(Stroke. 2000;31:151.)
© 2000 American Heart Association, Inc.
Original Contributions |
From the Department of Neurology, Medical University Lübeck, Lübeck, Germany.
Correspondence to Günter Seidel, MD, Department of Neurology, Medical University at Lübeck, Ratzeburger Allee 160, D-23538 Lübeck, Germany. E-mail seidel_g{at}neuro.mu-luebeck.de
| Abstract |
|---|
|
|
|---|
MethodsIn 12 healthy volunteers, 2 doses (0.5 and 1.5 mL) of Optison, a perfluoropropane-containing contrast agent, were injected intravenously and produced a strong increase of brightness in the brain parenchyma. We used harmonic imaging for quantification of ultrasound intensity in the thalamus, ipsilateral temporoparietal white matter (TPWM), and ipsilateral lateral fissure at both sides. Time-intensity curves were calculated, and peak increase (PI) of intensity and the area under the time-intensity curve (AUC) from baseline were compared.
ResultsWe found a significant dose dependence of the AUC in all regions at both sides. PI only showed a significant dose dependence in the TPWM but not in the ipsilateral thalamus and lateral fissure. No side differences for AUC and PI were detected in all regions and doses used. We found a significantly higher value of the PI insonating the thalamus from the ipsilateral side compared with the contralateral side. The same result was obtained for the AUC in the left thalamus for both doses and in the right thalamus for the high dose. Using 0.5 mL for insonation of the right thalamus AUC again showed a higher value for the insonation from the ipsilateral compared with the contralateral side but failed to show statistical significance (P=0.08, n=12).
ConclusionsHarmonic gray-scale imaging with Optison showed a strong enhancement effect in the brain parenchyma. A quantitative analysis of perfusion seems difficult because of the depth dependence of the effect. The most robust parameter is the AUC.
Key Words: contrast media perfluorocarbons perfusion ultrasonography
| Introduction |
|---|
|
|
|---|
The purpose of this human study was to investigate the contrast-enhancing effects of a new perfluoropropane-based UCA (Optison; Mallinckrodt) and to evaluate harmonic gray-scale imaging with respect to its potential for detection of contrast agent in the microcirculation of the brain. This study is of particular interest for those designing further investigations planning the visualization of perfusion defects in patients with acute ischemic stroke by means of ultrasound methods.
| Subjects and Methods |
|---|
|
|
|---|
Ultrasound Contrast Agent
Optison (provided by Mallinckrodt Medical GmbH, Hennef,
Germany) is a perfluoropropane-containing UCA based on a 1%
albumin solution.5 Octafluoropropane is chemically
characterized with a molecular weight of 188 and an empirical formula
of C3F8. The
microspheres in Optison injectable suspension are produced by
heat treatment and sonication of appropriately diluted human
albumin (USP) in the presence of octafluoropropane gas by a
proprietary process. The protein in the microsphere shell makes
up
5% to 7% of the total protein in the liquid. Each 1 mL of
Optison contains 5.0 to 8.0x108 human
albumin microspheres, 10 mg albumin human
(USP), 0.22±0.11 mg/mL octafluoropropane, 0.2 mg
N-acetyltryptophan, and 0.12 mg caprylic acid in 0.9%
aqueous sodium chloride. The headspace of the vial is filled with
octafluoropropane gas. The pH is adjusted with sodium hydroxide to 6.4
to 7.4. The microsphere particle size parameters
are mean diameter (range) 2.0 to 4.5 µm (maximum, 32.0
µm), 93% less than 10 µm. Octafluoropropane is a stable gas
that is not metabolized. The human albumin component of the
microsphere is expected to be handled by the normal
metabolic routes for human albumin. The UCA is
commercially available and was originally developed for
echocardiology (Generic FS069, Mallinckrodt Inc). The
solution was prepared following the manufacturers instructions. Two
intravenous bolus injections of 0.5 and 1.5 mL (injection
speed, 1 mL/s) were used. Each injection was followed immediately by a
second bolus of 3 mL of 0.9% NaCl solution to ensure clearance of the
residual UCA in the venous system. The time between 2 UCA bolus
injections was 5 to 10 minutes.
Transcranial Sonography
Harmonic gray-scale imaging was performed with an HP SONOS 5500
ultrasound system (Hewlett Packard) connected to a 1.8/3.6-MHz sector
transducer (S4 probe, Hewlett Packard) in an investigation depth of 10
cm (focus on 8 cm). For gray-scale imaging, we used the integrated
backscatter (IBS) mode and the study type T-INT (mechanical index, 1.0
to 1.1).
After each UCA injection, 62 digitized gray-scale images of the brain triggered by ECG were stored in continuous-loop-review memory and then recorded on an optical disk for later offline analysis. We used the transient response imaging mode6 with a frame rate of 1 image every 4 cardiac cycles. Gain and transmit power setting were optimized for each volunteer at the beginning of each investigation and were not changed throughout the procedure. The entire investigation was also recorded on videotape.
Harmonic Gray-Scale Imaging
For analysis of harmonic gray-scale data, the IBS
of brain tissue was measured offline with the acoustic densitometry
unit of the HP SONOS 5500. This unit assists in the quantification of
ultrasound images by measuring the scattered energy received by the
transducer. Because acoustic densitometry measurement is made upstream
in the imaging chain, it is less influenced by postprocessing functions
of the imaging chain. The IBS is a relative measure of the total
ultrasound energy scattered by a small volume of the interrogated
tissue. The IBS data measurements were displayed on a logarithmic scale
in decibels.7 8 We specified the regions of
interest (ROIs) to the thalamus region at both sides, ipsilateral
temporoparietal white matter, and ipsilateral lateral fissure, where
branches of the middle cerebral artery are located. Identification of
the anatomic sites has been described previously.9 The
sample volume of the ROIs was 21x21 pixels. The mean IBS in the ROIs
of the first 2 images served as baseline reference (noise floor). The
change in the IBS in the seconds after UCA injection was measured, and
the mean values were displayed graphically.
In this study, we compared the area under curve (AUC) and the peak increase (PI) from baseline in the brain parenchyma for increasing doses of Optison using a nonparametric test for related samples (Friedman ANOVA test). Baseline for each ROI was the mean value from the first 2 acoustic densitometry values after injection.
| Results |
|---|
|
|
|---|
|
|
With a maximum insonation depth of 10 cm, we could image the
ipsilateral and contralateral thalamus. In the healthy young volunteers
under examination, we assume similar perfusion in the thalamus at both
sides. In Figure 2
, washout curves for
the 2 doses of Optison in the ipsilateral and contralateral thalamus
are displayed to determine the depth dependence of UCA detection and of
the parameter characterizing the washout curves (PI and
AUC).
|
For 0.5 and 1.5 mL, we found a significantly higher value of the
PI insonating the thalamus from the ipsilateral compared with
contralateral side (Table 2
). We found
the same result for the AUC in the left thalamus for both doses and in
the right thalamus for the high dose. With 0.5 mL used for insonation
of the right thalamus, the AUC again showed a higher value for the
insonation from the ipsilateral compared with contralateral side but
failed to show statistical significance (P=0.08, n=12).
|
Ultrasound examinations and injections of UCAs were generally well tolerated. One subject reported a mild, transient headache after the investigation; another volunteer experienced a mild burning sensation in the right upper abdominal quadrant that resolved completely within an hour. No abnormalities in physical state or blood chemical values were noted on follow-up examination.
| Discussion |
|---|
|
|
|---|
Analyzing the PI, we found a significant dose dependence in only the temporoparietal white matter but not in the ipsilateral thalamus and lateral fissure; there were no side differences for all regions and doses. This effect could be explained by a nonlinear relationship between the high microbubble concentration in the tissue with higher capillary or artery density (thalamus/lateral fissure) and a PI in the IBS measured by the ultrasound system.10
For comparison of the washout curves of the thalamus at each side with ipsilateral and contralateral insonation, we found significant differences, indicating a significant decrease in signal intensity in investigations of similar structures at different insonation depths. Because of this significant depth dependence of the enhancing effect after UCA injection, a quantitative analysis of brain perfusion seems to be impossible through analysis of washout curves without a method for correction of this effect.
For qualitative visualization of brain perfusion, gray-scale imaging seems an appropriate imaging mode because of the homogenous echo pattern of the brain obtained after UCA injection. A comparison of our data with other human studies that used gray-scale harmonic imaging for the analysis of brain perfusion2 3 4 indicates that Optison has a reliable and stronger contrast-enhancing effect with less depth-dependent decrease in echo enhancement.
Ultrasound examinations and Optison injections were generally well tolerated. There were only minor adverse effects in 2 subjects, and no abnormalities in physical state or blood chemical values were noted on follow-up examination. These side effects are comparable with those published for Levovist.11
In conclusion, harmonic imaging is a useful technique for visualizing brain perfusion. Harmonic gray-scale imaging with Optison showed a strong enhancement effect in the parenchyma. A quantitative analysis of perfusion seems difficult because of the depth dependence of the effect. The most robust parameter is the intensityheart cycle AUC for harmonic gray-scale imaging.
This study indicates that it is possible to visualize and measure changes in ultrasound intensities in perfused areas of the brain through the intact skull. This observation is encouraging for further studies evaluating brain perfusion in patients with acute brain infarctions.
| Acknowledgments |
|---|
Received August 6, 1999; revision received September 27, 1999; accepted October 13, 1999.
| References |
|---|
|
|
|---|
2.
Postert T, Muhs A, Meves S, Federlein J, Przuntek H,
Büttner T. Transient response harmonic imaging: an ultrasound
technique related to brain perfusion. Stroke. 1998;29:19011907.
3. Seidel G, Greis C, Sonne J, Kaps M. Harmonic grey scale imaging of the human brain. J Neuroimaging. 1999;9:171174.[Medline] [Order article via Infotrieve]
4.
Postert T, Federlein J, Weber S, Przuntek H,
Büttner T. Second harmonic imaging in acute middle cerebral
artery infarction: preliminary results. Stroke. 1999;30:17021706.
5.
Hancock J, Dittrich H, Jewitt DE, Monaghan MJ.
Evaluation of myocardial, hepatic, and renal perfusion in a variety of
clinical conditions using an intravenous ultrasound
contrast agent (Optison) and second harmonic imaging. Heart. 1999;81:636641.
6.
Porter TR, Xie F. Transient myocardial contrast after
exposure to diagnostic ultrasound pressures with minute
doses of intravenously injected microbubbles: demonstration
and potential mechanisms. Circulation. 1995;92:23912395.
7. Acoustic densitometry: white paper. Boeblingen, Germany: Hewlett Packard GmbH; 1996.
8. Masuyama T, Nellessen U, Schnittger I, Tye T, Haskell WL, Popp RL. Ultrasonic tissue characterization with real time integrated backscatter imaging system in normal and aging human hearts. J Am Coll Cardiol. 1989;14:17021708.[Abstract]
9.
Bogdahn U, Becker G, Winkler J, Greiner K, Perez J,
Meurers B. Transcranial color-coded real-time sonography in
adults. Stroke. 1990;21:16801688.
10. Skyba DM, Camarano G, Goodman NC, Price RJ, Skalak TC, Kaul S. Hemodynamic characteristics, myocardial kinetics and microvascular rheology of FS-069, a second generation echocardiographic contrast agent capable of producing myocardial opacification from a venous injection. J Am Coll Cardiol. 1996;28:12921300.[Abstract]
11.
Otis S, Rush M, Boyajian R. Contrast-enhanced
transcranial imaging: results of an American phase-two
study. Stroke. 1995;26:203209.
This article has been cited by other articles:
![]() |
E. Bartels and H.-J. Bittermann Transcranial Contrast Imaging of Cerebral Perfusion in Patients With Space-Occupying Intracranial Lesions J. Ultrasound Med., April 1, 2006; 25(4): 499 - 507. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Eyding, W. Wilkening, C. Krogias, T. Holscher, H. Przuntek, S. Meves, and T. Postert Validation of the Depletion Kinetic in Semiquantitative Ultrasonographic Cerebral Perfusion Imaging Using 2 Different Techniques of Data Acquisition J. Ultrasound Med., August 1, 2004; 23(8): 1035 - 1040. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Kern, F. Perren, K. Schoeneberger, A. Gass, M. Hennerici, and S. Meairs Ultrasound Microbubble Destruction Imaging in Acute Middle Cerebral Artery Stroke Stroke, July 1, 2004; 35(7): 1665 - 1670. [Abstract] [Full Text] [PDF] |
||||
![]() |
J Eyding, C Krogias, W Wilkening, and T Postert Detection of cerebral perfusion abnormalities in acute stroke using phase inversion harmonic imaging (PIHI): preliminary results J. Neurol. Neurosurg. Psychiatry, June 1, 2004; 75(6): 926 - 929. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Seidel, K. Meyer-Wiethe, G. Berdien, D. Hollstein, D. Toth, and T. Aach Ultrasound Perfusion Imaging in Acute Middle Cerebral Artery Infarction Predicts Outcome Stroke, May 1, 2004; 35(5): 1107 - 1111. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Wiesmann, K. Meyer, T. Albers, and G. Seidel Parametric Perfusion Imaging With Contrast-Enhanced Ultrasound in Acute Ischemic Stroke Stroke, February 1, 2004; 35(2): 508 - 513. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. H. Meves, W. Wilkening, T. Thies, J. Eyding, T. Holscher, M. Finger, G. Schmid, H. Ermert, and T. Postert Comparison Between Echo Contrast Agent-Specific Imaging Modes and Perfusion-Weighted Magnetic Resonance Imaging for the Assessment of Brain Perfusion Stroke, October 1, 2002; 33(10): 2433 - 2437. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. U. Harrer and C. Klotzsch Second Harmonic Imaging of the Human Brain: The Practicability of Coronal Insonation Planes and Alternative Perfusion Parameters Stroke, June 1, 2002; 33(6): 1530 - 1535. [Abstract] [Full Text] [PDF] |
||||
![]() |
S.-J. Rim, H. Leong-Poi, J. R. Lindner, D. Couture, D. Ellegala, H. Mason, M. Durieux, N. F. Kassel, and S. Kaul Quantification of Cerebral Perfusion With "Real-Time" Contrast-Enhanced Ultrasound Circulation, November 20, 2001; 104(21): 2582 - 2587. [Abstract] [Full Text] [PDF] |
||||
![]() |
J Federlein, T. Postert, S Meves, S Weber, H Przuntek, and T. Buttner Ultrasonic evaluation of pathological brain perfusion in acute stroke using second harmonic imaging J. Neurol. Neurosurg. Psychiatry, November 1, 2000; 69(5): 616 - 622. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Wiesmann and G. Seidel Ultrasound Perfusion Imaging of the Human Brain Stroke, October 1, 2000; 31(10): 2421 - 2425. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Seidel, C. Algermissen, A. Christoph, T. Katzer, M. Kaps, and R. W. Baumgartner Visualization of Brain Perfusion With Harmonic Gray Scale and Power Doppler Technology : An Animal Pilot Study Editorial Comment: An Animal Pilot Study Stroke, July 1, 2000; 31(7): 1728 - 1734. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Stroke Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2000 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |