-TCCS. To
determinate the third plane (craniocaudal), it was necessary to use a
coronal insonation plane for 2D-TCCS. Coronal sections were possible
only if the aneurysms were located at the intracranial carotid
bifurcation, the middle cerebral artery trunk, or at the top of the
basilar artery. Other locations required an extreme inclination of the
probe.
Conclusions
The usefulness of 2D-TCCS as a diagnostic tool for
vascular alterations is unquestioned, but its dependence on operator
skill was contributed to conflicting results in some
trials.1 3 4 5 Compared with DSA, the demonstration of
vessel disease through use of screenshots or video clips of 2D-TCCS
examinations is often not clear for those who are unfamiliar with
the method. Multiple 2D-TCCS images must be integrated in the
sonographer's mind to develop a 3D impression of the intracranial
vessel anatomy. Various 2D-TCCS studies1 3 4 5
report that the minimal detectable diameter of an intracranial
aneurysm ranges between 3 to 6 mm, but the detection of
aneurysms may be impossible if 2 vascular segments are close
together. The restriction to only a few insonation windows (temporal,
nuchal, transorbital) with variable insonation conditions sometimes
makes it impossible to obtain optimal image planes for diagnosis of
vascular changes.
In the past few years computer technology has advanced sufficiently to allow the development of visualization techniques suitable for 3D-TCCS. Power-based TCCS is better suited than mean frequencybased TCCS for 3D reconstruction, because it is more sensitive in detecting low flow near the vessel wall and is less degraded by noise and clutter. Together with advances in acquisition, reconstruction techniques and the use of echo-contrast enhancer, this means that 3D images of high quality can be produced.
Using a power-based 3D-TCCS system, Delcker et al14 have reported a detection rate of 60% to 90% through the temporal bone in ipsilateral vessels (A1, M1, M2, P1 and P2, and posterior communicating artery) in 10 patients. The identification rate was increased to 100% with use of a transpulmonary stable contrast agent. The contralateral vessels were visualized in just 10% to 60% of cases, whereas after application of an echo-contrast enhancer identification rates of 80% to 100% could be obtained. Bauer et al15 compared frequency-based 3D-TCCS with power-based 3D-TCCS, using the echo contrast enhancer Levovist. They found significantly better vessel continuity as well as smoother vessel surfaces when using power mode.
To our knowledge, the present study is the first investigation of intracranial vessel disease on a large number of patients with use of 3D-TCCS. Compared with conventional 2D-TCCS, the differentiation between artifact and real changes of the vessel anatomy is much easier and yields an excellent correlation with the gold standard, DSA. The echo-contrast enhancer enables the 3D reconstruction of small vascular alterations, avoids vessel discontinuity, and significantly improves the differentiation from disturbing background signals. The low rate of detected aneurysms (30%) using nonenhanced 3D-TCCS underlines the necessity of applying contrast enhancers for sonographic 3D reconstruction.
3D-TCCS significantly improved the detection of intracranial aneurysms compared with 2D-TCCS (97% versus 67%) and revealed a better correlation of the estimated aneurysmal diameters with the angiographic data (0.95 versus 0.615). The relatively low detection rate of aneurysms with use of contrast-enhanced 2D-TCCS in the presented study was mainly caused by the large number of small aneurysms.
3D-TCCS enables the investigator to reconstruct virtually any arbitrary view angle and may allow the comparison with 3D techniques such as CT or MR angiography. Unfavorably located vessel segments, such as the carotid siphon, can be insonated in transversal planes and may be assessed after 3D- reconstruction. Because the same 3D-TCCS data can be postprocessed by different investigators, it may be possible to improve reproducibility and increase the objectivity in transcranial color-coded duplex sonography. Remote consultations of the sonographic examinations are possible because the complete data of the 3D examination are digitally stored and can be transferred by usual methods of digital data transfer.
Other techniques compete with 3D-TCCS in the noninvasive detection of intracranial vessel disease. MR angiography16 17 18 19 enables the detection of intracranial aneurysms with a sensitivity of up to 90% and a specificity of 75%, but in aneurysms with a diameter of <5 mm or a location in the carotid siphon, the sensitivity decreases to 60%.17 20 21 With CT angiography22 it is possible to detect aneurysms >5 mm with a sensitivity of 90%.23 24 Due to its vicinity to the skull, the sensitivity decreases if the aneurysms are located in the carotid siphon.25 Further disadvantages may be the interference of venous flow signals26 and allergic reactions against the contrast agent. The radiation exposure excludes repeated investigations with CT- angiography.
In summary, 3D-TCCS is not suitable as a screening method, because of its limited sensitivity in detecting small aneurysms and aneurysms in particular locations. The major limitation is an absent temporal bone window, which is found in only 5% of the population but in up to 30% of women aged >70 years.27 28
However, routine 3D-TCCS examinations for other reasons will increase the incidental detection of asymptomatic intracranial aneurysms and offer the chance to treat them before rupture.29 Its advantages lie in the possibility to perform the examination in a stroke unit or intensive care unit within a few minutes and to obtain a day-by-day follow-up.
| Acknowledgments |
|---|
Received April 1, 1999; accepted August 5, 1999.
| References |
|---|
|
|
|---|
2. Martin PJ, Gaunt ME, Naylor AR, Hope DT, Orpe V, Evans DH. Intracranial aneurysms and arteriovenous malformations: transcranial colour-coded sonography as a diagnostic aid. Ultrasound Med Biol. 1994;20:689698.[Medline] [Order article via Infotrieve]
3. Baumgartner RW, Mattle HP, Kothbauer K, Schroth G. Transcranial color-coded duplex sonography in cerebral aneurysms. Stroke. 1994;25:24292434.[Abstract]
4. Wardlaw JM, Cannon JC. Colour transcranial power Doppler ultrasound of intracranial aneurysms. J Neurosurg. 1996;84:459461.[Medline] [Order article via Infotrieve]
5. Klötzsch C, Nahser HC, Fischer B, Henkes H, Kühne D, Berlit P. Visualization of intracranial aneurysms by transcranial duplex sonography. Neuroradiology. 1996;38:555559.[Medline] [Order article via Infotrieve]
6. Lindner A, Shambal S, Georgiadis D, Becker G. Transcranial color-coded duplex ultrasound in interventional therapy of cerebral aneurysms. Ultraschall Med. 1997;18:148152.[Medline] [Order article via Infotrieve]
7. Griewing B, Motsch L, Piek J, Schminke U, Brassel F, Kessler C. Transcranial power mode Doppler duplex sonography of intracranial aneurysms. J Neuroimaging. 1998;8:155158.[Medline] [Order article via Infotrieve]
8. Bartels E, Rodiek SO, Lumenta C, Flügel KA. What advantages has power Doppler imaging in comparison with transcranial color-coded duplex ultrasonography in the evaluation of cerebral aneurysms? In: Klingelhöfer J, Bartels E, Ringelstein RB, eds. New Trends in Cerebral Hemodynamics. Amsterdam, Netherlands: Elsevier Science Publishing Co; 1997:320325.
9. Fischer B, Klötzsch C, Nahser HC, Henkes H, Kühne D, Berlit P. Klinische Anwendung der TCCD zum Nachweis intrakranieller Aneurysmen. Nervenarzt. 1998;69:671677.[Medline] [Order article via Infotrieve]
10. Bauer A, Becker G, Krone A, Fröhlich T, Bogdahn U. Transcranial duplex sonography using ultrasound contrast enhancers. Clin Radiol. 1996;51:1923.
11. Wardlaw JM, Cannon J, Sellar RJ. Use of color power transcranial Doppler sonography to monitor aneurysmal coiling. AJNR Am J Neuroradiol. 1996;17:864867.[Abstract]
12. Berger G, Roesch P, Steffgen L. 3D-Rekonstruktion intrakranieller Gefäße mit Hilfe des Colour-Doppler-Energy (CDE)-Verfahrens - Erste Versuche. Ultraschall Med. 1996;17:277280.[Medline] [Order article via Infotrieve]
13. Lyden PD, Nelson TR. Visualization of the cerebral circulation using three-dimensional transcranial power Doppler ultrasound imaging. J Neuroimaging. 1997;7:3539.[Medline] [Order article via Infotrieve]
14. Delcker A, Turowski B. Diagnostic value of three-dimensional transcranial contrast duplex sonography. J Neuroimaging. 1997;7:139144.[Medline] [Order article via Infotrieve]
15. Bauer A, Bogdahn U, Soldner R, Hutzelmeier A, Haase A. Dreidimensionale transkranielle und Echosignalverstärker-Farbduplex-Sonographie. In: Bogdahn U, Becker G, Schlachetzki F, eds. Echosignalverstärker und transkranielle Duplexsonographie. Berlin, Germany: Blackwell Wissenschaftsverlag; 1998.
16. Ikawa F, Sumida M, Uozumi T, Kuwabara S, Kiya K, Kurisu K, Arita K, Satoh H. Comparison of three-dimensional phase-contrast magnetic resonance angiography with three-dimensional time-of-flight magnetic resonance angiography in cerebral aneurysms. Surg Neurol. 1994;42:287292.[Medline] [Order article via Infotrieve]
17. Korogi Y, Takahashi M, Mabuchi N, Watabe T, Shiokawa Y, Shiga H, O'Uchi T, Nakagawa T, Miki H, Horikawa Y, Fujiwara S, Furuse M. MR-angiography of intracranial aneurysms: a comparison of 0.5 T and 1.5 T. Comput Med Imaging Graph. 1997;21:111116.[Medline] [Order article via Infotrieve]
18. Shankla SK, Gunawardena WJ, Coutinho CM, Jones AP, Keogh AJ. Magnetic resonance angiography in the management of aneurysmal subarachnoidal haemorrhage: a study of 51 cases. Neuroradiology. 1996;38:724729.[Medline] [Order article via Infotrieve]
19. Wilcock D, Jaspan T, Holland I, Cherryman G, Worthington B. Comparison of magnetic resonance angiography with conventional angiography in the detection of intracranial aneurysms in patients presenting with subarachnoid haemorrhage. Clin Radiol. 1996;51:330334.[Medline] [Order article via Infotrieve]
20. Falk A, Schmieder K, Hentsch A, Harders A, Heuser L. 3-D-TONE magnetic resonance angiography in the detection of intracranial aneurysms compared with digital subtraction angiography: a prospective study. Rofo Fortschr Geb Rontgenstr Neuen Bildgeb Verfahren. 1996;164:3137.
21. Houkin K, Aoki T, Takahashi A, Abe H, Koiwa M, Kashiwaba T. Magnetic resonance angiography (MRA) of ruptured cerebral aneurysms. Acta Neurochir (Wien). 1994;128:132136.[Medline] [Order article via Infotrieve]
22. Tacke J, Klein M, Bertalanffy H, Mayfrank L, Thron A, Gilsbach JM, Günther RW. Clinical significance of three-dimensional helical CT in neurosurgery. Minim Invasive Neurosurg. 1997;40:3035.[Medline] [Order article via Infotrieve]
23. Rohnert W, Hanig V, Hietschold V, Abolmaali N. Detection of aneurysm in subarachnoid hemorrhage: CT angiography vs. digital subtraction angiography. Aktuelle Radiol. 1998;8:6370.[Medline] [Order article via Infotrieve]
24. Eberhardt KE, Tomandl B, Nomayr A, Huk WJ. Value of CT-angiography in the diagnosis of cerebral artery aneurysms. Radiologe. 1997;37:905912.[Medline] [Order article via Infotrieve]
25. Dietrich J, Gunther L, Fried H. 3-dimensional computerized tomography in diagnosis and surgical planning of intracranial aneurysms. Zentralbl Neurochir. 1995;56:3439.[Medline] [Order article via Infotrieve]
26. Wilms G, Guffens M, Gryspeerdt S, Bosmans H, Maaly M, Boulanger T, Van-Hoe L, Marchal G, Baert A. Spiral CT of intracranial aneurysms: correlation with digital subtraction and magnetic resonance angiography. Neuroradiology. 1996;38:2025.[Medline] [Order article via Infotrieve]
27. Hennerici M, Rautenberg W, Sitzer G, Schwartz A. Transcranial Doppler Ultrasound for the assessment of intracranial arterial flow velocity, part 1: examination technique and normal values. Surg Neurol. 1987;27:439448.[Medline] [Order article via Infotrieve]
28. Widder B. Transkranielle Dopplersonographie bei zerebrovaskulären Erkrankungen. Berlin, Germany: Springer Publishing Co. 1987:3031.
29. Rinkel G, Djibutti M, Algra A, van Gijn J. Prevalence and risk of rupture of intracranial aneurysms: a systematic review. Stroke. 1998;229:251256.
This article has been cited by other articles:
![]() |
J. U. Harrer, T. Wessels, S. Poerwowidjojo, W. Moller-Hartmann, and C. Klotzsch Three-Dimensional Color-Coded Duplex Sonography for Assessment of the Vertebral Artery Origin and Vertebral Artery Stenoses J. Ultrasound Med., August 1, 2004; 23(8): 1049 - 1056. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Deklunder Role of ultrasound and contrast-enhanced ultrasound in patients with cerebrovascular disease Eur. Heart J. Suppl., March 1, 2002; 4(suppl_C): C51 - C55. [Abstract] [PDF] |
||||
![]() |
C. Klotzsch, A. Bozzato, G. Lammers, M. Mull, and J. Noth Contrast-Enhanced Three-Dimensional Transcranial Color-Coded Sonography of Intracranial Stenoses AJNR Am. J. Neuroradiol., February 1, 2002; 23(2): 208 - 212. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. M. White, J. M. Wardlaw, E. Teasdale, S. Sloss, J. Cannon, and V. Easton Power Transcranial Doppler Ultrasound in the Detection of Intracranial Aneurysms Stroke, June 1, 2001; 32(6): 1291 - 1297. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. A. Boyajian, S. M. Otis, D.W. Droste, R. Jurgens, R. Tietje, E.B. Ringelstein, and S. Weber Continuous Infusion Versus Bolus Injection Of Ultrasound Contrast Agents in Vascular Doppler Flow Imaging Response Stroke, September 1, 2000; 31 (9): 2266 - 2278. [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Stroke Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1999 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |