(Stroke. 2000;31:1342.)
© 2000 American Heart Association, Inc.
Original Contributions |
From the Department of Diagnostic Radiology and Organ Imaging, Prince of Wales of Hospital, Hong Kong, China.
Correspondence to Stella Ho, MPhil, Department of Diagnostic Radiology and Organ Imaging, Prince of Wales of Hospital, Shatin, Hong Kong, China. E-mail petella{at}hkstar.com
| Abstract |
|---|
|
|
|---|
MethodsThe common, internal, and external carotid arteries of 50 healthy subjects (22 men, 28 women, age range 19 to 54 years) were examined with CVI and SDI. The total blood flow volume of the internal and external carotid arteries was then compared with the ipsilateral common carotid artery flow. An accurate technique would demonstrate no difference. The difference (expressed as a percent inconsistency) was therefore a measure of the accuracy of the method.
ResultsThe mean±SD inconsistency was found to be 10.6±8.3% for CVI and 27.9±14.3% for SDI. The difference in inconsistency between CVI and SDI in measurement of carotid blood flow volume was statistically significant (P<0.01).
ConclusionsCVI is more accurate than SDI in the determination of blood flow volume in the carotid arteries. For noninvasive clinical estimation of cerebrovascular blood flow volume, CVI quantification should be the preferred technique.
Key Words: cerebral blood flow cerebral ischemia ultrasonography, Doppler
| Introduction |
|---|
|
|
|---|
Even though in vitro and in vivo tests have shown that color velocity imaging (CVI) is accurate in the measurement of BFV6 7 8 and that spectral Doppler imaging (SDI) tends to significantly overestimate it,9 SDI is still used by sonologists.
In this regard, the present study was undertaken to demonstrate the discrepancy of BFV estimations with CVI and SDI. An in vivo "internal" validation of these 2 techniques was performed to determine the carotid BFV in a group of healthy subjects, to compare the accuracies of these 2 techniques, and to justify their applications in clinical practice.
| Subjects and Methods |
|---|
|
|
|---|
This study was performed based on the assumption that total BFV of the internal carotid artery (ICA) and external carotid artery (ECA) was equal to that of the common carotid artery (CCA). Any measure of their difference would then reflect the inconsistency of the imaging technique applied. All 3 blood vessels (ie, CCA, ICA, and ECA) of the 100 groups of extracranial carotid arteries were examined with the high-resolution 7.5-MHz linear probe of the Philips SD800 ultrasound scanner (Philips Ultrasound International). This device has a standard feature with the capability of both CVI with flow quantification (CVIQ) and pulsed Doppler imaging. Each blood vessel was in turn interrogated with CVI and SDI in the straight segment at least 2 cm from the bifurcation. The measurements were repeated 3 times and averaged to provide the BFV estimates for each vessel.
With the technique of SDI, a large Doppler sample volume that
corresponds to the vessel diameter was used to determine the
approximate mean velocity. The anatomic vessel diameter was measured as
close to the line of interrogation as possible (Figure 1
). The BFV was calculated with the
equation BFV=mean velocityxvessel cross-sectional area=mean
velocityx
/4x(vessel diameter)2.
|
To determine the BFV with the CVI technique, a color box of a size that
covered the entire luminal cross section was used. The image was
adjusted with optimal color saturation with no aliasing or color
"bleeding" over the lumen. The color image was synchronized with
the M-mode, which can provide simultaneous information
about functional vessel diameter during the cardiac cycle and
multirange gated velocity information. The BFV was automatically
computed with the built-in system software after the color M-mode image
was frozen (Figure 2
). An identical angle
correction of 60° was applied to all vessels in both techniques.
|
With each technique, the sum of ICA and ECA BFVs on each side was compared with that of the ipsilateral CCA. The inconsistency of each technique was expressed as the percent difference between the sum of ICA and ECA BFV estimates and that of the CCA. It was calculated by subtracting the total ICA and ECA blood flow by the CCA flow. The absolute difference was then divided by the CCA flow and expressed as a percentage. The paired t test was used to compare the percentage inconsistency of the two techniques. The significance level was taken at P<0.01.
Because CVIQ is a relatively new technique for measurement of BFV in contrast to the more popular and conventional SDI technique, interobserver and intraobserver variabilities for this technique were determined. The BFV of 32 CCAs of an additional 16 normal volunteers (8 men and 8 women, age range 18 to 49 years, average age 33.2 years) were measured by 3 operators (A, B, and C). Ten readings of BFV for each CCA were made by each operator, blinded to the results of the others. The results were then analyzed with the Friedman test for intraobserver variability and Kendalls W test for interobserver variability. The significance level was set at P<0.01.
| Results |
|---|
|
|
|---|
|
Although there was a significant difference in the BFV between CCA and
its branches (ICA+ECA) obtained with either technique
(P<0.001), there was better correlation of the CCA BFV with
the sum of (ICA and ECA) BFV with the CVI technique than with the SDI
technique (Figures 3A
and 3B
). With both
techniques, the blood flow estimates were consistently greater
for CCA than for the sum of its branches.
|
The interobserver (P=0.8825) and intraobserver variabilities of the 3 operators (P=0.7619 for A, P=0.5610 for B, and P=0.5432 for C) in measurement of BFV with CVIQ was insignificant (P>0.01).
| Discussion |
|---|
|
|
|---|
Diameter Measurement
This study showed that both CVI and SDI techniques were
inconsistent in estimations of BFV, but the mean percent
inconsistency was less for CVI (
11%) than for SDI
(
28%). Other experimental studies have also shown that CVIQ
produces better consistency than SDI.9 10 The
error of SDI mainly comes from erroneous diameter measurement on a
static gray scale image, with the assumption of a stable vessel. It
must be emphasized that small errors in the diameter measurement will
result in large errors in the calculation of cross-sectional area and,
hence, flow volume. Because the physiological
anatomic diameter in systole or diastole varies and may
differ by as much as 10%,11 the diameter variation alone
can account for flow volume errors of up to 20%. The unique feature of
the acquisition of simultaneous information about
functional diameter and multirange gated velocities of the CVI
technique ameliorates the problem with SDI for measurement of BFV.
Angle Correction
Angle correction is essential for flow velocity and diameter
calculations that may affect the accuracy of BFV estimation with both
techniques. Unfortunately, a favorable insonation angle for diameter
measurement is adverse for flow velocity estimation. Normal incidence
of the ultrasound beam produces echoes of the shortest duration and
greatest amplitude, so true vessel diameter measurement is ideal at an
angle of 90° with little error.12 At this large angle of
incidence, even a 2° error can result in unacceptably high error in
the volume flow determination.13 Imaging at an angle of
<90°, however, will tend to underestimate the vessel diameter due to
the effect of beam width.12 As a compromise, an angle
correction of 60° was standardized for BFV
measurement with both techniques. Angle correction error was likely to
account in part for the internal inconsistency of each
technique but would be of similar magnitude for both.
Turbulent or Disturbed Flow
Often, helical or disturbed flow is present near the
bifurcating point, making the relationship between the angle of the
beam with the velocity vectors even more uncertain.14 This
inherent uncertainty will contribute to an inconsistent BFV
estimate. To reduce the uncertainty, a straight vessel segment at least
2 cm from the bifurcation should be interrogated. In practice, this
kind of error is sometimes unavoidable. This error particularly affects
the ECA BFV measurement when the prebranching segment of the ECA near
the carotid bulb had to be chosen to validate the test.
Off-Axis Sampling
Off-axis sampling is a significant error in subjects with
respiratory movement of the carotid arteries or in pulsatile vessels.
The off-axis error simultaneously causes a velocity error
by missing the central peak flow velocities and underestimating the
true vessel diameter, with resulting large errors in BFV
estimates.9
Of the many potential sources of error, most seem to be common to both
techniques (Table 2
). Temporal change in
diameter and improved quantification of both temporal changes in
velocity and differences of velocity across the vessel at any point in
time are better estimated with CVIQ, and this is probably one of the
major contributions to its greater accuracy for BFV. With SDI, if a
time-averaged M-mode measurement for the vessel diameter is used, the
error can be minimized.15 Where M-mode is not available,
the measurements can be repeated several times and averaged to reduce
the random error to an acceptable level.12 However, this
is time consuming and less accurate than the automated CVIQ method.
|
A further validation that CVIQ is more accurate than SDI can be derived
from the prediction of brain mass with these 2 techniques. It has been
shown that mean regional cerebral blood flow is
50 mL · 100
g-1 ·
min-1.16 If we assume that nearly
all of the CCA flow perfuses the brain, the brain weight predicted with
CVIQ (total CCA BFV was
680 mL/min) is 1.36 kg. The same calculation
with SDI (total CCA BFV was
1340 mL/min) yields 2.68 kg. Current
estimates of brain weight with CVIQ far more closely approximate the
values of human brain weight (men
1.5 kg, women
1.3 kg) quoted in
the literature.17 18
This study was performed on young normal subjects with promising results on CVIQ. Although difficulties may be encountered in elderly patients in reproducing similar results due to diseased arteries, vessel tortuosity, arrhythmia, or poor patient condition, CVIQ remains a suitable technique for noninvasive clinical estimation of cerebrovascular BFV and should be the preferred technique.
| Acknowledgments |
|---|
Received January 21, 2000; revision received March 15, 2000; accepted March 15, 2000.
| References |
|---|
|
|
|---|
2. Knappertz VA, Tegeler CH, Meyers LG. Clinical investigative studies. J Neuroimaging. 1996;6:17.[Medline] [Order article via Infotrieve]
3.
Eicke BM, Buss E, Bahr RR, Hajak G, Paulus W.
Influence of acetazolamide and CO on extracranial flow
volume and intracranial blood flow velocity. Stroke. 1999;30:7680.
4.
Uematsu S Yang A, Preziosi TJ, Kouba R, Toung TJ.
Measurement of carotid blood flow in man and its clinical application.
Stroke. 1983;14:256266.
5. Dalla CF, Signorini GP, Previato Schiesari A. Quantitative evaluation of carotid blood flow before and after surgery. Int Angiol. 1987;6:371373.[Medline] [Order article via Infotrieve]
6. Eicke BM, Tegeler CH, Howard G, Bennett JB, Myers LG, Meads D. In-vitro validation of color velocity imaging and spectral Doppler for velocity determination. J Neuroimaging. 1993;3:8992.[Medline] [Order article via Infotrieve]
7. Eicke BM, Tegeler CH. Ultrasonic quantification flow volumetry of the carotid arteries: initial experience with a color flow M-mode system. Cerebrovasc Dis. 1996;5:145149.
8. Maulik D, Kadado T, Downing G, Phillips C. In vitro and in vivo validation of time domain velocity and flow measurement technique. J Ultrasound Med. 1995;14:939947.[Abstract]
9. Eicke BM, Tegeler CH. Ultrasonic quantification of blood flow volume. In Tegeler CH, Babikian VL, Gomez CR, eds. Neurosonology. New York, NY: MosbyYear Book; 1996:101110.
10. Winkler AJ, Wu J, Case T, Ricci MA. An experimental study of the accuracy of volume flow measurements using commercial ultrasound systems. J Vasc Tech. 1995;19:175180.
11. Olson RM, Cooke JP. Human carotid artery diameter and flow by a noninvasive technique. Med Instrum. 1975;9:99102.[Medline] [Order article via Infotrieve]
12. Gill RW. Measurement of blood flow by ultrasound: accuracy and sources of error. Ultrasound Med Biol. 1985;11:625641.[Medline] [Order article via Infotrieve]
13. Kremkau FW. Doppler Ultrasound: Principles and Instruments, 2nd ed. Philadelphia, Pa: WB Saunders; 1995:89119.
14. Ku DN, Giddens DP. Hemodynamics of the normal human carotid bifurcation: in vitro and in vivo studies. Ultrasound Med Biol. 1985;11:1326.[Medline] [Order article via Infotrieve]
15. Landwehr P. Basic hemodynamics. In: Wolf KJ, Fobbe F, eds. Color Duplex Sonography: Principles and Clinical Applications. New York, NY: Thieme Medical Publishers, Inc: 1995:2036.
16. Fisher M. The pathophysiology of ischaemic stroke. In: Fisher M, ed. Clinical Atlas of Cerebrovascular Disorders. London, UK: MosbyYear Book; 1994:3(2)-3(23).
17. Pick TP, Howden R. Grays Anatomy, 19th ed. Philadelphia, Pa: Running Press; 1974:703705.
18. Orrison WW, Lewine JD, Sanders JA, Hartshorne MF, eds. Functional Brain Imaging. St Louis, Mo: MosbyYear Book; 1995.
This article has been cited by other articles:
![]() |
Y. Grad, H. Sievert, B. Nishri, G. W. Stone, B. T. Katzen, O. Yodfat, R. Higashida, D. Harris, A. K. Wakhloo, Y. Assaf, et al. A Novel Endovascular Device for Emboli Rerouting: Part I: Evaluation in a Swine Model Stroke, October 1, 2008; 39(10): 2860 - 2866. [Abstract] [Full Text] [PDF] |
||||
![]() |
S.O. Oktar, C. Yucel, D. Karaosmanoglu, K. Akkan, H. Ozdemir, N. Tokgoz, and T. Tali Blood-Flow Volume Quantification in Internal Carotid and Vertebral Arteries: Comparison of 3 Different Ultrasound Techniques with Phase-Contrast MR Imaging. AJNR Am. J. Neuroradiol., February 1, 2006; 27(2): 363 - 369. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Wintermark, M. Sesay, E. Barbier, K. Borbely, W. P. Dillon, J. D. Eastwood, T. C. Glenn, C. B. Grandin, S. Pedraza, J.-F. Soustiel, et al. Comparative Overview of Brain Perfusion Imaging Techniques Stroke, September 1, 2005; 36(9): e83 - e99. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Hendrikse, A. F. van Raamt, Y. van der Graaf, W. P. T. M. Mali, and J. van der Grond Distribution of Cerebral Blood Flow in the Circle of Willis Radiology, April 1, 2005; 235(1): 184 - 189. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. S. Y. Ho, W. W-m. Lam, S. C. P. Ng, M. K. Lam, M. T. V. Chan, W. S. Poon, and C. Metreweli Cerebral Vasoreactivity: A Comparison of Color Velocity Imaging Quantification and Stable Xenon-Enhanced CT Am. J. Roentgenol., March 1, 2005; 184(3): 948 - 952. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. W. M. Lam, S. S. Y. Ho, S. F. Leung, K. S. Wong, and C. Metreweli Cerebral Blood Flow Measurement by Color Velocity Imaging in Radiation-Induced Carotid Stenosis J. Ultrasound Med., October 1, 2003; 22(10): 1055 - 1060. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.F. Soustiel, T.C. Glenn, P. Vespa, B. Rinsky, C. Hanuscin, and N.A. Martin Assessment of Cerebral Blood Flow by Means of Blood-Flow-Volume Measurement in the Internal Carotid Artery: Comparative Study With a 133Xenon Clearance Technique Stroke, August 1, 2003; 34(8): 1876 - 1880. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. F. Soustiel, E. Levy, M. Zaaroor, R. Bibi, S. Lukaschuk, and D. Manor A New Angle-Independent Doppler Ultrasonic Device for Assessment of Blood Flow Volume in the Extracranial Internal Carotid Artery J. Ultrasound Med., December 1, 2002; 21(12): 1405 - 1412. [Abstract] [Full Text] [PDF] |
||||
![]() |
S.S.Y. Ho, C. Metreweli, and C.H. Yu Color Velocity Imaging Quantification in the Detection of Intracranial Collateral Flow Stroke, July 1, 2002; 33(7): 1795 - 1798. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. S. Y. Ho, Y. L. Chan, D. K. W. Yeung, and C. Metreweli Blood Flow Volume Quantification of Cerebral Ischemia: Comparison of Three Noninvasive Imaging Techniques of Carotid and Vertebral Arteries Am. J. Roentgenol., March 1, 2002; 178(3): 551 - 556. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Krejza, Z. Mariak, M. Huba, S. Wolczynski, and J. Lewko Effect of Endogenous Estrogen on Blood Flow Through Carotid Arteries Stroke, January 1, 2001; 32(1): 30 - 36. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Stroke Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2000 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |