From the Departments of Neurology (G.S.-A., V.P., L.C.) and Psychiatry (M.K.), Medical University of Debrecen, Debrecen, Hungary; the Department of Neurology, University of Münster, Münster, Germany (G.S.-A., D.W.D., D.G.N., E.B.R.); and the Department of Radiology, Zentralklinikum Augsburg, Augsburg, Germany (W.A.W.).
Correspondence to Gernot Schulte-Altedorneburg, MD, Department of Radiology, Zentralklinikum Augsburg, Stenglinstr. 2, D-86156 Augsburg, Germany. E-mail Gernot.SAD{at}gmx.de
| Abstract |
|---|
|
|
|---|
MethodsThe BA was prospectively studied by TCCS shortly before death (median 3 days) in 46 moribund neurological patients (mean±SD age 71.1±13.1 years). The length of the BA was determined by measuring the distance between the vertebrobasilar junction and the deepest available flow signal in the top of the BA. During autopsy, photos of the vertebrobasilar system were taken to evaluate the true anatomic length and variations of the course of BA in situ, eg, straight, curved, or S-shaped.
ResultsComparison of the in vivo ultrasound measurements of BA length and postmortem data was possible in 44 of 46 cases. In the 2 remaining patients, the BA was occluded. The mean insonation depth of the vertebrobasilar junction was found at 66.9±7.1 mm. The mean BA length was 21.5±6.8 mm by color-coded duplex and 32.9±6 mm anatomically (P<0.0001). The mean difference between color mode and anatomic findings was 11.3±6.4 mm in the case of a straight BA (35 cases) and 16.3±4.8 mm in an anatomically tortuous course of the BA (9 cases).
ConclusionsColor duplex imaging enables correct visualization of the proximal two thirds of the BA, but only exceptionally of its distal one third. A tortuous course of the BA leads to an underestimation of its anatomic length.
Key Words: basilar artery duplex scanning pathology ultrasonography
| Introduction |
|---|
|
|
|---|
The technique of transcranial color-coded sonography (TCCS) helps to overcome some of the diagnostic insufficiencies of "conventional" TCD. It allows for the simultaneous visualization of intracranial parenchymal structures by B-mode and cerebral artery blood flow by color-coded flow imaging.3 Vertebrobasilar arteries can be depicted directly and continuously. Measurements of vessel length, as well as the hemodynamic assessment, are possible.4 5 6
In this study, we focused on the ability of transnuchal TCCS to determine the length and the gross morphology of the BA by comparing in vivo ultrasound findings acquired shortly before death with postmortem findings. Furthermore, we wanted to identify the anatomic variations mostly affecting the suboccipital findings at the BA.
| Subjects and Methods |
|---|
|
|
|---|
Reasons for admission to hospital of these patients are given in
Table 1
; there were 16 cases of
intracerebral bleeding detected by initial CT scan.
Supratentorial parenchymal hemorrhage was
found in 12 cases, and infratentorial hemorrhage was observed
in 2. In another 2 cases, a supratentorial and
infratentorial hemorrhage occurred at the same time. The
intracerebral hemorrhage was accompanied by
intraventricular bleeding in 8 cases. An additional
subarachnoid hemorrhage occurred in 2 of 16 cases.
|
In every case, the cause of death was determined by an experienced pathologist during autopsy. Seventeen patients died of pneumonia. In 15 cases the cause of death was increased intracranial pressure; 7 patients died of pulmonary embolism. In 5 cases the cause of death was cardiopulmonary arrest, in 1 case the patient died of sepsis, and in 1 case the cause of death remained unknown, despite necropsy. The median interval between the ultrasound examination and death was 3 days (range 2 hours to 135 days).
TCCS
The intracranial segments of the VAs and the BA were assessed by
TCCS via the foramen magnum (suboccipital approach) with a dual
frequency 2.0/2.5-MHz 90° sector imaging transducer (HP SONOS 2000,
Hewlett-Packard). An emission frequency of 2 MHz was used for all
studies. The ultrasound beam was placed approximately 4 cm beneath the
occipital protuberance, while the patients were lying in a lateral
position with the neck semi-inclined. The majority of patients were
examined at the bedside in the intensive care unit. In these cases, the
examiners position was at the patients side. Patients with raised
intracranial pressure were investigated in supine position; the
examiner elevated the patients head by means of a pillow to place the
ultrasound beam at the nuchal region. The position of the head was
never changed to the same level as the body during ultrasound. The
ultrasound studies of patients with intracranial bleeding were included
only if an adequate flow signal could be detected. To optimize the
quality of the Doppler signal and consider a tortuous course of the
insonated arteries, the probe was moved slightly to the right or left
of the midnuchal position. The examination was started at an insonation
depth of 10 cm. By superimposing the color-coded image with the B-mode
gray scale image at this site, the typical Y-shaped vertebrobasilar
junction could easily be visualized. The axial sample volume of 0.29 cm
was kept constant during the measurements. At the beginning of each
examination, the color scale was set at 42 cm/s; during the course of
the examination, color scale and color gain settings were chosen
individually to provide optimal imaging conditions. Visualization of a
color signal at a distal site without a corresponding pulsed
Doppler signal was judged to be an artifact (so-called color
bleeding). The length of the BA was measured from the vertebrobasilar
junction to the most distant visible segment of the BA. In some cases,
the distal bifurcation of the BA into both posterior cerebral arteries
(ie, the top of the BA) could be visualized. The insonation depths of
these landmarks were noted by the investigator. The Doppler
waveform was continuously tracked and recorded along the course of
the artery. Per definition, the disappearance and/or change of flow
direction of the Doppler signal indicated the site of the most
distant segment of the BA.
By means of Doppler mode, blood flow velocity was determined in the presumed midbasilar segment of the depicted BA and was considered to be representative of this patient and his condition.
All examinations were performed by the same investigator (G.S.-A.) and were recorded onto videotape. Offline evaluation by videotape was performed in a blinded manner without knowledge of the postmortem findings. The investigator was aware of the clinical diagnoses when the patient was alive.
Anatomic Length and Course of the BA
During autopsy, special care was taken not to damage the basal
cerebral arteries, including the vertebrobasilar system. The vessels
remained attached to the basis of the brain to avoid artificial changes
such as stretching or distortion of the BA. Photos of the
vertebrobasilar system were taken immediately after the brain was
removed from the skull. A reference divided in millimeters was placed
next to the BA. Several photos were made of the entire BA, including
the vertebrobasilar junction and the top of BA, with a CE-50 camera
(Chinon) with a zoom objective 52 mm in diameter (Soligor C/D) and
a focus of 95 to 210 mm focus and a Sensia II Fujichrome 100 color
slide film (Fuji; Figure
); thereafter, the brain was
placed in formalin for fixation. One to 2 weeks later, photos of the
basal aspect of the fixed brain were made again. The length of the BA
was measured on the photos (ie, color slides) of the fresh and fixed
specimen at 12- to 20-fold magnification; the measurements were carried
out along the entire course of the BA.
|
To describe the variations in the course of the BA, a subdivision was made into the following 4 configuration types: (1) straight (curve of <20°), (2) slightly single-curved (ie, 20° to 45°), (3) severely single-curved (ie, >45°), and (4) S-shaped. This subdivision considered only lateral deviations of the BA from the midline in an anterior-posterior view, not sagittal deviations.
Comparison Between TCCS and Anatomic Findings
The distance between the depth of the vertebrobasilar junction
and the deepest reliably identified segment of the BA was compared with
the corresponding anatomic length measurement at the fresh specimen.
"Reliably identified" refers to the deepest available color signal
and spectral waveform that could be clearly distinguished from an
artifact and/or from a signal of a different vessel.
The difference between the anatomic and the TCCS-based length of the BA was evaluated in millimeters. Mean values, standard deviations, medians, and ranges were calculated. The nonparametric Wilcoxon signed rank test for matched pairs was performed to check the difference between anatomic and TCCS measurements for statistical significance. The same procedure was applied to the difference between color mode and Doppler mode. Statistical significance was declared at the 0.05 level.
In 11 cases, only photos of the fixed specimens were available. This is why measurements of the fresh and the fixed specimens were compared in advance to estimate the influence of the fixation procedure. The mean difference in BA length of 1.3 mm yielded a negligible effect of fixation in this gross pathological analysis.
The study was performed according to the Hungarian laws, and patients were studied in compliance with a protocol previously approved by the local ethics committee of the University Medical school.
| Results |
|---|
|
|
|---|
Comparison of in vivo and postmortem measurements of BA length was possible in 44 of the 46 cases. In these patients, an occlusion or a hemodynamically relevant stenosis of the BA had been excluded in advance by ultrasound. In the 2 remaining cases, the diagnosis of a BA occlusion had been made based on the absence of the Doppler signal in the BA and on a reduced flow velocity and increased pulsatility index of the Doppler signal in the VAs. In both patients, this diagnosis could be confirmed after death by sectioning the BA into 3-mm slices. Both occluded BAs had a straight course.
The mean values of the in situ findings and of the TCCS measurements
are shown in Table 2
.
|
The overall mean difference in the length of the BA between the color-coded mode and the anatomic measurement was 12.4±6.4 mm (median 12.5 mm, range -6 to 25 [negative values indicate that the length detected by TCCS was greater than that detected anatomically]) (P<0.0001). Furthermore, the overall mean difference in the length of the BA between the Doppler mode and the postmortem measurement was 9.8±6.1 mm (median 9.5 mm, range -10 to 23; P<0.0001). Length measurements determined by Doppler mode were always greater than or equal to those obtained on color mode: The mean difference between measurements evaluated by color-coded mode and Doppler mode was 3.0±2.6 mm (median 2 mm, range 0 to 11; P<0.0001).
In 35 cases with a straight BA course, the mean difference between color modebased measurements and anatomic measurements was 11.3±6.4 mm (median 11 mm, range -8 to 25). In the 9 patients with a tortuous BA configuration, the anatomic measurement revealed a greater length of the BA than could be evaluated during color mode imaging. The mean difference was 16.3±4.8 mm (median 16 mm, range 7 to 22).
In 5 cases, the BA length evaluated by ultrasound was greater than in the anatomic reality.
In 5 patients, the investigator visualized the distal basilar bifurcation by TCCS; in these cases, the difference between postmortem analysis and visualization by color mode were -6 mm, -4 mm, -1 mm, +5 mm, and +6 mm, respectively.
| Discussion |
|---|
|
|
|---|
In concordance with previous TCCS studies, we found a wide range for the insonation depth of the vertebrobasilar junction (52 to 83 mm); Martin et al7 and Droste et al8 identified the origin of the BA at an insonation depth of 54 to 88 mm and 62 to 84 mm, respectively. Schöning and Walter6 detected the junction site at a distance of 55 to 76 mm in women and 65 to 80 mm in men. The large variation of the thickness of the nuchal soft-tissue layer and the variable adaptation pressure of the probe used by the investigator to place the transducer in its optimal position are reasons for this broad range. In a pathoanatomic validation study, Ringelstein et al2 measured the distances of various vertebrobasilar landmarks from the nuchal skin surface in sagittal MRI images of the skull; they found that the distance from the suboccipital skin surface to the vertebrobasilar junction ranged between 70 and 110 mm.
Besides our study and that of Droste et al,8 no other studies have evaluated the diagnostic accuracy of TCCS for measurement of BA length. Similarly to our results, the authors of the latter series could image the vertebrobasilar junction and the proximal 2 cm of the BA completely or partially after the application of echocontrast agent in most patients with insufficient precontrast TCCS examination.
The postmortem length of the BA was as large as 32.9±6 mm; this value exactly agrees with previous anatomic studies, which reported an average BA length of 32 to 33 mm.9 10 Comparing ultrasonic (21.5±6.8 mm) and anatomic (32.9±6 mm) measurements, we evaluated a shorter BA by color imaging with a mean difference of 12.4±6.4 mm. Surprisingly, in 11% (5/44) of the cases, the BA was shorter in anatomically measured length compared with values determined by TCCS. We assume that the investigator erroneously insonated the proximal segment of one PCA in cases of overestimation by ultrasound and consequently could not depict the real superior bifurcation.
Our data demonstrate that an anatomically tortuous (curved or S-shaped) configuration of the BA leads to a systematic underestimation of TCCS in measuring vessel length. The mean difference between color mode and anatomic measurement was larger in case of tortuous course than it was in the remaining patients (16.3±5 mm versus 11.3±6 mm). We attribute this to 2 facts: First, a tortuous configuration of the BA introduces per se a higher possibility to lose the Doppler signal distally to the site of change in direction. Second, we did not trace a serpentine shape of the BA segment by segment by means of extremely different positions of the transducer, because the summation of such segmental measurements would have yielded an incorrect overall-length measurement.
The results of our postmortem validation study confirmed those of Brandt et al11 : In a clinical prospective study, the authors could show in 19 patients with clinically suspected acute BA occlusion that native extracranial and intracranial Doppler sonography is able to exclude proximal BA occlusion verified by CT angiography; however, in 2 cases of distal BA occlusion, Doppler sonography findings were false-negative.
In summary, we could show by means of a comparison between in vivo and postmortem findings that native transnuchal TCCS enabled us to visualize the proximal two thirds of the BA length. A tortuous BA might confound an accurate measurement, and variations of the superior bifurcation of BA may make it harder to identify distal segments of the BA transnuchally. This confirms a rule of sonography that the transtemporal insonation of the origin of both posterior cerebral arteries is an unavoidable step for a reliable judgment of the vertebrobasilar system.
|
| Acknowledgments |
|---|
Received December 10, 1999; revision received January 27, 2000; accepted February 7, 2000.
| References |
|---|
|
|
|---|
2. Ringelstein EB, Kahlscheuer B, Niggemeyer E, Otis SM. Transcranial Doppler sonography: anatomical landmarks and normal velocity values. Ultrasound Med Biol. 1990;16:745761.[Medline] [Order article via Infotrieve]
3.
Bogdahn U, Becker G, Winkler J, Greiner K, Perez J,
Meurers B. Transcranial color-coded real-time sonography in
adults. Stroke. 1990;21:16801688.
4. Becker G, Lindner A, Bogdahn U. Imaging of the vertebrobasilar system by transcranial color-coded real-time sonography. J Ultrasound Med. 1993;12:395401.[Abstract]
5.
Kaps M, Seidel G, Bauer T, Behrmann B. Imaging of the
vertebrobasilar system using color-coded ultrasound. Stroke. 1992;23:15771582.
6.
Schöning M, Walter J. Evaluation of the
vertebrobasilar-posterior system by transcranial color
duplex sonography in adults. Stroke. 1992;23:12801286.
7. Martin PJ, Evans DH, Naylor AR. Transcranial color-coded sonography of the basal cerebral circulation: reference data from 115 volunteers. Stroke. 1994;25:390396.[Abstract]
8. Droste DW, Nabavi DG, Kemény V, Schulte-Altedorneburg G, Ritter MA, Weber S, Ringelstein EB. Echocontrast enhanced transcranial colour coded duplex offers improved visualization of the vertebrobasilar system. Acta Neurol Scand. 1998;98:193199.[Medline] [Order article via Infotrieve]
9. Kretschmann HJ, Weinrich W. Hirnarterien und ihre vaskulären Territorien. In: Kretschmann HJ, Weinrich W, eds. Klinische Neuroanatomie und kranielle Bilddiagnostik: Computertomographie und Magnetresonanztomographie. Stuttgart; New York: Thieme; 1991:191192.
10. Huber P. Radiologic anatomy and topography of cerebral vessels. In: Huber P, Krayenbühl H, Yasargil MG, eds: Cerebral Angiography. Stuttgart, Germany: Thieme; 1982:152160.
11.
Brandt T, Knauth M, Wildermuth S, Winter R, von Kummer
R, Sartor K, Hacke W. CT angiography and Doppler sonography for
emergency assessment in acute basilar artery ischemia.
Stroke. 1999;30:606612.