Stroke. 2002;33:1975-1980
doi: 10.1161/01.STR.0000022809.46400.4B
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow Request Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Russell, D.
Right arrow Articles by Brucher, R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Russell, D.
Right arrow Articles by Brucher, R.
Related Collections
Right arrow Valvular heart disease
Right arrow Behavioral/psychosocial - stroke
Right arrow Carotid Stenosis
Right arrow Embolic stroke
Right arrow Doppler ultrasound, Transcranial Doppler etc.

(Stroke. 2002;33:1975.)
© 2002 American Heart Association, Inc.


Original Contributions

Online Automatic Discrimination Between Solid and Gaseous Cerebral Microemboli With the First Multifrequency Transcranial Doppler

David Russell, MD Rainer Brucher, PhD

From the Department of Neurology (D.R.), the National Hospital, Oslo, Norway, and the Department of Medical Engineering (R.B.), University of Applied Sciences, Ulm, Germany.

Correspondence to David Russell, Department of Neurology, The National Hospital, 0027 Oslo, Norway. E-mail david.russell{at}klinmed.uio.no


*    Abstract
up arrowTop
*Abstract
down arrowIntroduction
down arrowMaterials and Methods
down arrowResults
down arrowDiscussion
down arrowReferences
 
Background and Purpose The aim of this study was to assess the first multifrequency transcranial Doppler system that was specially developed to automatically detect and discriminate between solid and gaseous cerebral microemboli.

Methods The multifrequency transcranial Doppler instrumentation insonates simultaneously with 2.5 and 2.0 MHz. Differentiation between solid and gaseous microemboli is based on the principle that solid microemboli reflect more ultrasound at the higher than at the lower frequency, whereas the opposite is the case for gaseous microemboli. In the in vitro studies, 159 plastic spheres (50 or 80 µm in diameter) and 105 gas bubbles (8 to 25 µm) were studied in a pulsatile closed-loop system containing irodinium or pig blood. In vivo studies were carried out for 1 hour in 15 patients with mechanical heart valves and in 45 patients with carotid stenosis. This gave a total of 60 hours of online automatic monitoring in patients.

Results In the in vitro studies, 152 of the 159 (95.6%) plastic spheres were classified as solid, and 7 (4.4%) were classified as uncertain solid. Of the 105 gas bubbles, 99 (94.3%) were classified as gaseous and 6 (5.7%) as uncertain gaseous. Thus, correct classification was made for 251 (95.1%) of the 264 embolic events studied. A comparison between the automatic multifrequency discrimination and the known embolic classification gave a {kappa} value of 0.897 (P<0.0001). The multifrequency Doppler classified 433 (84.2%) of the 514 emboli detected in the mechanical heart valve patients as gaseous, 74 (14.4%) as solid, and 7 (1.4%) as uncertain (3 uncertain solid, 4 uncertain gas). Thirty-two emboli were detected in 17 (38%) of the 45 carotid stenosis patients; 30 (93.7%) were classified as solid and 2 (6.3%) as uncertain solid.

Conclusions This study has shown that multifrequency transcranial Doppler can be used to automatically differentiate between solid and gaseous microemboli online. Most detected microemboli in this initial study of mechanical heart valves were classified as gaseous, whereas most were classified as solid in the patients with carotid stenosis.


Key Words: carotid stenosis • cerebral embolism • mechanical heart valve protheses • ultrasonography, Doppler


*    Introduction
up arrowTop
up arrowAbstract
*Introduction
down arrowMaterials and Methods
down arrowResults
down arrowDiscussion
down arrowReferences
 
Cerebral microemboli have been detected with transcranial Doppler (TCD) in various patient groups and clinical situations in which there is an increased risk of stroke. These include carotid stenosis,1,2 heart disease,35 invasive cardiovascular examinations,68 and cardiovascular surgery.911 Assessment of the clinical significance of this method has been limited because differentiating between solid and gaseous microemboli has been impossible. The reason is that the detection of microemboli is based on the principle that they reflect an increased amount of ultrasound compared with the surrounding whole blood (embolus-to-blood ratio, EBR). However, the amount of ultrasound being reflected depends not only on the size but also on the acoustic impedance of the embolus, which is characterized by its density and propagation of ultrasound. The greater the difference is between the acoustic impedance of the embolus and that of the surrounding whole blood, the greater the reflection is.12 Air has an acoustic impedance that is <1/4000 that of whole blood and therefore causes an extremely large reflection. Solid microemboli, on the other hand, have acoustic impedances that are more similar to whole blood and therefore give a much smaller reflection. It is therefore impossible to determine the composition of an embolus by measuring the increase in reflected ultrasound power alone because a small gaseous embolus may cause an increase in reflected ultrasound power similar in size to that caused by a much larger solid embolus.13

Differentiation between solid and gaseous microemboli is theoretically possible by insonating an embolus simultaneously with 2 different ultrasound frequencies; the reflected ultrasound power depends on the insonating ultrasound frequency, and this dependency differs for solid and gaseous elements.14,15 Solid microemboli reflect more ultrasound power at a higher compared with a lower frequency, whereas the opposite is the case for gaseous microemboli. The aim of this study was to assess the first multifrequency TCD specially developed to automatically detect and discriminate between solid and gaseous cerebral microemboli. Studies were first carried out in vitro and then in the clinical situation.


*    Materials and Methods
up arrowTop
up arrowAbstract
up arrowIntroduction
*Materials and Methods
down arrowResults
down arrowDiscussion
down arrowReferences
 
Doppler Instrumentation
The multifrequency Doppler instrumentation (EmboDop, DWL) insonates simultaneously with 2.0- and 2.5-MHz frequencies.16 The broadband dual-frequency probe specially developed for this system allows insonation with a special but similar spacial intensity distribution using the 2 different transmitting frequencies. The probe may be used at a depth between 45 and 80 mm, which is covered by the longitudinally stretched focus area of the ultrasound beam. This results in minimal difference in ultrasound beam distortion caused by the tissue layers and the temporal bone and prevents inaccuracies in the simultaneous EBR measurements at the 2 transmitting frequencies.17

The frequency responses of the 2.5- and 2.0-MHz Doppler audio channels have been made as similar as possible (difference <0.1 dB) by use of special low-pass filters. The system has a 128 Fast Fourier Transform with 60% overlapping and Blackman time-weighted function. It has also a second 2.0-MHz gate that can be used as a reference gate for artifact rejection.18 Embolus monitoring with this system may be carried out bilaterally with 3 Doppler channels on each side. Data can be continuously stored digitally on the computer hard disk for up to 4 hours with the possibility of transferral to an archive system.

The multifrequency TCD classifies microemboli depending on the difference between the EBR at 2.5-MHz insonation frequency and the EBR at 2.0-MHz insonation (dEBR). Upper and lower classification limits were determined before this study by in vitro testing with plastic microspheres and gas bubbles. The lower dEBR limit for discrimination was set at -0.83±0.2 dB to take into account small variations in the background Doppler signal and in measurements resulting from beam distortion. The lower limit for solid emboli was therefore -0.83 dB. However, if the dEBR was within 0.2 dB of this limit (ie, between -0.63 and -0.83 dB), it was classified as uncertain solid. Microemboli with dEBR values below the -0.83-dB limit were classified as gaseous; if they were within -0.2 dB of this limit (ie, between -0.83 and -1.03 dB), they were classified as uncertain gaseous. Classification limits were introduced in an attempt to limit the possibility of errors resulting from resonance effects of very small gas bubbles, which cause a power increase of <7 dB. The higher dEBR limit for the classification of solid microemboli was 2.05±0.2 dB. This limit has a slight slope of y=0.02x+2.05 dB, where y=dEBR and x=2.0-MHz EBR. dEBR values above this level were therefore classified as gaseous; if they were within 0.2 dB of this limit (ie, between 2.05 and 2.25 dB), they were classified as uncertain gaseous. Solid microemboli were classified when the dEBR was below this limit and as uncertain solid if they were within -0.2 dB of this limit (ie, between 2.05 and 1.85 dB). Solid decisions were also classified as uncertain when the EBR value for 2.0-MHz insonation frequency was between 5 and 7 dB. This was also done to take into account possible resonance effects from gas bubbles with diameters of approximately 2.5 to 2.7 µm that could give dEBR values similar to approximately 80- to 86-µm solid microemboli.

In Vitro Studies
In vitro studies were carried out with a pulsatile closed-loop system (Figure 1). The silicone tubing had a diameter of 3 mm and was immersed in a water tank in which the water was kept constant at 35°C. The background Doppler signal was due to 95% volume fraction distilled water and a 5% volume fraction of 5- to 20-µm particles of irodinium-302 Pearl Luster Pigment Powder (Merck Co) for approximately half of the studies and pig blood for the remainder. The tubing was insonated through Plexiglas damping layers at constant 45° angle. The sample volume had an axial length of 10 mm, and the depth was kept at 50 mm.



View larger version (132K):
[in this window]
[in a new window]
 
Figure 1. Closed-loop system used in the in vitro studies. L indicates Plexiglas damping layers; P, pump.

We studied 119 plastic spheres19 with a diameter of 80 µm and 40 with a diameter of 50 µm. In addition, 105 gas bubbles with a diameter from 8 to 25 µm were generated by electrolysis. The plastic spheres and gas bubbles were introduced 20 cm upstream of the point of insonation.

Clinical Studies
Automatic transcranial embolus monitoring was carried out for 1 hour on the right middle cerebral artery (MCA) in 15 patients with a mechanical heart valve (MHV). None of the MHV patients had previous symptoms suggesting cerebrovascular disease, and none had evidence of atherosclerosis on color duplex examination of the major precerebral and intracranial arteries. Embolus monitoring was also carried out on the ipsilateral MCA in 45 patients with an internal carotid artery stenosis that reduced the diameter by >=70% on color duplex examination. Of these patients, 39 were symptomatic and 6 were asymptomatic.

This gave a total of 60 hours of online automatic TCD cerebral microembolus monitoring in patients. Some of the data from these studies were also used in a study to assess automatic embolus detection online with EmboDop.16

Embolus Detection
The threshold level for embolus detection in this study was defined as a relative Doppler energy increase of >20 dB · ms for which the Doppler power increase (EBR) is at least 5 dB and lasts >4 ms above the average background energy. Short-term power measurements were carried out every 2 ms on the basis of digital root-mean-square calculations and used to calculate a moving average for 30 ms of the background Doppler signal, which was updated every 2 ms. A relative Doppler energy threshold of 20 dB · ms was chosen to exclude normal Doppler energy variations resulting from Doppler speckle (random clustering of red blood cells).

Emboli were differentiated from artifacts by use of 4 parameters in an optimized binary decision tree: quarter Doppler shift, maximum duration limit, reference gate, and bidirectional enhancement.16 The hard-disk data were also assessed by an experienced observer blinded to automatic assessment using standard observer criteria for embolus detection.20 The times for each embolus detection by the instrumentation were compared with those for the experienced observer, and only those emboli on which both the automatic assessment and the observer agreed were used for the discrimination assessment.

Data Assessment
The accuracy of multifrequency TCD in differentiating the known gaseous and solid microembolic signals in the in vitro studies was assessed with a standard {kappa} calculation.21 For assessments of the results, an automatic classification of uncertain solid was considered a gaseous decision and an uncertain gaseous classification as a solid decision.


*    Results
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMaterials and Methods
*Results
down arrowDiscussion
down arrowReferences
 
In Vitro Study
In the in vitro studies, 152 (95.6%) of the 159 plastic spheres were correctly classified as solid (Figures 2 and 3), and 7 (4.4%) were classified as uncertain solid. None were classified as certain gaseous. The dEBR was 0.74±0.63 dB for the 80-µm plastic spheres and 0.63±0.56 dB for the 50-µm plastic spheres.



View larger version (60K):
[in this window]
[in a new window]
 
Figure 2. Plastic microsphere (80 µm) in the pulsatile closed-loop system with pig blood background. Reflected ultrasound at 2.5-MHz insonation (blue) is 0.6 dB greater compared with 2.0-MHz insonation (red).



View larger version (125K):
[in this window]
[in a new window]
 
Figure 3. In vitro study showing differentiation between 58 gas bubbles (8 to 25 µm) below the -0.83±0.2-dB dEBR lower limit and 58 plastic spheres (80 µm) between the lower -0.83±0.2-dB and upper 2.05±0.2-dB dEBR differentiation limits. Arrow, microsphere from Figure 2.

Of the 105 gas bubbles, 99 (94.3%) were classified as gaseous (Figures 3 and 4), and 6 (5.7%) were classified as uncertain gaseous. None were classified as certain solid. The 105 gas microemboli had a dEBR of -3.51±0.94 dB. Four of the 6 uncertain classification decisions were made when the pig blood background signal was used and 2 with the irodinium background. Therefore, correct classification was made for 251 (95.1%) of the 264 embolic events (plastic spheres and gas bubbles) studied. A comparison between the automatic multifrequency discrimination and the known embolic classification gave {kappa}=0.897 (P<0.0001).



View larger version (65K):
[in this window]
[in a new window]
 
Figure 4. Gas bubble in the pulsatile closed-loop system with pig blood background. Reflected ultrasound is 4.8 dB lower at 2.5 MHz (blue) compared with 2.0 MHz (red).

Clinical Studies
A total of 514 emboli were detected in the 15 MHV patients and 32 in 17 (38%) of the 45 carotid stenosis patients. In the MHV patients, multifrequency TCD classified 433 (84.2%) as gaseous (Figure 5) and 74 (14.4%) as solid (Figure 6). A total of 7 (1.4%) were not given a definite classification (3 uncertain solid, 4 uncertain gaseous).



View larger version (54K):
[in this window]
[in a new window]
 
Figure 5. MCA microembolus in an MHV patient that was classified as gaseous. Reflection is greater (13.9 dB) at 2 MHz (red) compared with 8.3 dB at 2.5 MHz (blue).



View larger version (61K):
[in this window]
[in a new window]
 
Figure 6. MCA microembolus in an MHV patient that was classified as solid. Reflection is greater (11.5 dB) at 2.5 MHz (blue) compared with 10.6 dB at 2.0 MHz (red).

In the carotid stenosis patients, 30 (93.7%) of the 32 emboli were classified as solid (Figure 7) and 2 (6.3%) as uncertain solid. None were classified as gaseous.



View larger version (113K):
[in this window]
[in a new window]
 
Figure 7. MCA microembolus in a carotid stenosis patient that was classified as solid. Reflection is greater (12.9 dB) at 2.5 MHz (blue) compared with 11.5 dB at 2.0 MHz (red).


*    Discussion
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMaterials and Methods
up arrowResults
*Discussion
down arrowReferences
 
This study has shown for the first time that multifrequency TCD can differentiate between solid and gaseous microemboli. This is possible because the reflection of ultrasound power is dependent not only on the size of the embolus but also on its composition and the insonating frequency used.14,15 This is the case for microemboli with a diameter <300 µm for which the principle of Raleigh scattering applies22 and the insonation frequency is higher than the resonance frequencies of the emboli.

The upper size limit for detection is otherwise dependent on the absence of overloading, which most often is due to larger gas emboli. This problem can be overcome in the near future by installation of an automatic gain in this Doppler instrumentation. The lower diameter limit for embolus detection is determined by the background signal and therefore by the diameter of the insonated artery and the axial sample volume. An assumed axial sample volume of 10 mm, a mean MCA diameter of 3 mm, and a hematocrit of 45% would give an approximate theoretical lower detection limit for gaseous microemboli of 2 to 3 µm and a lower limit for solid microemboli of 80 µm.14,15 Previous clinical experience with TCD-detected asymptomatic cerebral microemboli suggests that their size normally is within these limits.

Individual detection and differentiation will be difficult when many cerebral microemboli pass through the sample volume at the same time. This may be the case when multiple gaseous emboli are introduced into the cerebral circulation, eg, during invasive diagnostic or therapeutic cardiovascular examinations and heart surgery. Studies of ultrasound contrast agents have also shown that resonance effects cause errors when monitoring very small gas bubbles that have a diameter of <3 µm (ie, a Doppler power increase of <7 dB). Classification limits (see Materials and Methods) for solid microemboli have therefore been introduced in an attempt to reduce this potential problem. However, a higher Doppler energy threshold of >28 dB · ms (Doppler power >=7 dB times duration >4 ms) is more reliable for differentiation in clinical situations when very small microbubbles may be present.

There have been few reports to date on the problem of embolus differentiation. Devuyst et al23 recently reported that wavelet representation combined with dual-gate TCD had only fair accuracy in differentiating between solid and gaseous emboli. Wavelet analysis cannot at the moment be used online because of the considerable computer capacity required for data analysis.

A definite decision was made in this study for 94% of the emboli detected in the carotid stenosis patients, and all were classified as solid. In the remaining 6%, the decision was uncertain solid. None of the emboli in this patient group were classified as gaseous. These findings support the view that clinically silent cerebral microemboli in patients with carotid stenosis are predominantly solid in composition. Preliminary clinical studies have suggested that their frequency may be of prognostic significance24 and may be reduced by antiplatelet treatment.25 The most likely cause of solid microemboli in these patients would seem to be platelet aggregates or whole-blood clots formed at the site of the stenosis.

In the MHV patients, a definite decision was made for 98.6% of the 514 microemboli. Most (84.2%) were classified as gaseous, whereas 14.4% were solid and 1.4% were uncertain. The cause and clinical significance of gaseous and solid microemboli in MHV patients are uncertain. Indirect evidence suggesting that most are due to gas bubbles has previously been obtained through the use of oxygen inhalation, which reduced the number of microemboli by >50%, whereas their number increased under hyperbaric conditions.26,27 Cavitation just behind the valve leaflets may be at least partly responsible for gas bubble formation. However, cavitation bubbles do not normally last >1 ms, which is not long enough to reach the cerebral circulation.28 A possibility is that they become stabilized in some way, eg, by being coated with membrane fragments from platelets or other cells.29

Solid microemboli in MHV patients may be due to activation of platelets or plasma coagulation on foreign surfaces, incomplete "washout" of blood from valves, and increased shear rates with secondary platelet activation.30,31 Air-fluid interfaces also activate platelets, which may be another mechanism generating platelet aggregation.31,32 No relationship has been found between the degree of anticoagulation assessed on a single international normalized ratio measurement and the total frequency of microemboli in MHV patients.4,5 In a retrospective study, an increased number of microemboli signals, platelet microparticles, and procoagulation activity were found in those patients who had experienced cerebrovascular symptoms, whereas there was no difference with regard to thrombin or fibrin generation.29 Although the addition of aspirin to anticoagulation therapy may reduce the risk of systemic embolism,33 it has not been shown to cause a significant reduction in the total frequency of microemboli.34

Prospective follow-up studies are required to determine whether the relative frequency of solid and gaseous microemboli is a marker for increased stroke risk that could be used clinically to decide an optimal anticoagulation and antiplatelet treatment or whether it may provide information on the relative merits of different valve designs. This would represent an important advance in patient management because the incidence of thromboembolic events in MHV patients is in the range of 1% to 5% per year despite oral anticoagulation. In most cases ({approx}85%), the cerebral circulation is involved.33,3539 In industrialized countries, the replacement rate for heart valves is between 8 and 20 per 100 000.40 Because the risk is continuous and cumulative, thousands of strokes occur every year in this patient population, and {approx}5% to 10% of them are fatal.34,39

We hope that the ability to automatically detect and differentiate between gaseous and solid cerebral microemboli will lead to a better understanding of the cause of cerebral injury in several other clinical situations. Cerebral microemboli are frequent during cerebral and coronary intra-arterial angiographies.68 They often occur during flushing and contrast injection when it is assumed that gas bubbles may also be injected into the cerebral circulation. However, they may also be due to loosening of atheromatous material from the arterial wall or thrombus formation on the tip of the catheter. Although the risk of neurological complications associated with intra-arterial angiography is uncommon, with a range of 0.55% to 3.2%,41,42 there is evidence to suggest that asymptomatic cerebral lesions may be relatively common. A necropsy study reported embolism after angiography that was thought to be catheter related in 30% of patients.43 Using diffusion MRI, Bendszus et al44 have recently shown signs of an embolic pattern of ischemic lesions in the distal vascular territory of small cortical, subcortical, or perforating vessels. They carried out 100 intra-arterial angiographies of the cerebral vessels and found 42 new MRI bright lesions in 23 patients.

Although mortality and morbidity are low during cardiopulmonary bypass operations, postoperative cognitive impairment has been reported in relatively large numbers of patients. The incidence of cognitive decline has been reported to be at its highest (50% to 80%) at discharge, 20% to 50% at 6 weeks, and 10% to 30% at 6 months.4548 A late decline to >40% predicted by the presence of early postoperative decline has also been reported after 5 years.49 Cerebral microemboli have been implicated in the pathogenesis of this cognitive decline, and some studies have found a positive association between the number of intraoperative cerebral microemboli detected by TCD and postoperative neuropsychological outcome.5051

Differentiation between solid and gaseous asymptomatic cerebral microemboli would at least theoretically provide some information regarding embolus size. Microemboli that cause relative increases in Doppler power of 14 dB in the MCA are not uncommon in several clinical situations such as MHV patients. Theoretical calculations using an assumed MCA diameter of 3 mm and an axial sample volume of 10 mm suggest that this Doppler power increase may be due to a gas bubble that is 4 µm in diameter or a solid microembolus of 130 µm. This allows some assessment of the possible harmful effects because a 4-µm gas bubble should normally pass through the microcirculation (7 to 10 µm), whereas it may seem more difficult for a 130-µm solid microembolus. It is therefore a possibility that "asymptomatic" solid microemboli of this size may cause a progressive cognitive decline if they continuously enter the brain in large numbers.

In conclusion, this study has shown that it is possible to differentiate between solid and gaseous cerebral microemboli with multifrequency TCD. This new method may lead to new insights into the pathophysiology and prevention of stroke and may be of help in preventing cerebral injury during invasive cardiovascular examinations and surgery.

Received December 6, 2001; revision received March 14, 2002; accepted April 16, 2001.


*    References
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMaterials and Methods
up arrowResults
up arrowDiscussion
*References
 

  1. Siebler M, Kleinschmidt A, Sitzer M, Steinmetz H, Freund HJ. Cerebral microembolism in symptomatic and asymptomatic high-grade internal carotid artery stenosis. Neurology. 1994; 44: 615–618.[Abstract/Free Full Text]
  2. Markus HS, Thomson ND, Brown MM. Asymptomatic cerebral embolic signals in symptomatic and asymptomatic carotid artery disease. Brain. 1995; 118: 1005–1011.[Abstract/Free Full Text]
  3. Cullinane M, Wainwright R, Brown A, Monaghan M, Markus HS. Asymptomatic embolization in subjects with atrial fibrillation not taking anticoagulants: a prospective study. Stroke. 1998; 29: 1810–1815.[Abstract/Free Full Text]
  4. Braekken SK, Russell D, Brucher R, Svennevig J. Incidence and frequency of cerebral embolic signals in patients with a similar bileaflet mechanical heart valve. Stroke. 1995; 26: 1225–1230.[Abstract/Free Full Text]
  5. Georgiadis D, Grosset DG, Kelman A, Faichney A, Lees KR. Prevalence and characteristics of intracranial microemboli signals in patients with different types of prosthetic cardiac valves. Stroke. 1994; 25: 587–592.[Abstract]
  6. Braekken SK, Endresen K, Russell D, Brucher R, Kjekshus J. Influence of guidewire and catheter type on the frequency of cerebral microembolic signals during left heart catheterization. Am J Cardiol. 1998; 82: 632–637.[CrossRef][Medline] [Order article via Infotrieve]
  7. Dagirmanjian A, Davis DA, Rothfus WE, Deeb ZL, Goldberg AL. Silent cerebral microemboli occurring during carotid angiography: frequency as determined with Doppler sonography. AJR Am J Roentgenol. 1993; 161: 1037–1040.[Abstract/Free Full Text]
  8. Markus H, Loh A, Israel D, Buckenham T, Clifton A, Brown MM. Microscopic air embolism during cerebral angiography and strategies for its avoidance. Lancet. 1993; 341: 784–787.[CrossRef][Medline] [Order article via Infotrieve]
  9. Spencer MP. Transcranial Doppler monitoring and causes of stroke from carotid endarterectomy. Stroke. 1997; 28: 685–691.[Abstract/Free Full Text]
  10. Ackerstaff RG, Jansen C, Moll FL, Vermeulen FE, Hamerlijnck RP, Mauser HW. The significance of microemboli detection by means of transcranial Doppler ultrasonography monitoring in carotid endarterectomy. J Vasc Surg. 1995; 21: 963–969.[CrossRef][Medline] [Order article via Infotrieve]
  11. Braekken SK, Russell D, Brucher R, Abdelnoor M, Svennevig JL. Cerebral microembolic signals during cardiopulmonary bypass surgery: frequency, time of occurrence, and association with patient and surgical characteristics. Stroke. 1997; 28: 1988–1992.[Abstract/Free Full Text]
  12. Russell D. The detection of cerebral emboli using Doppler ultrasound. In: Newell DW, Aaslid R, eds. Transcranial Doppler. New York, NY: Raven Press; 1992: 207–213.
  13. Russell D, Madden KP, Clark WM, Sandset PM, Zivin JA. Detection of arterial emboli using Doppler ultrasound in rabbits. Stroke. 1991; 22: 253–258.[Abstract/Free Full Text]
  14. Lubbers J, van der Berg JW. An ultrasonic detector for microgas emboli in a blood flow line. Ultrasound Med Biol. 1976; 2: 301–310.[CrossRef]
  15. Moehring MA, Klepper JR. Pulse Doppler ultrasound detection, characterization and size estimation of emboli in flowing blood. IEEE Trans Biomed Eng. 1994; 41: 35–44.[CrossRef][Medline] [Order article via Infotrieve]
  16. Brucher R, Russell D. Automatic online embolus detection and artifact rejection with the first multifrequency transcranial Doppler. Stroke. 2002; 33: 1969–1974.[Abstract/Free Full Text]
  17. Brucher R, Russell D. Assessment of temporal bone beam distortion when using multifrequency Doppler to differentiate cerebral microemboli. Cerebrovasc Dis. 2002; 13 (suppl 4): 46.Abstract.
  18. Brucher R, Russell D, Frenkenberger H. Automatic emboli detection using dual-gate Doppler. In: Klingelhöfer J, Bartels E, Ringelstein EB, eds. New Trends in Cerebral Hemodynamics and Neurosonology. Amsterdam, Netherlands: Elsevier Science BV; 1997: 364–373.
  19. Ugelstad J. Preparation and application of new monosized polymer particles. In: Progress in Polymer Science. 17th ed. New York, NY: Pergamon Press; 1992: 87–161.
  20. Ringelstein EB, Droste DW, Babikian VL, Evans DH, Grosset DG, Kaps M, Markus HS, Russell D, Siebler M. Consensus on microembolus detection by TCD: International Consensus Group on Microembolus Detection. Stroke. 1998; 29: 725–729.[Abstract/Free Full Text]
  21. Altmann DG. Practical Statistics for Medical Research. London, UK: Chapman and Hall; 1991.
  22. Brucher R, Russell D. The Doppler power characteristics of solid and gaseous emboli. Cerebrovasc Dis. 1994; 4 (suppl. 3): 18.Abstract.[Medline] [Order article via Infotrieve]
  23. Devuyst G, Darbellay GA, Vesin J-M, Kemény V, Ritter M, Droste DW, Molina C, Serena J, Sztajzel R, Ruchat P, Lucchesi C, Dietler G, Ringelstein EB, Despland P-A, Bogousslavsky J. Automatic classification of HITS into artifacts or solid or gaseous emboli by a wavelet representation combined with dual-gate TCD. Stroke. 2001; 32: 2803–2809.[Abstract/Free Full Text]
  24. Siebler M, Nachtmann A, Sitzer M, Rose G, Kleinschmidt A, Rademacher J, Steinmetz H. Cerebral microembolism and the risk of ischaemia in asymptomatic high-grade internal carotid artery ischaemia. Stroke. 1995; 26: 2184–2186.[Abstract/Free Full Text]
  25. Goertler M, Baeumer M, Kross R, Blaser T, Lutze G, Jost S, Wallesch C-W. Rapid decline of cerebral microemboli of arterial origin after intravenous acetylsalicylic acid. Stroke. 1999; 30: 66–69.[Abstract/Free Full Text]
  26. Kaps M, Hansen J, Weiher M, Tiffert K, Kayser I, Droste DW. Clinically silent microemboli in patients with artificial prosthetic aortic valves are predominantly gaseous and not solid. Stroke. 1997; 28: 322–325.[Abstract/Free Full Text]
  27. Droste DW, Hansberg T, Kemeny V, Hammel D, Schulte-Altedorneburg G, Nabavi DG, Kaps M, Scheld HH, Ringelstein EB. Oxygen inhalation can differentiate gaseous from nongaseous microemboli detected by transcranial Doppler ultrasound. Stroke. 1997; 28: 2453–2456.[Abstract/Free Full Text]
  28. Graf T, Fischer H, Reul H, Rau G. Cavitation potential of mechanical heart valve prosthesis. Int J Artif Organs. 1991; 14: 169–174.[Medline] [Order article via Infotrieve]
  29. Geiser T, Sturzenegger M, Genewein U, Haeberli A, Beer JH. Mechanisms of cerebrovascular events as assessed by procoagulant activity, cerebral microemboli, and platelet microparticles in patients with prosthetic heart valves. Stroke. 1998; 29: 1770–1777.[Abstract/Free Full Text]
  30. Chow TW, Hellums DJ, Moake JL, Kroll MH. Shear stress-induced von Willebrand factor binding to platelet glycoprotein Ib initiates calcium influx associated with aggregation. Blood. 1992; 80: 113–120.[Abstract/Free Full Text]
  31. Moake JL, Turner NA, Stathopoulos NA, Nolasco L, Hellums JD. Shear-induced platelet aggregation can be mediated by vWF release from platelets, as well as by exogenous large or unusually large vWF multimers, requires adenosine diphosphate, and is resistant to aspirin. Blood. 1988; 71: 1366–1374.[Abstract/Free Full Text]
  32. Malmgren R, Thorsen T, Nordvik A, Holmsen H. Microbubble-induced phospholipase C activation does not correlate with platelet aggregation. Thromb Haemost. 1993; 69: 394–396.[Medline] [Order article via Infotrieve]
  33. Turpie AG, Gent M, Laupacis A, Latour Y, Gunstensen J, Basile F, Klimek M, Hirsh J. A comparison of aspirin with placebo in patients treated with warfarin after heart-valve replacement. N Engl J Med. 1993; 329: 524–529.[Abstract/Free Full Text]
  34. Sturzenegger M, Beer J, Rihs F. Monitoring combined antithrombotic treatments in patients with prosthetic heart valves using transcranial Doppler and coagulation markers. Stroke. 1995; 26: 63–69.[Abstract/Free Full Text]
  35. Cannegieter SC, Rosendaal FR, Wintzen AR, van der Meer FJ, Vandenbroucke JP, Briet E. Optimal oral anticoagulant therapy in patients with mechanical heart valves. N Engl J Med. 1995; 333: 11–17.[Abstract/Free Full Text]
  36. Cortelazzo S, Finazzi G, Viero P, Galli M, Remuzzi A, Parenzan L, Barbui T. Thrombotic and hemorrhagic complications in patients with mechanical heart valve prosthesis attending an anticoagulation clinic. Thromb Haemost. 1993; 69: 316–320.[Medline] [Order article via Infotrieve]
  37. Criscitiello MG, Levine HJ. Thromboembolism and prosthetic heart valves. Hosp Pract (Off Ed). 1992; 27: 69–79.
  38. Hanle DD, Harrison EC, Yoganathan AP, Allen DT, Corcoran WH. In vitro flow dynamics of four prosthetic aortic valves: a comparative analysis. J Biomech. 1989; 22: 597–607.[CrossRef][Medline] [Order article via Infotrieve]
  39. Stein B, Fuster V, Halperin JL, Chesebro JH. Antithrombotic therapy in cardiac disease: an emerging approach based on pathogenesis and risk. Circulation. 1989; 80: 1501–1513.[Free Full Text]
  40. Sapirstein JS, Smith PK. The "ideal" replacement heart valve. Am Heart J. 2001; 141: 856–860.[CrossRef][Medline] [Order article via Infotrieve]
  41. Horowitz MB, Dutton K, Purdy PD. Assessment of complication types and rates related to diagnostic angiography and interventional procedures. Intervent Neuroradiol. 1998; 4: 27–37.
  42. Noto TJ, Johnson LW, Krone R, Weaver WF, Clark DA, Kramer JR, Vetrovec GW. Cardiac catheterization 1990: a report of the Registry of the Society for Cardiac Angiography and Interventions. Cathet Cardiovasc Diagn. 1991; 24: 75–83.[Medline] [Order article via Infotrieve]
  43. Ramirez G, O’Neill WM, Lambert R, Bloomer HA. Cholesterol embolization: a complication of angiography. Arch Intern Med. 1978; 138: 1430–1432.[Abstract]
  44. Bendszus M, Koltzenburg M, Burger R, Warmuth-Metz M, Hofmann E, Solymosi L. Silent embolism in diagnostic cerebral angiography and neurointerventional procedures: a prospective study. Lancet. 1999; 354: 1594–1597.[CrossRef][Medline] [Order article via Infotrieve]
  45. Selnes OA, Goldsborough MA, Borowicz LM, McKhann GM. Neurobehavioural sequelae of cardiopulmonary bypass. Lancet. 1999; 353: 1601–1606.[CrossRef][Medline] [Order article via Infotrieve]
  46. Blumenthal JA, Mahanna EP, Madden DJ, White WD, Croughwell ND, Newman ME. Methodological issues in the assessment of neuropsychologic function after cardiac surgery. Ann Thorac Surg. 1995; 59: 1345–1350.[Abstract/Free Full Text]
  47. Newman MF, Croughwell ND, Blumenthal JA, Lowry E, White WD, Spillane W, Davis RD Jr, Glower DD, Smith LR, Mahanna EP, et al. Predictors of cognitive decline after cardiac operation. Ann Thorac Surg. 1995; 59: 1326–1330.[Abstract/Free Full Text]
  48. Murkin JM, Baird IL, Martzke JS, Adams SJ, Lok P. Long-term neurological and neuropsychological outcome 3 years after coronary artery bypass surgery. Anesth Analg. 1996; 82 (suppl): S328.Abstract.
  49. Newman MF, Kirchner JL, Phillips-Bute B, Gaver V, Grocott H, Jones RH, Mark DB, Reves JG, Blumenthal JA. Longitudinal assessment of neurocognitive function after coronary-artery bypass surgery. N Engl J Med. 2001; 344: 395–402.[Abstract/Free Full Text]
  50. Pugsley W, Klinger L, Paschalis C, Treasure T, Harrison M, Newman S. The impact of microemboli during cardiopulmonary bypass on neuropsychological functioning. Stroke. 1994; 25: 1393–1399.[Abstract]
  51. Stump DA, Newman S. Embolus detection during cardiopulmonary bypass. In: Tegeler C, ed. Neurosonology. St Louis, Mo: Mosby; 1996; 252–255.



This article has been cited by other articles:


Home page
Eur. J. Cardiothorac. Surg.Home page
N. Stroobant, G. van Nooten, D. De Bacquer, Y. Van Belleghem, and G. Vingerhoets
Neuropsychological functioning 3-5 years after coronary artery bypass grafting: does the pump make a difference?
Eur. J. Cardiothorac. Surg., August 1, 2008; 34(2): 396 - 401.
[Abstract] [Full Text] [PDF]


Home page
Anesth. Analg.Home page
M. Barak, M. Kabha, D. Norman, M. Soudry, Y. Kats, and S. Milo
Cerebral Microemboli During Hip Fracture Fixation: A Prospective Study
Anesth. Analg., July 1, 2008; 107(1): 221 - 225.
[Abstract] [Full Text] [PDF]


Home page
StrokeHome page
M. Skjelland, A. Michelsen, F. Brosstad, J. L. Svennevig, R. Brucher, and D. Russell
Solid Cerebral Microemboli and Cerebrovascular Symptoms in Patients With Prosthetic Heart Valves
Stroke, April 1, 2008; 39(4): 1159 - 1164.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
L. Guerrieri Wolf, B. P. Choudhary, Y. Abu-Omar, and D. P. Taggart
Solid and gaseous cerebral microembolization after biologic and mechanical aortic valve replacement: Investigation with multirange and multifrequency transcranial Doppler ultrasound
J. Thorac. Cardiovasc. Surg., March 1, 2008; 135(3): 512 - 520.
[Abstract] [Full Text] [PDF]


Home page
StrokeHome page
R. Dittrich and E. B. Ringelstein
Occurrence and Clinical Impact of Microembolic Signals During or After Cardiosurgical Procedures
Stroke, February 1, 2008; 39(2): 503 - 511.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
L. G. Wolf, Y. Abu-Omar, B. P. Choudhary, D. Pigott, and D. P. Taggart
Gaseous and solid cerebral microembolization during proximal aortic anastomoses in off-pump coronary surgery: The effect of an aortic side-biting clamp and two clampless devices
J. Thorac. Cardiovasc. Surg., February 1, 2007; 133(2): 485 - 493.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
Y. Abu-Omar, S. Cader, L. G. Wolf, D. Pigott, P. M. Matthews, and D. P. Taggart
Short-term changes in cerebral activity in on-pump and off-pump cardiac surgery defined by functional magnetic resonance imaging and their relationship to microembolization.
J. Thorac. Cardiovasc. Surg., November 1, 2006; 132(5): 1119 - 1125.
[Abstract] [Full Text] [PDF]


Home page
ICVTSHome page
D. Maselli, R. Pizio, G. Borelli, and F. Musumeci
Endovascular balloon versus transthoracic aortic clamping for minimally invasive mitral valve surgery: impact on cerebral microemboli
Interactive CardioVascular and Thoracic Surgery, April 1, 2006; 5(2): 183 - 186.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
A. Liebold, A. Khosravi, B. Westphal, C. Skrabal, Y.H. Choi, C. Stamm, A. Kaminski, A. Alms, T. Birken, D. Zurakowski, et al.
Effect of closed minimized cardiopulmonary bypass on cerebral tissue oxygenation and microembolization
J. Thorac. Cardiovasc. Surg., February 1, 2006; 131(2): 268 - 276.
[Abstract] [Full Text] [PDF]


Home page
ICVTSHome page
D. Maselli, R. Pizio, and F. Musumeci
Multifrequency transcranial Doppler for intraoperative automatic detection and characterisation of cerebral microemboli during port-access mitral valve surgery
Interactive CardioVascular and Thoracic Surgery, February 1, 2006; 5(1): 32 - 35.
[Abstract] [Full Text] [PDF]


Home page
StrokeHome page
D. Russell, R. Brucher, H. S. Markus, and M. Punter
Embolus Detection and Differentiation Using Multifrequency Transcranial Doppler * Response:
Stroke, February 1, 2006; 37(2): 340 - 342.
[Full Text] [PDF]


Home page
Am. J. Neuroradiol.Home page
M. Rosenkranz, J. Fiehler, W. Niesen, C. Waiblinger, B. Eckert, O. Wittkugel, T. Kucinski, J. Rother, H. Zeumer, C. Weiller, et al.
The Amount of Solid Cerebral Microemboli during Carotid Stenting Does Not Relate to the Frequency of Silent Ischemic Lesions
AJNR Am. J. Neuroradiol., January 1, 2006; 27(1): 157 - 161.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
M. Skjelland, J. Bergsland, R. Lundblad, P. S. Lingaas, K. A. Rein, S. Halvorsen, J. L. Svennevig, E. Fosse, R. Brucher, and D. Russell
Cerebral microembolization during off-pump coronary artery bypass surgery with the Symmetry aortic connector device
J. Thorac. Cardiovasc. Surg., December 1, 2005; 130(6): 1581 - 1585.
[Abstract] [Full Text] [PDF]


Home page
StrokeHome page
H. S. Markus and M. Punter
Can Transcranial Doppler Discriminate Between Solid and Gaseous Microemboli?: Assessment of a Dual-Frequency Transducer System
Stroke, August 1, 2005; 36(8): 1731 - 1734.
[Abstract] [Full Text] [PDF]


Home page
Eur Heart JHome page
C. Lund, R. B. Nes, T. P. Ugelstad, P. Due-Tonnessen, R. Andersen, P. K. Hol, R. Brucher, and D. Russell
Cerebral emboli during left heart catheterization may cause acute brain injury
Eur. Heart J., July 1, 2005; 26(13): 1269 - 1275.
[Abstract] [Full Text] [PDF]


Home page
SEMIN CARDIOTHORAC VASC ANESTHHome page
J. M. Murkin
Applied Neuromonitoring and Improving CNS Outcomes
Seminars in Cardiothoracic and Vascular Anesthesia, June 1, 2005; 9(2): 139 - 142.
[Abstract] [PDF]


Home page
ChestHome page
N. Stroobant, G. Van Nooten, Y. Van Belleghem, and G. Vingerhoets
Relation Between Neurocognitive Impairment, Embolic Load, and Cerebrovascular Reactivity Following On- and Off-Pump Coronary Artery Bypass Grafting
Chest, June 1, 2005; 127(6): 1967 - 1976.
[Abstract] [Full Text] [PDF]


Home page
StrokeHome page
D. Russell and R. Brucher
Embolus Detection and Differentiation Using Multifrequency Transcranial Doppler
Stroke, April 1, 2005; 36(4): 706 - 706.