(Stroke. 1999;30:1855-1861.)
© 1999 American Heart Association, Inc.
Original Contributions |
From the Klinik Berlin, Department Neurological Rehabilitation, Free University Berlin (Germany) (S.H., C.W., K.M.); Disablement Services Center, Addenbrookes Hospital NHS Trust, Cambridge, UK (K.S.); and Department of Physical Medicine and Rehabilitation, University of Bucharest (Romania) (M.B.).
| Abstract |
|---|
|
|
|---|
MethodsTwenty-one hemiparetic subjects who had been using an ankle-foot orthosis for equinovarus deformity for <1 week participated. Patients walked cued by a metronome at a comparable speed with and without the orthosis. Dependent variables were basic, limb-dependent cycle parameters, gait symmetry, vertical ground reaction forces, sagittal ankle excursions, and kinesiological electromyogram of several lower limb muscles.
ResultsThe use of the caliper was associated with more dynamic and balanced gait, characterized by longer relative single-stance duration of the paretic lower limb, better swing symmetry, better pivoting over the stationary paretic foot, and better ankle excursions (P<0.05). The functional activity of the paretic quadriceps muscles increased, while the activity of the paretic tibialis anterior muscle decreased (P<0.05).
ConclusionsThe orthosis led to a more dynamic and balanced gait, with enhanced functional activation of the hemiparetic vastus lateralis muscle. The study further supports the functional benefits of a rigid ankle-foot orthosis in hemiparetic subjects as an integral part of a comprehensive rehabilitation approach. However, the reduced activity in the tibialis muscle may lead to disuse atrophy and hence long-term dependence on the orthosis.
Key Words: gait hemiplegia motor activity orthotic devices rehabilitation
| Introduction |
|---|
|
|
|---|
With an AFO, hemiparetic patients walk faster, with a longer stride, diminished plantar flexion, and larger active ankle moment, and they consume less oxygen when walking compared with walking without a caliper.1 2 3 4 5 6 7 8 Beyond that, Lehmann and coworkers, who investigated double-stopped AFOs, described how the dorsiflexion stop created a bending moment and the plantar flexion stop created an extensor moment at the knee in hemiparetic patients.3 9
The present study investigated orthosis-related effects on hemiparetic gait that could not be explained by an increase in walking velocity alone. Patients usually walk more quickly with an AFO, and walking speed affects most gait variables.10
Furthermore, the study addressed the question of the muscular activity of the lower limb muscles in hemiparetic subjects walking with an AFO and barefoot. This question is relevant to subjects' daily clinical routine since therapists often hesitate to prescribe an AFO despite the obvious functional gains. They fear that the use of an orthosis might result in untimely overactivity of the plantar flexor and nonuse of the tibialis anterior muscle.
| Subjects and Methods |
|---|
|
|
|---|
Inclusion criteria for the study were as follows: the ability to walk 20 m barefoot without physical help by a therapist; use of a Valens AFO for <1 week; minimum modified Ashworth grade of 3 (tested for ankle dorsiflexion while supine; grades 0 to 5); no obvious ankle contracture (plantigrade posture after at least 10 minutes of standing in the standing bar); and no additional orthopedic or neurological deficits impairing ambulation.
The Valens caliper (a 1-bar rigid AFO, with a medial stainless steel upright, padded, and leather-covered posterior steel calf band and an anterior soft closure, with an outside T strap correcting varus position) was attached to a firm shoe (Bally, Oxford-type shoe with a leather outer sole and rubber heel, with touch-and-loop fasteners). The caliper had a double-stopped ankle, which was set by the technician to the range of 80° (-10° dorsiflexion) to 90° (neutral position). In addition to the plantar flexion stop, dorsiflexion was assisted by a spring, the tension of which could be adjusted with a screw. The Valens caliper recommended by Davies12 is frequently prescribed in Germany and Switzerland. The Valens caliper is comparable to a thermoplastic model by biomechanical principles. A thermoplastic AFO is rigid, encasing the ankle, and therefore is a biomechanical equivalent to an anterior and posterior pin stop as provided by the Valens caliper.13
Gait Analysis
Gait velocity, cadence, and mean stride length were calculated
after patients walked 10 m at maximum speed 3 times.
Angle of ankle dorsiflexion, vertical ground reaction forces at heel-on and at toe-off, and onset and durations of stance, swing, and double support were measured by biaxial goniometers (Penny & Giles, type 180) and overshoe slippers with 8 insole force sensors and recorded at 100 Hz by a portable Datalogger worn by the patient (Infotronic System).14
Surface electromyography (EMG) was recorded from the subjects'
hemiparetic and unaffected tibialis anterior, medial head of
gastrocnemius, vastus lateralis, and gluteus medius, with the use of
8-mm self-adhesive electrodes. The impedance was checked and kept below
5 k
. These signals (1000-Hz sampling rate) were
recorded by the same portable Datalogger for later offline
analysis.
Limb-dependent cycle parameters (stance, swing, and double-support durations), vertical ground reaction forces at heel-on and at toe-off (Fz1 and Fz2), trajectories of the force point of action under both feet, and the maximum ankle plantar flexion and dorsiflexion of the affected side were averaged over at least 15 strides. The limb-dependent cycle parameters were normalized with respect to the gait cycle. Symmetry ratios for stance, swing, and double-support durations (duration of the shorter side divided by that of the longer side) were calculated. The lengths of the trajectories of the force point of action under both feet were given as a percentage of the anatomic foot length (referred to as "gait line"). Furthermore, the loading rate of the first force crest was assessed from zero intercept to 80% of Fz1. Additionally, the mode of initial contact of the affected leg was noted (either forefoot, midfoot, or heel).
The EMG data were digitally filtered (bandpass, 10 to 300 Hz), rectified, averaged over at least 10 strides, and time-normalized to the mean cycle duration set to 100%. To quantify the physiological, functional activities of the recorded muscles, mean values of the nonlow-pass signals were calculated in the following intervals of the cycle duration, set to 100% (in accordance with the adult muscle phasic activity chart, Shriner's Hospital, San Francisco, Calif): from 60% to 100% (tibialis anterior), from 20% to 50% (gastrocnemius), from 90% to 120% (vastus lateralis), and from 90% to 140% (gluteus medius). For the gastrocnemius, so-called premature activity was calculated in the interval from 90% to 110%. Two previous kinematic EMG studies of equinovarus after stroke revealed premature calf muscle activation in the terminal swing as an important cause of excessive plantar flexion and varus in stroke patients.15 16 All data were gathered within 1 session, so that the comparison of absolute EMG activities within subjects was appropriate if stable recording conditions were assumed.
Experimental Procedure
The gait of all subjects was assessed (1) when subjects were
walking barefoot and (2) when subjects were using the Valens caliper
and a firm shoe. Any necessary use of a cane (in 5 subjects) was kept
constant.
All patients walked cued by a metronome at a comparable walking velocity during both experimental conditions to eliminate the effect of changes in gait speed, so that the effect of the orthosis alone could be analyzed. Before analysis, patients were able to become accustomed to the experimental protocol, and subjects chose a speed at which they could walk with or without the orthosis.
Statistical Analysis
Data were analyzed with the help of a
multivariate profile analysis. EMG activity was
logarithmically transformed because our own previous experience had
shown that the distribution was more normal and variances more uniform
after this transformation. Walking without and with orthosis was
compared across each type of dependent variable. Additionally, to
make analysis more comprehensive, certain types of
variables were analyzed simultaneously on a
multivariate basis, namely, basic and limb-dependent
cycle parameters, symmetry ratios, vertical ground reaction
forces, and muscles of affected and nonaffected sides. In the case of a
significant multivariate result (
=0.05), the
univariate tests for the different types of variables
were additionally considered.
| Results |
|---|
|
|
|---|
Limb-Dependent Cycle Parameters
The use of the caliper resulted in a significant increase
(+31.6%) of the relative single-stance period of the affected lower
limb (swing not affected) and of the relative terminal double-support
duration (-21.1%) with the paretic limb behind. The remaining
parameters (stance affected, stance not affected, swing
affected, and initial double-support times) did not change
significantly (Table 1
).
|
Symmetry Ratios
Swing symmetry improved with the orthosis (+19.1%), while the
stance and double-support symmetry remained unchanged (Table 1
).
Vertical Ground Reaction Forces
The gait line of the affected lower limb increased with caliper
use (+44.2%), and the loading rate of the nonparetic limb was greater
(+30.4%) (Figure 1
). The peak forces,
gait line of the nonaffected lower limb, and loading rate of the
affected lower limb did not change (Table 1
).
|
Ankle Excursions
When the subjects were wearing the orthosis, the ankle
dorsiflexion during stance became larger (+201.2%) and the plantar
flexion during swing became less (-71.2%).
Muscular Activity
Paretic tibialis anterior activity decreased by -48.0% (±24.2)
and the affected vastus lateralis increased by +35.0% (±39.1) with
caliper use (Figure 2
). The activity of
the remaining recorded muscles did not change significantly (Table 2
). The same applied to so-called
premature activity (90% to 110% of gait cycle) of the plantar flexors
when subjects walked with the caliper (although 7 of the 21 patients
exhibited less premature activity with the AFO) (Figure 3
).
|
|
|
| Discussion |
|---|
|
|
|---|
The use of the orthosis reduced double-stance duration and increased single-stance duration on the hemiparetic leg. This was also reflected in an increase in gait symmetry with the AFO. Obviously, the brace could provide a feeling of security, thereby reducing the subjects' need to quickly shift weight to the nonparetic limb.
In confirmation of previous reports,1 2 3 4 5 patients wearing the AFO cleared their toes better during swing and pivoted over the stationary foot better. The corresponding changes of the ankle angles were characterized by less plantar flexion during the swing (providing better toe clearance) and a larger dorsiflexion during the stance phase. These effects are mainly attributable to the mechanical properties of the shoes and the preset ankle excursions allowed by the double-stopped brace. A larger flexion of the knee and hip might have also contributed to the observed better toe clearance (as described by Lehmann et al3 ), but this was not assessed within the present study.
The use of the orthosis changed the muscle activity of the lower limbs. When the patients were wearing the AFO, the activity of the tibialis anterior of the affected side became less, confirming therapists' notion of diminished activity of the ankle dorsiflexor during swing. The brace with its plantar flexor stop probably reduced the patients' efforts to clear the toe during the swing. In the long term, this might render the patients' muscles inactive and might therefore prolong the patients' dependency on the mechanical device. On the other hand, therapists may aim at suppression of an abnormal flexor synergy in hemiparetic subjects triggered by the swing effort.12 Within this synergy, the tibialis anterior muscle is responsible for an unwanted foot inversion due to its medial insertion with respect to the subtalar joint.15 The resulting varus deformity impedes the initial contact and weight acceptance, with a risk of spraining of the ankle.
In addition, patients wearing the AFO exhibited a larger activity of the quadriceps muscle during the early stance, when it helps with weight acceptance, and during midstance, when it helps with load bearing.17 The main peripheral drives for this facilitation were most likely the aforementioned faster loading of the paretic limb and its prolongation of the relative single-stance period. Accordingly, animal and human studies showed that loading of the paretic limb had a potent facilitatory effect on the antigravity muscles.18 19
On the other hand, there was no overactivity of the plantar flexors, particularly not of the so-called premature activity (90% to 110% of the gait cycle) occurring during the terminal swing and initial stance. Two previous kinematic EMG studies of equinovarus after stroke revealed premature calf muscle activation in the terminal swing as an important cause of excessive plantar flexion and varus in stroke patients.15 16 The present result of no overall change of plantar flexor activity therefore does not confirm therapists' fear of increased muscle tone because of the caliper. The marked premature gastrocnemius activity of some patients was even reduced when they were wearing the AFO, as previously noted.15 This may have been due to a reduction in calf muscle stretch when the first contact with the ground is at the heel instead of at the forefoot.
The swing symmetry improved when the subjects used an AFO. In this context, one should keep in mind that a balanced gait is a hallmark of healthy subjects. "Neurodevelopmental" or Bobath therapists, aiming to restore a symmetrical gait pattern, often hesitate to prescribe an "artificial" caliper under this premise. A large outcome study on 156 hemiparetic subjects, however, failed to show any improvement of gait symmetry during a 4-week comprehensive rehabilitation program following the neurodevelopmental technique.20
Potential confounding factors were the use of a cane in 5 subjects and the firm shoes worn by all subjects together with the AFO. According to 2 previous studies, the gait of hemiparetic subjects walking with and without a cane, however, was not shown to differ significantly with respect to gait symmetry and the kinesiological EMG of thigh and shank muscles of the affected side.21 22 With regard to the potential effect of shoes, one study compared walking barefoot and with firm shoes. The use of shoes merely improved the gait line of the nonaffected limb, while the gait line of the nonaffected lower limb, relative cycle parameters, and symmetry ratios were comparable in both conditions.5 Ankle excursions and muscular activity, however, were not assessed.
In conclusion, the use of the caliper effected a more dynamic and balanced gait irrespective of any change of gait speed. Gait was characterized by better ankle excursions, faster loading of the paretic limb, longer relative single-stance duration of the paretic lower limb, better swing symmetry, and better pivoting over the stationary paretic foot. Furthermore, the functional activity of the paretic quadriceps muscles increased, while there was no overactivity of the plantar flexors. These effects are predominantly beneficial and support the correction of spastic ankle plantar flexion and inversion with the Valens caliper as an integral part of a comprehensive rehabilitation approach; however, the long-term effects, particularly on the hemiparetic tibialis anterior, need further study.
| Acknowledgments |
|---|
| Footnotes |
|---|
Received December 15, 1998; revision received May 3, 1999; accepted May 3, 1999.
| References |
|---|
|
|
|---|
2. Corcoran PJ, Jebsen RH, Brengelmann GL, Simons BC. Effects of plastic metal leg braces on speed and energy cost of hemiparetic ambulation. Arch Phys Med Rehabil. 1970;51:6977.[Medline] [Order article via Infotrieve]
3. Lehmann JF, Condon SM, Price R, de Lateur BJ. Gait abnormalities in hemiplegia: their correction by ankle-foot orthoses. Arch Phys Med Rehabil. 1987;68:763771.[Medline] [Order article via Infotrieve]
4. Wong AMK, Fuk-Tan T, Shin-Han W, Chang-Ming C. Clinical trial of a low temperature plastic anterior ankle-foot orthosis. Am J Phys Med Rehabil. 1992;71:4143.[Medline] [Order article via Infotrieve]
5. Hesse S, Luecke D, Jahnke MT, Mauritz KH. Gait function in spastic hemiparetic patients walking barefoot, with firm shoes, and with ankle-foot orthosis. Int J Rehabil Res. 1996;19:133141.[Medline] [Order article via Infotrieve]
6. Burdett RG, Borello-France D, Blatchly C, Potter C. Gait comparison of subjects with hemiplegia walking unbraced, with ankle-foot orthosis, and with Air-Stirrup brace. Phys Ther. 1988;68:11971203.
7. Miyazaki S, Yamamoto S, Kubota T. Effect of ankle-foot orthosis on active moment in patients with hemiparesis. Med Biol Eng Comput. 1997;35:381385.[Medline] [Order article via Infotrieve]
8. Lehmann JF. Biomechanics of ankle-foot orthosis: prescription and design. Arch Phys Med Rehabil. 1979;60:200207.[Medline] [Order article via Infotrieve]
9. Lehmann JF. Push-off and propulsion of the body in normal and abnormal gait. Clin Orthop. 1993;288:97108.
10. Murray MP, Mollinger LA, Gardner GM, Sepic SB. Kinematic and EMG patterns during slow, free, and fast walking. J Orthop Res. 1984;2:272280.[Medline] [Order article via Infotrieve]
11. Bohannon RW, Smith MB. Interrater reliability of a modified Ashworth spastic scale of muscle spasticity. Phys Ther. 1987;67:206207.
12. Davies PM. Steps to Follow. Berlin, Germany: Springer-Verlag; 1985.
13. Lehmann JF. Lower limb orthotics. In: Redford JB, ed. Orthotics Etcetera. 3rd ed. Baltimore, Md: Williams & Wilkins; 1986:278351.
14. Hermens HJ, de Waal CA, Buurke J, Zilvod G. A new gait analysis system for clinical use in a rehabilitation center. Orthopedics.. 1986;9:16691675.[Medline] [Order article via Infotrieve]
15. Perry J, Walters RL, Perrin T. Electromyographic analysis of equinovarus following stroke. Clin Orthop. 1978;131:4753.
16. Knuttsen E, Richards C. Different types of disturbed motor control in gait of hemiparetic patients. Brain. 1979;102:404430.
17. Perry J. Gait Analysis. Thorofare, NJ: Slack Inc; 1992.
18. Duysens J, Pearson KG. Inhibition of flexor burst generation by loading ankle extensor muscles in walking cats. Brain Res. 1980;187:321332.[Medline] [Order article via Infotrieve]
19. Barbeau H, Rossignol S. Enhancement of locomotor recovery following spinal cord injury. Curr Opin Neurol. 1994;7:517524.[Medline] [Order article via Infotrieve]
20. Hesse S, Jahnke MT, Bertelt C, Schreiner C, Luecke D, Mauritz KH. Gait outcome in ambulatory hemiparetic patients after a 4-week comprehensive rehabilitation program and prognostic factors. Stroke. 1994;25:19992004.[Abstract]
21. Tyson SF, Ashburn A. The influence of walking aids on hemiplegic gait. Physiother Theory Pract. 1994;10:7786.
22. Hesse S, Jahnke MT, Schaffrin A, Luecke D, Reiter F, Konrad M. Immediate effects of therapeutic facilitation on the gait of hemiparetic patients as compared with walking with and without a cane. Electroencephalogr Clin Neurophysiol. 1998;109:515522.[Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
M. Pohl and J. Mehrholz Immediate effects of an individually designed functional ankle-foot orthosis on stance and gait in hemiparetic patients Clinical Rehabilitation, April 1, 2006; 20(4): 324 - 330. [Abstract] [PDF] |
||||
![]() |
M. Franceschini, M. Massucci, L. Ferrari, M. Agosti, and C. Paroli Effects of an ankle-foot orthosis on spatiotemporal parameters and energy cost of hemiparetic gait Clinical Rehabilitation, April 1, 2003; 17(4): 368 - 372. [Abstract] [PDF] |
||||
![]() |
W. M. Landau and S. A. Sahrmann Preservation of Directly Stimulated Muscle Strength in Hemiplegia Due to Stroke Arch Neurol, September 1, 2002; 59(9): 1453 - 1457. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. T Wade Rehabilitation is a way of thinking, not a way of doing Clinical Rehabilitation, June 1, 2002; 16(6): 579 - 581. [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Stroke Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1999 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |