(Stroke. 2002;33:456.)
© 2002 American Heart Association, Inc.
Original Contributions |
From the Department of Pathology, University of Padua Medical School (A.A., M.D.B., M.V., G.T.), and Cardiology Department, Mirano Hospital, Mirano (Venice) (B.R., S.S., G.P., C.C., P.P.), Italy.
Correspondence to Gaetano Thiene, MD, FESC, Department of Pathology, University of Padua Medical School, Via A. Gabelli 61, 35121 Padua, Italy. E-mail cardpath{at}unipd.it
| Abstract |
|---|
|
|
|---|
Methods Elective carotid stent implantation with the use of a distal filter protection was attempted in 38 consecutive lesions (36 patients) of the internal carotid artery presenting >70% diameter stenosis (mean, 82.1±11.1%). Mean age of the patients was 70.7±7.7 years; 75% were men, and 50% of patients had previous neurological symptoms.
Results In 37 lesions (97.4%) it was possible to position the filter device, and in all lesions a stent was successfully implanted. The only complication occurring in the hospital and during the 30-day follow-up was 1 death due to acute myocardial infarction. Neurological sequelae did not occur. Histomorphometric analysis was performed on the filters. Presence of debris was detected in 83.7% of filters. The mean surface area of the polyurethane membrane filter covered with material was 53.2±19.8%. Particle size ranged from 1.08 to 5043.5 µm (mean, 289.5±512 µm) in the major axis and 0.7 to 1175.3 µm (mean, 119.7±186.7 µm) in the minor axis. Collected debris consisted predominantly of thrombotic material, foam cells, and cholesterol clefts.
Conclusions By the use of distal protection filters during carotid artery stenting, it was possible to collect particulate debris potentially leading to distal vessel occlusion in a high percentage of cases. Qualitative analysis of embolized material showed debris dislocated during the percutaneous intervention from atheromatous plaques.
Key Words: angioplasty carotid arteries embolism pathology stents
| Introduction |
|---|
|
|
|---|
Only a few studies give a detailed morphological evaluation of the material retrieved during percutaneous intravascular procedures.1015 From these studies it appears that most of the embolic dissemination occurs during iatrogenic manipulation of the atheromatous plaque.
Protection devices have the potential to reduce the incidence of intracranial debris embolization and render percutaneous carotid artery revascularization safer.1219 In light of novel observations made by Reimers et al18 suggesting the safety and feasibility of filter devices for carotid artery stenting, the present prospective study was designed to analyze histopathologically debris collected during intravascular percutaneous carotid artery stent implantation with the use of a recently available protection device.
| Subjects and Methods |
|---|
|
|
|---|
|
Angiographic Evaluation
Baseline and postprocedural quantitative angiography was performed online with the use of the automated analysis system coordinated with the angiographic equipment (Intergris 3000, Philips, Medical Instruments). The angiographic diameter stenosis was measured according to the North American Symptomatic Carotid Endarterectomy Trial (NASCET) criteria with the distal, nontapering portion of the internal carotid artery used as the reference segment.20 Qualitative angiographic evaluation was performed offline by 2 independent operators according to the criteria described by Mathur et al.21
Drug Regimen
All patients were taking aspirin (100 to 325 mg). Ticlopidine (500 mg) was started at least 48 hours before the procedure. Heparin 70 to 100 U/kg was given intra-arterially to achieve an activated clotting time of >250 seconds. In case of inadequate anticoagulation, additional boluses of 2000 U of heparin were given until the target value was achieved. Atropine 1 mg was administered only when required to treat bradycardia during balloon inflation. Arterial blood pressure was monitored during the procedure and, if needed, was modulated with nitrates or dopamine. After the procedure, patients received aspirin indefinitely and ticlopidine for 1 month. Glycoprotein IIb/IIIa inhibitors were not used.
Carotid Stenting Procedure
Percutaneous access was gained through the femoral or brachial (1 lesion) artery. Long sheaths (6F or 7F) or coronary guiding catheters (8F) were advanced into the common carotid artery on long 0.035-inch support wires previously placed in the external carotid artery with the use of 4F or 5F diagnostic catheters. Angiography of the carotid artery and the intracranial circulation was performed in angulated views (Figure 1). Stenoses were crossed with the filter protection device. After filter opening, predilatation with 3.5- to 4.0-mm-diameter coronary angioplasty balloons was performed in 60.5% of the lesions at medium-low pressure (6 to 10 atm) until complete expansion of the balloon was seen. Appropriately sized self-expandable stents (Carotid Wallstent, Boston ScientificSchneider; Smart 0.018 inch, J&J Cordis) were implanted, covering the carotid bifurcation in all cases. All stents were postdilated with the use of 5.5- to 7.0-mm (mean, 6.06±0.62 mm) balloons at medium-high pressures (8 to 12 atm; mean, 11.0±0.1). The arterial sheaths were removed on the same day. After the procedure, patients were transferred to the cardiology ward with continuous ECG monitoring for the following 12 hours, and noninvasive blood pressure measurements were taken every 2 hours for at least 12 hours.
|
Filter Device
The first-generation filter device (Angioguard, J&J Cordis) consists of an olive-shaped nitinol skeleton covered at its distal portion by a polyurethane membrane with pores of 100 µm in diameter. The filters used had diameters of 6, 7, and, in 1 case, 8 mm. The device is connected to a proximal, floppy wire tip and a distal 0.014-inch wire shaft 300 cm in length used as guidewire for the interventional procedure.
The closed filter, contained in a delivery sheath (diameter of 0.061 to 0.065 inch) with a maximum profile of 4F to 4.5F, was advanced through the lesion. The filter was opened in the internal carotid artery distal to the lesion by removal of the delivery sheath. A filter 0.5 to 1.0 mm larger than the visually estimated distal vessel diameter was chosen. This was to ensure complete apposition of the filter to the vessel wall. At the end of the procedure, a retrieval sheath was advanced, and the filter was closed and removed from the artery. Visual inspection of the filter was performed by the interventional cardiologist at the end of each procedure to evaluate the macroscopic presence of material.
Neurological Evaluation
A board-certified neurologist performed a complete neurological examination, including the National Institutes of Health Stroke Scale, before and after stent implantation.
Follow-Up
Clinical follow-up was performed 30 days after the procedure in all patients.
Histopathologic Evaluation
After removal, the filters were immediately fixed in 10% neutral, buffered formalin. Photographs of the device in its integrity and after removal of the metallic wire were taken under a stereo microscope. These images were used for the computerized morphometric analysis with Image-Pro Plus software (Media Cybernetics). The total area of the filter membrane and the area covered by particulate material were quantified. The percentage of membrane occupied by debris was expressed as percentage of covered surface area. According to the extension of the covered area, patients were divided into 4 groups: group 1 had a covered area of
25%, group 2 >25% and
50%, group 3 >50% and
75%, and group 4 >75%. If particles were entrapped in the scaffolding metallic wires, they were carefully removed and separately photographed. Particulate floating in the formalin was filtered and processed for histological characterization.
Particles identifiable as such were outlined by an experienced pathologist. The particles were subsequently counted and measured by the computer software. The 2 longest perpendicular diameters of the largest particles were recorded (Figure 2). All measurements were made at least twice.
|
For histology and tissue characterization, the material was removed from the filter, dehydrated, and embedded in paraffin, and the sections (5 µm thick) were stained with hematoxylin-eosin and Heidenhain trichrome.
Scanning electron microscopy was performed on the internal surface of the polyurethane membrane after it had been flattened. After removal from the 10% buffered formalin, the material was washed first in normal saline solution and then in distilled water. Subsequently it was dehydrated in crescent series of alcohol, with a final incubation in amyl acetate, and processed for the CO2 critical dry point and gold sputtering. The specimens were observed under a scanning electron microscope (Philips XL 30).
Statistical Analysis
The data were computerized and analyzed with the use of the GraphPad Prism Software package. Mean, SD, and lower and upper 95% CIs were calculated for all parameters. Results are expressed as mean±SD. P<0.05 was considered significant.
| Results |
|---|
|
|
|---|
|
Careful neurological examination, including the National Institutes of Health Stroke Scale, performed after the procedure, at discharge, and during the follow-up visit at 30 days did not reveal the occurrence of any neurological complications. Two patients required prolonged intravenous dopamine infusion for persistent hypotension. The 30-day mortality rate was 2.8% (1 patient). This patient with unstable coronary disease suffered massive anterior myocardial infarction 4 days after the procedure and died 3 days later.
Histopathological Results
Microscopic evaluation revealed that particles could be detected in 31 of 37 filters (83.7%) and that they were adherent mostly to the filter device (Figure 2). The mean number of particles in these filters was 33.7±5.6 (range, 24 to 46). In 13 of 37 filters (35%), additional material was entangled in the scaffolding metallic wire. Only a small number of immersed particles were retrieved from the fixative. The average covered area on the plastic filter was 53.2±19.8%. Mean particle size ranged from 1.08 to 5043.5 µm (mean, 289.5±512 µm) in the major axis and from 0.7 to 1175.3 µm (mean, 119.7±186.7) in the minor axis (Figure 2b). The distribution of the size of the particles is shown in Figure 3, and results of the histopathological analysis are shown in Table 3. Scanning electron microscopy was performed, and mean particulate size of 311.6±251.5 µm in the major axis and 257.76±247.9 µm in the minor axis was observed (Figure 4). In all 8 filters of patients in which slow flow or no flow was observed, embolic material was detected, with an average covered area of 53.2±14.8%. Moreover, no relations could be detected between particle size and average covered area occupied by the material on the filters (P=NS).
|
|
|
Qualitative analysis with light microscopy revealed that the particles consisted primarily of soft acellular and amorphous material characterized by lipid-rich macrophages and cholesterol clefts. To a minor degree, calcium particulates and platelets entrapped in fibrin strands were found (Figure 5). All detected material is typically identifiable in an atheromatous plaque whether complicated or uncomplicated. The single case treated with distal occlusive balloon protection showed similar findings.
|
| Discussion |
|---|
|
|
|---|
The present study shows that embolic particulates that are commonly released during carotid artery stent implantation3,4 can be captured by distal filter protection devices. This is the clinical confirmation of the experience with filter protection in the ex vivo model reported by Ohki et al.13 Debris was captured in 83.7% of procedures compared with 81%12 and 100%14,15 in previously published studies in which occlusive balloons were used for protection. The size of the particles detected in the present study (289.5±512 µm in diameter in the major axis; range, 1.08 to 5043.5 µm) was greater compared with the ex vivo filter evaluation (226±130 µm)13 and 1 of the distal occlusive balloon studies (54.4 mm crystal and 110.67 mm lipoid mass)5 and smaller compared with another distal balloon protection study (800 µm).14 In our experience, as well as in all previous observations, particles >300 µm were found in all filters with debris.1215,18 The microcirculation encompasses vessels with diameter <300 µm.22 The arterial network arborizes from the aorta to arteries up to 400 to 300 µm, small arteries with diameter between 300 and 100 µm, arterioles with diameter between 100 and 12 µm, and capillaries with diameter of approximately 12 µm. In the present study 100% of filters with materials presented particles >300 µm, and 52% of filters presented particles >1000 µm. It cannot be excluded that some of the retrieved thrombotic material could have been locally produced inside the filter, explaining the relatively high incidence (21.1%) of flow impairment and the high filter surface coverage of >50%, which was not correlated to the size of particles. As for the amorphous material, we cannot see any source other than the amorphous material dislocated from the atheromatous core of the plaque. All cases of transient flow impairment due to pore occlusion occurred immediately after postdilatation of the stent. Therefore, it appears that most emboli are released during this part of the procedure. Possible emboli release may occur during the passage of the closed filter device through the stenosis, and in the absence of intracranial Doppler we cannot exclude it. Protection systems such as low-profile distal occlusion balloons or proximal occlusion systems, which do not require any lesion crossing without protection, have the potential to reduce embolization during this procedural step.
It is, however, assumable that not all captured particles would have had acute clinical sequelae. The significance of clinically silent embolization during carotid artery interventions has not yet been established, and possibly microvascular obstructions are difficult to recognize.23,24 According to recent observations,22 emboli can potentially trigger platelet aggregation and may amplify microvascular obstructions. A shower of small particulate matter could cause microvascular obstruction. Furthermore, particles of vasoactive substances contained within plaque or thrombus could cause intense, prolonged vasospasm and subsequent cerebral infarction.25
The histopathologic analysis showed that retrieved material consists of dislocated debris featuring atheroembolism as a consequence of atheromatous plaque squeezing and rupture occurring during carotid stent implantation. This result confirms the data of previously reported studies evaluating different emboli protection devices during percutaneous intravascular interventions.1115
Our study confirms that this method of cerebral protection18,25 is feasible and safe and that it limits intracranial debris embolization. Therefore, it strongly supports the use of protection devices during percutaneous carotid stenting. However, we could not exclude the development of a stroke even in the setting of distal protection.25 In a recently published article, Roubin et al2 reported the important reduction of periprocedural complication rate over 5 years of experience from 9.3% in the first year to 4.3% in the fifth year. This was probably due to refinement of stenting techniques and newly developed equipment. We cannot rule out, as suggested by Qureshi et al,26 that the cardiac death occurring in our study within the 30-day window may be in some way related to the hemodynamic instability after carotid stenting. To achieve the target of
3% periprocedural complication rate recommended by the American Heart Association/Society of Vascular Surgery guidelines27 for the treatment of asymptomatic patients, cerebral protection may well be essential.
Limitations of the Study
Because of the absence of intracranial Doppler monitoring during the stent procedure, this study does not exclude the presence of emboli not captured by the filter. In a recent ex vivo study, the filter device captured 88% of embolized particles.13 Moreover, the small study size does not allow conclusions regarding the clinical efficacy of protection filters. Larger evaluations are warranted.
Conclusions
The present study demonstrates that protection filters collect debris of significant quantity and size during carotid artery stent implantation. Histopathological analysis showed that the collected material is dislocated from the atherosclerotic plaque during the intravascular procedure. Protection devices have the potential to further reduce neurological complications during the treatment of carotid artery stenosis.
| Acknowledgments |
|---|
Received August 15, 2001; revision received October 2, 2001; accepted October 15, 2001.
| References |
|---|
|
|
|---|
2.
Roubin GS, New G, Iyer SS, Vitek JJ, Al-Mubarak N, Liu MW, Yadav J, Gomez C, Kunts RE. Immediate and late clinical outcomes of carotid artery stenting in patients with symptomatic and asymptomatic carotid artery stenosis: a 5-year prospective analysis. Circulation. 2001; 103: 532537.
3.
Phatouros CC, Higashida RT, Malek AM, Mayers PM, Lempert TE, Dowd CF, Halbach VV. Carotid artery stent placement for atherosclerotic disease: rationale, technique, and current status. Radiology. 2000; 217: 2641.
4. Jordan WD, Voellinger DC, Doblar DD, Plyushcheva NP, Fisher WS, McDowell HA. Microemboli detected by transcranial Doppler monitoring in patients during carotid angioplasty versus carotid endarterectomy. Cardiovasc Surg. 1999; 7: 3338.[CrossRef][Medline] [Order article via Infotrieve]
5. Markus HS, Clifton A, Buckenham T, Brown MM. Carotid angioplasty: detection of embolic signals during and after the procedure. Stroke. 1994; 25: 24032406.[Abstract]
6. Ackerstaff RG, Jansen C, Moll FL, Vermeulen FE, Hammerlijinck RP, Mauser HW. The significance of microemboli detection by means of transcranial Doppler ultrasonography monitoring in carotid endarterectomy. J Vasc Surg. 1995; 21: 963969.[CrossRef][Medline] [Order article via Infotrieve]
7.
Ackerstaff RG, Moons KG, van de Vlasakker CJ, Moll FL, Vermeulen FE, Algra A, Spencer MP. Association of intraoperative transcranial Doppler monitoring variables with stroke from carotid endarterectomy. Stroke. 2000; 31: 18171823.
8. Gaunt ME, Martin PJ, Smith JL, Rimmer T, Cherryman G, Ratliff DA, Bell PR, Maylor AR. Clinical relevance of intraoperative embolization detected by transcranial Doppler ultrasonography during carotid endarterectomy: a prospective study of 100 patients. Br J Surg. 1994; 81: 14351439.[Medline] [Order article via Infotrieve]
9.
Krul JMJ, van Gijn J, Askerstaff GA, Eikelboom BC, Theodorides T, Vermeulen FE. Site and pathogenesis of infarcts associated with carotid endarterectomy. Stroke. 1989; 20: 324328.
10.
Manninen HI, Rasanen HT, Vanninen RL, Vainio P, Hippeläinen M, Kosma VM. Stent placement versus percutaneous transluminal angioplasty of human carotid arteries in cadaver in situ: distal embolization and findings at intravascular US, MR imaging and histopathological analysis. Radiology. 1999; 212: 483492.
11.
Webb JG, Carere RG, Virmani R, Baim D, Teistein PS, Whitlow P, McQueen C, Kolodgie FD, Buller E, Dodek A, Mancini GBJ, Oesterle S. Retrieval and analysis of particulate debris after saphenous vein grafts intervention. J Am Coll Cardiol. 1999; 34: 46875.
12.
Theron JG, Payelle GG, Coskun O, Huet HF, Guimaraens L. Carotid artery stenosis: treatment with protected balloon angioplasty and stent placement. Radiology. 1996; 201: 627636.
13. Ohki T, Roubin GS, Veith FJ, Iyer SS, Brady E. Efficacy of a filter device in the prevention of embolic events during carotid angioplasty and stenting: an ex vivo analysis. J Vasc Surg. 1999; 30: 10341044.[CrossRef][Medline] [Order article via Infotrieve]
14. Henry M, Amor M, Henry I, Klonaris C, Chati Z, Masson I, Kownator S, Luizy F, Hugel M. Carotid stenting with cerebral protection: first clinical experience using the PerceSurge GuardWire system. J Endovasc Surg. 1999; 6: 321331.[CrossRef][Medline] [Order article via Infotrieve]
15. Martin JB, Pache JC, Treggiar-Venzi M, Murphy KJ, Gailloud P, Puget E, Pizzolato G, Sugiu K, Giumaraens L, Théron J, Rüfenacht DA. Role of the distal balloon protection technique in the prevention of cerebral embolic events during carotid stent placement. Circulation. 2001; 32: 479484.
16. Ohki T, Veith FJ. Carotid stenting with and without protection devices: should protection be used in all patients? Semin Vasc Surg. 2000; 13: 144152.[Medline] [Order article via Infotrieve]
17. Albuquerque FC, Teitelbaum GP, Lavine SD, Larsen DW, Giannotta SL. Balloon-protected carotid angioplasty. Neurosurgery. 2000; 46: 918921.[CrossRef][Medline] [Order article via Infotrieve]
18.
Reimers B, Corvaja N, Moshiri S, Saccà S, Albiero R, Di Mario C, Pascotto P, Colombo A. Cerebral protection with filter devices during carotid artery stenting. Circulation. 2001; 104: 1215.
19. Parodi JC, La Mura R, Ferreira LM, Mendez MV, Cersosimo H, Schonholz C, Garelli G. Initial evaluation of carotid angioplasty and stenting with three different cerebral protection devices. J Vasc Surg. 2000; 32: 11271136.[CrossRef][Medline] [Order article via Infotrieve]
20. North American Symptomatic Carotid Endarterectomy Trial Collaboration. Beneficial effect of carotid endarterectomy in symptomatic patients with high grade carotid stenosis. N Engl J Med. 1991; 325: 445453.[Abstract]
21.
Mathur A, Roubin GS, Iyer S, Piamsonboon C, Liu MW, Gomez CR, Yadav JS, Chastain HD, Fox LM, Dean LS, Vitek JJ. Predictors of stroke complicating carotid artery stenting. Circulation. 1998; 97: 12391245.
22.
Topol EJ, Yadav JS. Recognition of the importance of embolization in atherosclerotic vascular disease. Circulation. 2000; 101: 570580.
23.
Jacobs A, Neveling M, Horst M, Ghaemi J, Kessler J, Eichstaedt H, Rudolf J, Model P, Bönner H, de Vivie ER, Heiss WD. Alterations of neuropsychological function and cerebral glucose metabolism after cardiac surgery are not related only to intraoperative microembolic events. Stroke. 1998; 29: 660667.
24.
Barth A, Remonda L, Loevblad K-O, Schroth G, Seiler RW. Silent cerebral ischemia detected by diffusion-weighted MRI after carotid endarterectomy. Stroke. 2000; 31: 18241828.
25.
Williams DO. Carotid filters: new additions to the interventionists toolbox. Circulation. 2001; 104: 23.
26.
Qureshi A, Luft AR, Sharma M, Janardhan V, Lopes DK, Khan J, Guterman LR, Hopkins LN. Frequency and determinants of postprocedural hemodynamic instability after carotid angioplasty and stenting. Stroke. 1999; 30: 20862093.
27.
Biller J, Feinberg WM, Castaldo JE, Whittemore AD, Harbaugh RE, Dempsey RJ, Caplan LR, Kresowik TF, Matchar DB, Toole JF, Easton JD, Adams HP Jr, Brass LM, Hobson RW, Brott TG, Sternau L. Guidelines for carotid endarterectomy: a statement for healthcare professionals from a special writing group of the Stroke Council, American Heart Association. Circulation. 1998; 97: 501509.
This article has been cited by other articles:
![]() |
P. Pinero, A. Gonzalez, E. Martinez, A. Mayol, E. Rafel, J.R. Gonzalez-Marcos, F. Moniche, A. Cayuela, and A. Gil-Peralta Volume and Composition of Emboli in Neuroprotected Stenting of the Carotid Artery AJNR Am. J. Neuroradiol., March 1, 2009; 30(3): 473 - 478. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. G. DeRubertis Embolization During Carotid Angioplasty and Stenting : What Is the Optimal Method for Detecting Embolic Debris and Its Sequelae? Perspectives in Vascular Surgery and Endovascular Therapy, September 1, 2008; 20(3): 260 - 269. [Abstract] [PDF] |
||||
![]() |
J. A. Bittl Damage Control for Renal Artery Stenting Circulation, May 27, 2008; 117(21): 2724 - 2726. [Full Text] [PDF] |
||||
![]() |
E. I. Levy, J. Mocco, R. M. Samuelson, R. D. Ecker, B. S. Jahromi, and L. N. Hopkins Optimal treatment of carotid artery disease. J. Am. Coll. Cardiol., March 11, 2008; 51(10): 979 - 985. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. M. Samuelson, J. Yamamoto, E. I. Levy, A. H. Siddiqui, and L. N. Hopkins The Argument to Support Broader Application of Extracranial Carotid Artery Stent Technology Circulation, October 2, 2007; 116(14): 1602 - 1610. [Full Text] [PDF] |
||||
![]() |
A. Angelini, M. Della Barbera, and G. Thiene Interventional procedures for atherothrombosis: pathology of retrieved material Heart, October 1, 2007; 93(10): 1301 - 1308. [Full Text] [PDF] |
||||
![]() |
I. P. Casserly, A. Abou-Chebl, R. B. Fathi, D. S. Lee, J. Saw, J. E. Exaire, S. R. Kapadia, C. T. Bajzer, and J. S. Yadav Slow-Flow Phenomenon During Carotid Artery Intervention With Embolic Protection Devices: Predictors and Clinical Outcome J. Am. Coll. Cardiol., October 18, 2005; 46(8): 1466 - 1472. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. D. Aronow, M. Shishehbor, D. A. Davis, I. L. Katzan, D. L. Bhatt, C. T. Bajzer, A. Abou-Chebl, K. W. Derk, P. L. Whitlow, and J. S. Yadav Leukocyte Count Predicts Microembolic Doppler Signals During Carotid Stenting: A Link Between Inflammation and Embolization Stroke, September 1, 2005; 36(9): 1910 - 1914. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Angelini, P. Rubartelli, F. Mistrorigo, M. Della Barbera, F. Abbadessa, M. Vischi, G. Thiene, and S. Chierchia Distal Protection With a Filter Device During Coronary Stenting in Patients With Stable and Unstable Angina Circulation, August 3, 2004; 110(5): 515 - 521. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. E. Cil, E. Akpinar, B. Peynircioglu, and S. Cekirge Utility of Covered Stents for Extracranial Internal Carotid Artery Stenosis AJNR Am. J. Neuroradiol., August 1, 2004; 25(7): 1168 - 1171. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. A. Gray A cardiologist in the carotids J. Am. Coll. Cardiol., May 5, 2004; 43(9): 1602 - 1605. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. D. Barr, J. J. Connors III, D. Sacks, J. C. Wojak, G. J. Becker, J. F. Cardella, B. Chopko, J. E. Dion, A. J. Fox, R. T. Higashida, et al. Quality Improvement Guidelines for the Performance of Cervical Carotid Angioplasty and Stent Placement: Developed by a Collaborative Panel of the American Society of Interventional and Therapeutic Neuroradiology, the American Society of Neuroradiology, and the Society of Interventional Radiology AJNR Am. J. Neuroradiol., November 1, 2003; 24(10): 2020 - 2034. [Full Text] [PDF] |
||||
![]() |
C. W. Kopp, S. Steiner, C. Nasel, D. Seidinger, I. Mlekusch, W. Lang, A. Bartok, R. Ahmadi, and E. Minar Abciximab Reduces Monocyte Tissue Factor in Carotid Angioplasty and Stenting Stroke, November 1, 2003; 34(11): 2560 - 2567. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Crane and N. Cheshire Recent developments in vascular surgery BMJ, October 18, 2003; 327(7420): 911 - 915. [Full Text] [PDF] |
||||
![]() |
J. E. Scoble Do protection devices have a role in renal angioplasty and stent placement? Nephrol. Dial. Transplant., September 1, 2003; 18(9): 1700 - 1703. [Full Text] [PDF] |
||||
![]() |
F. Liistro and C. Di Mario Carotid artery stenting Heart, August 1, 2003; 89(8): 944 - 948. [Full Text] [PDF] |
||||
![]() |
A. Kastrup, K. Groschel, H. Krapf, B. R. Brehm, J. Dichgans, and J. B. Schulz Early Outcome of Carotid Angioplasty and Stenting With and Without Cerebral Protection Devices: A Systematic Review of the Literature Stroke, March 1, 2003; 34(3): 813 - 819. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Macdonald and P. A Gaines Current concepts of mechanical cerebral protection during precutaneous carotid intervention Vascular Medicine, February 1, 2003; 8(1): 25 - 32. [Abstract] [PDF] |
||||
![]() |
G. S. Roubin Angioplasty and Stenting Should Not Be Restricted to Clinical Trials Stroke, October 1, 2002; 33(10): 2520 - 2522. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Stroke Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2002 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |