(Stroke. 2002;33:2402.)
© 2002 American Heart Association, Inc.
Original Contributions |
From the Departments of Neurology (G.S-A., L.C., V.P., K.H.), Anatomy, Histology and Embryology (S.F., L.M.), and Radiology (J.K.), University of Debrecen, Medical and Health Science Centre, Faculty of Medicine, Debrecen, Hungary, and Department of Neurology, University of Münster, Münster, Germany (G.S-A., E.B.R.).
Correspondence to Gernot Schulte-Altedorneburg, MD, Department of Radiology, Klinikum Augsburg, Stenglinstr 2, D-86156 Augsburg, Germany. E-mail Gernot.SAD{at}gmx.de
| Abstract |
|---|
|
|
|---|
Methods Shortly before their death, 59 critically ill neurological patients (mean age, 70 years) were prospectively examined by extracranial and intracranial Doppler sonography and color-coded duplex ultrasound. Carotid stenosis was classified by hemodynamic and morphological ultrasound criteria. Carotid specimens were removed in toto during autopsy. Under standardized conditions, specimens were redistended, sectioned, and histologically processed. Computerized planimetric measurements of the arteries were carried out and compared with the ultrasound findings. Correlation of the ultrasound and postmortem planimetric findings was available in 93 carotid bifurcations.
Results Through both techniques, 46 carotid arteries were found to be normal. Steno-occlusive carotid lesions ranged from 8.5% to 100% lumen reduction. Overall, r=0.96 and adjusted R2=0.90. For the steno-occlusive carotid lesions, r=0.91.
Conclusions Extracranial and intracranial Doppler and color-coded duplex ultrasound permits reliable detection and quantification of carotid artery stenoses and occlusions even under difficult examination conditions in critically ill patients.
Key Words: carotid arteries diagnosis pathology ultrasonography
| Introduction |
|---|
|
|
|---|
Carotid ultrasound has been used worldwide as a noninvasive diagnostic test for evaluating extracranial carotid artery disease. Besides its noninvasiveness, ultrasound is cost-effective and easy to perform (bedside technique). In the past, numerous articles confirmed the reliability of carotid ultrasonography in carotid artery disease, 59 particularly with respect to color-coded duplex ultrasound combining the techniques of high-resolution gray-scale B-mode pictures, color flow imaging, and frequency analysis of the Doppler spectrum. This battery of ultrasound information allows carotid artery stenosis to be quantified. In most previous articles, the ability of carotid ultrasound to predict carotid artery stenosis was validated by angiography, the presumed "gold standard." It is well known, however, that angiography itself is limited in its precision because of the restricted number of imaging planes (mostly 2) and because the frequent clinical practice of eyeballing the images leads to investigator dependency similar to ultrasound.
In this study, we evaluated the ability of extracranial and intracranial ultrasound to detect and to grade carotid artery stenosis by means of the true gold standard, the anatomic specimens of the affected arteries themselves. We compared in vivo ultrasound findings with the planimetric analysis of the corresponding postmortem histological results.
| Patients and Methods |
|---|
|
|
|---|
Ultrasound
In all patients, the neck arteries were investigated by color duplex ultrasound (7.5-MHz linear transducer, SONOS 2000, Hewlett-Packard), and the periorbital arteries were examined by continuous-wave Doppler (8-MHz probe, Multidop X, DWL). The intracranial arteries were assessed by transcranial color-coded duplex sonography with a dual-frequency, 2.0/2.5-MHz, 90° sector imaging transducer (HP SONOS 2000, Hewlett-Packard) or with a pulsed-wave, 2-MHz probe (Multidop X, DWL). The extracranial duplex examination included longitudinal and axial sections of the common carotid artery (CCA) and internal carotid artery (ICA) and was recorded on videotape. The angle-adjusted peak systolic and end-diastolic flow velocities in the CCA, in the jet of the stenosis, and in the most distal part of the ICA obtainable by duplex imaging were recorded. The degree of stenosis was measured by lumen diameter reduction in the longitudinal projection and by area reduction in the axial projection visible on the B-mode images. The peak systolic velocity of
120 cm/s was the threshold for the diagnosis of a stenosis graded
50%. Beyond this threshold, higher-degree stenoses were graded in categories of 5%, taking into account (1) the peak systolic velocity in the jet of the stenosis, (2) the peak systolic velocity in the most distal poststenotic ICA, (3) intrastenotic and poststenotic turbulence in the ICA (spectral broadening), (4) direction of the ophthalmic artery flow, (5) the presence of collateral flow via communicating arteries of the circle of Willis, (6) and the absolute asymmetry in flow velocity and pulsatility measured in the CCA and middle cerebral artery.10,11 In addition, continuous-wave Doppler measurements were carried out at the stenotic site. Subtotal ICA stenosis with a variable signal (eg, reduced peak systolic velocity without turbulence or venouslike flow) was defined as 97.5% stenosis. In the case of the indirect, ie, hemodynamic, criteria (fourth or fifth criteria described above), the stenosis was classified as
80%. Occlusion was diagnosed in the complete absence of detectable flow in and above the stenosis in conjunction with the above indirect hemodynamic criteria.12 If a contralateral steno-occlusive ICA lesion of
80% lumen narrowing was present in conjunction with collateral cross flow via the anterior communicating artery, intrastenotic peak systolic flow velocity of the ipsilateral carotid stenosis was correlated with area reduction and lumen diameter reduction visible on B mode. If these B-mode results for lumen narrowing and hemodynamic findings (ie, peak systolic flow velocity) of the ipsilateral carotid stenosis were inconsistent, local area reduction and lumen diameter reduction measured by B-mode were preferred for determining carotid stenosis to avoid overestimation of the ipsilateral carotid stenosis.
In low-grade ICA stenoses of <50% lumen narrowing, the intrastenotic area reduction measured by B mode was given preference over Doppler ultrasound parameters for comparison with the postmortem findings. However, if plaque calcification did not permit clear-cut axial visualization of the vessel walls, the lumen diameter reduction in the longitudinal projection measured by B mode was used for comparison with the pathoanatomic findings.
Stenoses with >50% lumen reduction were classified according to the above criteria. All ultrasound studies were performed by the same investigator (G.S.-A.).
Macropathological Examination of the Carotid Specimen
At autopsy, the CCA, ICA, and external carotid artery (ECA) were excised en bloc. The time between death and autopsy ranged from 1 to 110 hours (mean, 23.8 hours; median, 14 hours). ECA branches were ligated, and the ends of the 3 arterial segments (ie, CCA, ICA, and ECA) were tied to plastic or glass tubes for the injection of the tissue-embedding medium (Cryomatrix, Shandon Inc). This medium is fluid on room temperature but solid at -20°C. Cryomatrix was pressed manually into the plastic tube at the end of the CCA against an intraluminal pressure of 100 mm Hg. Then, the arterial tree was clamped at the ends with the lumen still under pressure. After the redistended carotid specimens were frozen at -21°C, they were cut into serial 3-mm cross slices. Accordingly, slices distal and proximal to the flow divider were labeled +3, +6, etc, and -3, -6, etc, respectively. Photographs with a 1-mm scale were taken from each slice. Technical details of the procedures for preparing and filling the arteries have been given previously.13
Histological Examination and Planimetry
The 3-mm slices containing the carotid stenosis were selected and slowly defrosted. A dental silicon rubber (President Micro System Jet Bite, Coltene AG) was put into the lumen to prevent shrinkage. The slices were placed for fixation into a 4:1 mixture of 100% ethanol and 40% formaldehyde for 24 to 48 hours. After dehydration, the silicon rubber was carefully removed, and the specimens were embedded in paraffin. Transversal sections 7 to 8 µm thick were cut by microtome and stained with hematoxylin and eosin, dimethyl methylene blue, periodic acidSchiff stain, and Verhoeffs elastica combined with picrosirius red.14
Neurolucida (Micro Bright Field, Colchester) software, 15 a Leitz Labor Lux S microscope (Leitz, Wetzlar) equipped with an 10x/0.25 numerical aperture objective (Leitz, Wetzlar), and associated instrumentation were used to calculate the original lumen vessel area and remaining lumen area. After system calibration, graphic overlays were made along the luminal surface of the endothelial and the internal elastic membrane according to the microscopic image. Afterward, these bit map format images were analyzed with IMAN1.4 software (KFKI). The area of these graphic overlays was calculated. The area bordered by the endothelium (ie, remaining lumen area) was subtracted from the area bordered by the internal elastic membrane (ie, original lumen area), resulting in a value quantifying the ICA or CCA area reduction at the site of the stenosis. This measurement of area reduction was determined to be the gold standard and was compared with the ultrasound findings. Planimetric analysis was carried out by an investigator who was not aware of the ultrasound results.
Statistical Analysis
The degree of carotid stenosis of the in vivo sonographic and postmortem planimetric measurements and the mean differences were determined in percent. Linear regression equation was used to compare carotid stenosis determined by either technique. The correlation coefficients r and R2 were calculated. The corresponding values obtained by ultrasound were plotted against those obtained by pathological investigation. In addition, a Bland-Altman plot was applied to illustrate the agreement between anatomic and sonographic measurements.16
This investigation is part of a more extended research project comparing ultrasound findings of the brain-supplying arteries with the corresponding pathological postmortem findings. The study was performed according to national laws, and patients were studied in compliance with a protocol previously approved by the local ethics committee of the University Medical School of Debrecen.
| Results |
|---|
|
|
|---|
|
The slice (vessel piece) with the most severe steno-occlusive carotid lesion was 3 to 9 mm below the flow divider in 5 cases, at the flow divider in 8 cases, and 3 to 15 mm above the flow divider (ie, ICA) in 33 cases.
In 14 carotid stenoses, a satisfactory B-mode measurement of the area reduction was not possible because of ultrasonic shadowing or a difficult-to-delineate vessel wall; in these cases, the lumen diameter reduction in the longitudinal projection was used for comparison with anatomy. Nine ICA occlusions were found by pathological analysis. Seven had also been detected by ultrasound. In the 2 remaining cases, a pseudoocclusion had been diagnosed sonographically as a result of the detection of poststenotic flow. The ICA could have been occluded in the time interval between ultrasound and death. One CCA occlusion and an ICA occlusion of the same vessel were found by both techniques. Sonographic findings and their corresponding pathoanatomical findings in stenoses
50% identified by both techniques are given in the Table. In 5 cases (cases 14, 17, and 19 through 21), the intrastenotic peak systolic velocity was <120 cm/s despite an anatomical area reduction of >50%. In 1 of these cases (case 14), the stenosis was found in the distal CCA 6 mm below the flow divider; in the remaining 4 cases, carotid stenosis was located above the flow divider in the ICA.
|
Figure 2a and 2b shows the scatterplot, Bland-Altman plot, and regression equation for both measurements. Overall, r=0.96, and the adjusted R2=0.90; for the 46 steno-occlusive CCAs and ICAs detected on histopathology only, r=0.91.
|
An elevated intracranial pressure resulting from space-occupying infarction or bleeding was found by CT in 34 patients; despite this, spectral waveform analysis of the blood flow in the brain-supplying arteries was diagnostic in all cases.
| Discussion |
|---|
|
|
|---|
70% and
82% angiographic diameter reductions, respectively. We investigated the ability of carotid ultrasound to both detect and grade carotid artery stenoses. Unlike previous studies, we did not assess preselected patients with known or probable carotid artery disease screened for endarterectomy but used unselected, critically ill neurological patients. Alexandrov et al5 have already pointed out that bidimensional angiographic measurements of carotid stenosis (eg, NASCET and ECST methods) differ significantly from the actual anatomical stenosis because these methods do not consider the area reduction of the residual, often asymmetrical, vessel lumen. In contrast, we used planimetry in this study to evaluate artery luminal area reduction as the gold standard instead of angiographic 1-plane measurements, which are only surrogates of the real anatomy. The histological examination allows identification of both the original and residual arterial lumen. Eccentric lumens pose no problems in estimating the degree of narrowing at necropsy but may be a source of error during angiography.
In accordance with previous studies,6,10 we could confirm that the sole acquisition of the intrastenotic peak systolic velocity and/or calculation of velocity ratios between the ICA and CCA, as proposed by other investigators,7,9,19,22 do not provide an adequate sonographic approach for reliable determination of the degree of carotid artery stenosis. All the above-mentioned accessible extracranial and intracranial ultrasound parameters have to be taken into account.10 For instance, the peak velocities in
80% ICA stenosis in conjunction with cross flow to the contralateral side via an anterior communicating artery require different interpretation than the same findings without anterior cross flow.
The Table shows that the synopsis of direct and indirect ultrasound findings evaluated by extracranial and intracranial sonography permits reliable quantification of high-grade carotid stenosis. For the quantification of moderate- and low-grade stenoses, however, the B-mode measurement of the area reduction at the stenotic site or, if not available for technical reasons, the measurement of the lumen diameter reduction provided a good match of in vivo and postmortem findings (Figure 2).
De Bray and Glatt10 indicated that the identification and measurement of tight stenoses by area ratio measurement is less accurate, and hemodynamic criteria are preferred. This is widely accepted and is valid for most patients. In 5 of our patients, however, we found a peak systolic velocity of <120 cm/s although an anatomic area reduction of >50% was present (the Table). This fact underlines the above-mentioned sonographers rule to consider as many available Doppler and B-mode findings as possible when determining the degree of carotid stenosis. The hemodynamic findings in distal CCA stenoses require an evaluation different from ICA stenosis (see case 14 in the Table).
In previous pathoanatomic studies, ultrasound had been validated through carotid plaques removed during carotid endarterectomy.57 These specimens, however, may be incomplete in terms of the full extension of the plaque, are frequently damaged by surgery, and because of shrinkage and tissue collapse, would not provide the true in vivo situation distal to, proximal to, or at the site of stenosis. In contrast to shortcomings with endarterectomy specimens, we assessed the complete redistended carotid specimens and evaluated systematically the entire length of the stenotic and nonstenotic arterial segment by performing serial slices. Furthermore, we avoided the collapse of the lumen by filling it with silicon rubber before histological processing.
Several recent studies had focused on the analysis of moderate to severe carotid artery stenosis.58 The diagnostic value of modern ultrasonography to determine low-grade stenosis and plaques has been investigated less extensively. In the present study, however, we assessed diseased carotid arteries with lumen narrowings ranging from 8.5% to 100%. As shown in Figure 2, B-mode ultrasound, combined with Doppler ultrasound, allows reliable measurement of all degrees of carotid artery lesions. Moreover, with 1 exception, all nonstenotic vessels were correctly identified by ultrasound.
In accordance with a previous pathoanatomic comparative study,6 we found r>0.9 for the
50% stenoses and overall r=0.96 for the agreement of anatomic and sonographic measurements as long as a combination of several criteria was used. Other investigators had similar results although they used either hemodynamic criteria or morphological measurements alone. In the study by Ranke et al, 22 exclusive use of the mean velocity ratio (ie, the intrastenotic mean flow velocity divided by the distally recorded mean blood flow velocity) revealed the closest correlation with angiography (R2=0.93). Griewing et al8 measured the cross-sectional area reduction and longitudinal diameter reduction by color flow Doppler and power Doppler, leading to an excellent correlation between power Doppler and angiography in high-grade stenosis with r=0.93.
In our study, the ultrasound examination was not carried out under optimal conditions because most patients had to be examined at bedside in the intensive care unit. The uncomfortable position of the sonographer, jugular vein catheters, deep breathing, unextended neck, and agitation of the patient limited the quality of the ultrasound investigation in some cases. Unlike previous investigators,23 we included all ultrasound examinations in the analysis, even those with poor quality, on an "intention-to-diagnose" basis. Furthermore, at the time of this study, there was no approval for ultrasound contrast agent. We could not use ultrasound contrast agents for a more accurate differentiation of total ICA occlusion from pseudoocclusion or for the investigation of intracranial pathways.24,25
In summary, using an unique in vivo and postmortem approach, we could show that color-coded duplex ultrasound permits a reliable detection and quantification of carotid artery steno-occlusive disease even under difficult examination conditions as long as the investigator considers the entire spectrum of the available extracranial and intracranial ultrasound parameters.
| Acknowledgments |
|---|