(Stroke. 2002;33:1308.)
© 2002 American Heart Association, Inc.
Original Contributions |
From the Department of Cardiology, The Cleveland Clinic Foundation, Cleveland, Ohio (P.L.W.); Department of Neuroradiology, ENERI, Buenos Aires, Argentina (P.L.); Department of Cardiology, Fundacion Favaloro, Buenos Aires, Argentina (H.L., O.A.M.); Department of Radiology, Stadtische Kliniken Dortmund Academic Teaching Hospital, Dortmund, Germany (K.M., H.J.); Department of Vascular Surgery, Instituto Cardiovascular de Buenos Aires, Buenos Aires, Argentina (J.P.); Department of Interventional Radiology, Clinica Sagrada Familia, Buenos Aires, Argentina (C.S.); and Department of Pathology, Cardiopsis-Conicet, Buenos Aires, Argentina (J.M.).
Correspondence to Patrick L. Whitlow, MD, Director, Interventional Cardiology, Department of Cardiology, F25, The Cleveland Clinic Foundation, 9500 Euclid Ave, Cleveland, OH 44195. E-mail whitlop{at}ccf.org
| Abstract |
|---|
|
|
|---|
Methods A multidisciplinary study group tested a balloon occlusion-aspiration emboli entrapment device in conjunction with carotid stenting. The device consists of an elastomeric balloon on a steerable wire with a detachable adapter that inflates and deflates the distal temporary occlusion balloon. An aspiration catheter is used to remove trapped emboli after stenting and before occlusion balloon deflation.
Results Seventy-five patients with severe internal carotid artery stenosis were treated with stents deployed with this cerebrovasculature protection system. All 75 patients (100%) had grossly visible particulate material aspirated, and all were treated successfully without major or minor stroke or death at 30 days. Preintervention stenosis was 81±10%, and residual stenosis was 5±7%. Nine patients (12%) had angiographic evidence of thrombus before intervention, but no patient had thrombus or vessel cutoff after the procedure. Four patients (5%) developed transient neurological symptoms during protection balloon occlusion, but symptoms resolved with balloon deflation. The 22 to 667 particles aspirated per patient ranged from 3.6 to 5262 µm in maximum diameter (mean, 203±256 µm). These particles included fibrous plaque debris, lipid or cholesterol vacuoles, and calcific plaque fragments.
Conclusions Protected carotid stenting was performed successfully and safely in this study early in the experience with cerebrovascular protection devices. Particulate emboli are frequent with stenting, and cerebral protection will likely be necessary to minimize stroke. Randomized trials comparing protected carotid stenting with endarterectomy are warranted.
Key Words: carotid artery diseases carotid endarterectomy stents stroke
| Introduction |
|---|
|
|
|---|
Because of the success of percutaneous stenting of atherosclerotic lesions in other arterial systems,710 several centers have investigated the less invasive alternative of stenting severe carotid stenoses.1116 However, fear of distal embolization of plaque fragments to the brain has generated concern regarding the safety and wisdom of this approach, especially considering the established low risk and durability of endarterectomy.17,18 The incidence of stroke generally reported with carotid stenting is 5% to 10%,1116 although 1 center reported stroke in 71% of an early series.19 Transcranial Doppler studies suggest the universal occurrence of emboli associated with this procedure.20 Therefore, most clinicians have avoided carotid stenting except in patients with high surgical risk or surgically inaccessible lesions.
Systems to guard against distal embolization during percutaneous intervention have recently been developed. The GuardWire (Percusurge Inc) emboli entrapment and aspiration system was tested in vitro, used in animals, and subsequently used in human saphenous vein graft interventions.2123 A randomized trial recently confirmed a 40% to 50% reduction in major complications during protected versus unprotected saphenous vein graft stenting.24 The first protected carotid artery intervention cases with this system began in 1998.25 A multidisciplinary group of physicians (a surgeon, an interventional neuroradiologist, 3 interventional radiologists, and 3 interventional cardiologists) treating patients with carotid disease agreed to evaluate this emboli containment and aspiration system in patients undergoing carotid artery stenting. These are the results of this investigation.
| Methods |
|---|
|
|
|---|
60% by angiography or asymptomatic patients with lesions
70% were considered for this study. Patients with a stroke within 1 week and patients with 100% occlusion of the ipsilateral carotid artery were excluded. The study protocol was approved by the institutional review board at each institution, and the patient signed informed consent to undergo this investigational procedure.
Carotid Stenting Procedure
Patients were treated with aspirin 81 to 325 mg/d for at least 24 hours and underwent an independent neurological examination at baseline. The patient was taken to the angiographic laboratory with an intravenous line infusing. A 7F to 9F sheath or guiding catheter was placed in the common carotid artery according to standard techniques1116 after local anesthesia. The patient was anticoagulated with 50 to 100 U/kg heparin until the activated clotting time was >250 seconds. Angiograms were recorded in at least 2 projections. Intracranial views were also documented.
The cerebral protection system used consists of 3 components: an exchange length guidewire, a MicroSeal adapter, and a monorail aspiration catheter. The wire is a 0.014- or 0.018-in angioplasty-style wire with a segment of hollow nitinol hypotube. The distal wire segment is shapeable, radiopaque, and steerable. Just proximal to this platinum distal segment is a compliant latex balloon capable of occluding blood flow when inflated. The proximal end of the hypotube wire incorporates a moveable seal allowing inflation and deflation of the balloon via the detachable adapter.
After the distal tip is shaped, the guidewire is advanced through the guide catheter under fluoroscopic guidance. The wire is steered through the carotid stenosis and advanced at least 3 cm beyond the target lesion (Figure 1). The intended predilatation balloon, stent, and postdilatation balloon are all prepped and available. The first device to be used in the lesion (predilatation balloon or stent, at operator discretion) is passed over the wire into the distal guiding catheter. The MicroSeal adapter is then attached, and the latex balloon is inflated to 5.5-mm diameter. Contrast is injected to confirm complete occlusion of distal flow. If flow is still present, then an additional 0.5 cm3 diluted contrast is added to enlarge the balloon until complete occlusion is achieved. The MicroSeal is then closed, and the adapter is removed. The predilatation balloon, if used, is passed across the lesion and inflated until any indentation of the balloon is resolved. The dilatation balloon is then deflated and removed. The stent is then placed over the guidewire and deployed in the lesion. The stent delivery system is removed, and the postdilatation balloon is then passed over the guidewire into the lesion. A single balloon inflation of 8 to 15 atm for 15 to 20 seconds is performed. The balloon is deflated and then removed. The aspiration catheter is placed up to the level of the distal occlusion balloon under fluoroscopy. Fifteen to 45 mL of blood is then aspirated from the area between the carotid artery bifurcation and the occlusion balloon. The aspiration catheter is removed. The occlusion balloon is then deflated, restoring antegrade flow.
|
Most procedures were completed with only 1 occlusion balloon inflation, aspiration, and deflation. In the first 3 patients and in 4 additional patients in the series, aspiration was performed and the occlusion balloon was deflated after 1 to 2 steps of a sequential procedure, ie, after predilatation followed by aspiration and occlusion balloon deflation, after stent placement followed by aspiration and occlusion balloon deflation, and after stent postdilatation followed by aspiration and occlusion balloon deflation. With the sequential approach, the occlusion time in each step could be minimized at operator discretion.
After final occlusion balloon deflation and removal, angiography was repeated, confirming adequate lumen diameter in at least 2 orthogonal views. Intracranial views were repeated to check for vessel cutoff or flow abnormality. The patient underwent frequent brief neurological assessments during the procedure (question and answers, hand and foot movement) and a detailed neurological evaluation at the end of the procedure.
The patient was taken to a recovery area where the sheath was removed when the activated clotting time was <170 seconds. The patient remained on bed rest for 6 to 9 hours. The following morning, a detailed neurological examination by an independent observer was repeated. Discharge medications included aspirin 325 mg/d indefinitely and ticlopidine 250 mg BID or clopidogrel 75 mg/d for at least 2 weeks. The patient returned at 30 days for a repeated independent neurological examination.
Particulate Analysis
Aspirated blood was injected from a syringe into a container through a 40-µm filter. The filter was rinsed by injection of 20 to 60 cm3 saline and then immersed in either 10% buffered formaldehyde or 2.5% glutaraldehyde in cacodylate buffer. Particulate debris from 31 consecutive patients was sent to a core pathology laboratory for analysis (J.M.).
After 24-hour fixation, the mesh of the filters was carefully cut from its frame and mounted on glass slides, stained with toluidine blue 1%, and covered with a coverslip. Digitized black-and-white images were used for measurement. The Optimas Analysis Program, version 5.0, was used. Each of 4215 particles was manually selected, and the maximum diameter of the particles was measured. In 2000 particles, area was also measured. In 7 patients, particles floating in the fixative were centrifuged, and a smear was stained with hematoxylin and eosin.
Data Analysis
The primary end point of the study was the composite of death or nonfatal stroke at 30 days. Secondary end points were balloon occlusion intolerance, transient ischemic attack, arrhythmia, congestive heart failure, myocardial infarction, pulmonary complications extending hospitalization, and access site complications. The study was designed as a prospective registry to determine the feasibility and complications of protected carotid stenting. Data are reported as mean±SD. Investigators agreed to compile the data and analyze the results after the first 50 to 75 cases had been completed. An independent group of monitors audited and source verified the data at each institution before analysis.
| Results |
|---|
|
|
|---|
|
|
Procedural details are compiled in Table 3. Four patients (5%) had balloon occlusion intolerance with transient neurological symptoms that resolved after balloon deflation. One of these intolerant patients had significant contralateral carotid stenosis. One patient had adjunctive thrombus aspiration with the AngioJet (Possis Inc). No angiographic evidence of vessel cutoff or flow abnormality was seen.
|
Neurological examinations (immediately after the procedure and the next day) confirmed no evidence of stroke after the procedure. One patient with sustained hypotension (systolic blood pressure
70 mm Hg for 4 hours) and intracranial ipsilateral and contralateral carotid stenoses >70% had a transient ischemic attack resolved with restoration of systolic blood pressure >100 mm Hg. No unstable angina, ventricular arrhythmia, or myocardial infarction occurred. One patient developed pulmonary edema after being given a rapid infusion of saline for relative hypotension (systolic blood pressure, 90 to 100 mm Hg) immediately after stenting. This patient responded to diuresis and medical management. Three patients required transfusion for bleeding. Two of these patients required surgical femoral artery repair.
At 30 days, repeated neurological examination revealed no incidence of cerebrovascular accident, and no patient reported symptoms consistent with transient ischemic attack after hospital discharge. No patients died or were rehospitalized at 30 days.
Particulate debris was visible in all aspiration samples (Figure 2). The number of particles analyzed per patient ranged from 22 to 667. The mean maximum diameter was 203±256 µm. Particles ranged from 3.6- to 5262-µm maximum diameter. Forty-one percent of particles were <100 µm in maximum dimension, 64% were <200 µm, and 35% were 200 to 1000 µm. Particles ranged from 41.2 to 5157 µm2 in area. The mean area of the particles was 279.4 µm2.
|
On microscopic examination, these particles consisted of lipid vacuoles with cholesterol clefts surrounded by fibrin and platelets, fibrous eosinophilic staining material consistent with connective tissue from the cap of a plaque, calcific plaque fragments, lymphocyte clusters, and platelet clumps (Figure 3).
|
| Discussion |
|---|
|
|
|---|
Several devices to trap emboli have been conceived and developed. Theron and colleagues reported their experience with the first cerebral protection device, but that device has not been universally adopted for carotid stenting because of its limited steerability and relatively large profile. This first-generation device also did not eliminate emboli or clinical complications.26,27
Since those pioneering efforts, emboli containment devices have evolved.25 We examined and scrutinized the system tested in this study and found it encouraging enough to warrant a clinical trial with carotid stenting. The data collected are strikingly positive. We judged the 100% success rate and the elimination of stroke in 75 consecutive cases of protected carotid stenting to be a major clinical advance. The presence of visible particles aspirated from all patients and the absence of stroke as a complication support the use of an emboli containment system in cases of percutaneous carotid intervention. The ease of use of the balloon on a wire and its adaptability into the clinical practices of interventionists in vascular surgery, radiology, neuroradiology, and cardiology were established in this study.
The emboli containment device tested does have the limitation of not stopping distal embolization until the occlusion balloon is inflated. Crossing the lesion with this balloon on a wire does create transcranial Doppler signals suggesting microemboli, but the number of emboli counts is only a fraction of those found with unprotected stenting.28 In addition, this device does transiently interrupt antegrade flow in the carotid artery. However, most patients (95% in this series) tolerate temporary balloon occlusion without symptoms, and all patients could be treated successfully with the serial inflation-deflation cycles used in those patients manifesting transient symptoms. Diffusion-weighted MRI of the brain was not done in this study. However, sensitive imaging techniques are needed in future studies to determine whether subclinical emboli are prevented by this and other cerebral protection devices. Likewise, diffusion-weighted MRI may be useful in determining whether prolonged occlusion times may cause subclinical defects. However, the long occlusion times used in this study did not cause clinical stroke.
Interestingly, significant contralateral carotid stenosis was not associated with balloon occlusion intolerance in any patient in this series. Online transcranial Doppler monitoring of middle cerebral artery flow velocity, used in only a small subset of patients in this series, might prove useful in predicting patients at high risk of balloon occlusion intolerance, but clinical parameters did not allow the operators to predict which patients might benefit from brief, sequential occlusion balloon cycles rather than 1 prolonged inflation.
Every new device used in percutaneous interventional procedures adds to the complexity and cost of the procedure. The cerebral protection system used in this study was added to the carotid stenting procedure without causing major clinical complications. However, the case involving a Guardwire balloon-induced dissection requiring an additional stent
4 to 5 cm above the carotid bifurcation illustrated that this device does have the potential for causing vascular complications even while limiting emboli to the brain. This complication, plus the price of the device ($1575 for the saphenous vein graft GuardWire System; Medtronic, Inc), underscores the importance of larger trials and cost-effectiveness analysis to determine the ultimate role of this new tool for the treatment of patients with carotid disease. The carotid protection device used in this study is not approved or available for routine clinical use in North America, but a similar device is approved and marketed for use in saphenous vein graft intervention.
Pathological analysis of the aspirated material suggests that emboli are an expected consequence of carotid stenting. The data imply that an effective, atraumatic protection device will be necessary to limit carotid stenting complications, possibly to the level expected with the established gold standard of carotid endarterectomy. Clearly, there will be further improvements in carotid stenting equipment and cerebrovascular protection devices over the next few years. However, the device studied in this trial has evolved sufficiently to warrant routine use as an adjunct to carotid stenting.
The emboli entrapment-aspiration system tested appears to provide adequate protection to the cerebral circulation to make carotid stenting a viable alternative to endarterectomy in symptomatic patients with high surgical risk, and perhaps even in asymptomatic patients who are routine surgical candidates. Randomized trials of protected carotid stenting versus carotid surgery should be expedited.
| Acknowledgments |
|---|
Received March 21, 2001; revision received January 7, 2002; accepted January 9, 2002.
| References |
|---|
|
|
|---|
2. European Carotid Surgery Trialists Collaborative Group. MRC European Carotid Surgery Trial: interim results for symptomatic patients with severe (7090%) or with mild (029%) carotid stenosis. Lancet. 1991; 337: 12351243.[CrossRef][Medline] [Order article via Infotrieve]
3.
Mayberg MR, Wilson ES, Yatsu F, Weiss DG, Messina L, Hershey LA, Colling C, Eskridge J, Deykin D, Winn HR. Carotid endarterectomy and prevention of cerebral ischemia in symptomatic carotid stenosis. JAMA. 1991; 266: 32893294.
4.
Executive Committee for the Asymptomatic Carotid Atherosclerosis Study. Endarterectomy for asymptomatic carotid artery stenosis. JAMA. 1995; 273: 14211428.
5. Baker JD. Recurrent stenosis of the carotid artery: incidence, diagnosis, prognosis, and management.In: Moore WS, ed. Surgery for Cerebrovascular Disease. 1st ed. New York, NY: Churchill Livingstone Inc; 1987: 703713.
6.
Moore WS, Barnett HJM, Beebe HG, Bernstein EF, Brener BJ, Brott T, Caplan LR, Day A, Goldstone J, Hobson RWII, Kempczinski RF, Matchar DB, Mayberg MR, Nicolaides AN, Norris JW, Ricotta JJ, Robertson JT, Rutherford RB, Thomas D, Toole JF, Trout HH III, Wiebers DO. Guidelines for carotid endarterectomy: a multidisciplinary consensus statement from the Ad Hoc Committee, American Heart Association. Circulation. 1995; 91: 566579.
7. Palmaz JC, Garcia OJ, Schatz RA, Rees CR, Roeren T, Richter GM, Noeldge G, Gardiner GA Jr, Becker GJ, Walker C. Placement of balloon-expandable intraluminal stents in iliac arteries: first 171 procedures. Radiology. 1990; 174: 969975.[Abstract]
8.
Becker GJ, Palmaz JC, Rees CR, Ehrman KO, Lalka SG, Dalsing MC, Cikrit DF, McLean GK, Burke DR, Richter GM. Angioplasty-induced dissections in human iliac arteries: management with Palmaz balloon-expandable intraluminal stents. Radiology. 1990; 176: 3138.
9.
Serruys PW, De Jaegere P, Kiemeneij F, Macaya C, Rutsch W, Heyndrickx G, Emanuelsson H, Marco J, Legrand V, Materne P, Belardi J, Sigwart U, Colombo A, Goy JJ, Van den Heuvel P, Delcan J, Morel MA. A comparison of balloon-expandable-stent implantation with balloon angioplasty in patients with coronary artery disease: Benestent Study Group. N Engl J Med. 1994; 331: 489495.
10.
Fischman DL, Leon MB, Baim DS, Schatz RA, Savage MP, Penn I, Detre K, Veltri L, Ricci D, Nobuyoshi M, Cleman M, Heuser R, Almond D, Teirstein PS, Fish RD, Colombo A, Brinker J, Moses J, Shaknovich A, Hirshfeld J, Bailey S, Ellis S, Rake R, Goldberg S. A randomized comparison of coronary-stent placement and balloon angioplasty in the treatment of coronary artery disease: Stent Restenosis Study Investigators. N Engl J Med. 1994; 331: 496501.
11.
Yadav JS, Roubin GS, Iyer S, Vitek J, King P, Jordan WD, Fisher WS. Elective stenting of the extracranial carotid arteries. Circulation. 1997; 95: 376381.
12.
Yadav JS, Roubin GS, King P, Iyer S, Vitek J. Angioplasty and stenting for restenosis after carotid endarterectomy: initial experience. Stroke. 1996; 27: 20752079.
13.
Mathur A, Roubin GS, Iyer SS, Piamsonboon C, Liu MW, Gomez CR, Yadav JS, Chastain HD, Fox LM, Dean LS, Vitek JJ. Predictors of stroke complicating carotid artery stenting. Circulation. 1998; 97: 12391245.
14. Shawl FA. Elective Stenting of extracranial carotid arteries: short- and mid-term results.In: Branchereau A, ed. New Trends and Developments in Carotid Artery Disease. 1st ed. Armonk, NY: Futura Publishing Co; 1998: 147160.
15. Wholey MH, Wholey M, Bergeron P, Diethrich EB, Henry M, Laborde JC, Mathias K, Myla S, Roubin GS, Shawl F, Theron JG, Yadav JS, Dorros G, Guimaraens J, Higashida R, Kumar V, Leon M, Lim M, Londero H, Mesa J, Ramee S, Rodriguez A, Rosenfield K, Teitelbaum G, Vozzi C. Current global status of carotid artery stent placement. Cathet Cardiovasc Diagn. 1998; 44: 16.[CrossRef][Medline] [Order article via Infotrieve]
16. Henry M, Amor M, Masson I, Henry I, Tzvetanov K, Chati Z, Khanna N. Angioplasty and stenting of the extracranial carotid arteries. J Endovasc Surg. 1998; 6: 293304.
17.
Beebe HG, Archie JP, Baker WH, Barnes RW, Becker GJ, Bernstein EF, Brener B, Clagett GP, Clowes AW, Cooke JP. Concern about safety of carotid angioplasty. Stroke. 1996; 27: 197198.
18.
Bettmann MA, Katzen BT, Whisnant J, Brant-Zawadzki M, Broderick JP, Furlan AJ, Hershey LA, Howard V, Kuntz R, Loftus CM, Pearce W, Roberts A, Roubin G. Carotid stenting and angioplasty: a statement for healthcare professionals from the Councils on Cardiovascular Radiology, Stroke, Cardio-Thoracic and Vascular Surgery, Epidemiology and Prevention, and Clinical Cardiology, American Heart Association. Circulation. 1998; 97: 121123.
19. Naylor AR, Bolia A, Abbott RJ, Pye IF, Smith J, Lennard N, Lloyd AJ, London NJ, Bell PR. Randomized study of carotid angioplasty and stenting versus carotid endarterectomy: a stopped trial. J Vasc Surg. 1998; 28: 326334.[CrossRef][Medline] [Order article via Infotrieve]
20. Jordan W Jr, Voellinger DC, Doblar DD, Plyushchevat NP, Fisher WS, McDowell HA. Microemboli detected by transcranial Doppler monitoring in patients during carotid angioplasty versus carotid endarterectomy. Cardiovasc Surg. 1999; 7: 3336.[CrossRef][Medline] [Order article via Infotrieve]
21. Oesterle SN, Hayase M, Baim DS, Teirstein PS, Ramee SR, Whitlow PL, Webb J, Virmani R. An embolization containment device. Cathet Cardiovasc Interv. 1999; 47: 243250.[CrossRef][Medline] [Order article via Infotrieve]
22.
Webb JG, Carere RG, Virmani R, Baim D, Teirstein PS, Whitlow P, McQueen C, Kolodgie FD, Buller E, Dodek A, Mancini GB, Oesterle S. Retrieval and analysis of particulate debris after saphenous vein graft intervention. J Am Coll Cardiol. 1999; 34: 468475.
23.
Carlino M, De Gregorio J, Di Mario C, Anzuini A, Airoldi F, Albiero R, Briguori C, Dharmadhikari A, Sheiban I, Colombo A. Prevention of distal embolization during saphenous vein graft lesion angioplasty. Circulation. 1999; 99: 32213223.
24. Baim DS. The SAFER Trial: SVG Angioplasty Free of Emboli Randomized. Presented at: the Transcatheter Cardiovascular Therapeutics Conference, October 1721, 2000; Washington, DC.
25. Henry M, Amor M, Henry I, Klonaris C, Chati Z, Masson I, Kownator S, Luizy F, Hugel M. Carotid stenting with cerebral protection: first clinical experience using the Percusurge Guardwire system. J Endovasc Surg. 1999; 6: 321331.[CrossRef][Medline] [Order article via Infotrieve]
26. Theron J, Raymond J, Casasco A, Courtheoux F. Percutaneous angioplasty of atherosclerotic and postsurgical stenosis of carotid arteries. Am J Neuroradiol. 1987; 8: 495500.[Abstract]
27. Theron J, Courtheoux P, Alachkar F, Bouvard G, Maiza D. New triple coaxial catheter system for carotid angioplasty with cerebral protection. Am J Neuroradiol. 1990; 11: 869874.[Medline] [Order article via Infotrieve]
28. Whitlow PL, Katzan IL, Dagirmanjian A, Davis DA, Silver M, Krieger DW, Yadav JS. Embolization during protected versus unprotected carotid stenting. Circulation. 2000; 102 (suppl II): II-2310.Abstract.
This article has been cited by other articles:
![]() |
N. Janjua, A. Alkawi, M.F.K. Suri, and A.I. Qureshi Impact of Arterial Reocclusion and Distal Fragmentation during Thrombolysis among Patients with Acute Ischemic Stroke AJNR Am. J. Neuroradiol., February 1, 2008; 29(2): 253 - 258. [Abstract] [Full Text] [PDF] |
||||
![]() |
M.D. Hill, W. Morrish, G. Soulez, A. Nevelsteen, G. Maleux, C. Rogers, K.E. Hauptmann, A. Bonafe, R. Beyar, L. Gruberg, et al. Multicenter evaluation of a self-expanding carotid stent system with distal protection in the treatment of carotid stenosis. AJNR Am. J. Neuroradiol., April 1, 2006; 27(4): 759 - 765. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. A. Gorog, R. A. Foale, and I. Malik Distal Myocardial Protection During Percutaneous Coronary Intervention: When and Where? J. Am. Coll. Cardiol., October 18, 2005; 46(8): 1434 - 1445. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. I. Paraskevas, S. S. Daskalopoulou, M. E. Daskalopoulos, and C. D. Liapis Secondary Prevention of Ischemic Cerebrovascular Disease. What Is the Evidence? Angiology, September 1, 2005; 56(5): 539 - 552. [Abstract] [PDF] |
||||
![]() |
R. Zahn, T. Ischinger, B. Mark, S. Gass, U. Zeymer, W. Schmalz, K. Haerten, K. E. Hauptmann, E.-R. von Leitner, W. Kasper, et al. Embolic Protection Devices for Carotid Artery Stenting: Is There a Difference Between Filter and Distal Occlusive Devices? J. Am. Coll. Cardiol., June 7, 2005; 45(11): 1769 - 1774. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. E. Rodriguez and L. Leon Current Status of Carotid Angioplasty and Stenting Perspectives in Vascular Surgery and Endovascular Therapy, June 1, 2004; 16(2): 111 - 119. [Abstract] [PDF] |
||||
![]() |
W. A. Gray A cardiologist in the carotids J. Am. Coll. Cardiol., May 5, 2004; 43(9): 1602 - 1605. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. Linfante, J. A. Hirsch, M. Selim, G. Schlaug, L. R. Caplan, and A. S. Reddy Safety of Latest-Generation Self-expanding Stents in Patients With NASCET-Ineligible Severe Symptomatic Extracranial Internal Carotid Artery Stenosis Arch Neurol, January 1, 2004; 61(1): 39 - 43. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Sabeti, M. Schillinger, W. Mlekusch, T. Nachtmann, W. Lang, R. Ahmadi, and E. Minar Contralateral High-Grade Carotid Artery Stenosis or Occlusion Is Not Associated with Increased Risk for Poor Neurologic Outcome after Elective Carotid Stent Placement Radiology, January 1, 2004; 230(1): 70 - 76. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. D. Barr, J. J. Connors III, D. Sacks, J. C. Wojak, G. J. Becker, J. F. Cardella, B. Chopko, J. E. Dion, A. J. Fox, R. T. Higashida, et al. Quality Improvement Guidelines for the Performance of Cervical Carotid Angioplasty and Stent Placement: Developed by a Collaborative Panel of the American Society of Interventional and Therapeutic Neuroradiology, the American Society of Neuroradiology, and the Society of Interventional Radiology AJNR Am. J. Neuroradiol., November 1, 2003; 24(10): 2020 - 2034. [Full Text] [PDF] |
||||
![]() |
P. H. Lin, R. L. Bush, and A. B. Lumsden Carotid Artery Stenting: Current Status and Future Directions Vascular and Endovascular Surgery, September 1, 2003; 37(5): 315 - 322. [Abstract] [PDF] |
||||
![]() |
B. Eckert and H. Zeumer Editorial Comment--Carotid Artery Stenting With or Without Protection Devices? Strong Opinions, Poor Evidence! Stroke, August 1, 2003; 34(8): 1941 - 1943. [Full Text] [PDF] |
||||
![]() |
A. Kastrup, K. Groschel, H. Krapf, B. R. Brehm, J. Dichgans, and J. B. Schulz Early Outcome of Carotid Angioplasty and Stenting With and Without Cerebral Protection Devices: A Systematic Review of the Literature Stroke, March 1, 2003; 34(3): 813 - 819. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Macdonald and P. A Gaines Current concepts of mechanical cerebral protection during precutaneous carotid intervention Vascular Medicine, February 1, 2003; 8(1): 25 - 32. [Abstract] [PDF] |
||||
![]() |
D. M. Pelz Advances in Interventional Neuroradiology Stroke, February 1, 2003; 34(2): 357 - 358. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Stroke Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2002 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |