(Stroke. 2002;33:293.)
© 2002 American Heart Association, Inc.
Short Communications |
From the Second Department of Medicine (H.W.S., W.W.), Department of Pediatrics (A.B., J-I.S., A.G.), and Division of Medical Engineering and Computing, Department of Surgery (P.R.), Landeskrankenhaus-Universitätsklinikum Graz, Graz, Austria.
Correspondence to Herwig W Schuchlenz, MD, 2 Medizinische Abteilung, LKH-Universitätsklinikum Graz, Auenbruggerplatz 15, A-8036 Graz, Austria. E-mail herwig.schuchlenz{at}klinikum-graz.at
Abstract
Background and Purpose Patent foramen ovale (PFO) is a risk factor for paradoxical embolism, and severe shunting and wide opening of PFO are risk factors for severe and recurrent cerebrovascular events. Neither contrast echocardiography nor 2-dimensional (2D) measurement of PFO size have been validated or compared with invasive balloon sizing.
Methods We performed transesophageal echocardiography (TEE) in 100 patients with cryptogenic stroke and catheter closure of PFO. The amount of contrast shunting through the PFO during cubital and femoral contrast delivery and the PFO size measured by 2D TEE were compared with balloon sizing.
Results There was a significant correlation (r2=0.8; P<0.0001) between 2D TEE measurement and invasive balloon sizing. Mean balloon-sized PFO diameter was significantly larger than mean PFO diameter measured by 2D TEE (8.3±2.6 versus 5.2±1.7 mm). Semiquantitative contrast TEE correlated with PFO size (r2=0.7; P<0.0001) only if the contrast agent was administered through a femoral vein. Correlation was poor when the contrast agent was administered via a cubital vein.
Conclusions We conclude that 2D TEE measurement of a PFO size is more accurate than the traditionally used contrast technique.
Key Words: cerebral ischemia contrast media echocardiography, transesophageal embolism, paradoxical foramen ovale, patent
The prevalence of patent foramen ovale (PFO) is higher in patients with unexplained cerebrovascular events than in the general population (up to 75% versus 25% to 30%).13 Additionally, patients with PFO carry risk for recurrent cerebrovascular events in the range of 1.7% to 4.7% per year.4 Recent studies have demonstrated that the size of a PFO as estimated by 2-dimensional (2D) transesophageal echocardiography (TEE) affects the risk of stroke.3,5,6 Mean PFO diameter was significantly larger (4±2 mm) in patients with cryptogenic cerebrovascular events than in controls (2±1 mm) or in patients with strokes of defined cause.3,6 These data emphasize the importance of 2D measurements of PFO size. Thus far, neither 2D measurement of PFO size by TEE nor semiquantitative assessment by contrast echocardiography has been compared with invasive balloon sizing. Usually, PFO size is assessed semiquantitatively by transthoracic echocardiography, by TEE, or by transcranial Doppler ultrasonography on the basis of the number of contrast microbubbles appearing in the left atrium or cerebral circulation after injection of contrast agent into a cubital vein. However, in addition to the size of a PFO, the magnitude of the contrast shunt is influenced by the position of the patient; the choice, dosage, and route of administration of the contrast agent; the provocative maneuvers used; and patient compliance in performing provocative maneuvers.710 Furthermore, anatomic and functional variations of the right atrial inflow area8 and pathological conditions influence the amount of contrast shunting through a PFO.11 Nonetheless, semiquantitative grading systems of PFO size are used for diagnosis, clinical decision making, and treatment recommendations.4,1215 Direct evaluation of the PFO size may be more independent of these factors and was shown to be an independent indicator of stroke.3 The purpose of this study was to investigate the diagnostic accuracy of TEE for assessing PFO size by 2D measurement or semiquantitative contrast echocardiography.
Subjects and Methods
We studied 100 consecutive patients (mean age, 37.4±11.7 years; range, 13 to 69 years; 62 males and 38 females) with a PFO and a presumed paradoxical cerebrovascular event who were undergoing cardiac catheterization for closure of the PFO with a transcatheter device. The indications for intervention were recurrent presumed paradoxical cerebrovascular events despite medical treatment, complications or contraindications for anticoagulation, or patient preference for this treatment.
Transesophageal Echocardiography
TEE was performed with a 5-MHz phased multiplane probe (Sonos 2500, Hewlett Packard, or Vingmed System Five, General Electric). In addition to the standard views, the region of the fossa ovalis was studied to detect a separation between the septum primum and the septum secundum. The maximum opening diameter of the communicating channel was measured at the entrance into the left atrium, online and offline, by 2 different observers using the equipment software. Two contrast studies (D-galactose, Echovist, Shering Berlin) were performed in each patient, administered into an antecubital vein, during normal inspiration and with a Valsalva maneuver. Right-to-left shunting was graded as minimal if only a few bubbles passed through the PFO, moderate if a cloud of contrast bubbles was documented in the left atrium, or severe if there was opacification of the left atrium (Figure 1).
|
Cardiac Catheterization (Intervention)
The right femoral vein was accessed with a standard 7F introducer, and the echo contrast agent was administered through this port. A Radiofocus guidewire (Terumo Corporation) was used to cross the PFO. Thereafter, the PFO was passed under TEE and fluoroscopic guidance with a 6F Lehman right heart catheter (Cook). Pressures were recorded in the right heart and the left atrium. An extra-stiff Amplatz exchange guidewire (Cook) was placed in the left upper pulmonary vein, and an occlusion balloon catheter (MEDI-TECH, Boston Scientific) was passed over the guidewire and inflated with dilute contrast material. The maximum volume that could be passed from right to left was calibrated with a sizing template. The observers who estimated PFO size by the invasive technique were blinded to PFO size measured by TEE.
Statistical Analysis
Correlations between PFO diameters determined invasively and by TEE were assessed with linear regression analysis; differences were compared with the paired Wilcoxon test. Semiquantitative contrast studies and PFO diameter measured by TEE or by balloon sizing were compared with nonparametric ANOVA (Kruskal-Wallis). Agreement of contrast TEE studies (femoral versus cubital contrast administration and interobserver variability) was analyzed by Cohenss weighted
. A value of P<0.05 was considered statistically significant. The relative influence of variables, including age, weight, size, sex, atrial septal aneurysm, cubital or femoral contrast TEE, and left or right atrial pressure, on PFO diameter measured by 2D TEE was analyzed with multiple linear regression.
Results
There was a strong correlation between 2D TEE measurements and balloon sizing (r2=0.8; P<0.0001 by linear regression). The 2D TEE measurement of PFO size was related to semiquantitative contrast TEE via the femoral route (r2=0.75; P<0.0001) (Figure 2A) but not to semiquantitative contrast TEE via the cubital route (Figure 2B). The mean PFO diameter measured by 2D TEE (5.2±1.7 mm; range, 2 to 10 mm) was significantly smaller than that measured by balloon (8.3±2.6 mm; range, 4 to 14 mm; P<0.0001, Wilcoxon test). Multiple linear regression analysis showed a significant association between PFO diameter measured by 2D TEE, atrial septal aneurysm (r2=0.6; P<0.0001), and femoral contrast TEE (r2=0.75; P<0.0001). There was no significant association between PFO size as measured with 2D TEE or balloon sizing and cubital contrast TEE, left or right atrial pressure, age, weight, or body size. Mean right atrial pressure was significantly lower than mean left atrial pressure (5.4±2.7 mm Hg; range, 0 to 14 mm Hg versus 7.6±2.9 mm Hg; range, 2 to 16 mm Hg).
|
Interobserver agreement for evaluation of PFO size was 0.9 for 2D TEE and 0.8 for semiquantitative contrast echocardiography (weighted
correlation coefficient).
Discussion
We found a significant correlation between PFO size measured by 2D TEE and measured by balloon sizing. There was no correlation between the results of these 2 modalities and the results obtained by the traditional semiquantitative contrast echo method. A significant correlation between the amount of right-to-left contrast shunting and PFO diameter was found only if the contrast agent was administered through a femoral vein.
Some investigators have stated that the opening diameter and the volume shunted through this valvelike structure are related mainly to an increase in right atrial pressure.16,17 This is consistent with the increased prevalence of PFO in patients with chronic obstructive lung disease and pulmonary hypertension.18 However, elevated right heart pressures were not present in our study population. Our observations confirm that the mode of administration is a major determinant of contrast agent shunting through a PFO. Because blood flow from the inferior vena cava is directed against the fossa ovalis8 and blood from the superior vena cava is directed primarily through the tricuspid valve, contrast agent entering the right atrium from the superior vena cava can bypass the interatrial septum or a PFO, thus causing a false-negative or weak contrast study.8
However, the methods we used (2D TEE and balloon sizing) also have their limitations. The patent foramen is a flaplike valve that is formed by overlapping of the relatively thick septum secundum and the relatively thin, mobile, and compliant septum primum. These 2 septa form a tunnel-like ellipsoid communication between the right and the left atrium. Measurement of this complex 3-dimensional oval channel by 2D TEE is usually performed in the shorter axis because it is difficult to obtain the long axis of this oval communication with this technique. Because of this limitation, the true orifice may be underestimated by 2D TEE. Invasive measurements of PFO and atrial septal defects with the use of relatively stiff guidewires and sizing balloons have demonstrated significant compliance of the septum primum and resolution of the overlap. This technique makes the oval defect more circular, causing an elongation of the shorter axis and a foreshortening of the longer axis. As a result of the softness and distensibility of the septum primum, there is a potential of overestimating the size of the PFO by the balloon sizing technique.19,20 Recently, 3-dimensional echocardiography has been shown to be superior to 2D echocardiography in measurement of atrial septal defects.21,22 It remains to be seen whether this technique can provide additional information on PFO quantification.
Most studies have used saline contrast with transthoracic echocardiography, transcranial Doppler,2,5,7,9,23 and transmitral Doppler24 to detect or quantify right-to-left shunts. A limitation of saline contrast is that bubble size is not standardized. Backscattered signal power is dependent on bubble size, which for agitated saline contrast varies from 24 to 144 µm, and its value for shunt quantification is questionable. If bubble size is not standardized, there is no simple relationship between returned signal power and bubble number.24,25 Moreover, the large bubbles of the widely used air/saline mixture entail a risk of adverse effects in patients with a right-to-left shunt because bubbles as small as 10 µm can obstruct the microcirculation.26,27 We performed the contrast studies with a commercially available contrast agent with an exactly defined bubble size (99% <12 µm, 95% <8 µm). Therefore, we do not know whether the results of the correlation between PFO size and the contrast agent we used are pertinent to saline.
Our study probably has a selection bias toward patients with large PFOs. The mean PFO diameter measured by 2D TEE in our series was 5.2 mm, which is more than twice the diameter found in control subjects without presumed paradoxical cerebrovascular events.5,7
In conclusion, multiplane 2D TEE sizing of PFO is as reliable as femorally administrated semiquantitative contrast TEE or balloon sizing. Because balloon sizing or femoral contrast administration is invasive and not applicable in routine practice, 2D TEE should be used to quantify a PFO. A small contrast shunt does not rule out a large PFO if the contrast agent is administrated through a cubital vein. Further studies of the risk of PFO or of treatment strategies in patients after presumed paradoxical embolism should include PFO diameter determined by 2D TEE instead of the traditionally used contrast method.
Received August 3, 2001; revision received October 11, 2001; accepted October 12, 2001.
References
1. Hagen PT, Scholz DG, Edwards WD. Incidence and size of patent foramen ovale during the first 10 decades of life: an autopsy study of 965 normal hearts. Mayo Clin Proc. 1984; 59: 1720.[Medline] [Order article via Infotrieve]
2. Lechat P, Mas JL, Lascault G, Loron P, Theard M, Klimczac M, Drobinski G, Thomas D, Grosgogeat Y. Prevalence of patent foramen ovale in patients with stroke. N Engl J Med. 1988; 318: 11481152.[Abstract]
3. Schuchlenz HW, Weihs W, Horner S, Quehenberger F. The association between the diameter of a patent foramen ovale and the risk of embolic cerebrovascular events. Am J Med. 2000; 109: 456462.[CrossRef][Medline] [Order article via Infotrieve]
4. Nendaz MR, Sarasin FP, Junod AF, Bogousslavsky J. Preventing stroke recurrence in patients with patent foramen ovale: antithrombotic therapy, foramen closure, or therapeutic abstention? A decision analytic perspective. Am Heart J. 1998; 135: 532541.[CrossRef][Medline] [Order article via Infotrieve]
5. Homma S, Di Tullio MR, Sacco RL, Mihalatos D, Mandri GL, Mohr JP. Characteristics of patent foramen ovale associated with cryptogenic stroke: a biplane transesophageal echocardiographic study. Stroke. 1994; 25: 582586.[Abstract]
6.
Steiner MM, Di Tullio MR, Rundeck T, Gan R, Chen X, Liguori C, Brainin M, Homma S, Ralph L, Sacco RL. Patent foramen ovale size and embolic brain imaging findings among patients with ischemic stroke. Stroke. 1998; 29: 944948.
7.
Schwarze JJ, Sander D, Kukla C, Wittich I, Babikian VL, Klingelhofer J. Methodological parameters influence the detection of right-to-left shunts by contrast transcranial Doppler ultrasonography. Stroke. 1999; 30: 12341239.
8. Gin KG, Huckell VF, Pollick C. Femoral vein delivery of contrast medium enhances transthoracic echocardiographic detection of patent foramen ovale. J Am Coll Cardiol. 1993; 22: 19942000.[Abstract]
9.
Droste DW, Kriete JU, Stypman J, Castrucci M, Wichter T, Tietje R, Weltermann B, Young P, Ringelstein EB. Contrast transcranial Doppler ultrasound in the detection of right-to-left shunts: comparison of different procedures and different contrast agents. Stroke. 1999; 30: 18271832.
10. Fan S, Nagai T, Luo H, Atar S, Naqvi T, Birnbaum Y, Lee S, Siegel RJ. Superiority of the combination of blood and agitated saline for routine contrast enhancement. J Am Soc Echocardiogr. 1999; 12: 9498.[CrossRef][Medline] [Order article via Infotrieve]
11. Keidar S, Grenadier E, Binenboim C, Palant A. Transient right to left atrial shunt detected by contrast echocardiography in the acute stage of pulmonary embolism. J Clin Ultrasound. 1984; 12: 417419.[Medline] [Order article via Infotrieve]
12.
Cheitlin MD, Alpert JS, Armstrong WF, Aurigemma GP, Beller GA, Bierman FZ, Davidson TW, Davis JL, Douglas PS, Gillam LD. ACC/AHA Guidelines for the Clinical Application of Echocardiography: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Committee on Clinical Application of Echocardiography) developed in collaboration with the American Society of Echocardiography. Circulation. 1997; 95: 16861744.
13.
Chambers J. Should percutaneous devices be used to close a patent foramen ovale after cerebral infarction? Heart. 1999; 82: 537538.
14. Alp N, Clarke N, Banning AP. How should patients with patent foramen ovale be managed? Heart. 2001; 88: 242244.
15.
Lock JE. Patent foramen ovale is indicted, but the case hasnt gone to trial. Circulation. 2000; 101: 838.
16. De Castro S, Cartoni D, Conti G, Beni S. Continuous monitoring by biplane transesophageal echocardiography of pulmonary and paradoxical embolism. J Am Soc Echocardiogr. 1995; 8: 217220.[CrossRef][Medline] [Order article via Infotrieve]
17.
Cheng TO. Impending paradoxical embolism. Circulation. 2000; 101: e226.
18. Soliman A, Shanoudy H, Liu J, Russell DC, Jarmukli NF. Increased prevalence of patent foramen ovale in patients with severe chronic obstructive pulmonary disease. J Am Soc Echocardiogr. 1999; 12: 99105.[CrossRef][Medline] [Order article via Infotrieve]
19.
Bridges ND, Hellenbrand W, Latson L, Filiano J, Newburger JW, Lock JE. Transcatheter closure of patent foramen ovale after presumed paradoxical embolism. Circulation. 1992; 86: 19021908.
20. Marshall AC, Lock JE. Structural and compliant anatomy of the patent foramen ovale in patients undergoing transcatheter closure. Am Heart J. 2000; 140: 303307.[CrossRef][Medline] [Order article via Infotrieve]
21. Maeno YV, Benson LN, Boutin C. Impact of dynamic 3D transesophageal echocardiography in the assessment of atrial septal defects and occlusion by the double-umbrella device (CardioSEAL). Cardiol Young. 1998; 8: 368378.[Medline] [Order article via Infotrieve]
22. Zhu W, Cao QL, Rhodes J, Hijazi ZM. Measurement of atrial septal defect size: a comparative study between three-dimensional transesophageal echocardiography and the standard balloon sizing methods. Pediatr Cardiol. 2000; 21: 465469.[CrossRef][Medline] [Order article via Infotrieve]
23.
Chimowitz MI, Nemec JJ, Marwick TH, Lorig RJ, Furlan AJ, Salcedo EE. Transcranial Doppler ultrasound identifies patients with right-to-left cardiac or pulmonary shunts. Neurology. 1991; 41: 19021904.
24.
Kerr AJ, Buck T, Chia K, Chow CM, Fox E, Levine RA, Picard MH. Transmitral Doppler: a new transthoracic contrast method for patent foramen ovale detection and quantification. J Am Coll Cardiol. 2000; 36: 19591966.
25.
Tegue SM, Sharma MK. Detection of paradoxical cerebral echo contrast embolization by transcranial Doppler ultrasound. Stroke. 1991; 22: 740745.
26.
Srivastava TN, Undesser EK. Transient ischemic attack after air-contrast echocardiography in patients with septal aneurysm. Ann Intern Med. 1995; 122: 731732.
27. Feinstein SB, Shah PM, Bing RJ, Meerbaum S, Corday E, Chang BL, Santillan G, Fujibayashi Y. Microbubble dynamics visualized in the intact capillary circulation. J Am Coll Cardiol. 1984; 4: 595600.[Abstract]
This article has been cited by other articles:
![]() |
S. R. Messe and S. E. Kasner Patent Foramen Ovale in Cryptogenic Stroke: Not to Close Circulation, November 4, 2008; 118(19): 1999 - 2004. [Full Text] [PDF] |
||||
![]() |
D. Saura, G. de la Morena, and M. Valdes-Chavarri Femoral versus antecubital vein contrast injection Eur J Echocardiogr, December 1, 2007; 8(6): 416 - 417. [Full Text] [PDF] |
||||
![]() |
M. R. Sukernik and E. Bennett-Guerrero The Incidental Finding of a Patent Foramen Ovale During Cardiac Surgery: Should It Always Be Repaired? A Core Review Anesth. Analg., September 1, 2007; 105(3): 602 - 610. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. C. Johansson, P. Eriksson, Y. Peker, J. Hedner, L. Rastam, and U. Lindblad The influence of patent foramen ovale on oxygen desaturation in obstructive sleep apnoea Eur. Respir. J., January 1, 2007; 29(1): 149 - 155. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Schwerzmann, K. Nedeltchev, F. Lagger, H. P. Mattle, S. Windecker, B. Meier, and C. Seiler Prevalence and size of directly detected patent foramen ovale in migraine with aura Neurology, November 8, 2005; 65(9): 1415 - 1418. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Homma and R. L. Sacco Patent Foramen Ovale and Stroke Circulation, August 16, 2005; 112(7): 1063 - 1072. [Full Text] [PDF] |
||||
![]() |
F J Pinto When and how to diagnose patent foramen ovale Heart, April 1, 2005; 91(4): 438 - 440. [Full Text] [PDF] |
||||
![]() |
O. K. Mohrs, S. E. Petersen, D. Erkapic, C. Rubel, R. Schrader, B. Nowak, W. A. Fach, H.-U. Kauczor, and T. Voigtlaender Diagnosis of Patent Foramen Ovale Using Contrast-Enhanced Dynamic MRI: A Pilot Study Am. J. Roentgenol., January 1, 2005; 184(1): 234 - 240. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A. Topcuoglu, H. P. Mattle, B. Meier, S. Windecker, for the PC Trial Group, S. R. Messe, S. Homma, and S. E. Kasner Practice Parameter: Recurrent stroke with patent foramen ovale and atrial septal aneurysm: Report of the Quality Standards Subcommittee of the American Academy of Neurology Neurology, December 14, 2004; 63(11): 2198 - 2199. [Full Text] [PDF] |
||||
![]() |
H. W. Schuchlenz, G.P. Anzola, E. Morandi, F. Casilli, and E. Onorato Contrast Ultrasound Techniques in the Detection and Quantification of Patent Foramen Ovale: Myth Versus Reality * Response: Stroke, December 1, 2004; 35(12): 2755 - 2756. [Full Text] [PDF] |
||||
![]() |
G.P. Anzola, E. Morandi, F. Casilli, and E. Onorato Does Transcatheter Closure of Patent Foramen Ovale Really "Shut the Door?": A Prospective Study With Transcranial Doppler Stroke, September 1, 2004; 35(9): 2140 - 2144. [Abstract] [Full Text] [PDF] |
||||
![]() |
N.R.A Clarke, J Timperley, A.D Kelion, and A.P Banning Transthoracic echocardiography using second harmonic imaging with Valsalva manoeuvre for the detection of right to left shunts Eur J Echocardiogr, June 1, 2004; 5(3): 176 - 181. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. P. Anzola Transcranial Doppler: Cinderella in the Assessment of Patent Foramen Ovale in Stroke Patients Stroke, June 1, 2004; 35(6): e137 - e137. [Full Text] [PDF] |
||||
![]() |
H. W. Schuchlenz Patent Foramen Ovale and Stroke Stroke, June 1, 2004; 35(6): e135 - e136. [Full Text] [PDF] |
||||
![]() |
M Braun, V Gliech, A Boscheri, S Schoen, G Gahn, H Reichmann, M Haass, R Schraeder, and R.H Strasser Transcatheter closure of patent foramen ovale (PFO) in patients with paradoxical embolism: Periprocedural safety and mid-term follow-up results of three different device occluder systems Eur. Heart J., March 1, 2004; 25(5): 424 - 430. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J Landzberg and P. Khairy Indications for the closure of patent foramen ovale Heart, February 1, 2004; 90(2): 219 - 224. [Full Text] [PDF] |
||||
![]() |
H. P. Adams Jr Patent Foramen Ovale: Paradoxical Embolism and Paradoxical Data Mayo Clin. Proc., January 1, 2004; 79(1): 15 - 20. [PDF] |
||||
![]() |
J. R. Corboy, H. W. Schuchlenz, G. Saurer, W. Weihs, W. Lalouschek, the Vienna Stroke Study Group, J.-L. Mas, L. Cabanes, J. Coste, and the Patent Foramen Ovale and Atrial Septal Aneurys Patent Foramen Ovale, Atrial Septal Aneurysm, and Recurrent Stroke N. Engl. J. Med., April 25, 2002; 346(17): 1331 - 1332. [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Stroke Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2002 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |