(Stroke. 2002;33:1803.)
© 2002 American Heart Association, Inc.
Original Contributions |
From the Departments of Neuroradiology (M.O., H.K.) and Neurosurgery (M.Y., H.N.), Nagatomi Neurosurgical Hospital, and the Department of Radiology (H.H., T.S., Y.S., H.M.), Oita Medical University, Oita, Japan.
Correspondence to Mika Okahara, MD, Department of Radiology, Nagatomi Neurosurgical Hospital, Omichi-machi 4-5-28, Oita 870-0822, Japan. E-mail okaharam{at}mb.infoweb.ne.jp
| Abstract |
|---|
|
|
|---|
Methods 3D-TOF MRA was performed in 82 patients with 133 cerebral aneurysms. Each patient underwent rotational angiography. Three-dimensional reconstructed images were obtained from data of the rotational angiography (as the gold standard). A blind study with 4 readers of different experiences was performed to evaluate the diagnostic accuracy of 3D-TOF MRA for cerebral aneurysms.
Results One hundred five (79%) of all 133 aneurysms were detected with MRA by a neuroradiologist, 100 (75%) were detected by an experienced neurosurgeon, 84 (63%) were detected by a general radiologist, and 80 (60%) were detected by a resident neuroradiologist. For each reader, the detectability was lower for small aneurysms (<3 mm in maximum diameter) and/or for those located at the internal carotid artery and anterior cerebral artery. False-positive aneurysms were 29 for the neuroradiologist, 19 for the neurosurgeon, 31 for the general radiologist, and 30 for the resident neuroradiologist; most of the aneurysms were at the internal carotid artery. Causes of the false-positive and false-negative results included complex flow in a tortuous artery and susceptibility artifacts.
Conclusions Although MRA is useful in the diagnosis of cerebral aneurysms, sufficient experience and careful attention are necessary for accurate diagnosis of aneurysms located at the internal carotid and anterior cerebral arteries.
Key Words: angiography, digital subtraction intracranial aneurysm magnetic resonance angiography
| Introduction |
|---|
|
|
|---|
In the present study, we determined the sensitivity of 3Dtime of flight (3D-TOF) MRA in the detection of cerebral aneurysms by using 3D-DSA as the gold standard. We also evaluated the effects of several factors on the sensitivity of 3D-TOF, including location of the aneurysms, number of the aneurysms, and experience of the image reader.
| Subjects and Methods |
|---|
|
|
|---|
All MRAs were performed with the use of 1.5-T magnetic resonance equipment (VISERT H1, Toshiba) with the following parameters: repetition time 40 ms, echo time 6.8 ms, flip angle 20°, field of view 160 mm, volume thickness 52 mm, matrix 160x256, number of acquisitions 1, and acquisition time 10 minutes 2 seconds. Maximal intensity projection (MIP) reconstructions were performed at the time of imaging. The data were reconstructed around the head-to-foot axis and right-to-left axis. If necessary, target MIP reconstructions were performed.
All angiographic procedures were performed by using a DSA unit (Advantx LCN Plus, GE Medical Systems). Three-dimensional DSA images were produced by a Workstation (Advantage Workstation 3.1, GE Medical Systems) with the use of data from rotational angiographies. Rotational angiography was performed with a 200° rotation of the C arm in 5 seconds. The matrix size of each frame was 512x512 pixels. Fifteen to 25 mL of the contrast medium (Iohexol, 300 mg/dL) was injected at a rate of 3 to 5 mL/s via a catheter positioned at the cervical portion of the carotid arteries and/or vertebral arteries. Both mask data and contrast data were automatically transferred to the workstation. Three-dimensional subtraction images were automatically produced on the workstation by a 3D-reconstruction algorithm based on the algebraic reconstruction technique. Reconstructed images, including MIP and shaded surface display were produced from the data. Minimum density threshold processing was automatically performed at the same levels. The maximum diameters of each aneurysm were calculated on the workstation.
Randomized MRA images were blindly interpreted by 4 readers who had different levels of experience: a neurosurgeon (NS, 20 years of experience), a neuroradiologist (NR, 12 years of experience), a general radiologist (GR, 11 years of experience), and a resident neuroradiologist (RR, 4 years of experience). On the first interpretation, all readers were blinded to clinical information. They reinterpreted the same cases again 12 months later after being provided with clinical information and CT. Two other neuroradiologists retrospectively compared the MRA findings with those of 3D-DSA. First, they discussed and interpreted 3D-DSA and excluded patients without aneurysms. Then they assessed the sensitivity of detection of the cerebral aneurysms and recorded differences in the sensitivity due to reader experience, location and size of the aneurysms, and the presence or absence of SAH and differences in the first and second interpretations. Furthermore, in definitely detected aneurysms, they assessed the correlation between MRA and 3D-DSA findings with regard to the following parameters: maximum diameter, neck size, shape, relationship of the aneurysm and neighboring branches (involved or not), and the presence or absence of blebs. Causes of false-negative were also investigated. These differences in sensitivity due to each factor were subjected to statistical analysis. Differences based on readers and locations were tested by the Scheffé test. The Mann-Whitney test was used for analysis of differences in the detectability according to size, amount (single or multiple), and existence of SAH. The Wilcoxon signed rank test was used for analysis of differences in the detectability according to the first and second interpretations. The statistical level of significance was set at P<0.05.
| Results |
|---|
|
|
|---|
The sensitivities of MRA in detection of cerebral aneurysms are demonstrated in Figures 1 to 6. The sensitivity in detection of aneurysms was 79% for NR, 75% for NS, 63% for GR, and 60% for RR (Figure 1). The difference in sensitivity between NR and GR was significant (P<0.05), as was the difference between NR and RR (P<0.05). With regard to aneurysm location, the sensitivity of all readers was lower in identifying aneurysms located at ICA (47% to 71%) and ACA (14% to 50%) than at other locations (Figure 2). Significant differences in sensitivity were noted between detecting ACA and MCA aneurysms by NS, GR, and RR (P<0.05, P<0.05, and P<0.05, respectively), ACA and posterior circulation by NS (P<0.05), ACA and ACoA by RR (P<0.001), and ICA and ACoA by RR (P<0.05). The sensitivity of all readers (38% to 55%) was significantly lower for small aneurysms (ie, those <3 mm in maximum diameter) than for large aneurysms (68% to 89%) (Figure 3, P
0.0001 for NS, NR, and GR and P<0.01 for RR). The sensitivity in detecting 10 ruptured small aneurysms (<3 mm in maximum diameter) was 60% for each reader. The sensitivity in detecting multiple aneurysms (51% to 73%) was significantly lower than that in detecting single aneurysms (77% to 90%) (Figure 4, P<0.05 for NS, P<0.05 for NR, P<0.05 for GR, and P<0.001 for RR). Sensitivity was lower in the detection of aneurysms with SAH (54% to 79%) than in the detection of aneurysms without SAH (65% to 79%) (Figure 5), albeit statistically insignificant (P=0.29 for NS, P=0.98 for NR, P=0.13 for GR, and P=0.24 for RR). Eighteen false-positive aneurysms were encountered by NS, 29 were encountered by NR, 31 were encountered by GR, and 30 were encountered by RR. The majority of false-positive aneurysms were located in the ICA.
|
|
|
|
|
At the second interpretation performed with clinical information made available, sensitivity in the detection of aneurysms was 79% for NR, 81% for NS, 67% for GR, and 74% for RR (Figure 6). The sensitivity significantly improved in RR (P<0.05) compared with the initial interpretation. In NS and GR, the sensitivity slightly improved, although it was statistically insignificant (P=0.059 for NS, P=0.25 for GR).
|
Diameters of the neck and dome measured on MRA were significantly correlated with the diameters calculated on 3D-DSA. The correlation coefficients of neck diameter and dome diameter, respectively, were 0.764 and 0.863 for NR and 0.764 and 0.886 for NS. The coincidence ratios of shape, relationship of the aneurysm and neighboring branches, and the presence or absence of blebs were 78% (73 of 93), 85% (79 of 93), and 77% (72 of 93) for NR and 74% (63 of 85), 55% (47 of 85), and 70% (59 of 85) for NS, respectively.
In retrospective analysis, 71% of the false-negative aneurysms were detectable. The causes of false-negative interpretations are listed in the Table. Thirty-one of 67 false-negative aneurysms were simply overlooked. Ten were caused by inhomogeneous signal intensity due to atherosclerotic changes11 or turbulent flow.3 Nine were not visible because of the small size, 6 were not distinguishable from a branch artery, and 5 were due to an overlapping of artery and aneurysm. The main factors that contributed to false-positive interpretations included tortuous origin of the branch, infundibular widening of the posterior communicating artery, and large and/or tortuous ACoA.
|
| Discussion |
|---|
|
|
|---|
3 mm in size.46 The sensitivity in the present study was 60% to 79%, which is lower than the percentages reported in previous studies. Conventional DSA was used as the gold standard in previous studies. Small aneurysms can be overlooked on conventional DSA because of overlapping cerebral arteries. Recently, 3D-DSA has emerged as a useful technique for visualizing intracranial vascular lesions. Compared with the conventional DSA, 3D-DSA can detect intracranial aneurysms with greater levels of sensitivity and accuracy because it allows the examiner to observe cerebral arteries from any projection.1618 The high detectability of cerebral aneurysms on 3D-DSA should affect the sensitivity reported in the present study. In clinical practice, however, a diagnostic study is usually observed by a single reader, whereas in some reports, the sensitivity in the detection of cerebral aneurysms was the result of examination by
2 readers, which could have resulted in a positive bias toward MRA. Some authors have shown that sensitivity improved with combined readings of MIP and source images.13,21 In clinical practice, however, routine use of the source images may be difficult in some institutions because of the necessary time and cost. In the present study, aneurysms of the ICA and ACA were difficult to detect with the use of MRA. Because of signal loss due to the saturation effect, detection of aneurysms in the distal ACA located at the margin of the slab may be difficult. Moreover, aneurysms of the ACA were often small. Flow signal loss at the siphon of the ICA is well known.1 In addition to flow signal loss, overlap of the artery and aneurysm, atherosclerotic changes in the parent artery, and turbulent flow will make detection of aneurysms at the ICA very difficult (Figure 7). On the other hand, aneurysms of the MCA and BA were easily detected. The relatively simple structure of the MCA and BA bifurcation is well evident on MRA.
|
The size of the aneurysms is one of the important factors affecting sensitivity. Aprile6 reported that the sensitivity of detecting aneurysms <3 mm in size (25%) was lower than that of aneurysms >3 mm (92%). In the present study, the sensitivity of detection of aneurysms >3 mm in size was 68% to 89%, and that of aneurysms
3 mm in size was only 38% to 55%. In agreement with previous reports, the results of the present study support the notion that the 3-mm size is a practical cutoff level for the detection of aneurysms. In this regard, several studies have reported that these small aneurysms are at less risk of rupture (0.05% to 0.7%).22,23 In our cases, 10 (25%) of 40 aneurysms, which were <3 mm in size, were ruptured. Detectability of these small ruptured aneurysms was 60% by each reader. Because of the lower sensitivity for small aneurysms, it is probably risky to use only MRA in patients with SAH (Figure 8).
|
In patients with multiple cerebral aneurysms, the observers tended to be more watchful for large aneurysms, while overlooking coexistent smaller aneurysms, which were frequently <3 mm in size. One should be careful in examining MRA and search for >1 aneurysm because of the high frequency of multiple cerebral aneurysms.
A number of investigators have indicated that there is little or no difficulty in the identification of aneurysms irrespective of the presence of SAH.12,24 In the present study, the sensitivity of detecting aneurysms with SAH (54% to 79%) was slightly lower than that of detecting aneurysms without SAH (65% to 79%). In our retrospective analysis, we found no difficulty due to the signal intensity of the SAH and parenchymal hematoma in the detection of cerebral aneurysms with SAH. Therefore, the low sensitivity of detection of aneurysms with SAH may be due to motion artifact based on the often poor clinical condition of these patients.
With regard to interreader differences, the sensitivities of NR and experienced NS were higher than those of the others. This suggests that the experience of the observer is also an important factor in improving the sensitivity in the detection of cerebral aneurysms. In the second interpretation performed after providing relevant clinical information, sensitivity in the detection of aneurysms was significantly improved for RR (P<0.05). We believe that the improvement of detectability by RR depended not only on the availability of the clinical information but also on the improved experience of RR in diagnostic neuroradiology.
The morphological risk factors of rupture of cerebral aneurysms include size of the dome and location and presence of the bleb. Regarding the size of aneurysm (neck and dome), both MRA findings and 3D-DSA findings correlated well; however, detectability of the bleb was relatively lower (70% to 77%). Therefore, we believe that it is difficult to assess the angioarchitecture of aneurysms and predict the risk of aneurysm by MRA only.
We used MIP images in the present study. Recent developments in computed technology have included the availability of 3D-reconstructed images of MRA within a relatively short time. These 3D magnetic resonance images will contribute to improvement in the detection of cerebral aneurysms. In conclusion, although MRA is a useful tool for the diagnosis of cerebral aneurysms, we consider that the actual sensitivity in the detection of cerebral aneurysms is lower than that described in previous reports. In this regard, experience and careful attention are important factors that could improve sensitivity and accurate detection of single and multiple cerebral aneurysms, especially those located at the ICA and ACA.
Received November 14, 2001; revision received January 16, 2002; accepted March 13, 2002.
| References |
|---|
|
|
|---|
2.
Keller PJ, Drayer BP, Fram EK, Williams KD, Dumoulin CL, Souza SP. MR angiography with two-dimensional acquisition and three-dimensional display. Radiology. 1989; 173: 527532.
3.
Dumoulin CL, Hart HR. Magnetic resonance angiography. Radiology. 1986; 161: 717720.
4. Huston III J, Nicholas D, Luetmer P, Goodwin KD, Meyer FB, Wieber DO, Weaver AL. Blinded prospective evaluation of sensitivity of MR angiography to known intracranial aneurysms: importance of aneurysm size. AJNR Am J Neuroradiol. 1994; 15: 16071614.[Abstract]
5. Ross JS, Masaryk TJ, Modic MT, Ruggieri PM, Haacke EM, Selman WR. Intracranial aneurysms: evaluation by MR angiography. AJNR Am J Neuroradiol. 1990; 11: 449456.[Abstract]
6. Aprile I. Evaluation of cerebral aneurysms with MR-angiography. Rev Neuroradiol. 1996; 9: 541550.
7. White PM, Teadale EM, Wardlaw JM, Easton V. Intracranial aneurysms: CT angiography and MR angiography for detection: prospective blinded comparison in a large patient cohort. Radiology. 2001; 210: 739749.
8.
Ronkainen A, Puranen MI, Hernesniemi JA, Vanninen RL, Partanen LKP, Saari TJ, Vaninio PA, Ryynanen M. Intracranial aneurysms: MR angiographic screening in 400 asymptomatic individuals with increased familial risk. Radiology. 1995; 195: 3540.
9.
White PM, Wardlaw JM, Easton V. Can noninvasive imaging accurately depict intracranial aneurysms?: a systematic review. Radiology. 2000; 217: 361370.
10.
Stock KW, Radue EW, Jacob AL, Bao XS, Steinbrich W. Intracranial arteries: prospective blinded comparative study of MR angiography and DSA in 50 patients. Radiology. 1995; 195: 451456.
11. Schierer G, Huk WJ, Laub G. Magnetic resonance angiography of intracranial aneurysms: comparison with intracranial digital subtraction angiography. Neuroradiology. 1992; 35: 5054.[CrossRef][Medline] [Order article via Infotrieve]
12. Sankhla SK, Gunawardena WJ, Coutinho CMA, Jones AP, Keogh AJ. Magnetic resonance angiography in the management of aneurysmal subarachnoid hemorrhage: a study of 51 cases. Neuroradiology. 1996; 38: 724729.[CrossRef][Medline] [Order article via Infotrieve]
13. Wilcock D, Japsan T, Holland I, Cherryman G, Worthington B. Comparison of magnetic resonance angiography with conventional angiography in the detection of intracranial aneurysms in patients presenting with subarachnoid haemorrhage. Clin Radiol. 1996; 51: 330334.[CrossRef][Medline] [Order article via Infotrieve]
14. Horikoshi T, Fukamachi A, Nishi H, Fukasawa. Detection of intracranial aneurysms by three-dimensional time-of-flight magnetic resonance angiography. Neuroradiology. 1994; 36: 203207.[CrossRef][Medline] [Order article via Infotrieve]
15. Ikawa F, Sumida M, Uozumi T, Kuwabara S, Kiya K, Kurisu K, Arita K, Satoh H Comparison of three-dimensional phase-contrast magnetic resonance angiography with three-dimensional time-of-flight magnetic resonance angiography in cerebral aneurysms. Surg Neurol. 1994; 42: 287292.[CrossRef][Medline] [Order article via Infotrieve]
16. Unger B, Link J, Bohm-Jurkovic. Digital 3D-rotational angiography in preoperative and preinterventional diagnosis of intracranial aneurysms [in German]. Rofo Fortschr Geb Rontgenstr Neuen Bildgeb Verfahr. 1999; 170: 482491.[Medline] [Order article via Infotrieve]
17. Tanoue S, Kiyosue H, Kenai H, Nakamura T, Yamashita M, Mori H. Three-dimensional reconstructed images after rotational angiography in the evaluation of intracranial aneurysms: surgical correlation. Neurosurgery. 2000; 47: 866871.[CrossRef][Medline] [Order article via Infotrieve]
18.
Anxionnat R, Bracard S, Ducrocq X, Trousset Y, Launary L, Kerrien E, Braun M, Valliant R, Scomazzoni F, Lebedinsky A, et al. Intracranial aneurysms: clinical value of 3D digital subtraction angiography in the therapeutic decision and endovascular treatment. Radiology. 2001; 218: 799808.
19.
Metens T, Rio F, Baleriaux D, Roger T, David P, Rodesch G. Intracranial aneurysms: detection with gadolinium-enhanced dynamic three-dimensional MR angiography: initial results. Radiology. 2000; 216: 3946.
20.
Jager HR, Ellamushi H, Moore EA, Grieve JP, Kitchen ND, Taylor WJ. Contrast-enhanced MR angiography of intracranial giant aneurysms. AJNR Am J Neuroradiol. 2000; 21: 19001907.
21.
Korogi Y, Takahashi M, Mabuchi N, Nakagawa T, Fujiwara S, Horikawa Y, Miki H, Ouchi T, Shiga H, Shiokawa Y, et al. Intracranial aneurysm: diagnostic accuracy of MR angiography with evaluation of maximum intensity projection and source images. Radiology. 1996; 199: 199207.
22.
International Study of Unruptured Intracranial Aneurysms Investigators. Unruptured intracranial aneurysms: risk of rupture and risks of surgical intervention. N Engl J Med. 1998; 339: 17251733.
23.
Rinkel GJ, Djibuti M, van Gijn J. Prevalence and risk of rupture of intracranial aneurysms: a systematic review. Stroke. 1998; 29: 251256.
24. Anzalone N, Triulzi F, Scotti G. Acute subarachnoid haemorrhage: 3D time-of-flight MR angiography versus intra-arterial digital angiography. Neuroradiology. 1995; 37: 257261.[Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
M.-H. Li, Y.-S. Cheng, Y.-D. Li, C. Fang, S.-W. Chen, W. Wang, D.-J. Hu, and H.-W. Xu Large-Cohort Comparison Between Three-Dimensional Time-of-Flight Magnetic Resonance and Rotational Digital Subtraction Angiographies in Intracranial Aneurysm Detection Stroke, September 1, 2009; 40(9): 3127 - 3129. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. D. Burns, J. Huston III, K. F. Layton, D. G. Piepgras, and R. D. Brown Jr Intracranial Aneurysm Enlargement on Serial Magnetic Resonance Angiography: Frequency and Risk Factors Stroke, February 1, 2009; 40(2): 406 - 411. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. J. Kaufmann and D. F. Kallmes Diagnostic Cerebral Angiography: Archaic and Complication-Prone or Here to Stay for Another 80 Years? Am. J. Roentgenol., June 1, 2008; 190(6): 1435 - 1437. [Full Text] [PDF] |
||||
![]() |
A. Kemmling, I. Noelte, L. Gerigk, S. Singer, C. Groden, and J. Scharf A Diagnostic Pitfall for Intracranial Aneurysms in Time-of-Flight MR Angiography: Small Intracranial Lipomas Am. J. Roentgenol., January 1, 2008; 190(1): W62 - W67. [Abstract] [Full Text] [PDF] |
||||
![]() |
H.A. Deutschmann, M. Augustin, J. Simbrunner, B. Unger, H. Schoellnast, G.A. Fritz, and G.E. Klein Diagnostic Accuracy of 3D Time-of-Flight MR Angiography Compared with Digital Subtraction Angiography for Follow-Up of Coiled Intracranial Aneurysms: Influence of Aneurysm Size AJNR Am. J. Neuroradiol., April 1, 2007; 28(4): 628 - 634. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. L. Brisman, J. K. Song, and D. W. Newell Cerebral aneurysms. N. Engl. J. Med., August 31, 2006; 355(9): 928 - 939. [Full Text] [PDF] |
||||
![]() |
J.-Y. Gauvrit, X. Leclerc, S. Caron, C. A. Taschner, J.-P. Lejeune, and J.-P. Pruvo Intracranial Aneurysms Treated With Guglielmi Detachable Coils: Imaging Follow-Up With Contrast-Enhanced MR Angiography Stroke, April 1, 2006; 37(4): 1033 - 1037. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Satoh, M. Omi, C. Ohsako, K. Fujiwara, K. Tsuno, W. Sasahara, K. Onoda, K. Tokunaga, K. Sugiu, and I. Date Differential Diagnosis of the Infundibular Dilation and Aneurysm of Internal Carotid Artery: Assessment with Fusion Imaging of 3D MR Cisternography/Angiography. AJNR Am. J. Neuroradiol., February 1, 2006; 27(2): 306 - 312. [Abstract] [Full Text] [PDF] |
||||
![]() |
D W Anderson Bilateral absence of the internal carotid artery: MR angiography and ultrasound findings Br. J. Radiol., June 1, 2005; 78(930): 569 - 572. [Abstract] [Full Text] [PDF] |
||||
![]() |
H.-S. Kang, M. H. Han, B. J. Kwon, O-K. Kwon, S. H. Kim, and K.-H. Chang Evaluation of the Lenticulostriate Arteries with Rotational Angiography and 3D Reconstruction AJNR Am. J. Neuroradiol., February 1, 2005; 26(2): 306 - 312. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. C. Goehde, P. Hunold, F. M. Vogt, W. Ajaj, M. Goyen, C. U. Herborn, M. Forsting, J. F. Debatin, and S. G. Ruehm Full-Body Cardiovascular and Tumor MRI for Early Detection of Disease: Feasibility and Initial Experience in 298 Subjects Am. J. Roentgenol., February 1, 2005; 184(2): 598 - 611. [Abstract] [Full Text] [PDF] |
||||
![]() |
H.-S. Kang, M. H. Han, B. J. Kwon, S. I. Jung, C.-W. Oh, D. H. Han, and K.-H. Chang Postoperative 3D Angiography in Intracranial Aneurysms AJNR Am. J. Neuroradiol., October 1, 2004; 25(9): 1463 - 1469. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. W. Schrier, M. M. Belz, A. M. Johnson, W. D. Kaehny, R. L. Hughes, D. Rubinstein, and P. A. Gabow Repeat Imaging for Intracranial Aneurysms in Patients with Autosomal Dominant Polycystic Kidney Disease with Initially Negative Studies: A Prospective Ten-Year Follow-up J. Am. Soc. Nephrol., April 1, 2004; 15(4): 1023 - 1028. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Niijima An aneurysm is born J R Soc Med, November 1, 2003; 96(11): 550 - 551. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Stroke Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2002 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |