Stroke. 2001;32:2803-2809
doi: 10.1161/hs1201.099714
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow Request Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Devuyst, G.
Right arrow Articles by Bogousslavsky, J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Devuyst, G.
Right arrow Articles by Bogousslavsky, J.
Related Collections
Right arrow Doppler ultrasound, Transcranial Doppler etc.
Right arrow Pathology of Stroke
Right arrow Primary and Secondary Stroke Prevention

(Stroke. 2001;32:2803.)
© 2001 American Heart Association, Inc.


Original Contributions

Automatic Classification of HITS Into Artifacts or Solid or Gaseous Emboli by a Wavelet Representation Combined With Dual-Gate TCD

G. Devuyst, MD; G.A. Darbellay, PhD; J.-M. Vesin, PhD; V. Kemény, MD; M. Ritter, MD; D.W. Droste, MD; C. Molina, MD; J. Serena, MD; R. Sztajzel, MD; P. Ruchat, MD; C. Lucchesi, PhD; G. Dietler, PhD; E.B. Ringelstein, MD; P.-A. Despland, MD J. Bogousslavsky, MD

From the Department of Neurology, CHUV (G. Devuyst, V.K., P.R., P.-A.D., J.B.), the Signal Processing Laboratory, Swiss Federal Institute of Technology (G.A.D., J.-M.V.), and the Department of Applied Physics, UNIL (C.L., G. Dietler), Lausanne, and the Department of Neurology, HCUG (R.S.), Geneva, Switzerland; the Department of Neurology, University Hospital, Münster, Germany (M.R., D.W.D., E.B.R.); and the Department of Neurology, University Hospital, Barcelona (C.M.), and the Department of Neurology, University Hospital, Girona (J.S.), Spain.

Correspondence to Gérald Devuyst, Department of Neurology, CHUV, rue du Bugnon 15, 1011 Lausanne, Switzerland. E-mail gerald.devuyst{at}chuv.hospvd.ch or Georges Darbellay, Signal Processing Laboratory, Swiss Federal Institute of Technology, 1015 Lausanne, Switzerland. E-mail georges.darbellay@epfl.ch


*    Abstract
up arrowTop
*Abstract
down arrowIntroduction
down arrowSubjects and Methods
down arrowResults
down arrowDiscussion
down arrowReferences
 
Background and Purpose Transcranial Doppler (TCD) can detect high-intensity transient signals (HITS) in the cerebral circulation. HITS may correspond to artifacts or solid or gaseous emboli. The aim of this study was to develop an offline automated Doppler system allowing the classification of HITS.

Methods We studied 600 HITS in vivo, including 200 artifacts from normal subjects, 200 solid emboli from patients with symptomatic internal carotid artery stenosis, and 200 gaseous emboli in stroke patients with patent foramen ovale. The study was 2-fold, each part involving 300 HITS (100 of each type). The first 300 HITS (learning set) were used to construct an automated classification algorithm. The remaining 300 HITS (validation set) were used to check the validity of this algorithm. To classify HITS, we combined dual-gate TCD with a wavelet representation and compared it with the current "gold standard," the human experts.

Results A combination of the peak frequency of HITS and the time delay makes it possible to separate artifacts from emboli. On the validation set, we achieved a sensitivity of 97%, a specificity of 98%, a positive predictive value (PPV) of 99%, and a negative predictive value (NPV) of 94%. To distinguish between solid and gaseous emboli, where positive refers now to the solid emboli, we used the peak frequency, the relative power, and the envelope symmetry of HITS. On the validation set, we achieved a sensitivity of 89%, a specificity of 86%, a conditional PPV of 89%, and a conditional NPV of 89%.

Conclusions An automated wavelet representation combined with dual-gate TCD can reliably reject artifacts from emboli. From a clinical standpoint, however, this approach has only a fair accuracy in differentiating between solid and gaseous emboli.


Key Words: HITS • solid/gaseous emboli • automated system • dual-gate TCD • wavelets


*    Introduction
up arrowTop
up arrowAbstract
*Introduction
down arrowSubjects and Methods
down arrowResults
down arrowDiscussion
down arrowReferences
 
In recent years, the attention of physicians caring for stroke patients has been drawn to the ability of transcranial Doppler ultrasound (TCD) to detect high-intensity transient signals (HITS) said to be due to microemboli.1 There is increasing evidence that HITS do have clinical significance,211 but the value of HITS in determining patient care and the clinical usefulness of HITS detection are still under discussion.1 Additional information is needed about HITS, mainly on their nature: are they artifacts or solid or gaseous emboli?12

Previous attempts at producing an automated system for classifying HITS into artifacts or emboli have failed to achieve the same level of sensitivity (embolus detection) and specificity (artifact rejection) as the human expert, considered the current "gold standard."1324 Without an automated embolic signal detection system that is as effective as the human expert in terms of sensitivity and specificity, embolus detection remains a time-consuming process, restricted to research studies and unsuitable for clinical practice.25 To separate artifacts from emboli, the best compromise is dual-gate TCD, which consists of tracing the embolus at 2 different depths in the same artery (time delay) (Figure 1). Lately, Cullinane et al21 have shown that the use of a frequency-filtering approach improves the differentiation of artifacts from emboli and that the performance of an online automated system is only slightly below the mean performance of that of a panel of human experts.21,24 Automated systems that can distinguish between solid and gaseous emboli are currently being developed.26,27



View larger version (46K):
[in this window]
[in a new window]
 
Figure 1. The pair of signals belonging to an artifact is recognizable by its low frequency and absence of time delay. For the pair of signals belonging to a gaseous embolus, the time delay between the 2 signals is clearly visible. Note asymmetry of the envelope of the 2 signals. For the pair of signals belonging to a solid embolus, the time delay between the 2 signals is again apparent.

In the present study, we describe the evaluation of a new automated system for the classification of HITS into artifacts or solid or gaseous emboli, based on HITS time-frequency distribution and dual-gate TCD. The rationale for our approach is based on increasing evidence that the fast Fourier transform (FFT), universally used in Doppler devices for spectral analysis of Doppler signals, has notable drawbacks, especially in terms of its poor time-frequency resolution and the large variation in the Doppler spectrum in clinical practice.24,25,28 Consequently, several authors have argued that new time-frequency techniques that avoid the limitations of FFT are needed, especially in the field of embolus detection and classification.24,28,29 In addition, recent pilot studies have suggested that wavelet approaches20,26,27 are particularly well suited to the analysis of short-duration transient signals, such as HITS, and that these new techniques are better than FFT in describing embolic signals. On the basis of our previous pilot study,27 we have tested one such new approach on a large number of HITS in 2 different clinical situations, while a series of artifacts was produced in normal subjects.


*    Subjects and Methods
up arrowTop
up arrowAbstract
up arrowIntroduction
*Subjects and Methods
down arrowResults
down arrowDiscussion
down arrowReferences
 
TCD Recording
All 5 neurology departments involved in this study used the same Doppler device and diagnostic criteria for embolic signal detection as described in this section. TCD signals were recorded with a Doppler device (Multi-Dop X 4, DWL). Each middle cerebral artery (MCA) was simultaneously insonated through the temporal window at 2 different depths (50 and 60 mm). Two 2-MHz pulsed probes (1 for each MCA) were fixed on a special headset connected to the Doppler machine. A setting guaranteeing optimal embolus discrimination from the background spectrum was used according to the criteria recommended in the International Consensus Group on Microembolus Detection.29 The parameters used were TCD-8 software for MDX (DWL), version 8.00K, a scale between -100 and +150 cm/s, a sample volume ranging from 5 to 8 mm in length, a 64-point FFT, an FFT length of 2 ms, an FFT overlap of 60%, and a high-pass filter set at 100 Hz. The separation between the 2 sample volumes was 5 mm (dual-gate TCD). We selected an intensity detection threshold of 9 dB for HITS after this criterion had been validated in each center according to the guidelines previously published,29 whereas a low gain was used to allow the HITS to be completely displayed within the dynamic range of the machine. The pulse repetition frequency was set at 6000 Hz. We wish to stress that we used recordings only of the raw Doppler signals, and not the FFT-transformed signals; in particular, we did not use the relative intensity calculated after the FFT transform.

Role of Human Experts
The experienced observer’s offline analysis of HITS was taken as a standard reference, because the human expert is currently the "gold standard" for embolus detection.1324 In our study, the signals were reviewed by 2 observers, and for inclusion, both observers had to agree that the signal was a true embolic signal. The following definition for embolic signals was used: (1) typical visible, (2) short duration, (3) high-intensity signal, (4) within the Doppler spectrum, and (5) occurring at random within the cardiac cycle. In addition,6 the observers analyzed and assessed the time delay provided by the dual-gate Doppler to the signal of interest. For the differentiation between gaseous and solid emboli, the situation is different: because no expert can decide on the gaseous or solid nature of Doppler signals, we had to preselect patients to know the nature of embolic signals, solid emboli being obtained from high-grade symptomatic internal carotid artery (ICA) stenosis patients and gaseous emboli being collected during the diagnosis of a patent foramen ovale in stroke patients.

Patients and Subjects
Artifacts
Two hundred artifacts were recorded in 2 normal subjects (44±4 years old) who showed normal color Duplex flow imaging (CDFI) of their carotid arteries and MCAs. They did not have cardiac or cerebrovascular disease. In them, we performed a 6-minute series of provoked artifacts,16 including movement of the probe, coughing, sneezing, and head rotation.

Gaseous Emboli
Two hundred gaseous emboli were recorded in 4 patients admitted to our department (CHUV, Lausanne) for stroke associated with a patent foramen ovale. These 4 patients (49±7 years old) had normal carotid arteries and MCAs by both CDFI and MR angiography and had no thrombus lodged in the heart on transesophageal echocardiography. All were taking aspirin at the time of contrast TCD, which was performed 1 month after the stroke to evaluate the size of the patent foramen ovale. The contrast used was 1 mL of air vigorously mixed with 9 mL of saline solution and rapidly injected into an antecubital vein via a 3-way stopcock immediately after contrast preparation.30

Solid Emboli
A pure source of solid emboli was obtained from patients with high-grade symptomatic ICA stenosis without any other disclosed source of embolism. The emboli originating from ICA stenoses are all particulate by default, because no air is able to enter the arterial system.17,19,21,3133 Two hundred solid emboli were recorded in 24 patients (67±9 years old) admitted to the 5 neurology departments for stroke attributed to a severe ICA stenosis (>70% lumen narrowing) diagnosed by both CDFI and MR angiography, in whom no coexisting cause of stroke was found after a complete diagnostic examination, including transesophageal echocardiography. At the time of TCD embolus detection, they were taking aspirin.

All patients had given informed consent.

Automated HITS Analysis
In signal processing, one often reconstructs a signal as a sum of certain elementary functions, called the basis functions. The aim is to represent a fairly complex signal as the sum of just a few basis functions, so as to retain the most relevant parts of the signals and reject superfluous details. Usually, a set of nonredundant basis functions is chosen a priori. Pursuit algorithms are an attempt to free ourselves from this a priori choice and proceed by selecting the elementary functions from redundant dictionaries of time-frequency atoms.34,35 These techniques are sufficiently flexible to build compact representations of complex signals, such as music recordings or biomedical signals.

Selection of HITS
From the recordings, 600 pairs of signals were selected by 2 human experts (the use of the dual-gate technique means that a pair of signals is associated with every HITS). Because of our experimental setup (see the section on patients), it is clear that the selection consisted only of eliminating artifacts from the recordings of patients with circulating emboli. We took care to include a wide variety of signals, ranging from noisy to high-intensity signals. In quantitative terms, the relative intensity (also known as the relative power), expressed in decibels (dB), ranged from 2 to 34 (Table 1). We used the classic definition of relative intensity as 10 log10(mean power of the embolus part of the signal/mean power of the part of the signal without embolus), rather than 10 log10(peak power of the embolus part of the signal/mean power of the part of the signal without embolus). The classic definition is more robust. Its use explains why we have observed HITS intensity of <9 dB (Table 1).


View this table:
[in this window]
[in a new window]
 
Table 1. Relative Power of Artifacts and Gaseous or Solid Emboli

Distinction Between Artifacts and Emboli
To distinguish between artifacts and emboli, we used 2 features: the time delay and the peak frequency. It is widely accepted that the time delay provides useful information but is not absolutely reliable.16,22,23,33 From our extensive experience with TCD signals, we have learned that artifacts usually have lower frequencies than emboli, and we therefore decided to use a novel time-frequency method to calculate the time delay and the main frequency of the signal pairs. For this, we chose to use the matching pursuit performed on a dictionary of Gabor functions. Matching pursuit is related to the now famous wavelet transform34,35 and provides a remarkably compact time-frequency description of a signal. It has a better time-frequency compromise than FFT, ie, it produces a better joint estimate of time and frequency. Gabor functions or atoms are localized sinusoidal waveforms with a symmetrical Gaussian envelope. We found that 4 Gabor atoms were sufficient to represent a signal (Figure 2). To characterize the frequency of artifacts or emboli, we used the frequency of the first Gabor atom. This is a way of estimating what is called the main frequency or peak frequency, ie, the frequency when the signal reaches its peak. For each pair of signals, we then calculated the average peak frequency by averaging over the 2 channels.



View larger version (44K):
[in this window]
[in a new window]
 
Figure 2. Signal of a gas bubble and its time-frequency representation obtained from its Gabor decomposition, whose 4 atoms appear in black. The first atom is fitted to the signal (gray in upper left rectangle). In the next 3 steps of the decomposition, the approximation error, or residual signal, of the previous step appears in gray.

Distinction Between Gaseous and Solid Emboli
The distinction between gaseous and solid emboli is more challenging. The criteria used in a preliminary study with a small data set, namely the instantaneous frequency variation index and envelope variation index,27 did not produce sufficiently good results. We then considered 3 new features: the peak frequency (as above), the relative power (as above), and the envelope symmetry index; in all 3 cases, the value used was the average of the 2 signals forming a pair. The envelope was obtained with a moving window in the time domain. To get the envelope symmetry index, we calculated the average difference between the positive and negative parts of the envelope and then divided this number by the difference between the positive and negative peaks.

Data Analysis
The 600 pairs of signals were divided into 2 sets of 300, each consisting of 100 artifacts, 100 gaseous emboli, and 100 solid emboli. The first data set was used in the elaboration of our classification model, and the second was used to test the adequacy of our model; this division of a data set into a learning set and a validation set is standard statistical practice. The goal of this double-set approach is to prevent overfitting of the model to a "small world," the learning set. As the term implies, "overfitting" of a model is a sure recipe for a disappointing performance outside the learning set. Another way of understanding this important point is to realize that the purpose of our model is to predict the nature of future events, ie, events never seen before, so a prediction model must be tested on new examples, in our case, the validation test. The {kappa} values were determined to measure the level of agreement between the human experts and the automated Doppler system in terms of the classification of HITS into artifacts or gaseous or solid emboli. The {kappa} values were calculated with the formula {kappa}=(observed agreement-expected agreement)/(1-expected agreement).


*    Results
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowSubjects and Methods
*Results
down arrowDiscussion
down arrowReferences
 
Distinction Between Artifacts and Emboli
The average peak frequency is a good indicator of whether a pair of signals is produced by an artifact or an embolus. As can be seen in the histogram (Figure 3), in general, artifacts have a lower frequency than emboli; however, there is a zone of overlap. In this zone, we used the time delay to distinguish between artifacts and emboli. To produce a reasonably robust model, we chose a fairly wide overlap region ranging from 250 to 500 Hz. In this region, we used the time delay, ie, the difference between the start time of the proximal signal and the start time of the distal signal. A pair of signals with a time delay <4 ms was classified as an artifact, otherwise as an embolus. Above 500 Hz, the signals were declared to be emboli, and below 250 Hz they were declared to be artifacts. These critical values (250 Hz, 500 Hz, and 4 ms) were chosen by human judgment over some reasonable range (see Figure 3 for peak frequency and <10 ms for time delay) so as to achieve the best possible differentiation on the learning set before applying them in the validation set. The {kappa} values for the learning set are shown in Table 2, and those for the validation set in Table 3. The latter table shows that 98% of artifacts were recognized as artifacts, which means that the specificity is 98% (Table 4). The sensitivity is 194/200=97% (Table 4). The positive predictive value (PPV) is 194/196=99%, where positive refers to the emboli, and the negative predictive value (NPV) is 98/104=94% (Table 4).



View larger version (27K):
[in this window]
[in a new window]
 
Figure 3. Histograms of peak frequency, in Hz, averaged over the 2 gates. The total number of counts is 100 for each of the 6 histograms.


View this table:
[in this window]
[in a new window]
 
Table 2. Cross-Tabulation of HITS Classification Into Artifacts or Gaseous or Solid Emboli by the Automated Doppler System and Human Experts (Learning Set)


View this table:
[in this window]
[in a new window]
 
Table 3. Cross-Tabulation of HITS Classification Into Artifacts or Gaseous or Solid Emboli by the Automated Doppler System and Human Experts (Validation Set)


View this table:
[in this window]
[in a new window]
 
Table 4. Key Values of the Algorithm to Classify HITS Into Artifacts or Solid or Gaseous Emboli

Distinction Between Gaseous and Solid Emboli
In general, gaseous emboli tend to have a higher relative power and a higher frequency. We thus divided the plane defined by the peak frequency and the relative power into 3 regions. The high-frequency (>1600 Hz) or high-relative-power (>18 dB) region was associated with gaseous emboli, and the low-frequency (<1600 Hz) and low-relative-power (<14 dB) region was associated with solid emboli. The nature of points falling into the third region was decided as follows. If the envelope asymmetry was high (>2%), the corresponding point was classified as belonging to a gaseous embolus, whereas if the asymmetry was low (<1%), it was classified as belonging to a solid embolus; points with an intermediate symmetry were classified as undefined. Again, the critical values (1600 Hz, 14 dB, 18 dB, 1%, and 2%) were chosen by human judgment over some reasonable range (see Figure 4 for peak frequency/relative power and <5% for envelope asymmetry) so as to achieve the best possible differentiation on the learning set before applying them in the validation set. On the validation set, as shown in Table 4, the sensitivity is 89% and the specificity 86%, where the sensitivity is the probability that a solid embolus is declared as such. Conditional on their being classified as emboli, the predictive value for solid emboli (PPV) is 89% and the predictive value for gaseous emboli (NPV) is 89% (for calculating these last 2 numbers, we regarded the outcome "undefined" as "gaseous"). The {kappa} values for the entire table are 0.84 with 89% agreement (P<0.0001), again on the validation set (Table 3). The results on the validation set, ie, on examples that were not used in building the model, are only slightly worse than those on the learning set (Table 2). This shows that the model is not overfitted to the examples of the learning set.



View larger version (32K):
[in this window]
[in a new window]
 
Figure 4. Peak frequency and relative power plane. Circles represent gaseous emboli; crosses, solid emboli. For the learning set, the 3 regions discussed in the text are separated by solid lines.


*    Discussion
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowSubjects and Methods
up arrowResults
*Discussion
down arrowReferences
 
Because there is increasing evidence that time-frequency techniques other than FFT are necessary to classify HITS into artifacts or solid or gaseous emboli, we evaluated a new offline automated system based on a very general approach, the matching pursuit, the usefulness of which was suggested in our pilot study.27 With only 4 characteristic features—peak frequency, time delay, relative power, and envelope symmetry index (Figure 5)—the classification model presented above is fairly efficient. Our results show a high level of agreement between the human experts, still considered the gold standard,1324 and the automated classification of HITS into artifacts or solid or gaseous emboli ({kappa} value >0.8, P<0.0001).



View larger version (15K):
[in this window]
[in a new window]
 
Figure 5. Decision tree with the key parameters to classify HITS into artifacts or solid or gaseous emboli (peak frequency in Hz; time delay in ms; relative power in dB; envelope asymmetry in %). For emboli with envelope asymmetry between 1% and 2%, see text (Results).

To distinguish between artifacts and emboli, our automated system uses only 2 characteristic features, the peak frequency and the time delay (Figure 5), and achieved an excellent correct classification rate of 98% compared with human experts. With a sensitivity of 97% and a specificity of 98% for emboli, the level of artifact elimination obtained is on the order of that required if automated analysis is to replace the trained human observer in the detection of embolic signals. To distinguish between gaseous and solid emboli, our automated system uses 3 features: the peak frequency, the relative power, and the symmetry of the signal envelope (Figure 5). Given the large variability of TCD signals, it is clear that a combination of features is necessary, and this is all the more true for the distinction between gaseous and solid emboli. From a statistical perspective, however, excessive multiplication of features should be resisted,21 because it leads to overfitting. In this initial attempt at building an offline automated system for the classification of HITS into solid and gaseous emboli, we achieved correct classification rates of 87% for solid emboli and 83% for gaseous emboli in the validation set. To discriminate solid emboli from gaseous emboli, we obtained a sensitivity of 89% and a specificity of 86% for solid emboli. Although these results are promising, this performance in discriminating between solid and gaseous emboli can still be improved, but it seems evident that this latter task is more difficult than the separation of emboli from artifacts.17,27

To distinguish between gaseous and solid emboli, we introduced a novel feature, envelope symmetry. One may ask why the envelope of weak gaseous emboli should be less symmetrical than that of strong solid emboli ("weak" and "strong" refer to the intensity of the embolic signals). We offer the following explanation for this empirical observation, an example of which is shown in Figure 1. The attenuation of the incident and reflected ultrasound waves through the tissues is a complex process. If attenuation is strong, the operator will try to compensate for it by increasing the acoustic pressure, and if this is not small relative to the nondisturbed pressure of the medium, nonlinear propagation will occur, resulting in distortion of the pulsed waveforms, including envelope asymmetry.36,37 This distortion is likely to occur with weak emboli, whether gaseous or solid. Fortunately, gaseous emboli have a higher relative power than solid emboli, and overlap in relative power is therefore seen only between weak gaseous signals and strong solid signals, hence the relevance of the envelope symmetry in this overlap region.

In conclusion, our study shows that an automated system combining a time-frequency technique and dual-gate TCD and using only 4 features can classify HITS into artifacts or solid or gaseous emboli. It also demonstrates that the solid/gas classification is more challenging than the embolus/artifact classification.

Received May 21, 2001; revision received August 24, 2001; accepted August 28, 2001.


*    References
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowSubjects and Methods
up arrowResults
up arrowDiscussion
*References
 

  1. Easton JD. Cerebral embolism and Doppler ultrasound. Cerebrovasc Dis. 1999; 9: 188–192.[Medline] [Order article via Infotrieve]
  2. Levi CR, O’Malley HM, Fell G, Roberts AK, Hoare MC, Royle JP, Chan A, Beiles BC, Chambers BR, Bladin CF, Donnan GA. Transcranial Doppler detected cerebral microembolism following carotid endarterectomy: high microembolic signals loads predict postoperative cerebral ischemia. Brain. 1997; 120: 621–629.[Abstract/Free Full Text]
  3. Ackerstaff RG, Moons KG, van de Vlasakker CJ, Moll FL, Vermeulen FE, Algra A, Spencer MP. Association of intraoperative transcranial Doppler monitoring variables with stroke from carotid endarterectomy. Stroke. 2000; 31: 1817–1823.[Abstract/Free Full Text]
  4. Gallagher EG, Pearson DT. Ultrasonic identification of sources of gaseous microemboli during open heart surgery. Thorax. 1973; 28: 295–305.[Medline] [Order article via Infotrieve]
  5. Markus H. Doppler embolus detection: stroke treatment and prevention.In: Neurosonology. New York, NY: Mosby; 1996: 239–251.
  6. Molloy J, Markus HS. Asymptomatic embolization predicts stroke and TIA risk in patients with carotid artery stenosis. Stroke. 1999; 30: 1440–1443.[Abstract/Free Full Text]
  7. Siebler M, Nachtmann A, Sitzer M, Rose G, Kleischmidt A, Rademacher J, Steinmetz H. Cerebral microembolism and the risk of ischemia in asymptomatic high-grade internal carotid artery stenosis. Stroke. 1995; 26: 2184–2186.[Abstract/Free Full Text]
  8. Valton L, Larrue V, Pavy le Traon A, Massabuau P, Geraud G. Microembolic signals and risk of early recurrence in patients with stroke or transient ischemic attack. Stroke. 1998; 29: 2125–2128.[Abstract/Free Full Text]
  9. Sitzer M, Muller W, Siebler M, Hort W, Kniemeyer HW, Jancke L, Steinmetz H. Plaque ulceration and lumen thrombus are the main source of cerebral microemboli in high-grade internal carotid artery stenosis. Stroke. 1995; 26: 1231–1233.[Abstract/Free Full Text]
  10. Sitzer M, Siebler M, Steinmetz H. Silent emboli and their relation to clinical symptoms in extracranial carotid artery disease. Cerebrovasc Dis. 1995; 5: 121–123.
  11. Ries S, Schminke U, Daffertshofer M, Hennerici M. High intensity transient signals and carotid artery disease. Cerebrovasc Dis. 1995; 5: 124–127.
  12. Babikian VL, Cantelmo NL. Cerebrovascular monitoring during carotid endarterectomy. Stroke. 2000; 31: 1799–1801.[Free Full Text]
  13. Siebler M, Rose G, Sitzer M, Bender A, Steinmetz H. Real-time identification of cerebral microemboli with US feature detection by a neural network. Radiology. 1994; 192: 739–742.[Abstract/Free Full Text]
  14. Van Zuilen EV, Mess WH, Jansen C, Van Der Tweel I, Van Gijn J, Ackerstaff RGA. Automatic embolus detection compared with human experts: a Doppler ultrasound study. Stroke. 1996; 27: 1840–1843.[Abstract/Free Full Text]
  15. Markus HS, Ackerstaff R, Babikian V, Bladin C, Droste D, Grosset D, Levi C, Russel D, Siebler M, Tegeler C. Intercenter agreement in reading Doppler embolic signals: a multicenter international study. Stroke. 1997; 28: 1307–1310.[Abstract/Free Full Text]
  16. Droste DW, Hagedorn G, Notzold A, Siemens HJ, Sievers HH, Kaps M. Bigated transcranial Doppler for the detection of clinically silent circulating emboli in normal persons and patients with prosthetic cardiac valves. Stroke. 1997; 28: 588–592.[Abstract/Free Full Text]
  17. Smith J, Evans DH, Bell PRF, Naylor R. A comparison of four methods for distinguishing Doppler signals from gaseous and particulate emboli. Stroke. 1998; 29: 1133–1138.[Abstract/Free Full Text]
  18. Kemény V, Droste DW, Hermes S, Nabavi DG, Schulte-Altedorneburg G, Siebler M, Ringelstein EB. Automatic embolus detection by a neural network. Stroke. 1999; 30: 807–810.[Abstract/Free Full Text]
  19. Markus HS, Reid G. Frequency filtering improves ultrasonic embolic signal detection. Ultrasound Med Biol. 1999; 25: 857–860.[Medline] [Order article via Infotrieve]
  20. Aydin N, Padayachee S, Markus HS. The use of the wavelet transform to describe embolic signals. Ultrasound Med Biol. 1999; 25: 953–958.[Medline] [Order article via Infotrieve]
  21. Cullinane M, Reid G, Dittrich R, Kaposzta Z, Ackerstaff R, Babikian V, Droste DW, Grosset D, Siebler M, Valton L, Markus HS. Evaluation of new online automated embolic signal detection algorithm, including comparison with panel of international experts. Stroke. 2000; 31: 1335–1341.[Abstract/Free Full Text]
  22. Mess WH, Titulaer BM, Ackerstaff RG. Discrimination and characterization of emboli: old and new aspects.In: Klingelhöfer J, ed. New Trends in Cerebral Hemodynamics and Neurosonology. New York, NY: Elsevier Science; 1997: 355–363.
  23. Mess WH, Titulaer BM, Ackerstaff RGA. A new algorithm for on-line automated emboli detection based on the pseudo-Wigner distribution and the dual gate TCD technique. Ultrasound Med Biol. 2000; 26: 413–418.[Medline] [Order article via Infotrieve]
  24. Markus H, Culliname M, Reid G. Improved automated detection of embolic signals using a novel frequency filtering approach. Stroke. 1999; 30: 1610–1615.[Abstract/Free Full Text]
  25. Markus HS. Monitoring embolism in real time. Circulation. 2000; 102: 826–828.[Free Full Text]
  26. Brucher R, Russell D. Improved discrimination of microembolic events by combining dual frequency Doppler with wavelet transformation. Cerebrovasc Dis. 1999; 9: 55.Abstract.
  27. Devuyst G, Vesin J-M, Despland P-A, Bogousslavsky J. The matching pursuit: a new method of characterizing microembolic signals? Ultrasound Med Biol. 2000; 6: 1051–1056.
  28. Evans DH. Doppler signal analysis. Ultrasound Med Biol. 2000; S1: S13–S15.
  29. Ringelstein EB, Droste DW, Babikian VL, Evans DH, Grosset DJ, Kaps M, Markus HS, Russell D, Siebler M. Consensus on microembolus detection by TCD. Stroke. 1998; 29: 725–729.[Abstract/Free Full Text]
  30. Teague SM, Sharma MK. Detection of paradoxical cerebral echo contrast embolization by transcranial Doppler ultrasound. Stroke. 1991; 22: 740–745.[Abstract/Free Full Text]
  31. Georgiadis D, Mackay TG, Kelman AW, Grosset DG, Wheatley DJ, Lees KR. Differentiation between gaseous and formed embolic materials in vivo: application in prosthetic heart valve patients. Stroke. 1994; 25: 1559–1563.[Abstract]
  32. Markus SH, Tegeler CH. Experimental aspects of high-intensity transient signals in the detection of emboli. J Clin Ultrasound. 1995; 23: 81–87.[Medline] [Order article via Infotrieve]
  33. Molloy J, Markus HS. Multigated Doppler ultrasound in the detection of emboli in a flow model and embolic signals in patients. Stroke. 1996; 27: 1548–1552.[Abstract/Free Full Text]
  34. Mallat S, Zhang Z. Matching pursuit with time-frequency dictionaries. IEEE Trans Signal Processing. 1993; SP-41: 3397–3415.
  35. Mallat S. A wavelet tour of signal processing. San Diego, Calif: Academic Press; 1998.
  36. Evans DH, McDicken WN. Doppler Ultrasound: Physics, Instrumentation and Signal Processing. 2nd ed. New York, NY: Wiley; 2000.
  37. Jensen JA. Estimation of Blood Velocities Using Ultrasound: A Signal Processing Approach. Cambridge, UK: Cambridge University Press; 1996.



This article has been cited by other articles:


Home page
PerfusionHome page
J. Lynch and J. Riley
Microemboli detection on extracorporeal bypass circuitsa
Perfusion, January 1, 2008; 23(1): 23 - 32.
[Abstract] [PDF]


Home page
StrokeHome page
M. Saqqur, N. Dean, M. Schebel, M. D. Hill, A. Salam, A. Shuaib, and A. M. Demchuk
Improved Detection of Microbubble Signals Using Power M-Mode Doppler
Stroke, January 1, 2004; 35 (1): e14 - e17.
[Abstract] [Full Text] [PDF]


Home page
StrokeHome page
R. Brucher and D. Russell
Automatic Online Embolus Detection and Artifact Rejection With the First Multifrequency Transcranial Doppler
Stroke, August 1, 2002; 33(8): 1969 - 1974.
[Abstract] [Full Text] [PDF]


Home page
StrokeHome page
D. Russell and R. Brucher
Online Automatic Discrimination Between Solid and Gaseous Cerebral Microemboli With the First Multifrequency Transcranial Doppler
Stroke, August 1, 2002; 33(8): 1975 - 1980.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow Request Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Devuyst, G.
Right arrow Articles by Bogousslavsky, J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Devuyst, G.
Right arrow Articles by Bogousslavsky, J.
Related Collections
Right arrow Doppler ultrasound, Transcranial Doppler etc.
Right arrow Pathology of Stroke
Right arrow Primary and Secondary Stroke Prevention