(Stroke. 1999;30:827-833.)
© 1999 American Heart Association, Inc.
Original Contributions |
From the Department of Clinical Radiology, Kuopio University Hospital (Finland).
| Abstract |
|---|
|
|
|---|
MethodsTOF MRA at 1.5 T, IVUS, and selective digital subtraction angiography were performed on 31 extracranial carotid arteries of 27 patients (mean age, 52 years; age range, 17 to 75 years) undergoing neuroendovascular interventions. The atherosclerotic lesions were registered, and quantitative measurements of plaque thickness, luminal diameters, and diameter stenosis were independently performed for the imaging modalities.
ResultsAmong 170 arterial segments analyzed, IVUS revealed a total of 48 atherosclerotic lesions (mean diameter stenosis, 17%; range, 4% to 40%), only 25 of which were depicted on digital subtraction angiography. Analysis of the axial source images of TOF MRA resulted in sensitivity of 77% to 83% and specificity of 71% to 80% in lesion depiction for the 2 readers with reference to IVUS. The values of diameter stenosis measured from MRA and IVUS were closely interrelated (r=0.53 to 0.61, P<0.001).
ConclusionsThree-dimensional TOF MRA is feasible and moderately accurate for evaluation of mild atherosclerotic changes of carotid arteries.
Key Words: angiography carotid artery diseases magnetic resonance angiography ultrasonography, interventional
| Introduction |
|---|
|
|
|---|
Despite its well-known limitations, contrast angiography is widely preferred as the standard method in evaluating the severity of atherosclerotic carotid stenosis. Angiography offers only indirect information on atherosclerotic lesions as a change in the luminal wall profile but provides no further information on composition of plaque material unless ulcerated or heavily calcified. Because of compensatory enlargement of the vessel, mild atherosclerotic plaques do not encroach on the lumen until the lesion occupies up to 40% of the combined arterial wall and lumen volume.1 2
Intravascular ultrasound (IVUS) is currently verified as the standard of reference for imaging normal and atherosclerotic coronary and peripheral vessel wall morphology.3 4 5 The utility of IVUS in carotid artery imaging in vivo has recently been reported.6 However, in addition to its limited availability and some technical disadvantages, IVUS is invasive, and there is a definite need for low-risk noninvasive imaging.
Three-dimensional time-of-flight magnetic resonance angiography (3-D TOF MRA) is widely available and a commonly used method to determine the severity of atherosclerotic stenosis of the carotid artery. MRA has proven to be accurate in assessing degree of stenosis,7 8 9 10 and promising preliminary results in characterization of heavy atherosclerotic lesions have been published.11 The utility of 3-D TOF MRA in direct visualization of mild atherosclerotic lesions and intraplaque morphology is still not well known.
The purpose of this study was to evaluate the usefulness of 3-D TOF MRA in detection and quantification of mild atherosclerotic changes of carotid arteries, with reference to IVUS imaging and contrast angiography.
| Subjects and Methods |
|---|
|
|
|---|
Magnetic Resonance Angiography
MRA (1.5 T, Siemens Magneton SP4000 and Vision) was
performed with a neck coil within 2 days of the endovascular procedure.
Axial 3-D TOF MRA sequences (repetition time 33 ms, echo time 8.0 ms,
flip angle 20°, effective slice thickness 0.8 to 1.5 mm, field
of view 200 mm, 256x512 matrix) were performed covering both the
carotid bifurcations for a segment extending
30 mm cranially and
caudally from the flow divider. Acquisition time for TOF MRA was 9
minutes. Multiplanar-reformatted (MPR) and
maximum-intensity-projection images through the carotid bifurcation
were routinely obtained, but only the axial source images were used for
the evaluation of the present study. Theoretical in-plane
resolution was 0.78x0.39 mm.
IVUS and Contrast Angiography
We performed IVUS with a 3.5-F, 30-MHz imaging catheter (Sonos;
Hewlett Packard) using a mechanical transducer (Medi-tech/Boston
Scientific) through a 7F multipurpose or right coronary guiding
catheter. By the use of digital road mapping control, the IVUS catheter
was introduced into the internal carotid artery (ICA) over a 0.018-inch
guide wire. Imaging was registered on a super-VHS tape while the
catheter was slowly pulled backward manually from the ICA to the
proximal part of the common carotid artery (CCA). Gentle manipulation
of the catheter was applied to keep the probe positioned as centrally
as possible. To register the exact location of the probe within the
artery, 7 to 10 single x-ray exposures were obtained in the same
projection during the imaging, and the corresponding time points
were marked with annotations in the IVUS videotape for later
analysis. Selective digital subtraction angiography (DSA) of
the carotid artery was performed in
2 projections during the
intervention. Figure 1
shows the imaging
components used in the study.
|
Image Reading
The DSA images and the single x-ray exposures were first read to
determine the longitudinal position of the individual axial levels of
IVUS. All IVUS videotapes were interpreted by 2 vascular radiologists
with prior experience of peripheral and coronary
artery IVUS. Stenotic lesions were registered, and
intralesional areas with brighter echogenicity than the
reference adventitia and clear acoustic shadowing were interpreted as
calcified.12 13 14 A detailed description of IVUS reading is
presented elsewhere.6 One of the radiologists also
read the DSA films 6 months later, measured the diameter
stenosis, and registered calcifications of the lesions. The MRA
studies were independently reviewed by a neuroradiologist and a
vascular radiologist. Stenotic lesions were registered, and
focal areas with very low signal intensity were interpreted as
calcification.11 To determine the intraobserver
reproducibility, one radiologist read the MRA axial source images twice
with an interval of 3 months. The image quality for both IVUS and MRA
was graded as good, satisfactory, or poor, and the reason for this, if
evident, was defined.
Measurements
The imaged portion of the extracranial carotid artery was
divided into 4 segments: (1) the ICA above the bulbar area, ie, the
distal ICA; (2) the ICA bulb, while in cases without unequivocal
anatomic bulbar widening, the ICA bulb was defined as covering a
segment 0 to 20 mm above the flow divider; (3) the carotid
bifurcation 0 to 10 mm below the flow divider; and (4) the CCA 10
to 50 mm below the flow divider. Image analysis and the
measurements for each artery were routinely performed at 1 or 2 axial
levels within each segment.
The exact longitudinal position of the IVUS probe was determined by calculating the distance of the probe from the upper edge of carotid bifurcation (flow divider) in the single x-ray image. The corresponding axial levels of MRA source images were identified on the basis of the distance from the position of the flow divider within error limits of 0.9 mm. Quantitative measurements of IVUS and MRA source images were accomplished in these defined levels.
Luminal diameters and thickness of the plaques were measured on IVUS during the systolic phase at a single freeze video image. The luminal caliber was defined as the mean value of 2 perpendicular inner diameters, and the plaque thickness was gauged at the thickest site of the lesion on both IVUS and MRA images. The smallest bisecting luminal diameter at the site of tightest stenosis and the reference diameter from the outer edge of the arterial wall at the same site were measured. For IVUS, measurement of the percent stenosis and the thickness of plaques was performed with the equipment software; for MRA axial source images and for DSA, measurement was performed with a digital micrometer at a conventional view box. For DSA, the stenosis was measured according to the European Carotid Surgery Trial criteria,15 analogous to the IVUS and MRA measurements.
Statistical Analysis
For statistical comparison of the continuous MRA and IVUS
variables, the Pearson correlation coefficient was calculated. To
test intraobserver and interobserver consistency of
interpretation, the mean difference of the paired observations was
calculated. To assess the diagnostic performance of
MRA, sensitivity, specificity, and overall accuracy of the dichotomized
stenosis measurements were calculated with IVUS used as a
standard. The correspondence between MRA and IVUS classifications with
nominal scale variables was given as percent agreement.
| Results |
|---|
|
|
|---|
|
Atherosclerotic Findings on IVUS and DSA
Of 170 levels analyzed in the 31 carotid arteries, a total
of 48 atherosclerotic lesions were recognized on IVUS (Table 1
).
In 7 patients, no atherosclerotic changes were registered. Forty-three
of the lesions were eccentric, and 5 were concentric. The mean diameter
stenosis of the atherosclerotic lesions was 17% (range, 4% to
40%). Thirteen of the lesions caused stenosis <10%, and 14
lesions caused stenosis >20%.
DSA revealed 25 atherosclerotic lesions with mean diameter stenosis of 17% (range, 10% to 32%).
Image Quality
The overall image quality of IVUS was good in all patients except
in one case of compromised quality due to technical reasons. In 7
arteries there was difficulty in visualization of the fading opposite
side of the arterial wall at the level of carotid
bifurcation, causing some uncertainty in the assessment of
arterial wall morphology. There was no specific artifact
that degraded study analysis. The flow divider was accurately
identified at IVUS in every carotid artery.
The quality of 3-D TOF MRA axial source images was assessed as good in 64% (109/170), satisfactory in 22% (38/170), and poor in 14% (23/170) of the images. The poor quality was caused by signal loss and inhomogeneity in slices located at the periphery of the imaging slab (13/23), by motion artifacts (8/23), or by turbulent flow (2/23). All the images were included in the study.
Reproducibility of MRA Image Reading
MRA diameter stenosis measurements (n=170) performed by 2
readers correlated closely (r=0.61, P<0.001).
The mean difference between the measured diameter stenosis by
the 2 readers was 2.9% (SD, 9.0%). Two repeated readings of the
diameter stenosis on MRA by the same radiologist denoted
similar correlation (r=0.58, P<0.001), with a
mean difference of 4.1% (SD, 9.4%) The plaque thickness measured by
the 2 readers also correlated statistically very significantly
(r=0.55), with a mean difference of 0.07 mm (SD,
0.86 mm) between the results of the 2 observers. The readers
agreed in regard to 65% of the lesions for the presence of
calcification.
Correlation Between MRA, DSA, and IVUS Measurements
All stenoses detected by DSA were also diagnosed on MRA.
Pearson correlation coefficients between DSA and MRA for these lesions
were 0.38 and 0.37 (P<0.001) for the 2 readers.
Measurements of the luminal diameters on MRA and IVUS (n=170)
correlated closely (r=0.91, P<0.001), and the
mean lumen diameters obtained by the 2 modalities were almost equal
(mean±SD difference, 0.01±0.7 mm). The correlations did not
differ remarkably at the various arterial segments.
Measurements of plaque thickness from MRA and IVUS also correlated
statistically very significantly: correlation coefficients for the 2
readers were 0.28 to 0.64 (P<0.001) (Table 2
).
|
Diameter stenosis measured from IVUS and MRA axial source
images correlated closely (r=0.53 to 0.61,
P<0.001) for the 2 readers. The diagnostic
performance of MRA is shown in Table 3
, in which we calculated
sensitivity, specificity, and overall accuracy for stenosis
detection using various cutoff points and using IVUS as a standard.
Table 3
indicates that up to 100% sensitivity was reached by 1
reader for detection of stenoses >20% on IVUS at the expense
of an increasing number of false-positive findings. Analysis of
the false-positive findings on MRA showed that these were primarily due
to artifacts caused by turbulent flow near the arterial
wall (Figure 2
).
|
|
Assessment of Plaque Calcification
Altogether 29 lesions with calcification were registered on IVUS,
none of which were detected on DSA. Specificity of MRA for detection of
calcifications with reference to IVUS was high (79% to 95%) for the 2
readers, but sensitivity was poor (17% to 52%). The platelike, heavy
calcifications were constantly detected on TOF MRA axial source images,
while the smaller, disseminated lesions were frequently missed (Figures 1
, 3
, 4
, and 5
).
|
|
|
| Discussion |
|---|
|
|
|---|
Benefits of 3-D TOF MRA techniques in the extracranial carotid imaging include simultaneous ability to assess the degree and the extension of the stenosis and to gain information about the arterial wall. On the other hand, its sensitivity to motion artifacts and progressive in-plane spin saturation artifacts occasionally degrades image quality.25 Despite these limitations, good diagnostic performance of 3-D TOF MR angiography with the use of longitudinal MPR and maximum-intensity-projection reconstructions in the diagnosis of stenotic lesions of carotid arteries has been reported in several studies.7 8 9 10 In epidemiological studies, MRA may offer better postprocessing facilities and better observer-related repeatability than conventional US, especially since the anatomic location of the lesion is more accurately defined on MR than on external US imaging. In clinical practice, 3-D TOF MRA is also very useful as a complementary technique for gadolinium-enhanced MRA of carotid arteries.26
IVUS is widely used in coronary and peripheral
arteries for diagnosis and characterization of atherosclerotic lesions.
In a recent study,6 IVUS of the carotid arteries provided
good intraobserver and interobserver repeatability; 2 readers disagreed
in their assessment of atherosclerotic changes in 9%, and
intraobserver variation was found in 7%. The 2 readers disagreed about
28% of the segments in the assessment of IVUS morphology of the
lesions, correspondingly. Histopathologic and IVUS classifications of
the plaque were in agreement in the majority of plaques (82%), and
IVUS reliably depicted histopathologic "fibrous cap" lesions
(Figures 3
, 5
, and 6
). It
has also been firmly established by several studies, in both
coronary and peripheral arteries, that IVUS
frequently reveals atherosclerotic changes in angiographically normal
segments.3 6 27 The present study showed that IVUS is
feasible and safe in carotid arteries with mild
atherosclerosis, which confirms previous
results.6 Further evidence of safety is mandatory before
it is routinely used in severely atherosclerotic carotid arteries.
|
Three-Dimensional TOF in Quantification of Carotid
Atherosclerosis
While most of the MRA literature evaluates the accuracy of
identification of severely stenosed carotid arteries, the present
study concentrates on identification of incipient atherosclerotic
lesions. Diameter stenosis was <20% in more than two thirds
(34/48) of the lesions, and only 52% (25/48) of the lesions were
detected on DSA. We selected axial source images of 3-D TOF MR
angiography for the principal analysis of these mild
atherosclerotic changes instead of image reconstructions to maximize
sensitivity and to facilitate comparison with IVUS, which provides
basically 2-dimensional axial data.
Three-dimensional TOF MRA depicted all stenotic lesions of the
carotid arteries evident on DSA. Moreover, analysis of the
axial source images revealed several additional changes not seen on
DSA. Sensitivity of 3-D TOF MRA in the diagnosis of even the mildest
atherosclerotic changes with reference to IVUS was good: 86% and 83%
for the 2 readers for detection of lesions of
10% diameter
stenosis. The usefulness of MRA suffered from an excess of
false-positive findings (40% to 49%). This was mostly due to flowing
artifacts next to the luminal wall, which were mistaken for
arterial wall thickening. These phenomena usually appeared
eccentrically at the outer curve of the vessel wall and frequently had
a relatively high signal intensity. The most severe misinterpretation
due to flow artifact in MRA was as high as 40% and 31% for the 2
readers in the same ICA, where no stenosis was detected on
IVUS. A meticulous analysis of MPR reconstructions can
occasionally reveal the artifactual nature of this kind of finding
(Figure 2
).
Three-Dimensional TOF MRA in Characterization of Morphology of
Atherosclerotic Lesions
Few previous studies provide information about the morphological
classification of atherosclerotic plaques in TOF MRA, especially with
correlation to histopathologic analysis. Wildy et
al11 used 3-D TOF MRA to analyze plaque
composition of severely diseased arteries before
endarterectomy by using contrast-to-noise ratios
(CNRs), in which the signal intensity of sternocleidomastoid muscle and
the signal intensity of air were given as references in the equation.
The histological findings of
postendarterectomy plaque composition were compared
with CNR values with the result, although with a certain amount of
disagreement, that heterogeneous plaques (mixed plaques)
had the same signal intensity as surrounding muscle tissue (CNR value
of zero). Furthermore, tissues having negative CNR values, such as
calcium, were hypointense, and those with positive CNR values, such as
intraplaque hemorrhage, were hyperintense. Wildy et al noted
that these 3 types of plaque contentscalcification,
hemorrhage, and mixed compositionand flowing blood all had
distinct ranges of CNR that had no overlap.
In a preliminary analysis of the present study, we compared
plaque characterization of MRA with reference to IVUS. Classification
of intraplaque morphology was based on knowledge of MRA axial source
images11 and T1-weighted spin-echo images,28
and analysis of IVUS was based on criteria established by
several authors.5 12 13 14 In our experience, visual
analysis of MRA source images provided poor and
inconsistent differentiation of fibrotic and soft constituents
of mild atherosclerotic lesions with reference to IVUS. In our previous
study we found
82% agreement in morphology of carotid plaques
between IVUS and histopathology,6 and in the absence of
absolute histopathologic reference in the present study we
restrained from further detailed evaluation of plaque characterization
by 3-D TOF MRA.
It is firmly established that IVUS reliably detects macroscopic intralesional calcifications with reference to histopathologic findings.6 12 13 We interpreted focal intralesional hypointense areas at MR axial source images as calcifications, similar to other studies on MRI.11 28 In our study, MRA diagnosed very specifically (79% to 95%) intralesional calcifications with reference to IVUS, but at the expense of a large count of missed ones. While IVUS is sensitive even for punctate calcifications,6 small calcifications were missed on MRA because of the more limited spatial resolution. On the other hand, none of the calcifications was detected on DSA.
We found that some fibrous intimal thickenings seen on IVUS were very hypointense on MRA and were mistakenly interpreted as calcifications. According to the preliminary studies of von Ingersleben et al28 on MRI, a fibrous plaque appeared darker on T1-weighted and proton-densityweighted images when the cells had a layered appearance. This might suggest a similar explanation for the varying signal intensity on TOF MRA. On the other hand, the identification of calcified and fibrous plaques on TOF MRA might be as valuable because they probably are more stable.29 30 31 32 O'Holleran et al32 used external US and found no strokes in patients with calcified plaques that resulted in stenosis of the carotid bifurcation to <75% in diameter, whereas the risk for stroke was 10% if the plaque was soft.
Two of the patients with intracranial arteriovenous malformations died
within a short period after imaging. Carotid arteries of these patients
were excised and available for histopathologic studies.
Histopathological findings of arterial wall appearance and
plaque morphology were equal to in vivo IVUS findings. However, MRA
images were more demonstrative at the carotid bifurcation than IVUS
images, where the whole arterial wall circumference is not
always totally gained in a single freeze image (Figure 5
).
In addition to use in epidemiological follow-up studies and for verification of the efficacy of different conservative treatment interventions, a potential application of 3-D TOF MRA might be the observation of patients with calcified but only mildly or moderately stenosed carotid arteries. Extensive draping calcification often jeopardizes evaluation of accompanying plaque formation and impairs reliability of stenosis determination on Doppler US imaging. In contrast, macroscopic calcification of an atherosclerotic lesion offers a good landmark for TOF MRA for serial imaging, without impairing accurate measurements of the plaque burden.
To summarize, quantitative analysis of axial source images of 3-D TOF MRA seems accurate and reproducible enough to facilitate serial follow-up of mild atherosclerotic changes of human carotid arteries in vivo. Visual analysis may suggest the presence of intralesional calcifications, but the ability of the imaging modality to characterize other intralesional morphology requires further studies with histopathologic correlation.
| Acknowledgments |
|---|
| Footnotes |
|---|
Received October 30, 1998; revision received January 26, 1999; accepted January 26, 1999.
| References |
|---|
|
|
|---|
2.
Polak JF, Kronmal RA, Tell GS, O'Leary DH, Savage PJ,
Gardin JM, Rutan GH, Borhani NO. Compensatory increase in common
carotid artery diameter: relation to blood pressure and artery
intima-media thickness. Stroke. 1996;27:20122015.
3.
Waller B, Pinkerton C, Slack J. Intravascular
ultrasound: a histological study of vessel during life:
the new "gold standard" for vascular imaging.
Circulation. 1992;85:23052310.
4. Lockwood GR, Ryan LK, Gotlieb AI, Lonn E, Hunt JW, Liu P, Foster FS. In vitro high resolution intravascular imaging in muscular and elastic arteries. J Am Coll Cardiol. 1992;20:153160.[Abstract]
5. Nishimura RA, Edwards WD, Warnes CA, Reeder GS, Holmes DR Jr, Tajik AJ, Yock PG. Intravascular ultrasound imaging: in vitro validation and pathological correlation. J Am Coll Cardiol. 1990;16:145154.[Abstract]
6.
Manninen HI, Räsänen H, Vanninen RL, Berg
M, Hippeläinen M, Saari T, Yang X, Karkola K, Kosma V-M. Human
carotid arteries: correlation of intravascular US with angiographic and
histopathologic findings. Radiology. 1998;206:6574.
7. Chiesa R, Melissano G, Castellano R, Triulzi Z, Anzalone N, Veglia F, Scotti G, Grossi A. Three dimensional time-of-flight magnetic resonance angiography in carotid artery surgery: a comparison with digital subtraction angiography. Eur J Vasc Surg. 1993;7:171176.[Medline] [Order article via Infotrieve]
8.
Anderson CM, Lee RE, Leven DL, de la Torre Alonso S,
Saloner D. Measurement of internal carotid artery stenosis from
source MR angiograms. Radiology. 1994;193:219226.
9.
De Marco JK, Nesbit GM, Wesbey GE, Richardson D.
Prospective evaluation of extracranial stenosis: MR angiography
with maximum-intensity projections and multiplanar reformation
compared with conventional angiography. AJR Am J
Roentgenol. 1994;163:12051212.
10. Huston J, Nichols DA, Luetmer PH, Rydberg CH, Lewis BD, Meyer FB, Brown RD, Schleck CD. MR angiographic and sonographic indications for endarterectomy. AJNR Am J Neuroradiol. 1998;19:309315.[Abstract]
11. Wildy KS, Yuan C, Tsuruda JS, Ferguson MS, Wen N, Subramaniam DS, Strandness DE Jr. Atherosclerosis of carotid artery: evaluation by magnetic resonance angiography. J Magn Reson Imaging. 1996;6:726732.[Medline] [Order article via Infotrieve]
12. Gussenhoven EJ, Essed CE, Lancee CT, Mastik F, Frietman P, van Egmond FC, Reiber J, Bosch H, van Urk H, Roelandt J, Bom N. Arterial wall characteristics determined by intravascular ultrasound imaging: an in vitro study. J Am Coll Cardiol. 1989;14:947952.[Abstract]
13.
Potkin BN, Bartorelli AL, Gessert JM, Neville RF,
Almagor Y, Roberts WC, Leon MB. Coronary artery imaging with
intravascular high-frequency ultrasound. Circulation. 1990;81:15751585.
14. Hodgson JM, Reddy KG, Suneja R, Nair RN, Lesnefsky EJ, Sheehan HM. Intracoronary ultrasound imaging: correlation of plaque morphology with angiography, clinical syndrome and procedural results in patients undergoing coronary angioplasty. J Am Coll Cardiol. 1993;21:3544.[Abstract]
15. Vanninen RL, Manninen HI, Koivisto K, Tulla H, Partanen K, Puranen M. Carotid stenosis by digital subtraction angiography: reproducibility of the European Carotid Surgery Trial and the North American Symptomatic Carotid Endarterectomy Trial measurement methods and visual interpretation. AJNR Am J Neuroraadiol. 1994;15:16351641.
16.
Toussaint J-F, la Muraglia GM, Southern JF, Fuster V,
Kantor HL. Magnetic resonance images: lipid, fibrous, calcified,
hemorrhagic, and thrombotic components of human
atherosclerosis in vivo. Circulation. 1996;94:932938.
17. Bluth EI. Evaluation and characterization of carotid plaque. Semin Ultrasound CT MR. 1997;18:5765.[Medline] [Order article via Infotrieve]
18.
Beletsky VY, Kelley RE, Fowler M, Phifer T. Ultrasound
densitometric analysis of carotid plaque composition.
Stroke. 1996;27:21732177.
19. Anderson CM, Saloner D, Lee RE, Griswold VJ, Shaperoo LG, Rapp JH, Nagarkar S, Pan X, Gooding GA. Assessment of carotid artery stenosis by MR angiography: comparison with x-ray angiography and color-coded Doppler ultrasound. AJNR Am J Neuroradiol. 1992;13:9891003.[Abstract]
20.
Polak JF, Kalina P, Donaldson MC, O'Leary DH,
Whittemore AD, Mannick JA. Carotid endarterectomy:
preoperative evaluation of candidates with combined Doppler
sonography and MR angiography: work in progress. Radiology. 1993;186:333338.
21.
Cumming MJ, Morrow IM. Carotid artery stenosis:
a prospective comparison of CT angiography and conventional
angiography. AJR Am J Roentgenol. 1994;163:517523.
22. Link J, Brossmann J, Grabener M, Mueller-Huelsbeck S, Steffens JC, Brinkmann G, Heller M. Spiral CT angiography and selective digital subtraction angiography of internal carotid artery stenosis. AJNR Am J Neuroradiol. 1996;17:8994.[Abstract]
23. Yuan C, Tsuruda JS, Beach KN, Hayes CE, Ferguson MS, Alpers CE, Foo TK, Strandness DE. Techniques for high resolution MR imaging of atherosclerotic plaque. J Magn Reson Imaging. 1994;4:4349.[Medline] [Order article via Infotrieve]
24. Görtler M, Goldman A, Mohr W, Widder B. Tissue characterization of atherosclerotic carotid plaques by MRI. Neuroradiology. 1995;27:212215.
25. Dagirmanjian A, Ross JS, Obuchowski N, Lewin JS, Tkach JA, Ruggieri PM, Masaryk TJ. High resolution, magnetization transfer saturation, variable flip angle, time-of flight MRA in detection of intracranial vascular stenoses. J Comput Assist Tomogr. 1995;19:700706.[Medline] [Order article via Infotrieve]
26. Cloft HJ, Murphy KJ, Prince MR, Brunberg JA. 3D gadolinium-enhanced MR angiography of the carotid arteries. Magn Reson Imaging. 1996;14:593600.[Medline] [Order article via Infotrieve]
27. Hermiller JB, Buller CE, Tenaglia AN, Kisslo KB, Phillips HR, Bashore TM, Stack RS, Davidson CJ. Unrecognized left main coronary artery disease in patients undergoing interventional procedures. Am J Cardiol. 1993;71:173176.[Medline] [Order article via Infotrieve]
28. Von Ingersleben G, Schmiedl UP, Hatsukami TS, Nelson JA, Subramaniam DS, Ferguson MS, Yuan C. Characterization of atherosclerotic plaques at carotid bifurcation: correlation of high-resolution MR imaging with histologic analysis: preliminary study. Radiographics. 1997;17:14171423.[Abstract]
29. Scwartz CJ, Valente AJ, Sprague EA, Kelley JL, Nerem RM. The pathogenesis of atherosclerosis: an overview. Clin Cardiol. 1991;14(suppl 2):I1I16.
30. Davies MJ, Thomas AC. Thrombosis and acute coronary artery lesions in sudden cardiac ischemic death. N Engl J Med. 1984;310:11371140.[Abstract]
31.
Levin DC, Fallon JT. Significance of the angiographic
morphology of localized coronary stenoses:
histopathologic correlations. Circulation. 1982;66:316320.
32. O'Holleran LW, Kennelly MM, McClurken M, Johnson JM. Natural history of asymptomatic carotid plaque: five-year follow-up study. Am J Surg. 1987;154:659662.[Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
R. E. Latchaw, M. J. Alberts, M. H. Lev, J. J. Connors, R. E. Harbaugh, R. T. Higashida, R. Hobson, C. S. Kidwell, W. J. Koroshetz, V. Mathews, et al. Recommendations for Imaging of Acute Ischemic Stroke: A Scientific Statement From the American Heart Association Stroke, November 1, 2009; 40(11): 3646 - 3678. [Full Text] [PDF] |
||||
![]() |
D Mitra, D Connolly, S Jenkins, P English, D Birchall, C Mandel, K Shrikanth, B Gregson, and A Gholkar Comparison of image quality, diagnostic confidence and interobserver variability in contrast enhanced MR angiography and 2D time of flight angiography in evaluation of carotid stenosis. Br. J. Radiol., March 1, 2006; 79(939): 201 - 207. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.-M. Serfaty, E. Atalar, J. Declerck, P. Karmakar, H. H. Quick, K. A. Shunk, A. W. Heldman, and X. Yang Real-time Projection MR Angiography: Feasibility Study Radiology, October 1, 2000; 217(1): 290 - 295. [Abstract] [Full Text] |
||||
![]() |
T. S. Hatsukami, R. Ross, N. L. Polissar, and C. Yuan Visualization of Fibrous Cap Thickness and Rupture in Human Atherosclerotic Carotid Plaque In Vivo With High-Resolution Magnetic Resonance Imaging Circulation, August 29, 2000; 102(9): 959 - 964. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Stroke Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1999 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |