Donate Help Contact The AHA Sign In Home
American Heart Association
Stroke
Search: search_blue_button Advanced Search
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow Request Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Yang, Y.
Right arrow Articles by Lees, K. R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Yang, Y.
Right arrow Articles by Lees, K. R.
Related Collections
Right arrow Echocardiography
Right arrow Acute Cerebral Infarction
Right arrow Doppler ultrasound, Transcranial Doppler etc.
Right arrow Risk Factors for Stroke

(Stroke. 2000;31:907.)
© 2000 American Heart Association, Inc.


Original Contributions

Identification Of Echocardiographic "Smoke" in a Bench Model With Transcranial Doppler Ultrasound

Yi Yang, MD, PhD, RVT; Donald G. Grosset, MD; Qiu Li, MD Kennedy R. Lees, MD

From the Neurology Division, Department of Medicine, University of Alberta, Alberta, Canada (Y.Y., Q.L.); Department of Neurology, Southern General Hospital, Glasgow, UK (D.G.G.); and the Department of Medicine and Therapeutics, University of Glasgow, Glasgow, Scotland, UK (K.R.L.).

Correspondence to Dr Yi Yang, MD, PhD, Rm 530, HMRC, Cerebrovascular Laboratory, University of Alberta, Edmonton, Alberta, Canada T6G 2S2. E-mail yiy{at}ualberta.ca


*    Abstract
up arrowTop
*Abstract
down arrowIntroduction
down arrowMethods
down arrowResults
down arrowDiscussion
down arrowReferences
 
Background and Purpose—Spontaneous echo contrast in cardiac chamber has been indicated as a source of cerebral embolism. The nature of the echocardiographic smokelike signal is still not fully understood. This study was designed to regenerate spontaneous echo contrast and verify its thromboembolic characters in an in vitro model.

Methods—Spontaneous echo contrast was reproduced in an expansion chamber under low flow conditions in a close circulation system. The spontaneous echo contrast was monitored and recorded with a 2-dimensional cardiosonography system and a transcranial Doppler device. Meanwhile, clinically commonly encountered embolic materials such as whole-blood clots, platelet aggregate–rich plasma, air bubbles, and 100-mL normal saline were injected into this monitored circuit. The differentiation of spontaneous echo contrast from emboli was performed by both visual observations of the echo images and offline Doppler signal intensity analysis. Average signal intensities produced by spontaneous contrast and injection of embolic materials and saline were compared. Furthermore, the effect of Doppler-detected flow velocity on generation of spontaneous contrast was also evaluated.

Results—Spontaneous echo contrast was reproduced at low flow settings (90 to 120 mL/min) in this model. There was no significant difference in average signal intensity between the flow with spontaneous echo contrast and that without the echo (P=0.71). However, injection of embolic materials or normal saline did not generate smokelike image but caused much higher average signal intensity than the flow with spontaneous contrast (P<0.001). Injection of normal saline also increased average signal intensity.

Conclusions—Our results suggest that smokelike echo is a special echo phenomenon occurring at low flow situations and does not itself produce material capable of embolizing into the systemic circulation.


Key Words: cerebral embolism • spontaneous echo contrast • ultrasound, Doppler, transcranial


*    Introduction
up arrowTop
up arrowAbstract
*Introduction
down arrowMethods
down arrowResults
down arrowDiscussion
down arrowReferences
 
The echocardiographic appearance of spontaneous contrast, or echocardiographic "smoke," is recognized by mobile increased intensity of the blood within left and right heart chambers, great vessels, and veins,1 2 which was first described by Feigenbaum,3 using M-mode echocardiography in patients with coronary artery disease. This phenomenon is occasionally noted in the state of regional stasis of blood, such as larger left atrial or ventricle diameter,1 4 and it may disappear when contractility of the heart is improved.5

Previous clinical investigations have shown that the presence of "smoke" is associated with a greater incidence of left atrial thrombi1 4 6 7 and has been considered one of independent predictors for cardiac thrombus development after mitral valve replacement.8 To determine the incidence of smoke and its relation to the intracardiac thrombus, transesophageal and transthoracic echocardiography have been commonly used.9 10 It has been suggested that patients with left atrium and appendage smoke were at higher risk for intracardiac thrombus,9 and the latter play an important role as a cardiac source of cerebral embolism. Knowledge of the relationship between echocardiographic smoke and intracardiac thrombus and the nature of the smoke may help the management of patients with cerebral embolism when the need for anticoagulant or antiplatelet aggregatory agent is to be considered.11

While the association between ischemic stroke and cardiac thrombi and spontaneous echo contrast has been under investigation,12 13 14 15 several methods have been used to investigate the pathogenesis of spontaneous contrast.2 5 16 In the previous studies, although blood stasis,16 platelet aggregation,17 red blood cell (RBC) coagulation,18 and flow-related interaction of RBCs and plasma have been implicated in such echogenicity, the mechanism responsible for the formation of echocardiographic smoke is still not fully understood.

To understand the putative cellular elements responsible for spontaneous echocardiographic contrast in the blood pool, Turakhia et al19 recently examined, using a 7.5-MHz mechanical sector scanner, the roles of blood components, including packed RBCs, plasma, washed RBCs, and exogenous fibrinogen, and found that ultrasonic echo gray scale was an exponential function of fibrinogen and hematocrit (HCT) but was not influenced by leukocyte count, platelet count, or serum protein level. In another experimental study,20 spontaneous contrast was imaged from the rabbit heart after administration of xylazine and ketamine while RBC aggregation was prevented by a photometric method. In addition to the factors of pathological cellular components, the shear rate of circulating blood has also been considered a possible contributor to the echocardiographic smoke. To evaluate this assumption, a variety of models have been designed to mimic the dilated left atrium and examine spontaneous contrast at different flow rate and shear rate conditions.21 In an experiment by Merino et al,21 ultrasound imaging was performed with a fixed 7-MHz linear probe in a tubular chamber with circulating heparinized whole blood. It was observed that characteristic swirling waves of spontaneous contrast were seen in the circulating blood at low flow rate and low shear rate.

The ability of transcranial Doppler ultrasound (TCD) to detect embolic particles such as blood clots, platelet clots, and air bubbles has been recently confirmed both experimentally22 23 24 and clinically.22 25 26 An embolus causes a Doppler signal of increased intensity because its size and acoustic impedance are different from blood.26

In this study, using an in vitro model to generate spontaneous contrast, we aimed to (1) identify whether smokelike signals result in production of embolus signals, (2) investigate whether whole-blood clots or platelet clots or microbubbles can produce smokelike signals, and (3) evaluate the effect of the smoke state on the formation of thrombi.


*    Methods
up arrowTop
up arrowAbstract
up arrowIntroduction
*Methods
down arrowResults
down arrowDiscussion
down arrowReferences
 
Definition of Spontaneous Contrast
Echogenic smoke was defined according to the previous description2 5 of (1) low-amplitude echogenic haze; (2) slow, repetitive movement in the cavity; and (3) dissipation and disappearance of the image when blood flow increases and reappearance as flow decreases.

In Vitro Models for the Reproduction of Smokelike Echoes and TCD Evaluation
The model used to generate spontaneous contrast was introduced by Merino et al.2 It was constructed with a 28-mm-diameter plastic cylindrical chamber containing a 4-mm-diameter tube. A 5-MHz linear transducer (L518, Acuson) was fixed longitudinally to the chamber. In this study, we added a plastic tube of the same internal diameter that bypassed the expansion chamber to help us to identify the nature of smoke. A middle cerebral artery (MCA) model for TCD detection was formed with a plastic box filled with normal saline and inlaid with a piece of human temporal bone in one of the walls and a plastic tube with internal diameter of 4.5 mm. A 2-MHz TCD probe was placed on the temporal bone to record the Doppler signal produced by blood flow passing through the tube inside the MCA model. The models were connected together, and whole human blood (anticoagulated with 1:7 citrate phosphate dextrose adenine) was circulated by a constant pump, with the flow rate constantly monitored with a magnetic flowmeter (Figure 1Down).



View larger version (24K):
[in this window]
[in a new window]
 
Figure 1. Illustration of the circuit model for reproduction of spontaneous contrast and TCD examination.

Ultrasonic Signal Recording and Offline Study
Two-dimensional echography was performed with an ACUSON 128 Computed Sonography System (Acuson, Inc), and the images were recorded onto videotape for subsequent examination. Continuous Doppler monitoring was performed with a TC2000 TCD device (Nicolet). All parameters, including gain, power, and depth for both ultrasonic devices were maintained constant throughout. Echography and Doppler signals were synchronously recorded for 1 minute before and after each processing. TCD signal intensities were calculated with an intensity analysis software developed according to the validated and published method of Muller et al.27

TCD Detection During Appearance of Smokelike Echoes
The TCD signal was continuously monitored for 30 minutes. Ten TCD frames were consecutively saved for later offline intensity analysis during the period when typical smokelike echoes were imaged in the expansion chamber. At the same time, the blood flow was also allowed to travel through the bypass tube, and the same TCD manipulation was repeated.

Preparation and Injection of Embolic Materials
Whole-Blood Clots
One milliliter of 0.2 mol/L CaCl2 solution was added to 10 mL whole blood obtained from a donor and then gently shaken. This blood was incubated at 37°C for 24 hours to clot. To prepare clots with a volume of 1x1x1 mm3, 1 g of whole-blood clot was then sliced with a McIlwain tissue chopper (Mickle Laboratory Engineering) setting at a cutting distance of 1 mm. To determine the clot size and the number of clots, microscopy with a calibrated eyepiece was used to measure the clots dropped on the glass slide. The mean clot size was obtained by measuring 50 clots, and the concentration was determined by counting the number of clots in 0.1 mL blood clot–rich saline. Prepared clots were stored in 10 mL Dulbecco’s phosphate buffer.

Platelet Aggregate–Rich Plasma
9 mL fresh human blood was used for preparation of platelet aggregate-rich plasma. Ten milliliters of whole blood from a donor was anticoagulated with 0.5 mL 0.1 mol/L citrate anticoagulant and centrifuged with a refrigerated centrifuge (Sorvall RT6000B, Refrigerated Centrifuge, Du Pont) at 100g for 15 minutes to prepare platelet-rich plasma. One milliliter of platelet-rich plasma was added to 2 mL of 10 U/mL bovine thrombus (Sigma Chemical Co) and stored at 20°C for 30 minutes. Platelet-rich thrombus, containing >99% platelets, was then formed.23 After the platelet aggregate–rich plasma was formed, the same method as that used in evaluation of blood clots was applied to determine the mean size and concentration of platelet clots. We assumed that whole-blood and platelet clots used in this study were cubic.

Air Bubbles
Microbubbles were prepared in 15 mL of Ultravist 370, a nonionic contrast medium (Schering Health Care Ltd) by stirring with a polytron for 40 seconds with power set at 10 W. With this setting, microbubbles with a diameter of approximately 30 µm were generated. The size of prepared microbubbles was further controlled by transferring into a size control set. This set was constructed with a 20-mL plastic syringe, which was inlaid by 2 microfilter membranes, the first with a pore size of 40 µm (AF-10250 arterial blood filter, Baxter Healthcare) and the second with a pore size of 30 µm (AN 3H, Millipore). A side port was located between the 2 microfilter membranes. Once the agitated, bubble-rich medium was transferred into the set through the bottom of the syringe, certain prompting pressure was given manually. This allowed microbubbles with a diameter of >40 µm to stay outside the first filter and those with a diameter of <30 µm to pass through the second filter. Thus, the filtrate containing microbubbles with diameters ranging from 30 µm to 40 µm stayed between the 2 microfilters and were obtained through a side port of this set. A blood cell count chamber was used to estimate the bubble concentration. The method to determine bubble size was the same as that for blood clots.

Degassed Normal Saline
A 100-mL infusion of normal saline was degassed in the sonicator bath for 15 minutes before injection; 2.5 mL emboli-suspended solution or degassed normal saline was then manually injected into the circuit through a 3-way tap within 2 seconds.

Correlation of Smokelike Echo With Doppler Flow Velocity
In addition to echographic detection of smokelike signals, a pulsed-wave Doppler device was combined to record the Doppler flow velocity from the central part of outflow jet, the area in the presence of smokelike echo, and the area near the border of the chamber. Mean velocity was calculated from 10 peaks of the flow velocities.

Investigation of the Role of Smoke State in Thrombus Formation
A calcium chloride solution was used to supply Ca2+ for whole blood to coagulate: 0.2 mol/L CaCl2 was added into blood at the ratio of 1:20 in volume. The time required for formation of blood thrombus was recorded. We investigated the coagulation time during the following 3 conditions: (1) 270 mL blood flow passing through the cylindrical expansion chamber after the addition of 13.5 mL 0.2 mol/L CaCl2, (2) the same volume of blood flow passing through the bypass tube after the addition of the same amount of CaCl2 solution, and (3) 9.5 mL blood in a 25-mL glass beaker after the addition of 0.5 mL 0.2 mol/L CaCl2 solution. The formation of solid thrombus was confirmed with 2-dimensional echographies detected from the chamber, bypass tube, and the beaker.

The RBC concentration and HCT of sampled circulating whole blood were examined with Sysmex NE 8000 (TOA Medical Electronic Co).

Statistical Analysis
Statistical analyses were performed with 1-way analysis of variance for more than 2 groups of animals, with subsequent individual comparisons by the Scheffé test. All the data in this study were expressed as mean ±1 SD. A value of P<0.05 (2-sided) was considered statistically significant.


*    Results
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
*Results
down arrowDiscussion
down arrowReferences
 
RBC and HCT
RBC concentrations and HCTs of the circulating blood used in this study ranged from 2.97 to 4.19x1012/L and from 0.27 to 0.42, respectively.

Reproduction of Smokelike Echoes
At a flow rate of 0 mL/min, static blood in the chamber displayed the echogenic image of grainy haze with equal intensity at all areas within the chamber (Figure 2ADown). At a flow rate of 30 to 60 mL/min, blood entering the chamber showed an echolucent jet stream line (Figure 2BDown). The typical swirling smokelike images of flow lines appeared between the outflow jet stream and the static blood near the border of chamber when flow rates were 90 to 120 mL/min (Figures 2CDown and 2DDown). This spontaneous contrast also characterized as a higher echodensity than the surroundings but without definite margins. At flow rate of <60 or >180 mL/min (Figure 2EDown), smokelike echoes disappeared but reappeared when the flow rate was reduced to 90 mL/min or 120 mL/min.



View larger version (76K):
[in this window]
[in a new window]
 
Figure 2. Flow rate and smokelike signal. A, At a flow rate of 0 mL/min, blood that was already in the chamber was highly echogenic; B, at a flow rate of 30 mL/min, blood entering the chamber is echolucent; C and D, at 90 mL/min and 120 mL/min, repetitively swirling smokelike images of flow lines appeared between outflow jet stream and the static blood near the side of chamber (as denoted by arrows); and E, at 180 mL/min, smokelike signal has disappeared.

TCD Examination of Smokelike Echoes
During the presence of echogenic smoke, the blood passing through the chamber caused no significant difference in average TCD signal intensity (49.9±6.53 U) compared with that produced by the blood passing through the bypass tube (49.0±4.4, P=0.71; Figure 3Down). Additionally, no visible embolic signals were found from TCD recordings that corresponded to the appearance of smokelike echoes or the flow bypassing the chamber (Figures 4ADown and 4BDown).



View larger version (32K):
[in this window]
[in a new window]
 
Figure 3. Comparison of average Doppler signal intensity between smokelike flow and flow with injection of different materials. The average intensity of the Doppler signal registered from blood flow through the bypass did not show any significant difference from that obtained from flow with presence of smokelike signals. However, injection of whole-blood clots, platelet-rich clots, and air bubbles caused a significant increase in an average Doppler signal intensity. Air bubbles produced the highest average signal intensity among them. In contrast, injection of normal saline decreased average signal intensity.



View larger version (21K):
[in this window]
[in a new window]
 
Figure 4. TCD detection of embolic materials. No embolic signal was found from the blood flow with the presence of smokelike echoes (A) or from the bypassed blood flow (B). Injection of normal saline at the reported rate did not change the TCD imaging (C) However, injection of whole-blood clots (D), platelet-rich clots (E), and microbubbles (F) produced visible embolic signals (denoted by arrows), and the signal intensity increased by microbubbles was higher than for the clots.

Injection of Embolic Particles and Normal Saline
Injection of whole-blood clots (mean size 1.3±0.5 mm, concentration 67 clots/mL), platelet clots (0.9±0.5 mm, 47 clots/mL), microbubbles (30.2±9.8 µm, 1.4x103 bubbles/mL) and degassed normal saline (2.5 mL) all produced flame-shaped echoes with a high echodensity when they were initially pumped into chamber (Figure 5Down). Figure 5ADown showed that along with this high-density echo of the inflow jet flow from in the middle of the chamber, injection of saline into the chamber at the given volume and rate also generated a transient smokelike echo for several seconds inside the expansion chamber (Figure 5ADown). The emboli in the chamber produced a general increase in average echodensity as well as starlike, high-echodensity points with clear margins (Figures 5BDown, 5CDown, and 5DDown.). They floated inside the chamber and finally were exempted from chamber with the blood flow. The echographic appearance of emboli was quite distinct from the smokelike echoes. However, compared with the flow with presence of smokelike echoes (49.9±6.53 U), injection of whole-blood clots, platelet clots, and microbubbles caused a significant increase in average Doppler signal intensity (whole-blood clots 80.6±23.6 U, P<0.05; platelet clots 72.4±14.4 U, P=0.05; and microbubbles 2453.7±885.4 U, P=0.001). In contrast, injection of an equivalent volume of normal saline decreased the average signal intensity (14.8±0.68 U, P<0.05; Figure 3Up).



View larger version (113K):
[in this window]
[in a new window]
 
Figure 5. Two-dimensional echo images of different materials in the chamber. A, At a flow rate of 40 mL/min (without smokelike echo), injection of normal saline temporarily produced smokelike imaging (arrow). B through D, Injection of whole-blood clots, platelet-rich clots, and microbubbles generated starlike echographies (arrows) with clear margins and floated in the expansion chamber.

Moreover, injection of embolic materials but not normal saline (Figure 4CUp) produced high average signal intensity and visible embolic signals in the TCD velocity waveforms (Figures 4DUp, 4EUp, and 4FUp).

Correlation of Smokelike Echo With Doppler Velocity
When smokelike echoes occurred, mean Doppler flow velocity recorded in the middle of the outflow in the expansion chamber was higher (148.7±17.9 cm/s)(Figure 6aDown) than that obtained in the smokelike echo area (32.8±0.18 cm/s)(Figure 6bDown) (P<0.01). Both showed bi-directional wave patterns. However, the Doppler flow velocity was near to zero in the area near the side of this chamber (Figure 6cDown).



View larger version (50K):
[in this window]
[in a new window]
 
Figure 6. Doppler velocity related to smokelike echo. Mean peak velocity in the center of outflow stream was higher (148.7±17.9 cm/s) than that in the smokelike signal area (32.8±0.18 cm/s; P<0.01). Both waveforms were in a biphasic wave pattern. The velocity in the area near the side of chamber was too low to record.

Smokelike Echoes and Thrombus Formation
To examine whether coagulation of the blood could increase under smoke state, CaCl2 was added into the circuit, and the time taken for thrombus formation was observed at different settings. At a flow rate of 90 mL/min, the time taken for thrombus formation was 4 minutes for undisturbed blood, 4.5 minutes for the blood passing through the expansion chamber with the presence of smokelike echoes, and 6 minutes for blood passing through bypass tube.


*    Discussion
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
*Discussion
down arrowReferences
 
Two-dimensional echocardiography is a well-established technique in the assessment of left atrial spontaneous echocardiographic contrast or smokelike echo and thrombi.28 29 30 Using Yorkshire albino pig blood, Merino et al2 mimicked spontaneous echocardiographic contrast in an expansion chamber by pumping whole blood at a low flow velocity. This in vitro flow model was thought to emulate pulmonary vein flow into the left atrium or mitral flow into the left ventricle. With a similar model and flow velocity, we used human whole blood to generate a similar pattern of echoes, but produced pulsed flow because this is closer to the clinical situation.

We also noted that the mean Doppler peak velocity obtained in the smokelike echoes area in the chamber is close to that recorded in the human left atrium with the presence of spontaneous echo contrast and/or thrombi.31 32 This suggests that the model is adequate to mimic the clinical situation.

In this study, we used Doppler ultrasound, which has been proved to be very sensitive in detection of embolic materials in blood flow,18 23 to investigate whether smokelike echoes resulted in the production of embolic signals. Embolic particles produce a visibly high amplitude signal on the Doppler waveform as well as a remarkable increase in signal intensity. However, we found that the components that caused intracavitary smokelike echoes failed to produce any embolic Doppler signals and did not increase Doppler signal intensity compared with the signal intensity produced by blood passing through the bypass tube. This result is in accord with those of a previous experimental report5 that the components of spontaneous echo contrast are not solid but are liquid materials.

Although it has been reported that spontaneous echo contrast was acoustically distinct from endocardiac mural thrombus attached to the ventricular apex,5 no previous study has attempted to differentiate echographic smoke signals from the emboli. Furthermore, we have identified that the blood flow with smoke-like signals dose not produce embolic signals detectable with TCD. We also tested the reverse situation, whether emboli caused smokelike signals, because microthrombi,33 red blood and platelet aggregates,34 and air bubbles35 have been implicated in the development of smokelike echo. We found that, as expected, these emboli produced TCD signals after they circulated within the echo chamber. However, there was no evidence that they caused smoke echoes.

The cardiac cavity is normally echo free on the echocardiogram because the ultrasound reflection from intracardiac blood is not dense enough to appear on the screen.36 This was confirmed as a flow-dependent phenomenon in the present study. The mechanism for the echogenicity of static blood as well as spontaneous echo contrast has been investigated by Sigel et al.37 Whole blood suspended in plasma produces visible echoes in vitro. However, plasma serum or erythrocytes do not produce echoes. The echogenicity of blood components may also depend on factors related to blood flow, including rheological and biochemical properties, and changes in the physical alignment of potentially echogenic blood components. Our findings that smokelike echoes usually occurred in the area between the inflow jet with higher flow velocity and static blood located in the side of the chamber without Doppler-detectable flow velocity may support this hypothesis. Thrombi were detected more frequently in patients with reduced left ventricular global systolic function.7 It has been suggested that shear stress, defined as the product of the velocity gradients between parallel flow lines located in the center and the periphery of blood vessels times the blood viscosity, contributed to the echogenicity of spontaneous contrast by exerting a mechanical force to change the physical layering.2

It has been reported that direct shear-induced platelet aggregation may develop under certain pathologic flowing conditions,38 39 40 especially at a high shear stress condition in which stenosis of vessel usually existed,41 and such shear-induced platelet aggregates could be inhibited by administration of GPIIb/IIIa receptor antagonist.38 From a view of biofluid mechanics and blood biorheology, the formation of microparticles by platelet aggregation could influence the reflected Doppler ultrasound waveform by increasing the scattering intensity, since it increased the reflector size.42 Considering the scattering factor, in this study we analyzed the average intensity of total Doppler signal reflected (instead of individual signal intensity at any interested points) at different testing protocols and found that the blood flow associated with smokelike echo did not increase the ultrasound backscattered power. This may further indicate that platelet aggregates are formed in the chamber when the spontaneous echo contrast is presented at the current settings. This may be due to the fact that the smokelike echo and shear-induced platelet aggregates are generated in different levels of flow shear.

Interestingly, injection of normal saline produced short-lived spontaneous contrast. This phenomena has also been seen in an animal model.5 This provides further evidence that smokelike echo is generated by fluid components. This may result from a change in physical alignment of blood components by temporally altered regional shear force or acoustic impedance between blood and normal saline from which more echoes are expected to be reflected from the blood-saline interfaces than those from the scattering effect of RBCs. Therefore, this lasts only for a short time before complete admixture of normal saline and blood.

How static blood with smokelike echographic signal plays a role in thrombus formation has not been understood, although in previous echocardiographic investigations1 36 43 it has been found that most left atrial thrombi were accompanied by spontaneous echo contrast. Theoretically, the blood in the presence of smokelike signals needs a shorter time to develop thrombus than the blood without smokelike signals after thrombin is added. Our study demonstrated that blood in the presence of spontaneous echo contrast needs a shorter time to form thrombi than that in the bypass but longer than that in the stationary glass. It suggests that thrombus development is inversely related to the mobility (flow velocity) of blood in that location, and the blood with the presence of spontaneous echo contrast has an increased risk of thrombus formation. However, this effect does not seem to be specific because the blood with smoke echo still takes a longer time to develop thrombus than that in the stationary glass.

There were some limitations in this study. First, smokelike echo reproduction and its differentiation from embolic signals were conducted in an in vitro circulation model (circulated by the anticoagulated blood), which is different from the human physiological situation and the clinical setting. Second, Doppler signal intensity is an index affected by multiple factors. For example, the acoustic impedance for skull bone is variable among individuals and even over a life span in the same individual. Accordingly, the intensity measured from the Doppler blood flow signal as well as embolic signals may be incomparable between different persons. As a result, attention must be carefully paid in applying current in vitro research into clinical situations.

In conclusion, this in vitro study suggests that smokelike echo does not produce an increase in TCD signal intensity such as is achieved by embolic materials. The echographic appearances of whole-blood clots, platelet clots, and air bubbles are distinct from smokelike echo. Combined with our observation that smokelike echo could also be temporally generated during injection of saline into the blood, our results further suggest that smokelike echo is a special echo phenomenon produced not by embolic particles but instead by slow-motion whole blood, possibly with physic layering. Moreover, our results show clearly that smokelike echo occurs at low flow situations and does not itself produce material capable of embolizing into the systemic circulation. Thrombus formation is related to the low flow velocity of regional blood. Thus, smokelike echo may be helpful to predict a high risk of thrombus development.


*    Acknowledgments
 
This study was supported by British Council and Postgraduate Research Funding of the University of Glasgow. We would like to thank Prof John L. Reid for his kindly help and encouragement with this project and Dr A. Shuaib for his support for the preparation of this manuscript. We would also like to express our gratitude to Dr Andrew Kelman for his technical help.

Received November 17, 1999; revision received January 27, 2000; accepted January 27, 2000.


*    References
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
up arrowDiscussion
*References
 

  1. Castello R, Pearson AC, Labovitz AJ. Prevalence and clinical implications of atrial spontaneous contrast in patients undergoing transesophageal echocardiography. Am J Cardiol. 1990;65:1149–1153.[Medline] [Order article via Infotrieve]
  2. Merino A, Hauptman P, Badimon L, Badimon JJ, Cohen M, Fuster V, Goldman M. Echocardiographic "smoke" is produced by an interaction of erythrocytes and plasma proteins modulated by shear forces. J Am Coll Cardiol. 1992;20:1661–1668.[Abstract]
  3. Feigenbaum H. Echocardiography. 2nd ed. Philadelphia. Pa: Lea & Febiger; 1976.
  4. Daniel WG, Nellessen U, Schroder E, Nonnast-Daniel B, Bednarski P, Nikutta P, Lichtlen PR. Left atrial spontaneous echo contrast in mitral valve disease: an indicator for an increased thromboembolic risk. J Am Coll Cardiol. 1988;11:1204–1211.[Abstract]
  5. Mikell FL, Asinger RW, Elsperger KJ, Anderson WR, Hodges M. Regional stasis of blood in the dysfunctional left ventricle: echocardiographic detection and differentiation from early thrombosis. Circulation. 1982;66:755–763.[Free Full Text]
  6. Archer SL, Kvernern LR, James K, Ezekowitz M, Gornick C. Does warfarin reduce the prevalence of left atrial thrombus in chronic atrial fibrillation? A double blind, placebo-controlled study. Circulation. 1991;84(suppl II):II-693. Abstract.
  7. Stollberger C. Chnupa P. Kronik G. Brainin M. Finsterer J. Schneider B. Slany J. Transesophageal echocardiography to assess embolic risk in patients with atrial fibrillation: ELAT Study Group: Embolism in Left Atrial Thrombi. Ann Intern Med. 1998;128:630–638.[Abstract/Free Full Text]
  8. Ozkan M, Kaymaz C, Kirma C, Civelek A, Cenal AR, Yakut C, Deligonul U. Predictors of left atrial thrombus and spontaneous echo contrast in rheumatic valve disease before and after mitral valve replacement. Am J Cardiol. 1998;82:1066–1070.[Medline] [Order article via Infotrieve]
  9. Fisher EA, Stahl JA, Budd JH, Goldman RA, Klapper A, Lee Y, Goldman ME. Left atrial "smoke" is a major risk factor for thrombus formation. Circulation. 1991;84(suppl II):II-692. Abstract.
  10. Tsai LM, Chen JH, Tsao CJ. Relation of left atrial spontaneous echo contrast with prethrombotic state in atrial fibrillation associated with systemic hypertension, idiopathic dilated cardiomyopathy, or no identifiable cause (lone). Am J Cardiol. 1998;81:1249–1252.[Medline] [Order article via Infotrieve]
  11. Hata JS, Biller J, Stuhlmuller JE, Kerber RE, Vandenberg BF. Impact of transesophageal echocardiography on the management of patients admitted with focal cerebral ischemia. Circulation. 1991;84(suppl II):II-693. Abstract.
  12. Pop G, Sutherland GR, Koudstaal PJ, Sit TW, de Jong G, Roelandt JR. Transesophageal echocardiography in the detection of intracardiac embolic sources in patients with transient ischemic attacks. Stroke. 1990;21:560–565.[Abstract/Free Full Text]
  13. Albers GW, Comess KA, DeRook FA, Bracci P, Atwood JE, Bolger A, Hotson J. Transesophageal echocardiographic findings in stroke subtypes. Stroke. 1994;25:23–28.[Abstract]
  14. Zabalgoitia M, Halperin JL, Pearce LA, Blackshear JL, Asinger RW, Hart RG. Transesophageal echocardiographic correlates of clinical risk of thromboembolism in nonvalvular atrial fibrillation: Stroke Prevention in Atrial Fibrillation III Investigators. J Am Coll Cardiol. 1998;31:1622–1626.[Abstract/Free Full Text]
  15. Censori B, Colombo F, Valsecchi MG, Clivati L, Zonca A, Camerlingo M, Casto L, De Tommasi MS, Mamoli A. Early transoesophageal echocardiography in cryptogenic and lacunar stroke and transient ischaemic attack. J Neurol Neurosurg Psychiatry. 1998;64:624–627.[Abstract/Free Full Text]
  16. Meltzer RS, Tickner EG, Sahines TP, Popp RL. The source of ultrasound contrast effect. J Clin Ultrasound. 1980;8:121–127.[Medline] [Order article via Infotrieve]
  17. Wessler S, Stehbens WE. Thrombosis. In: Thrombosis and Bleeding Disorders. New York, NY: Academic Press; 1971:489.
  18. Markus HS, Brown MM. Differentiation between different pathological cerebral embolic materials using transcranial Doppler in an in vitro model. Stroke. 1993;24:1–5.[Abstract/Free Full Text]
  19. Turakhia AK, Teague SM, Lawler B, Harris J. Hemorrheologic determinants of the echocardiographic "smoke" phenomenon: an in vitro simulation. Circulation. 1991;84(suppl II):II-692. Abstract.
  20. Mahony C, Ayappa IA, Ferguson J, Brown LV, Lai-Fook SJ. Spontaneous contrast (SC) in the rabbit. Circulation. 1991;84(suppl II):II-692. Abstract.
  21. Merino A, Hauptman PJ, Badimon L, Badimon JJ, Goldman ME. Echogenic smoke is shear rate dependent. Circulation. 1991;84(suppl II):II-692. Abstract.
  22. Spencer MP, Campbell SD, Sealey JL, Henry FC, Lindbergh J. Experiments on decompression bubbles in the circulation using ultrasonic and electromagnetic flowmeters. J Occup Med. 1969a;11:238–244.
  23. Russell D, Madden KP, Clark WM, Sandset PM, Zivin JA. Detection of arterial emboli using Doppler ultrasound in rabbits. Stroke. 1991;22:253–258.[Abstract/Free Full Text]
  24. Yang Y, Grosset DG, Lees KR, Kelman A. Ultrasonic characteristics of embolic particles in an in vitro model. In: Proceedings of the 2nd International Conference on Stroke, May 12–15, 1993, Geneva, Switzerland. Abstract, p 138.
  25. Svartling N. Detection of embolized material in the right atrium during cementation in hip arthroplasty. Acta Anaesthesiol Scand. 1988;32:203–208.[Medline] [Order article via Infotrieve]
  26. Evans DH, McDicken WN, Skidmore R, Woodcock JP. Doppler Ultrasound: Physics, Instrumentation and Clinical Applications. Chichester, UK: John Wiley & Sons Inc; 1989.
  27. Muller, HR, Lampl Y, Haefele M. A TCD test for clinical assessment of cerebral autoregulation. Ultraschall Med. 1991;12:218–221.[Medline] [Order article via Infotrieve]
  28. Garcia-Fernandez MA, Torrecilla EG, Roman DS, Azevedo J, Bueno H, Moreno MM, Delcan JL. Left atrial appendage Doppler flow patterns: implications on thrombus formation. Am Heart J. 1992;124:955–961.[Medline] [Order article via Infotrieve]
  29. Muge A, Daniel WG, Hausmann D, Godke J, Wagenbreth I, Lichtlen PR. Diagnosis of left atrial appendage thrombi by transesophageal echocardiography: clinical implications and follow-up. Am J Card Imaging. 1990;4:173–179.
  30. Aschenberg W, Schlutter M, Kremer R, Schroeder E, Siglow V, Bleifeld W. Transesophageal two-dimensional echocardiography for the detection of left atrial appendage thrombus. J Am Coll Cardiol. 1986;7:163–166.[Abstract]
  31. Jolly N, Mohan JC, Arora R. Transoesophageal Doppler pulmonary venous flow pattern and left atrial spontaneous contrast in mitral stenosis. Int J Cardiol. 1992;36:357–360.[Medline] [Order article via Infotrieve]
  32. Pollick C, Taylor D. Assessment of left atrial appendage function by transesophageal echocardiography: implications for the development of thrombus. Circulation. 1991;84:223–230.[Abstract/Free Full Text]
  33. Vuilliomenet A, Bertel O, Kaufmann U. Spontanous left atrial echo contrast in transesophageal echocardiography. Schweiz Med Wochenschr. 1992;122:549–553.[Medline] [Order article via Infotrieve]
  34. Rohmann S, Erbel R, Darius H, Makowski T, Jensen P, Fisher T, Meyer J. Spontaneous echo contrast imaging in infective endocarditis: a predictor of complication. Int J Card Imaging. 1992;8:197–207.[Medline] [Order article via Infotrieve]
  35. Marcus RH, Weinert L, Neumann A, Borow KM, Lang RM. Venous air embolism: diagnosis by spontaneous right-sided contrast echocardiography. Chest. 1991;99:784–785.[Abstract/Free Full Text]
  36. Beppu S, Nimura Y, Sakakibara H, Nagata S, Park Y-D, Izumi S. Smoke-like echo in the left atrial cavity in mitral valve disease: its features and significance. J Am Coll Cardiol. 1985;6:774–779.
  37. Sigel B, Coelho JCU, Spigos DG, Flanigan DP, Schuler JJ, Kasprisin DO, Nyhus LM, Capec V. Ultrasonography of blood during stasis and coagulation. Invest Radiol. 1981;16:71–76.[Medline] [Order article via Infotrieve]
  38. Kamat SG, Turner NA, Konstantopoulos K, Hellums JD, McIntire LV, Kleiman NS, Moake JL. Effects of Integrelin on platelet function in flow models of arterial thrombosis. J Cardiovasc Pharmacol. 1997;29:156–163.[Medline] [Order article via Infotrieve]
  39. Sakariassen KS, Holme PA, Orvim U, Barstad RM, Solum NO, Brosstad FR. Shear-induced platelet activation and platelet microparticle formation in native human blood. Thromb Res. 1998;92(suppl 2):S33-S41.
  40. Goto S, Sakai H, Goto M, Ono M, Ikeda Y, Handa S, Ruggeri ZM. Enhanced shear-induced platelet aggregation in acute myocardial infarction. Circulation. 1999;99:608–613.[Abstract/Free Full Text]
  41. Holme PA, Orvim U, Hamers MJ, Solum NO, Brosstad FR, Barstad RM, Sakariassen KS. Shear-induced platelet activation and platelet microparticle formation at blood flow conditions as in arteries with a severe stenosis. Arterioscler Thromb Vasc Biol. 1997;17:646–653.[Abstract/Free Full Text]
  42. Cloutier G, Qin Z. Ultrasound backscattering from non-aggregating and aggregating erythrocytes: a review. Biorheology. 1997;34:443–470. Review.[Medline] [Order article via Infotrieve]
  43. Black IW, Hopkins AP, Lee LCL, Walsh WF, Jacobson BM. Left atrial spontaneous echo contrast: a clinical and echocardiographic analysis. J Am Coll Cardiol. 1991;18:398–404.[Abstract]



This article has been cited by other articles:


Home page
EuropaceHome page
Writing Committee Members, V. Fuster, L. E. Ryden, D. S. Cannom, H. J. Crijns, A. B. Curtis, K. A. Ellenbogen, J. L. Halperin, J.-Y. Le Heuzey, G. N. Kay, et al.
ACC/AHA/ESC 2006 guidelines for the management of patients with atrial fibrillation: full text: A report of the American College of Cardiology/American Heart Association Task Force on practice guidelines and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Revise the 2001 Guidelines for the Management of Patients With Atrial Fibrillation) Developed in collaboration with the European Heart Rhythm Association and the Heart Rhythm Society
Europace, September 1, 2006; 8(9): 651 - 745.
[Full Text] [PDF]


Home page
J Am Coll CardiolHome page
V. Fuster, L. E. Ryden, D. S. Cannom, H. J. Crijns, A. B. Curtis, K. A. Ellenbogen, J. L. Halperin, J.-Y. Le Heuzey, G. N. Kay, J. E. Lowe, et al.
ACC/AHA/ESC 2006 Guidelines for the Management of Patients With Atrial Fibrillation--Executive Summary: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Revise the 2001 Guidelines for the Management of Patients With Atrial Fibrillation) Developed in Collaboration With the European Heart Rhythm Association and the Heart Rhythm Society
J. Am. Coll. Cardiol., August 15, 2006; 48(4): 854 - 906.
[Full Text] [PDF]


Home page
J Am Coll CardiolHome page
V. Fuster, L. E. Ryden, D. S. Cannom, H. J. Crijns, A. B. Curtis, K. A. Ellenbogen, J. L. Halperin, J.-Y. Le Heuzey, G. N. Kay, J. E. Lowe, et al.
ACC/AHA/ESC 2006 Guidelines for the Management of Patients With Atrial Fibrillation: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Revise the 2001 Guidelines for the Management of Patients With Atrial Fibrillation) Developed in Collaboration With the European Heart Rhythm Association and the Heart Rhythm Society
J. Am. Coll. Cardiol., August 15, 2006; 48(4): e149 - e246.
[Full Text] [PDF]


Home page
CirculationHome page
V. Fuster, L. E. Ryden, D. S. Cannom, H. J. Crijns, A. B. Curtis, K. A. Ellenbogen, J. L. Halperin, J.-Y. Le Heuzey, G. N. Kay, J. E. Lowe, et al.
ACC/AHA/ESC 2006 Guidelines for the Management of Patients With Atrial Fibrillation: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Revise the 2001 Guidelines for the Management of Patients With Atrial Fibrillation): Developed in Collaboration With the European Heart Rhythm Association and the Heart Rhythm Society
Circulation, August 15, 2006; 114(7): e257 - e354.
[Full Text] [PDF]


Home page
CirculationHome page
V. Fuster, L. E. Ryden, D. S. Cannom, H. J. Crijns, A. B. Curtis, K. A. Ellenbogen, J. L. Halperin, J.-Y. Le Heuzey, G. N. Kay, J. E. Lowe, et al.
ACC/AHA/ESC 2006 Guidelines for the Management of Patients With Atrial Fibrillation--Executive Summary: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Revise the 2001 Guidelines for the Management of Patients With Atrial Fibrillation): Developed in Collaboration With the European Heart Rhythm Association and the Heart Rhythm Society
Circulation, August 15, 2006; 114(7): 700 - 752.
[Full Text] [PDF]


Home page
Eur Heart JHome page
Authors/Task Force Members, V. Fuster, L. E. Ryden, D. S. Cannom, H. J. Crijns, A. B. Curtis, K. A. Ellenbogen, J. L. Halperin, J.-Y. Le Heuzey, G. N. Kay, et al.
ACC/AHA/ESC 2006 guidelines for the management of patients with atrial fibrillation executive summary: A report of the American College of Cardiology/American Heart Association Task Force on practice guidelines and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Revise the 2001 Guidelines for the Management of Patients with Atrial Fibrillation) Developed in collaboration with the European Heart Rhythm Association and the Heart Rhythm Society
Eur. Heart J., August 2, 2006; 27(16): 1979 - 2030.
[Full Text] [PDF]


Home page
J Am Coll CardiolHome page
R. Rastegar, D. J. Harnick, P. Weidemann, V. Fuster, B. Coller, J. J. Badimon, J. Chesebro, and M. E. Goldman
Spontaneous echo contrast videodensity isflow-related and is dependent on the relative concentrations of fibrinogen and red blood cells
J. Am. Coll. Cardiol., February 19, 2003; 41(4): 603 - 610.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Neuroradiol.Home page
Y. Yang, D. G. Grosset, Q. Li, A. Shuaib, and K. R. Lees
Turbulence and Circulating Cerebral Emboli Detectable at Doppler Ultrasonography: A Differentiation Study in a Stenotic Middle Cerebral Artery Model
AJNR Am. J. Neuroradiol., August 1, 2002; 23(7): 1229 - 1236.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow Request Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Yang, Y.
Right arrow Articles by Lees, K. R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Yang, Y.
Right arrow Articles by Lees, K. R.
Related Collections
Right arrow Echocardiography
Right arrow Acute Cerebral Infarction
Right arrow Doppler ultrasound, Transcranial Doppler etc.
Right arrow Risk Factors for Stroke