(Stroke. 2002;33:2211.)
© 2002 American Heart Association, Inc.
Original Contributions |
From the Departments of Neurology (D.W.D., S.L., R.D., M.R., E.B.R.), Cardiology and Angiology (T.W., J.S.), and Clinical Radiology (M.M., M.F.), University of Münster, Münster, Germany.
Correspondence to Priv-Doz Dr Dirk W. Droste, Klinik und Poliklinik für Neurologie, Universitätsklinikum Münster, Albert-Schweitzer-Str 33, D-48129, Münster, Germany. E-mail droste2{at}uni-muenster.de
| Abstract |
|---|
|
|
|---|
Methods Eighty-one patients were investigated by both transesophageal echocardiography and bilateral TCD of the middle cerebral arteries. The following protocol with injections of 10 mL agitated saline was applied in a randomized way: (1) no VM, (2) VM for 5 seconds starting 5 seconds after the beginning of contrast injection, and (3) repetition of the test with VM if the first test with VM was negative. The VM was performed for 5 seconds starting exactly 5 seconds after the beginning of saline injection. Thereafter, the same protocol was repeated with 10 mL Echovist-300 instead of saline.
Results Thirty-one patients had a cardiac RLS. The Echovist-300 investigation disclosed all these 31 shunts, but saline disclosed only 29 of them. Twenty-two had an RLS only in at least 1 of the above TCD tests, some of them even with a considerable shunt volume.
Conclusions Contrast TCD performed with Echovist-300 but not with saline yields a 100% sensitivity to identify transesophageal echocardiography-proven cardiac RLSs. The TCD test should be repeated if negative the first time. This article gives detailed information for the optimization of the contrast TCD technique. Extracardiac shunts detected only during contrast TCD can have a considerable shunt volume and may allow for paradoxical embolism.
Key Words: cerebral embolism cerebrovascular disorders foramen ovale, patent ultrasonography
| Introduction |
|---|
|
|
|---|
50%.5,6 This fact corroborates the importance of paradoxical embolism in stroke origin. Extracardiac shunts, mainly pulmonary ones, can also deliver a substantial shunt volume comparable to the shunt volume of cardiac shunts. These shunts occur in an additional
11% of the population and may be the pathway of paradoxical embolism.712 Precise disclosure of an RLS is a keystone in the diagnosis of paradoxical embolism. Contrast-enhanced transesophageal echocardiography (TEE) is thought to be the method of choice to detect intracardiac RLSs and is generally accepted as the gold standard.1315 The performance of a VM enhances the RLS, unveils a latent RLS, or even leads to flow inversion of an otherwise left-to-right shunt.1618 Contrast-enhanced transcranial Doppler sonography (TCD) is a complementary method to detect cardiac and extracardiac RLSs. We used contrast agents with microbubbles (MBs), which do not survive pulmonary passage. Injected into a cubital vein, these bubbles appear only in the cerebral circulation, where they can be detected by TCD, if an RLS exists. Contrast TCD is as sensitive as TEE in the detection of cardiac RLS; moreover, it allows the detection of extracardiac shunts, which also can be the pathway of paradoxical embolism.812 In a consensus meeting and in several recent publications, recommendations have been set up to optimize the TCD technique.7,1923 A major point of discussion is whether agitated saline containing small air bubbles is as sensitive as a commercially available contrast agent (Echovist-300, Schering AG). In this study, we apply the elaborated modalities to a large patient cohort to test their usefulness and to compare both contrast agents.
| Patients and Methods |
|---|
|
|
|---|
In all 81 patients, TEE was performed to detect or rule out an intracardiac shunt. Twenty-three additional patients had been screened but were excluded from the study for the following reasons: TEE either was not available or had already been performed in an external hospital with an unknown protocol in 17 patients; the TCD recording had to be stopped because of noncompliance in 1 patient; there was no bilateral temporal window suitable for TCD in 4 patients7; and 1 patient was excluded because of an implanted mechanical cardiac valve prosthesis.
Echocardiography
All patients underwent TEE, which was performed by cardiologists well trained in this technique. The investigators used a Hewlett Packard Sonos 2500 or 5500 imaging system and a 4- to 7-MHz wide-band multiplane transducer. After informed consent had been obtained, patients were examined in the fasting state and received local pharyngeal anesthesia with 10% topical lidocaine spray. Additional intravenous sedation (midazolam) was given only if the probe was not well tolerated. For the diagnosis of an interatrial shunt, 10 mL contrast agent (Echovist-300) was injected as 2 boluses of 5 mL into a large antecubital vein during multiplane TEE. The first injection was made in a horizontal level (90°) showing vena cava superior, interatrial septum with the fossa ovalis, right atrium, and left atrium. The second bolus was given at a nearly transverse level (30°) giving a good view of the interatrial septum in the area of the fossa ovalis and of apical parts of the interatrial septum. Thus, the major manifestations of interatrial cardiac shunts such as a patent foramen ovale, defects of the interatrial septum of the primum and secundum types, or sinus venosus defects could easily be detected.
The VM was trained with the patients before the procedure. The duration of the pressing lasted 5 seconds, followed by deep expiration and inspiration and again expiration. The effectiveness of the VM was verified by a reduction in ventricular and atrial size and by bulging of the interatrial septum into the left atrium. The presence of an interatrial shunt was assumed if an MB transit from the right to the left atrium occurred during or within 5 seconds of the VM (first Echovist-300 application with VM) or spontaneously within the first 10 seconds of the first appearance of MBs in the right atrium (second Echovist-300 application without VM), respectively. In case of any doubt, the whole procedure was repeated. Intrapulmonary shunts were not systematically investigated in this study because this includes time-consuming Doppler investigation of the pulmonary arteries and veins.
Vascular Ultrasound Investigations
All subjects underwent a full color duplex investigation of their neck arteries (Sonos 2500, Hewlett Packard) and a continuous-wave Doppler investigation of the periorbital arteries. Subjects were also examined by TCD, including the intracranial segments of the internal carotid arteries, middle cerebral arteries (MCAs), and anterior and posterior cerebral arteries. Three patients had an extracranial internal carotid artery occlusion, and 1 patient had a high-grade internal carotid artery stenosis. No further high-grade stenoses or occlusions were detected in the anterior circulation.
For TCD embolus detection, the MCA was insonated bilaterally through the temporal bone windows. Two 2-MHz transducers were mounted on the temporal planes and secured with a head ribbon. A small sample volume of 8 mm in length and a low gain provided a setting optimal for embolus discrimination from the background spectrum. The bigate technique was used, placing 2 sample volumes a distance of 10 mm into each MCA main stem. The patients were lying comfortably on a stretcher. The investigations were well tolerated by the subjects without major side effects.
The same transcranial pulsed Doppler ultrasound device (TC4040, EME/Nicolet, software version 2.30) was used for all studies. The machine used a 128-point fast Fourier transform analysis and a graded color-scale to display the intensity of the Doppler signals received. In addition to online recording onto the hard disk, the Doppler audio signal was recorded by an 8-channel digital audio tape deck recorder (TA-88, TEAC Corporation) with normal speed. An experienced observers analysis of MB comprised listening to each of the software-recorded signals, watching each signal on the screen, and evaluating the tapes. The following definition for MB was used: typically visible and audible (click, chirp, whistle) short-duration, high-intensity signal within the Doppler flow spectrum with a time delay in the 2 channels of each side.24 Single MBs within clusters were discriminated by reducing the amplification during offline analysis. The test was considered positive when at least 1 MB was present within 40 seconds of the injection.20
The following protocol of contrast agent injection (10 mL each) was applied in a randomized way: (1) saline without VM, (2) saline with VM for 5 seconds starting 5 seconds after the beginning of contrast injection, (3) repetition of test 2 if it was negative. Afterward, independently from the previous results, the procedures were repeated following the same scheme but with Echovist-300 instead of saline. Each of these tests required at least 2 minutes, with bolus injection of the contrast agent starting at 0 second, Valsalva strain for 5 seconds, and resting phase until 120 seconds. Microcavitation saline contrast was generated by agitating a mixture of 10 mL normal saline and 1 mL air between two 12-mL syringes connected by a 3-way stopcock. Once the contrast was prepared, 10 mL was immediately injected as a bolus into a right cubital vein, which had previously been cannulated with a 21-gauge in-dwelling intravenous catheter. Echovist-300 was prepared following the instructions of the manufacturer. About 10 mL Echovist-300 was injected accordingly. The VM started with deep inspiration, followed by pressing against the closed glottis and expiration 5 seconds later. Before the procedure, the patients were meticulously trained to perform the VM correctly.
Occasionally, MB could still be detected 80 to 120 seconds after the injection. In these cases, the resting time preceding the next test was prolonged until an MB-free period at least 40 seconds long was documented. In each test, only the first 40 seconds after injection was used for MB analysis. Data from the right and left MCA were pooled.
Statistical Analysis
Patients were divided into 3 groups: (1) patients with a RLS in at least 1 TCD investigation but without demonstration of a shunt on TEE; (2) patients with a shunt on TEE only after VM, but not spontaneously, and with an RLS in at least 1 TCD investigation; and (3) patients with an RLS on TEE without VM, ie, with a spontaneous shunt, and with an RLS proven in at least 1 TCD investigation. With TEE used as the gold standard to detect intracardiac RLS, the sensitivity of TCD was calculated as follows: percentage of true positives (RLS confirmed by both methods) compared with true positives plus false negatives (TCD negatives and TEE positives). Specificity is not a meaningful parameter in this context because TCD detects shunts not noted by TEE. Therefore, specificity was not assessed. We used the following MB numbers for further calculation: total number of MBs with saline without VM, total number of MBs with saline with VM (in case of a negative first test with VM, only the result of the second performance was used ), and the corresponding numbers from the Echovist-300 tests. The absolute numbers of MB were assessed during the procedure without and with VM (sum from the right and left MCAs). A provisional, arbitrary categorization (a through d) was also used: a, no MBs, b, 1 to 20 MBs, c, >20 MB but no shower, and d, MB shower (temporarily, single MBs cannot be differentiated from each other).22,25 The Figure shows examples of this categorization.
|
For statistical analysis, the numbers of MBs seen during the 4 procedures were compared by use of 1-way ANOVA with Duncans multiple range test (Echovist-300 without VM, saline without VM, Echovist-300 with VM, and saline with VM). Statistical significance was determined at P=0.05.
| Results |
|---|
|
|
|---|
Table 1 gives the detailed numbers of MBs detected during all tests and their spread, as well as their allocation to the 4 categories. Taking all the patients together, the mean number of MBs was 82.6 with Echovist-300 with VM, 30.1 with Echovist-300 without VM, 15.0 with saline with VM, and 9.8 with saline without VM. In the 1-way ANOVA with post hoc multiple comparisons (Duncans multiple range test), there were more MBs detected during the TCD test with Echovist-300 with VM compared with each of the other 3 tests. In general, most MBs were detected by TCD in patients with a spontaneously open cardiac RLS, less MBs in those patients with a cardiac shunt patent only during VM, and only a few MBs in patients without a cardiac shunt. However, there was a considerable overlap of MB numbers, and in single cases, there were large amounts of bubbles also in the group of patients without an RLS during TEE (cf the maximum of 235 in a particular patient with Echovist-300 during VM in Table 1).
|
| Discussion |
|---|
|
|
|---|
Our findings are in concordance with 3 previous studies and with the majority of recommendations of a consensus meeting, which emphasized the need to perform the tests with a provocation maneuver twice, when the first test was negative.7,19,20,22 The performance of a VM for 5 s starting 5 s after the beginning of the injection using Echovist®-300 is advisable.
TCD cannot replace TEE because there are other possible cardiac embolic sources detectable only on TEE, such as left atrial thrombus or severe atherothrombosis of the aortic arch. Furthermore, TCD can give additional information on RLSs not noted during TEE. Possible causes of these discrepancies are poor VM, loss of the right plane of insonation during and because of the VM, or the presence of pulmonary shunts. Similar to intracardiac shunts, pulmonary shunts can produce a considerable transit of contrast bubbles. There is evidence that these RLSs can also allow for paradoxical embolism and stroke.812 Those patients with a shunt detected only during TCD warrant further interest in future trials concerning stroke origin. This may include an extended TEE investigation with the assessment of pulsatility in the pulmonary arteries and thoracic CT scans to possibly detect pulmonary shunts. Therefore, both techniques must not be seen as exchangeable but as complementary.
The fact that 1 MB was found during TCD in a patient with a cardiac RLS patent on TEE only during VM is in concordance with a previous report in which the threshold for a positive TCD examination was set at 1 MB.20 Shunt quantification remains a problem. The differentiation into shunts patent only under VM and those spontaneously patent is not dichotomous because under physiological breathing different cardiac right-to-left pressure gradients do occur. The findings of TEE and TCD are not congruent (Table 1): a patient with a shunt on TEE only under VM may have a spontaneously patent shunt in the TCD investigation and vice versa. The following TCD categorization should be used: (1) RLS patent only under VM and (2) RLS spontaneously open, although there is an overlap when the tests are repeated and these categories are not distinct. It is also worth performing the test with VM when the test without VM has already been positive to better evaluate the increase in shunting induced by VM. For the echocardiographer, quantification of a patent foramen ovale is not easy. The 2 septa of the foramen ovale separate from each other inconsistently, they overlap with a larger or smaller surface, and the pressure gradients differ independently from the size of the defect. Eventually, both the echocardiographers using TEE and the neurologists using TCD quantify the shunt by the number of MBs that trespassed. Stratifying the size of a RLS size makes sense as long as it indicates different paradoxical embolic risk and as long as it results in specific medical or interventional treatment. So far, only a few reports in the literature address the size of the RLS in relation to the risk of first or recurring embolic stroke.2529 The evidence for stratified therapeutic consequences is even more scarce.30 The above 4 categories of MB quantity cover quite evenly the quantity range of MBs found in a typical cohort of interest. Moreover, they give a rough estimate of the severity of the shunt. For TCD, we currently recommend it to indicate the absolute number of MBs in addition to the arbitrary categories as suggested in the literature.25 Furthermore, test modalities used should be indicated precisely because the number of MBs differs considerably with the methods used.7,19,20 This approach allows post hoc analysis once clinically useful categories are established. Table 2 summarizes recommendations based on this article and the current literature2,7,1923,3143 on how to optimize RLS detection by TCD.
|
In conclusion, TEE and properly performed contrast TCD are complementary diagnostic procedures that are helpful in the assessment of stroke and stroke-prone patients.
| Acknowledgments |
|---|
Received January 7, 2002; revision received April 29, 2002; accepted April 30, 2002.
| References |
|---|
|
|
|---|
2. Ward R, Jones D, Haponik EF. Paradoxical embolism: an underrecognized problem. Chest. 1995; 108: 549558.
3. Kessel-Schaefer A, Lefkovits M, Zellweger MJ, Brett W, Ruter F, Pfisterer ME, Buser P. Migrating thrombus trapped in a patent foramen ovale. Circulation. 2001; 103: 19281928.
4. Ofori CS, Moore LC, Hepler G. Massive cerebral infarction caused by paradoxical embolism: detection by transesophageal echocardiography. J Am Soc Echocardiogr. 1995; 8: 563566.[CrossRef][Medline] [Order article via Infotrieve]
5. Cabanes L, Mas JL, Cohen A, Amarenco P, Cabanes PA, Oubary P, Chedru F, Guerin F, Bousser MG, de Recondo J. Atrial septal aneurysm and patent foramen ovale as risk factors for cryptogenic stroke in patients less than 55 years of age: a study using transesophageal echocardiography. Stroke. 1993; 24: 18651873.
6. Di Tullio M, Sacco RL, Gopal A, Mohr JP, Homma S. Patent foramen ovale as a risk factor for cryptogenic stroke. Ann Intern Med. 1992; 117: 461465.
7. Droste DW, Reisener M, Kemény V, Dittrich R, Schulte-Altedorneburg G, Stypmann J, Wichter T, Ringelstein EB. Contrast transcranial Doppler ultrasound in the detection of right-to-left shunts: reproducibility, comparison of 2 agents, and distribution of microemboli. Stroke. 1999; 30: 10141018.
8. Kimura K, Minematsu K, Wada K, Yasaka M, Tagaya M, Kuribayashi S, Yamaguchi T. Transcranial Doppler of a paradoxical brain embolism associated with a pulmonary arteriovenous fistula. Am J Neuroradiol. 1999; 20: 18811884.
9. Marquez J, Sladen A, Gendell H, Boehnke M, Mendelow H. Paradoxical cerebral air embolism without an intracardiac septal defect. J Neurosurg. 1981; 55: 9971000.[Medline] [Order article via Infotrieve]
10. Natarajan K, Khabiri H, Jung S. Case 3: pulmonary arteriovenous malformation (PAVM): paradoxical embolism through the arteriovenous fistula can cause brain abscess and infarct. Am J Roentgenol. 2000; 175: 854854.
11. Wingen MJ, Günther RW. Pulmonale arteriovenöse Fisteln. Dt Ärztebl. 2001; 98: 13261330.
12. Yeung M, Khan KA, Antecol DH, Walker DR, Shuaib A. Transcranial Doppler ultrasonography and transesophageal echocardiography in the investigation of pulmonary arteriovenous malformation in a patient with hereditary hemorrhagic telangiectasia presenting with stroke. Stroke. 1995; 26: 19411944.
13. Belkin RN, Pollack BD, Ruggiero ML, Alas LL, Tatini U. Comparison of transesophageal and transthoracic echocardiography with contrast and color flow Doppler in the detection of patent foramen ovale. Am Heart J. 1994; 128: 520525.[CrossRef][Medline] [Order article via Infotrieve]
14. Hausmann D, Mügge A, Becht I, Daniel WG. Diagnosis of patent foramen ovale by transesophageal echocardiography and association with cerebral and peripheral embolic events. Am J Cardiol. 1992; 70: 668672.[CrossRef][Medline] [Order article via Infotrieve]
15. Pearson AC, Labovitz AJ, Tatineni S, Gomez CR. Superiority of transesophageal echocardiography in detecting cardiac source of embolism in patients with cerebral ischemia of uncertain etiology. J Am Coll Cardiol. 1991; 17: 6672.[Abstract]
16. Chen WJ, Kuan P, Lien WP, Lin FY. Detection of patent foramen ovale by contrast transesophageal echocardiography. Chest. 1992; 101: 15151520.
17. Lynch JJ, Schuchard GH, Gross CM, Wann LS. Prevalence of right-to-left atrial shunting in a healthy population: detection by Valsalva maneuver contrast echocardiography. Am J Cardiol. 1984; 53: 14781480.[CrossRef][Medline] [Order article via Infotrieve]
18. Nishimura RA, Tajik AJ. The Valsalva maneuver and response revisited. Mayo Clin Proc. 1986; 61: 211217.[Medline] [Order article via Infotrieve]
19. Droste DW, Kriete JU, Stypmann J, Castrucci M, Wichter T, Tietje R, Weltermann B, Young P, Ringelstein EB. Contrast transcranial Doppler ultrasound in the detection of right-to-left shunts: comparison of different procedures and different contrast agents. Stroke. 1999; 30: 18271832.
20. Droste DW, Silling K, Stypmann J, Grude M, Kemény V, Wichter T, Kühne K, Ringelstein EB. Contrast transcranial Doppler ultrasound in the detection of right-to-left shunts: time window and threshold in microbubble numbers. Stroke. 2000; 31: 16401645.
21. Droste DW, Jekentaite R, Stypmann J, Grude M, Hansberg T, Ritter MA, Nabavi DG, Nam E-M, Dittrich R, Wichter T, Ringelstein EB. Contrast transcranial Doppler ultrasound in the detection of right-to-left shunts: comparison of Echovist-200 and Echovist-300, timing of the Valsalva maneuver, and general recommendations for the performance of the test. Cerebrovasc Dis. 2002; 13: 235-241.[Medline] [Order article via Infotrieve]
22. Jauss M, Zanette EM. Detection of right-to-left shunt with ultrasound contrast agent and transcranial Doppler sonography. Cerebrovasc Dis. 2000; 10: 490496.[CrossRef][Medline] [Order article via Infotrieve]
23. Schwarze JJ, Sander D, Kukla C, Wittich I, Babikian VL, Klingelhofer J. Methodological parameters influence the detection of right-to-left shunts by contrast transcranial Doppler ultrasonography. Stroke. 1999; 30: 12341239.
24. Ringelstein EB, Droste DW, Babikian VL, Evans DH, Grosset DG, Kaps M, Markus HS, Russell D, Siebler M. Consensus on microembolus detection by transcranial Doppler ultrasound. Stroke. 1998; 29: 725729.
25. Serena J, Segura T, Perez Ayuso MJ, Bassaganyas J, Molins A, Davalos A. The need to quantify right-to-left shunt in acute ischemic stroke: a case-control study. Stroke. 1998; 29: 13221328.
26. Hausmann D, Mügge A, Daniel WG. Identification of patent foramen ovale permitting paradoxic embolism. J Am Coll Cardiol. 1995; 26: 10301038.[Abstract]
27. Homma S, Di Tullio MR, Sacco RL, Mihalatos D, Li Mandri G, Mohr JP. Characteristics of patent foramen ovale associated with cryptogenic stroke: a biplane transesophageal echocardiographic study. Stroke. 1994; 25: 582586.[Abstract]
28. Mas JL, Zuber M. Recurrent cerebrovascular events in patients with patent foramen ovale, atrial septal aneurysm, or both and cryptogenic stroke or transient ischemic attack: French Study Group on Patent Foramen Ovale and Atrial Septal Aneurysm. Am Heart J. 1995; 130: 10831088.[CrossRef][Medline] [Order article via Infotrieve]
29. Steiner MM, Di Tullio MR, Rundek T, Gan R, Chen X, Liguori C, Brainin M, Homma S, Sacco RL. Patent foramen ovale size and embolic brain imaging findings among patients with ischemic stroke. Stroke. 1998; 29: 944948.
30. Nendaz MR, Sarasin FP, Junod AF, Bogousslavsky J. Preventing stroke recurrence in patients with patent foramen ovale: antithrombotic therapy, foramen closure, or therapeutic abstention? A decision analytic perspective. Am Heart J. 1998; 135: 532541.[CrossRef][Medline] [Order article via Infotrieve]
31. Droste DW, Freund M, Wichter T, Stypmann J, Mesters RM, Ritter M, Henningsen H, Nabavi DG, Dittrich R, Heindel W, Ringelstein EB. Paradoxe Hirnembolien als Ursache des Schlaganfalls. Dt Ärztebl. 2002; 99: 281291.
32. Khan, K-A., Yeung M, and Shuaib A. Comparative study of 18 gauge and 20 gauge intravenous catheters during transcranial Doppler ultrasonography with saline solution contrast. J Ultrasound Med. 1997; 16: 341344.[Abstract]
33. Zanette EM, Mancini G, De Castro S, Solaro M, Cartoni D, Chiarotti F. Patent foramen ovale and transcranial Doppler: comparison of different procedures. Stroke. 1996; 27: 22512255.
34. Anzola GP, Renaldini E, Magoni M, Costa A, Cobelli M, Guindani M. Validation of transcranial Doppler sonography in the assessment of patent foramen ovale. Cerebrovasc Dis. 1995; 5: 194198.[CrossRef]
35. Horner S, Ni XS, Weihs W, Harb S, Augustin M, Duft M, Niederkorn K. Simultaneous bilateral contrast transcranial Doppler monitoring in patients with intracardiac and intrapulmonary shunts. J Neurol Sci. 1997; 150: 4957.[CrossRef][Medline] [Order article via Infotrieve]
36. Jauss M, Kaps M, Keberle M, Haberbosch W, Dorndorf W. A comparison of transesophageal echocardiography and transcranial Doppler sonography with contrast medium for detection of patent foramen ovale. Stroke. 1994; 25: 12651267.[Abstract]
37. Job FP, Ringelstein EB, Grafen Y, Flachskampf FA, Doherty C, Stockmanns A, Hanrath P. Comparison of transcranial contrast Doppler sonography and transesophageal contrast echocardiography for the detection of patent foramen ovale in young stroke patients. Am J Cardiol. 1994; 74: 381384.[CrossRef][Medline] [Order article via Infotrieve]
38. Klötzsch C, Janssen G, Berlit P. Transesophageal echocardiography and contrast TCD in the detection of a patent foramen ovale: experiences with 111 patients. Neurology. 1994; 44: 16031606.
39. Nemec JJ, Marwick TH, Lorig RJ, Davison MB, Chimowitz MI, Litowitz H, Salcedo EE. Comparison of transcranial Doppler ultrasound and transesophageal contrast echocardiography in the detection of interatrial right-to-left shunts. Am J Cardiol. 1991; 68: 14981502.[CrossRef][Medline] [Order article via Infotrieve]
40. Schminke U, Ries S, Daffertshofer M, Staedt U, Hennerici M. Patent foramen ovale: a potential source of cerebral embolism? Cerebrovasc Dis. 1995; 5: 133138.[CrossRef]
41. Stendel R, Schröder K, Funk T, Brock M. Is transcranial Doppler reliable in the detection of a patent foramen ovale? In: Klingelhöfer J, Bartels EM, Ringelstein EB, eds. New Trends in Cerebral Hemodynamics. Amsterdam, the Netherlands: Elsevier Science; 1998: 483489.
42. Venketasubramanian N, Sacco RL, Di Tullio M, Sherman D, Homma S, Mohr JP. Vascular distribution of paradoxical emboli by transcranial Doppler. Neurology. 1993; 43: 15331535.
43. Horner S, Schuchlenz S, Harb S, Weihs W, Ni XS, Fazekas F, Duft M, Kleinert G, Niederkorn K. Contrast transcranial Doppler monitoring in stroke patients with pulmonary right-to-left shunts. Cerebrovasc Dis. 1998; 8 (suppl 3): 7.Abstract.[Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
T. Holscher, J.A. Sattin, R. Raman, W. Wilkening, C.V. Fanale, S.E. Olson, R.F. Mattrey, and P.D. Lyden Real-Time Cerebral Angiography: Sensitivity of a New Contrast-Specific Ultrasound Technique AJNR Am. J. Neuroradiol., April 1, 2007; 28(4): 635 - 639. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Koch, A. Forteza, C. Lavernia, J. G. Romano, I. Campo-Bustillo, N. Campo, and S. Gold Cerebral Fat Microembolism and Cognitive Decline After Hip and Knee Replacement Stroke, March 1, 2007; 38(3): 1079 - 1081. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Souteyrand, P. Motreff, J.-R. Lusson, R. Rodriguez, E. Geoffroy, C. Dauphin, J.-Y. Boire, D. Lamaison, and J. Cassagnes Comparison of transthoracic echocardiography using second harmonic imaging, transcranial Doppler and transesophageal echocardiography for the detection of patent foramen ovale in stroke patients Eur J Echocardiogr, March 1, 2006; 7(2): 147 - 154. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Homma and R. L. Sacco Patent Foramen Ovale and Stroke Circulation, August 16, 2005; 112(7): 1063 - 1072. [Full Text] [PDF] |
||||
![]() |
O. K. Mohrs, S. E. Petersen, D. Erkapic, C. Rubel, R. Schrader, B. Nowak, W. A. Fach, H.-U. Kauczor, and T. Voigtlaender Diagnosis of Patent Foramen Ovale Using Contrast-Enhanced Dynamic MRI: A Pilot Study Am. J. Roentgenol., January 1, 2005; 184(1): 234 - 240. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. W. Schuchlenz, G.P. Anzola, E. Morandi, F. Casilli, and E. Onorato Contrast Ultrasound Techniques in the Detection and Quantification of Patent Foramen Ovale: Myth Versus Reality * Response: Stroke, December 1, 2004; 35(12): 2755 - 2756. [Full Text] [PDF] |
||||
![]() |
M. A. Sloan, A. V. Alexandrov, C. H. Tegeler, M. P. Spencer, L. R. Caplan, E. Feldmann, L. R. Wechsler, D. W. Newell, C. R. Gomez, V. L. Babikian, et al. Assessment: Transcranial Doppler ultrasonography: Report of the Therapeutics and Technology Assessment Subcommittee of the American Academy of Neurology Neurology, May 11, 2004; 62(9): 1468 - 1481. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Stroke Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2002 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |