(Stroke. 2001;32:2055.)
© 2001 American Heart Association, Inc.
Original Contributions |
From the Department of Neurosurgery, Nagoya City University Medical School, Nagoya, Japan.
Correspondence to Atsushi Umemura, MD, Department of Neurosurgery, Nagoya City University Medical School, 1 Kawasumi, Mizuho-ku, Nagoya 467-8601, Japan. E-mail aume{at}med.nagoya-cu.ac.jp
| Abstract |
|---|
|
|
|---|
Methods Sixty patients with ischemic cerebrovascular disease were examined. Conventional B-mode imaging and blood flow images by power Doppler imaging (PDI) and B-flow were obtained.
Results B-flow enabled simultaneous imaging of tissue and blood flow. Compared with PDI, B-flow provided higher spatial resolution and higher frame rate hemodynamic imaging. Consequently, a clear definition of the vessel lumen was obtained even in the stenotic portion of the carotid artery. In evaluating the degree of carotid stenosis, B-flow measurement agreed with digital subtraction angiography as well as PDI measurement. In addition, bloodstream swirl at the site of wall ulceration or vascular elongation was observed.
Conclusions B-flow was prominent in visualizing hemodynamic flow and detection of stenotic lesion in the cervical carotid artery. Combined with conventional B-mode technique, B-flow seems to be useful in evaluating carotid stenosis.
Key Words: carotid arteries carotid stenosis diagnostic imaging ultrasonography
| Introduction |
|---|
|
|
|---|
Recently, a new method for the detection of blood flow by ultrasound (B-flow) has been developed.68 With this technology, direct visualization of blood reflectors has made B-mode flow imaging possible without the limitations of Doppler technology such as aliasing, signal dropout at orthogonal detection angles, and wall filter limitations. B-flow visualizes real-time hemodynamic flow in relation to stationary tissue. We evaluated the efficacy of B-flow in examining the cervical carotid artery in patients with ischemic cerebrovascular disease.
| Subjects and Methods |
|---|
|
|
|---|
Cervical carotid arteries were examined in all patients using LOGIC 700 ultrasound device (GE Medical Systems) with a 6.5-MHz probe focusing on the level around the carotid bifurcation. Routine B-mode and blood flow images by PDI and B-flow were obtained. Sequential parallel longitudinal views of flow were displayed. We closely watched vessel walls and hemodynamic flow simultaneously. We calculated the degree of stenosis by measurements of luminal diameters contrasted by power Doppler color signals or B-flow signals.
Selective digital subtraction angiography (DSA) of the carotid artery was performed in 15 patients who were suspected of having carotid stenosis on ultrasonography. The degree of carotid stenosis was derived from the minimum internal carotid and the common carotid luminal diameters, because the common carotid artery is more easily visualized in ultrasonography.9 Simple regression analysis was used for statistical evaluation of the correlation between measurements of stenosis by sonographic and angiographic methods.
| Results |
|---|
|
|
|---|
|
B-flow provided clearer definition of the vessel lumen even in the carotid stenosis (Figure 2). PDI tended to overestimate the normal vessel lumen and underestimate the vessel lumen of the severely stenotic portion.
|
Fifteen patients were suspected of having carotid stenosis and underwent DSA. Measurement of the percentage of stenosis in both sonographic techniques (B-flow and PDI) and DSA correlated significantly (P<0.0001; Figure 3). However, the coefficient of correlation was higher for DSA versus B-flow (r=0.977) than for DSA versus PDI (r=0.954). Compared with B-flow, PDI has a tendency to underestimate the percentage of stenosis in severely stenotic lesions and overestimate the percentage of stenosis in mildly stenotic lesion.
|
Bloodstream swirl was clearly observed at the site of wall ulceration (Figure 4) or vascular tortuousness (Figure 5) on B-flow in 6 patients with symptomatic or asymptomatic cerebral infarction. Such complex hemodynamics seems to relate to the formation of thrombus and the occurrence of stroke.
|
|
| Discussion |
|---|
|
|
|---|
B-flow technology was achieved by General Electrics digitally encoded ultrasound. Coded sound waves are transmitted into the body and vasculature, and the returning signals are then decoded and displayed, as in B-mode. This technology enables boosting of weak flow reflector signals from blood reflectors and suppresses unwanted signals and frequencies from tissue. Consequently, B-flow can visualize real-time hemodynamic flow in relation to stationary tissue. In fact, B-flow provided higher spatial resolution for demonstrations of vessel anatomy and higher frame rate hemodynamic imaging than PDI.
In evaluating the degree of carotid stenosis, B-flow measurement agreed with DSA as well as PDI measurement. Both methods obtained much better correlation from longitudinal views than results shown in other studies.35 Although this good correlation may be influenced by the small sample bias, we feel it is mainly due to meticulous examination. In addition, the coefficient of correlation was slightly higher for DSA versus B-flow than for DSA versus PDI. This result is due to clearer definition of the vessel lumen without the overwriting seen with B-flow. In fact, the B-flow hemodynamic image resembles that of angiography. In PDI, overestimation of the normal vessel lumen is caused by overwriting, and underestimation of the stenotic vessel lumen is caused by limitations in detecting flow. Consequently, PDI underestimated the percentage of stenosis in severely stenotic lesions and overestimated the percentage of stenosis in mildly stenotic lesions.
It must be emphasized that B-flow provides a detailed hemodynamic image of phenomena such as bloodstream swirl. Visualization of such a complex flow pattern has not been achieved by CDFI or PDI. Thrombus formed by atherosclerosis is important in considering a cause of cerebral infarction. In addition, B-flow visualized bloodstream swirl at the site of wall ulceration or vascular tortuousness. This complex flow pattern might be related to shear stressinduced platelet aggregation and become another cause of thrombus formation in stroke patients.10 Careful analysis of flow patterns at the carotid bifurcation in relation to the pathogenesis of ischemic cerebrovascular disease will be the subject of further study.
In conclusion, B-flow is highly effective in visualizing hemodynamic flow and in detecting stenotic lesions in the cervical carotid artery. Combined with conventional B-mode technique, B-flow seems to be useful in evaluating carotid stenosis, especially in patients with ischemic cerebrovascular disease.
Received November 1, 2000; accepted June 8, 2001.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. Magnoni, S. Coli, M. M. Marrocco-Trischitta, G. Melisurgo, D. De Dominicis, D. Cianflone, R. Chiesa, S. B. Feinstein, and A. Maseri Contrast-enhanced ultrasound imaging of periadventitial vasa vasorum in human carotid arteries Eur J Echocardiogr, August 29, 2008; (2008) jen221v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Reiter, R. Horvat, S. Puchner, W. Rinner, P. Polterauer, J. Lammer, E. Minar, and R.A Bucek Plaque Imaging of the Internal Carotid Artery--Correlation of B-Flow Imaging with Histopathology AJNR Am. J. Neuroradiol., January 1, 2007; 28(1): 122 - 126. [Abstract] [Full Text] [PDF] |
||||
![]() |
S.O. Oktar, C. Yucel, D. Karaosmanoglu, K. Akkan, H. Ozdemir, N. Tokgoz, and T. Tali Blood-Flow Volume Quantification in Internal Carotid and Vertebral Arteries: Comparison of 3 Different Ultrasound Techniques with Phase-Contrast MR Imaging. AJNR Am. J. Neuroradiol., February 1, 2006; 27(2): 363 - 369. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Yurdakul, M. Tola, and T. Cumhur B-flow Imaging for Assessment of 70% to 99% Internal Carotid Artery Stenosis Based on Residual Lumen Diameter J. Ultrasound Med., February 1, 2006; 25(2): 211 - 215. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Tola, M. Yurdakul, and T. Cumhur B-Flow Imaging in Low Cervical Internal Carotid Artery Dissection J. Ultrasound Med., November 1, 2005; 24(11): 1497 - 1502. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Yucel, S. O. Oktar, Y. Erten, A. Bursali, and H. Ozdemir B-Flow Sonographic Evaluation of Hemodialysis Fistulas: A Comparison With Low- and High-Pulse Repetition Frequency Color and Power Doppler Sonography J. Ultrasound Med., November 1, 2005; 24(11): 1503 - 1508. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. G. Grant, C. B. Benson, G. L. Moneta, A. V. Alexandrov, J. D. Baker, E. I. Bluth, B. A. Carroll, M. Eliasziw, J. Gocke, B. S. Hertzberg, et al. Carotid Artery Stenosis: Gray-Scale and Doppler US Diagnosis--Society of Radiologists in Ultrasound Consensus Conference Radiology, November 1, 2003; 229(2): 340 - 346. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|