(Stroke. 2002;33:1975.)
© 2002 American Heart Association, Inc.
Original Contributions |
From the Department of Neurology (D.R.), the National Hospital, Oslo, Norway, and the Department of Medical Engineering (R.B.), University of Applied Sciences, Ulm, Germany.
Correspondence to David Russell, Department of Neurology, The National Hospital, 0027 Oslo, Norway. E-mail david.russell{at}klinmed.uio.no
| Abstract |
|---|
|
|
|---|
Methods The multifrequency transcranial Doppler instrumentation insonates simultaneously with 2.5 and 2.0 MHz. Differentiation between solid and gaseous microemboli is based on the principle that solid microemboli reflect more ultrasound at the higher than at the lower frequency, whereas the opposite is the case for gaseous microemboli. In the in vitro studies, 159 plastic spheres (50 or 80 µm in diameter) and 105 gas bubbles (8 to 25 µm) were studied in a pulsatile closed-loop system containing irodinium or pig blood. In vivo studies were carried out for 1 hour in 15 patients with mechanical heart valves and in 45 patients with carotid stenosis. This gave a total of 60 hours of online automatic monitoring in patients.
Results In the in vitro studies, 152 of the 159 (95.6%) plastic spheres were classified as solid, and 7 (4.4%) were classified as uncertain solid. Of the 105 gas bubbles, 99 (94.3%) were classified as gaseous and 6 (5.7%) as uncertain gaseous. Thus, correct classification was made for 251 (95.1%) of the 264 embolic events studied. A comparison between the automatic multifrequency discrimination and the known embolic classification gave a
value of 0.897 (P<0.0001). The multifrequency Doppler classified 433 (84.2%) of the 514 emboli detected in the mechanical heart valve patients as gaseous, 74 (14.4%) as solid, and 7 (1.4%) as uncertain (3 uncertain solid, 4 uncertain gas). Thirty-two emboli were detected in 17 (38%) of the 45 carotid stenosis patients; 30 (93.7%) were classified as solid and 2 (6.3%) as uncertain solid.
Conclusions This study has shown that multifrequency transcranial Doppler can be used to automatically differentiate between solid and gaseous microemboli online. Most detected microemboli in this initial study of mechanical heart valves were classified as gaseous, whereas most were classified as solid in the patients with carotid stenosis.
Key Words: carotid stenosis cerebral embolism mechanical heart valve protheses ultrasonography, Doppler
| Introduction |
|---|
|
|
|---|
Differentiation between solid and gaseous microemboli is theoretically possible by insonating an embolus simultaneously with 2 different ultrasound frequencies; the reflected ultrasound power depends on the insonating ultrasound frequency, and this dependency differs for solid and gaseous elements.14,15 Solid microemboli reflect more ultrasound power at a higher compared with a lower frequency, whereas the opposite is the case for gaseous microemboli. The aim of this study was to assess the first multifrequency TCD specially developed to automatically detect and discriminate between solid and gaseous cerebral microemboli. Studies were first carried out in vitro and then in the clinical situation.
| Materials and Methods |
|---|
|
|
|---|
The frequency responses of the 2.5- and 2.0-MHz Doppler audio channels have been made as similar as possible (difference <0.1 dB) by use of special low-pass filters. The system has a 128 Fast Fourier Transform with 60% overlapping and Blackman time-weighted function. It has also a second 2.0-MHz gate that can be used as a reference gate for artifact rejection.18 Embolus monitoring with this system may be carried out bilaterally with 3 Doppler channels on each side. Data can be continuously stored digitally on the computer hard disk for up to 4 hours with the possibility of transferral to an archive system.
The multifrequency TCD classifies microemboli depending on the difference between the EBR at 2.5-MHz insonation frequency and the EBR at 2.0-MHz insonation (dEBR). Upper and lower classification limits were determined before this study by in vitro testing with plastic microspheres and gas bubbles. The lower dEBR limit for discrimination was set at -0.83±0.2 dB to take into account small variations in the background Doppler signal and in measurements resulting from beam distortion. The lower limit for solid emboli was therefore -0.83 dB. However, if the dEBR was within 0.2 dB of this limit (ie, between -0.63 and -0.83 dB), it was classified as uncertain solid. Microemboli with dEBR values below the -0.83-dB limit were classified as gaseous; if they were within -0.2 dB of this limit (ie, between -0.83 and -1.03 dB), they were classified as uncertain gaseous. Classification limits were introduced in an attempt to limit the possibility of errors resulting from resonance effects of very small gas bubbles, which cause a power increase of <7 dB. The higher dEBR limit for the classification of solid microemboli was 2.05±0.2 dB. This limit has a slight slope of y=0.02x+2.05 dB, where y=dEBR and x=2.0-MHz EBR. dEBR values above this level were therefore classified as gaseous; if they were within 0.2 dB of this limit (ie, between 2.05 and 2.25 dB), they were classified as uncertain gaseous. Solid microemboli were classified when the dEBR was below this limit and as uncertain solid if they were within -0.2 dB of this limit (ie, between 2.05 and 1.85 dB). Solid decisions were also classified as uncertain when the EBR value for 2.0-MHz insonation frequency was between 5 and 7 dB. This was also done to take into account possible resonance effects from gas bubbles with diameters of approximately 2.5 to 2.7 µm that could give dEBR values similar to approximately 80- to 86-µm solid microemboli.
In Vitro Studies
In vitro studies were carried out with a pulsatile closed-loop system (Figure 1). The silicone tubing had a diameter of 3 mm and was immersed in a water tank in which the water was kept constant at 35°C. The background Doppler signal was due to 95% volume fraction distilled water and a 5% volume fraction of 5- to 20-µm particles of irodinium-302 Pearl Luster Pigment Powder (Merck Co) for approximately half of the studies and pig blood for the remainder. The tubing was insonated through Plexiglas damping layers at constant 45° angle. The sample volume had an axial length of 10 mm, and the depth was kept at 50 mm.
|
We studied 119 plastic spheres19 with a diameter of 80 µm and 40 with a diameter of 50 µm. In addition, 105 gas bubbles with a diameter from 8 to 25 µm were generated by electrolysis. The plastic spheres and gas bubbles were introduced 20 cm upstream of the point of insonation.
Clinical Studies
Automatic transcranial embolus monitoring was carried out for 1 hour on the right middle cerebral artery (MCA) in 15 patients with a mechanical heart valve (MHV). None of the MHV patients had previous symptoms suggesting cerebrovascular disease, and none had evidence of atherosclerosis on color duplex examination of the major precerebral and intracranial arteries. Embolus monitoring was also carried out on the ipsilateral MCA in 45 patients with an internal carotid artery stenosis that reduced the diameter by
70% on color duplex examination. Of these patients, 39 were symptomatic and 6 were asymptomatic.
This gave a total of 60 hours of online automatic TCD cerebral microembolus monitoring in patients. Some of the data from these studies were also used in a study to assess automatic embolus detection online with EmboDop.16
Embolus Detection
The threshold level for embolus detection in this study was defined as a relative Doppler energy increase of >20 dB · ms for which the Doppler power increase (EBR) is at least 5 dB and lasts >4 ms above the average background energy. Short-term power measurements were carried out every 2 ms on the basis of digital root-mean-square calculations and used to calculate a moving average for 30 ms of the background Doppler signal, which was updated every 2 ms. A relative Doppler energy threshold of 20 dB · ms was chosen to exclude normal Doppler energy variations resulting from Doppler speckle (random clustering of red blood cells).
Emboli were differentiated from artifacts by use of 4 parameters in an optimized binary decision tree: quarter Doppler shift, maximum duration limit, reference gate, and bidirectional enhancement.16 The hard-disk data were also assessed by an experienced observer blinded to automatic assessment using standard observer criteria for embolus detection.20 The times for each embolus detection by the instrumentation were compared with those for the experienced observer, and only those emboli on which both the automatic assessment and the observer agreed were used for the discrimination assessment.
Data Assessment
The accuracy of multifrequency TCD in differentiating the known gaseous and solid microembolic signals in the in vitro studies was assessed with a standard
calculation.21 For assessments of the results, an automatic classification of uncertain solid was considered a gaseous decision and an uncertain gaseous classification as a solid decision.
| Results |
|---|
|
|
|---|
|
|
Of the 105 gas bubbles, 99 (94.3%) were classified as gaseous (Figures 3 and 4), and 6 (5.7%) were classified as uncertain gaseous. None were classified as certain solid. The 105 gas microemboli had a dEBR of -3.51±0.94 dB. Four of the 6 uncertain classification decisions were made when the pig blood background signal was used and 2 with the irodinium background. Therefore, correct classification was made for 251 (95.1%) of the 264 embolic events (plastic spheres and gas bubbles) studied. A comparison between the automatic multifrequency discrimination and the known embolic classification gave
=0.897 (P<0.0001).
|
Clinical Studies
A total of 514 emboli were detected in the 15 MHV patients and 32 in 17 (38%) of the 45 carotid stenosis patients. In the MHV patients, multifrequency TCD classified 433 (84.2%) as gaseous (Figure 5) and 74 (14.4%) as solid (Figure 6). A total of 7 (1.4%) were not given a definite classification (3 uncertain solid, 4 uncertain gaseous).
|
|
In the carotid stenosis patients, 30 (93.7%) of the 32 emboli were classified as solid (Figure 7) and 2 (6.3%) as uncertain solid. None were classified as gaseous.
|
| Discussion |
|---|
|
|
|---|
The upper size limit for detection is otherwise dependent on the absence of overloading, which most often is due to larger gas emboli. This problem can be overcome in the near future by installation of an automatic gain in this Doppler instrumentation. The lower diameter limit for embolus detection is determined by the background signal and therefore by the diameter of the insonated artery and the axial sample volume. An assumed axial sample volume of 10 mm, a mean MCA diameter of 3 mm, and a hematocrit of 45% would give an approximate theoretical lower detection limit for gaseous microemboli of 2 to 3 µm and a lower limit for solid microemboli of 80 µm.14,15 Previous clinical experience with TCD-detected asymptomatic cerebral microemboli suggests that their size normally is within these limits.
Individual detection and differentiation will be difficult when many cerebral microemboli pass through the sample volume at the same time. This may be the case when multiple gaseous emboli are introduced into the cerebral circulation, eg, during invasive diagnostic or therapeutic cardiovascular examinations and heart surgery. Studies of ultrasound contrast agents have also shown that resonance effects cause errors when monitoring very small gas bubbles that have a diameter of <3 µm (ie, a Doppler power increase of <7 dB). Classification limits (see Materials and Methods) for solid microemboli have therefore been introduced in an attempt to reduce this potential problem. However, a higher Doppler energy threshold of >28 dB · ms (Doppler power
7 dB times duration >4 ms) is more reliable for differentiation in clinical situations when very small microbubbles may be present.
There have been few reports to date on the problem of embolus differentiation. Devuyst et al23 recently reported that wavelet representation combined with dual-gate TCD had only fair accuracy in differentiating between solid and gaseous emboli. Wavelet analysis cannot at the moment be used online because of the considerable computer capacity required for data analysis.
A definite decision was made in this study for 94% of the emboli detected in the carotid stenosis patients, and all were classified as solid. In the remaining 6%, the decision was uncertain solid. None of the emboli in this patient group were classified as gaseous. These findings support the view that clinically silent cerebral microemboli in patients with carotid stenosis are predominantly solid in composition. Preliminary clinical studies have suggested that their frequency may be of prognostic significance24 and may be reduced by antiplatelet treatment.25 The most likely cause of solid microemboli in these patients would seem to be platelet aggregates or whole-blood clots formed at the site of the stenosis.
In the MHV patients, a definite decision was made for 98.6% of the 514 microemboli. Most (84.2%) were classified as gaseous, whereas 14.4% were solid and 1.4% were uncertain. The cause and clinical significance of gaseous and solid microemboli in MHV patients are uncertain. Indirect evidence suggesting that most are due to gas bubbles has previously been obtained through the use of oxygen inhalation, which reduced the number of microemboli by >50%, whereas their number increased under hyperbaric conditions.26,27 Cavitation just behind the valve leaflets may be at least partly responsible for gas bubble formation. However, cavitation bubbles do not normally last >1 ms, which is not long enough to reach the cerebral circulation.28 A possibility is that they become stabilized in some way, eg, by being coated with membrane fragments from platelets or other cells.29
Solid microemboli in MHV patients may be due to activation of platelets or plasma coagulation on foreign surfaces, incomplete "washout" of blood from valves, and increased shear rates with secondary platelet activation.30,31 Air-fluid interfaces also activate platelets, which may be another mechanism generating platelet aggregation.31,32 No relationship has been found between the degree of anticoagulation assessed on a single international normalized ratio measurement and the total frequency of microemboli in MHV patients.4,5 In a retrospective study, an increased number of microemboli signals, platelet microparticles, and procoagulation activity were found in those patients who had experienced cerebrovascular symptoms, whereas there was no difference with regard to thrombin or fibrin generation.29 Although the addition of aspirin to anticoagulation therapy may reduce the risk of systemic embolism,33 it has not been shown to cause a significant reduction in the total frequency of microemboli.34
Prospective follow-up studies are required to determine whether the relative frequency of solid and gaseous microemboli is a marker for increased stroke risk that could be used clinically to decide an optimal anticoagulation and antiplatelet treatment or whether it may provide information on the relative merits of different valve designs. This would represent an important advance in patient management because the incidence of thromboembolic events in MHV patients is in the range of 1% to 5% per year despite oral anticoagulation. In most cases (
85%), the cerebral circulation is involved.33,3539 In industrialized countries, the replacement rate for heart valves is between 8 and 20 per 100 000.40 Because the risk is continuous and cumulative, thousands of strokes occur every year in this patient population, and
5% to 10% of them are fatal.34,39
We hope that the ability to automatically detect and differentiate between gaseous and solid cerebral microemboli will lead to a better understanding of the cause of cerebral injury in several other clinical situations. Cerebral microemboli are frequent during cerebral and coronary intra-arterial angiographies.68 They often occur during flushing and contrast injection when it is assumed that gas bubbles may also be injected into the cerebral circulation. However, they may also be due to loosening of atheromatous material from the arterial wall or thrombus formation on the tip of the catheter. Although the risk of neurological complications associated with intra-arterial angiography is uncommon, with a range of 0.55% to 3.2%,41,42 there is evidence to suggest that asymptomatic cerebral lesions may be relatively common. A necropsy study reported embolism after angiography that was thought to be catheter related in 30% of patients.43 Using diffusion MRI, Bendszus et al44 have recently shown signs of an embolic pattern of ischemic lesions in the distal vascular territory of small cortical, subcortical, or perforating vessels. They carried out 100 intra-arterial angiographies of the cerebral vessels and found 42 new MRI bright lesions in 23 patients.
Although mortality and morbidity are low during cardiopulmonary bypass operations, postoperative cognitive impairment has been reported in relatively large numbers of patients. The incidence of cognitive decline has been reported to be at its highest (50% to 80%) at discharge, 20% to 50% at 6 weeks, and 10% to 30% at 6 months.4548 A late decline to >40% predicted by the presence of early postoperative decline has also been reported after 5 years.49 Cerebral microemboli have been implicated in the pathogenesis of this cognitive decline, and some studies have found a positive association between the number of intraoperative cerebral microemboli detected by TCD and postoperative neuropsychological outcome.5051
Differentiation between solid and gaseous asymptomatic cerebral microemboli would at least theoretically provide some information regarding embolus size. Microemboli that cause relative increases in Doppler power of 14 dB in the MCA are not uncommon in several clinical situations such as MHV patients. Theoretical calculations using an assumed MCA diameter of 3 mm and an axial sample volume of 10 mm suggest that this Doppler power increase may be due to a gas bubble that is 4 µm in diameter or a solid microembolus of 130 µm. This allows some assessment of the possible harmful effects because a 4-µm gas bubble should normally pass through the microcirculation (7 to 10 µm), whereas it may seem more difficult for a 130-µm solid microembolus. It is therefore a possibility that "asymptomatic" solid microemboli of this size may cause a progressive cognitive decline if they continuously enter the brain in large numbers.
In conclusion, this study has shown that it is possible to differentiate between solid and gaseous cerebral microemboli with multifrequency TCD. This new method may lead to new insights into the pathophysiology and prevention of stroke and may be of help in preventing cerebral injury during invasive cardiovascular examinations and surgery.
Received December 6, 2001; revision received March 14, 2002; accepted April 16, 2001.
| References |
|---|
|
|
|---|
2. Markus HS, Thomson ND, Brown MM. Asymptomatic cerebral embolic signals in symptomatic and asymptomatic carotid artery disease. Brain. 1995; 118: 10051011.
3. Cullinane M, Wainwright R, Brown A, Monaghan M, Markus HS. Asymptomatic embolization in subjects with atrial fibrillation not taking anticoagulants: a prospective study. Stroke. 1998; 29: 18101815.
4. Braekken SK, Russell D, Brucher R, Svennevig J. Incidence and frequency of cerebral embolic signals in patients with a similar bileaflet mechanical heart valve. Stroke. 1995; 26: 12251230.
5. Georgiadis D, Grosset DG, Kelman A, Faichney A, Lees KR. Prevalence and characteristics of intracranial microemboli signals in patients with different types of prosthetic cardiac valves. Stroke. 1994; 25: 587592.[Abstract]
6. Braekken SK, Endresen K, Russell D, Brucher R, Kjekshus J. Influence of guidewire and catheter type on the frequency of cerebral microembolic signals during left heart catheterization. Am J Cardiol. 1998; 82: 632637.[CrossRef][Medline] [Order article via Infotrieve]
7. Dagirmanjian A, Davis DA, Rothfus WE, Deeb ZL, Goldberg AL. Silent cerebral microemboli occurring during carotid angiography: frequency as determined with Doppler sonography. AJR Am J Roentgenol. 1993; 161: 10371040.
8. Markus H, Loh A, Israel D, Buckenham T, Clifton A, Brown MM. Microscopic air embolism during cerebral angiography and strategies for its avoidance. Lancet. 1993; 341: 784787.[CrossRef][Medline] [Order article via Infotrieve]
9. Spencer MP. Transcranial Doppler monitoring and causes of stroke from carotid endarterectomy. Stroke. 1997; 28: 685691.
10. Ackerstaff RG, Jansen C, Moll FL, Vermeulen FE, Hamerlijnck RP, Mauser HW. The significance of microemboli detection by means of transcranial Doppler ultrasonography monitoring in carotid endarterectomy. J Vasc Surg. 1995; 21: 963969.[CrossRef][Medline] [Order article via Infotrieve]
11. Braekken SK, Russell D, Brucher R, Abdelnoor M, Svennevig JL. Cerebral microembolic signals during cardiopulmonary bypass surgery: frequency, time of occurrence, and association with patient and surgical characteristics. Stroke. 1997; 28: 19881992.
12. Russell D. The detection of cerebral emboli using Doppler ultrasound. In: Newell DW, Aaslid R, eds. Transcranial Doppler. New York, NY: Raven Press; 1992: 207213.
13. Russell D, Madden KP, Clark WM, Sandset PM, Zivin JA. Detection of arterial emboli using Doppler ultrasound in rabbits. Stroke. 1991; 22: 253258.
14. Lubbers J, van der Berg JW. An ultrasonic detector for microgas emboli in a blood flow line. Ultrasound Med Biol. 1976; 2: 301310.[CrossRef]
15. Moehring MA, Klepper JR. Pulse Doppler ultrasound detection, characterization and size estimation of emboli in flowing blood. IEEE Trans Biomed Eng. 1994; 41: 3544.[CrossRef][Medline] [Order article via Infotrieve]
16. Brucher R, Russell D. Automatic online embolus detection and artifact rejection with the first multifrequency transcranial Doppler. Stroke. 2002; 33: 19691974.
17. Brucher R, Russell D. Assessment of temporal bone beam distortion when using multifrequency Doppler to differentiate cerebral microemboli. Cerebrovasc Dis. 2002; 13 (suppl 4): 46.Abstract.
18. Brucher R, Russell D, Frenkenberger H. Automatic emboli detection using dual-gate Doppler. In: Klingelhöfer J, Bartels E, Ringelstein EB, eds. New Trends in Cerebral Hemodynamics and Neurosonology. Amsterdam, Netherlands: Elsevier Science BV; 1997: 364373.
19. Ugelstad J. Preparation and application of new monosized polymer particles. In: Progress in Polymer Science. 17th ed. New York, NY: Pergamon Press; 1992: 87161.
20. Ringelstein EB, Droste DW, Babikian VL, Evans DH, Grosset DG, Kaps M, Markus HS, Russell D, Siebler M. Consensus on microembolus detection by TCD: International Consensus Group on Microembolus Detection. Stroke. 1998; 29: 725729.
21. Altmann DG. Practical Statistics for Medical Research. London, UK: Chapman and Hall; 1991.
22. Brucher R, Russell D. The Doppler power characteristics of solid and gaseous emboli. Cerebrovasc Dis. 1994; 4 (suppl. 3): 18.Abstract.[Medline] [Order article via Infotrieve]
23. Devuyst G, Darbellay GA, Vesin J-M, Kemény V, Ritter M, Droste DW, Molina C, Serena J, Sztajzel R, Ruchat P, Lucchesi C, Dietler G, Ringelstein EB, Despland P-A, Bogousslavsky J. Automatic classification of HITS into artifacts or solid or gaseous emboli by a wavelet representation combined with dual-gate TCD. Stroke. 2001; 32: 28032809.
24. Siebler M, Nachtmann A, Sitzer M, Rose G, Kleinschmidt A, Rademacher J, Steinmetz H. Cerebral microembolism and the risk of ischaemia in asymptomatic high-grade internal carotid artery ischaemia. Stroke. 1995; 26: 21842186.
25. Goertler M, Baeumer M, Kross R, Blaser T, Lutze G, Jost S, Wallesch C-W. Rapid decline of cerebral microemboli of arterial origin after intravenous acetylsalicylic acid. Stroke. 1999; 30: 6669.
26. Kaps M, Hansen J, Weiher M, Tiffert K, Kayser I, Droste DW. Clinically silent microemboli in patients with artificial prosthetic aortic valves are predominantly gaseous and not solid. Stroke. 1997; 28: 322325.
27. Droste DW, Hansberg T, Kemeny V, Hammel D, Schulte-Altedorneburg G, Nabavi DG, Kaps M, Scheld HH, Ringelstein EB. Oxygen inhalation can differentiate gaseous from nongaseous microemboli detected by transcranial Doppler ultrasound. Stroke. 1997; 28: 24532456.
28. Graf T, Fischer H, Reul H, Rau G. Cavitation potential of mechanical heart valve prosthesis. Int J Artif Organs. 1991; 14: 169174.[Medline] [Order article via Infotrieve]
29. Geiser T, Sturzenegger M, Genewein U, Haeberli A, Beer JH. Mechanisms of cerebrovascular events as assessed by procoagulant activity, cerebral microemboli, and platelet microparticles in patients with prosthetic heart valves. Stroke. 1998; 29: 17701777.
30. Chow TW, Hellums DJ, Moake JL, Kroll MH. Shear stress-induced von Willebrand factor binding to platelet glycoprotein Ib initiates calcium influx associated with aggregation. Blood. 1992; 80: 113120.
31. Moake JL, Turner NA, Stathopoulos NA, Nolasco L, Hellums JD. Shear-induced platelet aggregation can be mediated by vWF release from platelets, as well as by exogenous large or unusually large vWF multimers, requires adenosine diphosphate, and is resistant to aspirin. Blood. 1988; 71: 13661374.
32. Malmgren R, Thorsen T, Nordvik A, Holmsen H. Microbubble-induced phospholipase C activation does not correlate with platelet aggregation. Thromb Haemost. 1993; 69: 394396.[Medline] [Order article via Infotrieve]
33. Turpie AG, Gent M, Laupacis A, Latour Y, Gunstensen J, Basile F, Klimek M, Hirsh J. A comparison of aspirin with placebo in patients treated with warfarin after heart-valve replacement. N Engl J Med. 1993; 329: 524529.
34. Sturzenegger M, Beer J, Rihs F. Monitoring combined antithrombotic treatments in patients with prosthetic heart valves using transcranial Doppler and coagulation markers. Stroke. 1995; 26: 6369.
35. Cannegieter SC, Rosendaal FR, Wintzen AR, van der Meer FJ, Vandenbroucke JP, Briet E. Optimal oral anticoagulant therapy in patients with mechanical heart valves. N Engl J Med. 1995; 333: 1117.
36. Cortelazzo S, Finazzi G, Viero P, Galli M, Remuzzi A, Parenzan L, Barbui T. Thrombotic and hemorrhagic complications in patients with mechanical heart valve prosthesis attending an anticoagulation clinic. Thromb Haemost. 1993; 69: 316320.[Medline] [Order article via Infotrieve]
37. Criscitiello MG, Levine HJ. Thromboembolism and prosthetic heart valves. Hosp Pract (Off Ed). 1992; 27: 6979.
38. Hanle DD, Harrison EC, Yoganathan AP, Allen DT, Corcoran WH. In vitro flow dynamics of four prosthetic aortic valves: a comparative analysis. J Biomech. 1989; 22: 597607.[CrossRef][Medline] [Order article via Infotrieve]
39. Stein B, Fuster V, Halperin JL, Chesebro JH. Antithrombotic therapy in cardiac disease: an emerging approach based on pathogenesis and risk. Circulation. 1989; 80: 15011513.
40. Sapirstein JS, Smith PK. The "ideal" replacement heart valve. Am Heart J. 2001; 141: 856860.[CrossRef][Medline] [Order article via Infotrieve]
41. Horowitz MB, Dutton K, Purdy PD. Assessment of complication types and rates related to diagnostic angiography and interventional procedures. Intervent Neuroradiol. 1998; 4: 2737.
42. Noto TJ, Johnson LW, Krone R, Weaver WF, Clark DA, Kramer JR, Vetrovec GW. Cardiac catheterization 1990: a report of the Registry of the Society for Cardiac Angiography and Interventions. Cathet Cardiovasc Diagn. 1991; 24: 7583.[Medline] [Order article via Infotrieve]
43. Ramirez G, ONeill WM, Lambert R, Bloomer HA. Cholesterol embolization: a complication of angiography. Arch Intern Med. 1978; 138: 14301432.
44. Bendszus M, Koltzenburg M, Burger R, Warmuth-Metz M, Hofmann E, Solymosi L. Silent embolism in diagnostic cerebral angiography and neurointerventional procedures: a prospective study. Lancet. 1999; 354: 15941597.[CrossRef][Medline] [Order article via Infotrieve]
45. Selnes OA, Goldsborough MA, Borowicz LM, McKhann GM. Neurobehavioural sequelae of cardiopulmonary bypass. Lancet. 1999; 353: 16011606.[CrossRef][Medline] [Order article via Infotrieve]
46. Blumenthal JA, Mahanna EP, Madden DJ, White WD, Croughwell ND, Newman ME. Methodological issues in the assessment of neuropsychologic function after cardiac surgery. Ann Thorac Surg. 1995; 59: 13451350.
47. Newman MF, Croughwell ND, Blumenthal JA, Lowry E, White WD, Spillane W, Davis RD Jr, Glower DD, Smith LR, Mahanna EP, et al. Predictors of cognitive decline after cardiac operation. Ann Thorac Surg. 1995; 59: 13261330.
48. Murkin JM, Baird IL, Martzke JS, Adams SJ, Lok P. Long-term neurological and neuropsychological outcome 3 years after coronary artery bypass surgery. Anesth Analg. 1996; 82 (suppl): S328.Abstract.
49. Newman MF, Kirchner JL, Phillips-Bute B, Gaver V, Grocott H, Jones RH, Mark DB, Reves JG, Blumenthal JA. Longitudinal assessment of neurocognitive function after coronary-artery bypass surgery. N Engl J Med. 2001; 344: 395402.
50. Pugsley W, Klinger L, Paschalis C, Treasure T, Harrison M, Newman S. The impact of microemboli during cardiopulmonary bypass on neuropsychological functioning. Stroke. 1994; 25: 13931399.[Abstract]
51. Stump DA, Newman S. Embolus detection during cardiopulmonary bypass. In: Tegeler C, ed. Neurosonology. St Louis, Mo: Mosby; 1996; 252255.
This article has been cited by other articles:
![]() |
M. Skjelland, K. Krohg-Sorensen, B. Tennoe, S. J. Bakke, R. Brucher, and D. Russell Cerebral Microemboli and Brain Injury During Carotid Artery Endarterectomy and Stenting Stroke, January 1, 2009; 40(1): 230 - 234. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Stroobant, G. van Nooten, D. De Bacquer, Y. Van Belleghem, and G. Vingerhoets Neuropsychological functioning 3-5 years after coronary artery bypass grafting: does the pump make a difference? Eur. J. Cardiothorac. Surg., August 1, 2008; 34(2): 396 - 401. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Barak, M. Kabha, D. Norman, M. Soudry, Y. Kats, and S. Milo Cerebral Microemboli During Hip Fracture Fixation: A Prospective Study Anesth. Analg., July 1, 2008; 107(1): 221 - 225. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Skjelland, A. Michelsen, F. Brosstad, J. L. Svennevig, R. Brucher, and D. Russell Solid Cerebral Microemboli and Cerebrovascular Symptoms in Patients With Prosthetic Heart Valves Stroke, April 1, 2008; 39(4): 1159 - 1164. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Guerrieri Wolf, B. P. Choudhary, Y. Abu-Omar, and D. P. Taggart Solid and gaseous cerebral microembolization after biologic and mechanical aortic valve replacement: Investigation with multirange and multifrequency transcranial Doppler ultrasound J. Thorac. Cardiovasc. Surg., March 1, 2008; 135(3): 512 - 520. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Dittrich and E. B. Ringelstein Occurrence and Clinical Impact of Microembolic Signals During or After Cardiosurgical Procedures Stroke, February 1, 2008; 39(2): 503 - 511. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. G. Wolf, Y. Abu-Omar, B. P. Choudhary, D. Pigott, and D. P. Taggart Gaseous and solid cerebral microembolization during proximal aortic anastomoses in off-pump coronary surgery: The effect of an aortic side-biting clamp and two clampless devices J. Thorac. Cardiovasc. Surg., February 1, 2007; 133(2): 485 - 493. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Abu-Omar, S. Cader, L. G. Wolf, D. Pigott, P. M. Matthews, and D. P. Taggart Short-term changes in cerebral activity in on-pump and off-pump cardiac surgery defined by functional magnetic resonance imaging and their relationship to microembolization. J. Thorac. Cardiovasc. Surg., November 1, 2006; 132(5): 1119 - 1125. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Maselli, R. Pizio, G. Borelli, and F. Musumeci Endovascular balloon versus transthoracic aortic clamping for minimally invasive mitral valve surgery: impact on cerebral microemboli Interactive CardioVascular and Thoracic Surgery, April 1, 2006; 5(2): 183 - 186. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Liebold, A. Khosravi, B. Westphal, C. Skrabal, Y.H. Choi, C. Stamm, A. Kaminski, A. Alms, T. Birken, D. Zurakowski, et al. Effect of closed minimized cardiopulmonary bypass on cerebral tissue oxygenation and microembolization J. Thorac. Cardiovasc. Surg., February 1, 2006; 131(2): 268 - 276. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Maselli, R. Pizio, and F. Musumeci Multifrequency transcranial Doppler for intraoperative automatic detection and characterisation of cerebral microemboli during port-access mitral valve surgery Interactive CardioVascular and Thoracic Surgery, February 1, 2006; 5(1): 32 - 35. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Russell, R. Brucher, H. S. Markus, and M. Punter Embolus Detection and Differentiation Using Multifrequency Transcranial Doppler * Response: Stroke, February 1, 2006; 37(2): 340 - 342. [Full Text] [PDF] |
||||
![]() |
M. Rosenkranz, J. Fiehler, W. Niesen, C. Waiblinger, B. Eckert, O. Wittkugel, T. Kucinski, J. Rother, H. Zeumer, C. Weiller, et al. The Amount of Solid Cerebral Microemboli during Carotid Stenting Does Not Relate to the Frequency of Silent Ischemic Lesions AJNR Am. J. Neuroradiol., January 1, 2006; 27(1): 157 - 161. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Skjelland, J. Bergsland, R. Lundblad, P. S. Lingaas, K. A. Rein, S. Halvorsen, J. L. Svennevig, E. Fosse, R. Brucher, and D. Russell Cerebral microembolization during off-pump coronary artery bypass surgery with the Symmetry aortic connector device J. Thorac. Cardiovasc. Surg., December 1, 2005; 130(6): 1581 - 1585. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. S. Markus and M. Punter Can Transcranial Doppler Discriminate Between Solid and Gaseous Microemboli?: Assessment of a Dual-Frequency Transducer System Stroke, August 1, 2005; 36(8): 1731 - 1734. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Lund, R. B. Nes, T. P. Ugelstad, P. Due-Tonnessen, R. Andersen, P. K. Hol, R. Brucher, and D. Russell Cerebral emboli during left heart catheterization may cause acute brain injury Eur. Heart J., July 1, 2005; 26(13): 1269 - 1275. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Murkin Applied Neuromonitoring and Improving CNS Outcomes Seminars in Cardiothoracic and Vascular Anesthesia, June 1, 2005; 9(2): 139 - 142. [Abstract] [PDF] |
||||
![]() |
N. Stroobant, G. Van Nooten, Y. Van Belleghem, and G. Vingerhoets Relation Between Neurocognitive Impairment, Embolic Load, and Cerebrovascular Reactivity Following On- and Off-Pump Coronary Artery Bypass Grafting Chest, June 1, 2005; 127(6): 1967 - 1976. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Russell and R. Brucher Embolus Detection and Differentiation Using Multifrequency Transcranial Doppler Stroke, April 1, 2005; 36(4): 706 - 706. [Full Text] [PDF] |
||||
![]() |
N. Saidi and J. M. Murkin Applied Neuromonitoring in Cardiac Surgery: Patient Specific Management Seminars in Cardiothoracic and Vascular Anesthesia, March 1, 2005; 9(1): 17 - 23. [Abstract] [PDF] |
||||
![]() |
F. Kantarci, I. Adaletli, G. Yalcin, D. Selcuk, S. Kurugoglu, and I. Mihmanli Lower Extremity Arterial Air Embolism Secondary to Clostridial Gas Gangrene: Diagnosis by Doppler Sonography J. Ultrasound Med., November 1, 2004; 23(11): 1531 - 1534. [Full Text] [PDF] |
||||
![]() |
O. Wolf, P. Heider, M. Heinz, H. Poppert, D. Sander, O. Greil, W. Weiss, M. Hanke, and H.-H. Eckstein Microembolic Signals Detected by Transcranial Doppler Sonography During Carotid Endarterectomy and Correlation With Serial Diffusion-Weighted Imaging Stroke, November 1, 2004; 35(11): e373 - e375. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Lund, P. K. Hol, R. Lundblad, E. Fosse, K. Sundet, B. Tennoe, R. Brucher, and D. Russell Reply Ann. Thorac. Surg., October 1, 2004; 78(4): 1514 - 1515. [Full Text] [PDF] |
||||
![]() |
Y. Abu-Omar, A. Cifelli, P. M. Matthews, and D. P. Taggart The role of microembolisation in cerebral injury as defined by functional magnetic resonance imaging Eur. J. Cardiothorac. Surg., September 1, 2004; 26(3): 586 - 591. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. D. Doblar Intraoperative Transcranial Ultrasonic Monitoring for Cardiac and Vascular Surgery Seminars in Cardiothoracic and Vascular Anesthesia, June 1, 2004; 8(2): 127 - 145. [Abstract] [PDF] |
||||
![]() |
J. M. Murkin Perioperative Multimodality Neuromonitoring: An Overview Seminars in Cardiothoracic and Vascular Anesthesia, June 1, 2004; 8(2): 167 - 171. [Abstract] [PDF] |
||||
![]() |
Y. Abu-Omar, L. Balacumaraswami, D. W. Pigott, P. M. Matthews, and D. P. Taggart Solid and gaseous cerebral microembolization during off-pump, on-pump, and open cardiac surgery procedures J. Thorac. Cardiovasc. Surg., June 1, 2004; 127(6): 1759 - 1765. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Lund, P. K. Hol, R. Lundblad, E. Fosse, K. Sundet, B. Tennoe, R. Brucher, and D. Russell Comparison of cerebral embolization during off-pump and on-pump coronary artery bypass surgery Ann. Thorac. Surg., September 1, 2003; 76(3): 765 - 770. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Stroke Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2002 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |