From the Department of Neurology, Medical Centre Chemnitz, Chemnitz, Germany (B.S., J.J.S., J.K.); Academic Neurosurgical Unit and Wolfson Brian Imaging Centre, Addenbrookes Hospital, Cambridge, UK (M.C.); Department of Neurosurgery, Johann Wolfgang von Goethe University, Frankfurt, Germany (A.R., H.Y.); Department of Neurosurgery, Academic Hospital München-Schwabing, Germany (D.S.); and Department of Neurology, Technical University, München, Germany (D. Sander).
Correspondence to Bernhard Schmidt, Pembaurstr 14a, 81243 München, Germany. E-mail BSchmidt{at}lrz.tum.de
| Abstract |
|---|
|
|
|---|
Methods We studied 145 patients after severe head injuries or stroke. All patients had direct ICP, arterial blood pressure (ABP), and transcranial Doppler middle cerebral artery blood flow velocity (FV) monitored. The SCA was assessed by moving correlation (Mx index) of cerebral perfusion pressure (CPP=ABP-ICP) and cerebral blood flow velocity and correlation of ABP and ICP (PRx index). nICP was calculated from ABP and FV waveforms. When nICP was used instead of ICP, the SCA was continuously estimated, and the model was dynamically adapted to the SCA.
Results High and moderate correlations between invasively (Mx, PRx) and noninvasively (nMx, nPRx) estimated autoregulation indexes were observed (Mx: R=0.90, P<0.001; PRx: R=0.62, P<0.001). Values of Mx and nMx indicated contradictory SCA in 4 of 167 evaluated recordings; values of PRx and nPRx were contradictory in 27 recordings. When the model was adapted to the SCA, the mean error of ICP estimation decreased significantly (P<0.005).
Conclusions Continuous adaptation of the model to SCA improves the accuracy of noninvasive estimation of ICP and ICP dynamics. The same model provides a noninvasive and continuous assessment of SCA.
Key Words: autoregulation cerebral blood flow intracranial pressure ultrasonography, Doppler, transcranial
| Introduction |
|---|
|
|
|---|
Earlier, we presented a procedure for a noninvasive assessment of ICP (nICP) that used FV and ABP signals.1316 This procedure previously was shown to depend on the patients SCA. The objective of the present multicenter study was to combine the methods of continuous estimation of autoregulation9,10 with nICP assessment to achieve a continuous, minimally invasive assessment of SCA. The second aim was to check whether adaptation to a fluctuating SCA could improve the accuracy of nICP monitoring. Finally, the reliability of the noninvasive assessment of autoregulation was compared with the reliability of a method that avoided invasive ICP measurement by using ABP instead of CPP.5
| Materials and Methods |
|---|
|
|
|---|
Monitoring
Transcranial Doppler (TCD) measurements were taken by various 2-MHz pulsed Doppler devices (TC 2-64B, EME, in Munich-Bogenhausen and Chemnitz; Multidop-P, DWL, in Munich-Schwabing; Multiflow, DWL, in Frankfurt; and PCDop842, Scimed, or Neuroguard, Medasonics, in Cambridge). Flow patterns of the MCA were continuously recorded in the hemisphere ipsilateral to the ICP measurements. The ultrasound probe was fixed mechanically with a holder frame or an elastic band. Blood pressure was measured with a standard manometer line inserted into the radial, femoral, or dorsalis pedis artery.
ICP was measured with either implanted intraparenchymal or intraventricular microsensors (Camino Laboratories in Cambridge and Munich-Bogenhausen; Codman Group Inc in Munich; Rehau AG in Munich-Schwabing and Chemnitz), a sensor with air pouch probes (Spiegelberg Plc/Ltd/Co in Munich and Frankfurt), or an external ventricular drainage (in Chemnitz, Munich-Schwabing, and Frankfurt; only periods with closed drainage were evaluated.)
The study was approved by the ethics committee of the Technical University of Munich. In other centers, such a method of monitoring was routine clinical practice and did not require separate local ethics committee approval.
Computer-Assisted Recording
Personal computers were used for recording and analyzing FV, ABP, and ICP signals. The computers for signal recording were fitted with data acquisition systems (in Munich and Chemnitz, Daq 112B, Iotech, Inc; in Frankfurt, DTA300, Data Translation; in Cambridge, DTA2814, Data Translation). The sampling frequencies used ranged from 25 to 50 Hz. Signals were recorded daily for 20 to 120 minutes. Home-written data recording software was used in Munich and Chemnitz,13 as well as in Cambridge and Frankfurt.17
Assessment of SCA
Time-averaged values of ICP, ABP, CPP (CPP=ABP-ICP), and FV were calculated from time integration for 10-second intervals. An autoregulation index (Mx) was calculated as a Pearson correlation coefficient of 36 consecutive samples of CPP and FV mean values, ie, every 6 minutes. Positive association between CPP and FV (Mx >0) indicates passive dependence of blood flow on CPP. A zero or negative value of Mx indicates that cerebral flood flow was independent of CPP, ie, intact (or even overregulating) CA. Index Mx therefore describes autoregulation using its definition as a vascular reflex allowing maintenance of constant cerebral blood flow regardless of varying ABP or CPP.
Similarly, the pressure-reactivity index (PRx) was calculated as a Pearson correlation coefficient of 36 consecutive samples of ABP and ICP. PRx is similar to Mx but takes into account active regulation in cerebral blood volume.18,19 If CA works, a decrease in ABP provokes compensatory vasodilatation, an increase in cerebral blood volume, and hence an increase in ICP. Therefore, with intact autoregulation, PRx is negative. If CA fails, a rise in ABP produces no active vascular response but passive vascular distension, an increase in blood volume, and a rise in ICP. Therefore, in this state, changes in ICP and ABP will take place in the same direction, rendering PRx positive. Mx and PRx are correlation coefficients, not ratios of signal changes, and thus are standardized in value (1 to -1).
In previous studies, Mx was significantly correlated with the leg-cuff test8 and the transient hyperemic response test.6 However, grading of CA by monitoring the association between slow waves of FV and CPP has more in common with the classic definition of autoregulation1 than with the concept of dynamic autoregulation.4 The index Mx has been shown to correlate with PRx, and both indexes9,10 are strong predictors of outcome.3,10 Autoregulation indexes were calculated if dynamics of slow waves in either ABP or FV signals (measured as relative SD of 10-second signal averages) exceeded definite thresholds: >0.03 for FV and 0.015 for ABP.
Noninvasive ICP Assessment
The procedure uses parameters derived from FV and ABP waveforms. They are called TCD characteristics and essentially consist of coefficients of a linear ABP
FV transformation. The TCD characteristics are used to calculate a dynamic transformation formula connecting ICP and ABP (ABP
ICP).
The model assumes a linear relationship between TCD characteristics and the coefficients of the ABP
ICP transformation. This linear function (nICP matrix) constitutes the kernel of the nICP assessment procedure and may be constructed using the reference data consisting of FV, ABP, and direct ICP recordings from a well-defined group of patients. Figure 1 is a flow chart of the nICP assessment.
|
Noninvasive ICP Assessment With Feedback-Controlled Adaptation to SCA
Previously, we used data from Cambridge from 113 reference TBI patients with either intact (AUT group, n=56) or disturbed (NAUT group, n=69) CA to generate 2 nICP procedures specifically suitable for nICP assessment in patients with corresponding SCA. We observed a clear benefit in the accuracy of assessment of nICP and its dynamics (A and B waves, ICP trends) if specific nICP procedures were used.
These results encouraged the following construction. For calculation of nICP, we used a linear combination of 2 CA-specific nICP procedures. During analysis, the state of autoregulation was regularly estimated by correlation between FV and nCPP (nMx) and between ABP and nICP (nPRx). These indexes were used as feedback to fit the nICP procedure to the current SCA. A linear combination of both above-mentioned CA-specific nICP assessment procedures was computed each time from the currently estimated SCA to create the best fitting procedure and to calculate final nICP values (Figure 2 and the Appendix).
|
The simultaneous use of Cambridge patients for construction of nICP procedure and for nICP assessment was avoided by excluding each tested patient from the construction of procedure; ie, each time, the remaining 112 patients were used to calculate the reference transformation that was used to calculate nICP in the 113th patient.
For evaluation of the accuracy of noninvasive SCA assessment, noninvasive autoregulation indexes nMx and nPRx, corresponding to Mx and PRx, were calculated for each patient. Furthermore, MxABP was calculated as the correlation coefficient between FV and ABP (instead of CPP in Mx).8
Although Mx, PRx >0.0 indicated impaired CA and Mx, PRx <0.0 indicated intact CA, it is clear from general properties of physiological systems that the value 0.0 (or any other value) does not constitute a sharp threshold between intact and impaired CA. It seems appropriate to assume an area (surrounding 0.0) of continuous transition from intact to impaired CA, indicating nondefinite states somewhere in between. These considerations motivated the following definitions. An autoregulatory index was defined to indicate a definite SCA if its value was either <-0.2 or >0.2 and a weak definite SCA if its value was >-0.1 or >0.1. Two indexes were classified as contradictory if 1 indicated a definite SCA and the other was signed differently (eg, with this terminology, Mx=0.25 and nMx=-0.1 were contradictory, whereas PRx=-0.6 and nPrx=-0.01 were not). Cases of agreement and contradiction between noninvasive and invasive indexes were counted, and the sensitivity (ratio of invasively and noninvasively concurring definite impaired CA to definite impaired CA) and specificity (ratio of invasively and noninvasively concurring definite intact CA to definite intact CA) of such an estimation were calculated. Fuzzy sensitivity and fuzzy specificity were defined accordingly using weakly definite noninvasive indexes.
nICP was quantitatively compared with ICP using mean absolute errors (
ICP) defined as the absolute value of 10-second averaged differences of ICP-nICP.
| Results |
|---|
|
|
|---|
|
Significant correlations (P<0.001) with coefficients R=0.90 between Mx and nMx, R=0.62 between PRx and nPRx, and R=0.86 between Mx and aMx were attained (Figure 3 and Table 1). In 125 recordings of 106 patients, Mx was definite (>0.2 or <-0.2). In 2 of these recordings, the value of nMx was contradictory to Mx; ie, nMx and Mx had different signs. aMx and Mx were contradictory in 6 recordings; in 4 of these, ICP was increasing >40 mm Hg. The sensitivity of both nMx and aMx for detection of impaired CA was 0.73, and the fuzzy sensitivity of nMx and aMx was 0.84 and 0.83. Subsequently, nMx indicated definite SCA in 121 recordings of 104 patients. nMx was contradictory to Mx in 3 cases, and aMx was contradictory to Mx in 5 cases. The specificity of nMx for Mx estimation was 0.92, the specificity of aMx was 0.79, and the fuzzy values were 0.97 and 0.92. The sensitivity of the nPRx index for predicting definite PRx values was 0.61, and its specificity was 0.67. The corresponding fuzzy values were 0.77 and 0.78 (Table 1).
|
The median
ICP was 6.0 mm Hg. In 75% of the evaluated recordings,
ICP was <9.1 mm Hg (Table 2). A paired t test comparison to nICP assessments with a fixed nICP procedure showed significantly lower ICP estimation errors if the nICP procedure was adapted to SCA (mean errors, 7.6 and 6.9 mm Hg; P<0.005). Restricted to TBI patients, the mean
ICP was 7.1 mm Hg; it was 4.3 mm Hg in patients with stroke. Examples of nICP assessments during ICP A and B waves20,21 are shown in Figure 4.
|
|
| Discussion |
|---|
|
|
|---|
If arguing with the sensitivity and specificity of SCA detection, the reader should keep in mind that we identify "real" SCA with certain values of its indicative parameters Mx and PRx. Although a strong relationship between Mx and other methods for CA assessment, demonstrated by earlier studies,6,8 appears to be obvious, an exact relationship between Mx, PRx, and CA may yet be unknown. In particular, this became evident in cases when the values of Mx and PRx indicated different SCA. A complete parameterized assessment of SCA might not be possible because of the complexity and variety of autoregulatory mechanisms. It is probable that CA after head injury and stroke is regionally disturbed; therefore, its assessment with MCA FV may be weighted by the volumes of brain tissue with preserved and disturbed autoregulation supplied by MCA.
Although our model was constructed from data of TBI patients, we additionally tested 10 stroke patients to check its broader applicability. Despite the small number of tested stroke patients, the results encourage further attempts to validate this method in stroke patients with intracerebral, subarachnoid, and cerebellar hemorrhages, whereas subarachnoid hemorrhages with manifestations of vasospasm and patients with infarcts are expected to be problematic because of their vessel abnormalities. Especially in stroke patients, the danger of intracranial hypertension is frequently underestimated. In many cases, ICP is not monitored, and it may cause severe secondary insults by unrecognized ICP elevation. Therefore, nICP assessment could be a helpful tool, particularly in these patients.
It was shown that the shifted correlation coefficient between ABP and FV (aMx) was a reasonable estimator of Mx. Although in most cases the contribution of ICP to CPP dynamics was low, in some patients when ICP was strongly increased or of high dynamics, CPP could not be replaced by ABP for SCA assessment.7,8 nMx was a reliable estimator of Mx even in patients with increased ICP, in which case monitoring of SCA was of particular importance.2224
The study of SCA-specific nICP estimation was motivated by the observation that interactions of FV, ABP, and ICP signals are affected by CA. Improved accuracy of these procedures led to the idea of a dynamic adaptation to current SCA. In our model, SCA was assessed with nICP, which was calculated from FV and ABP signals. SCA in turn was used to control the nICP calculation rules. This closed control loop is in accordance with the physiological situation: ABP and FV signals cannot completely assess nICP dynamics, which additionally are influenced by such autoregulatory system changes as dilatation or constriction of cerebral vessels causing intracranial volume changes. The SCA-controlled nICP assessment procedure showed higher accuracy in ICP estimation than the corresponding constant procedure. It generally predicted ICP dynamics more accurately, ie, less dampened, than the constant method did (Figure 4a and 4b). Furthermore, it allowed reliable computation of nICP in patients with highly dynamic changes in autoregulatory behavior, eg, in patients with ICP plateau waves (Figure 4c and 4d)20 in whom on the top plateau level of the patients autoregulatory reserve was exhausted and the plateau wave generation was initiated by an autoregulatory response.18
The SCA-controlled nICP assessment may be used without additional tests and is therefore easy to handle. Moreover, it provides continuous monitoring of SCA without clinical intervention. For a first approximation, SCA may be estimated by the correlation between ABP and FV signals. Use of nICP assessment improves the reliability of Mx estimation in cases of increased ICP dynamics and provides the additional index nPRx.
| Acknowledgments |
|---|
| Footnotes |
|---|
| Appendix 1 |
|---|
|
|
|---|
ICP). This linear transformation is familiarly called impulse response function (IRF). If input and output signals are known, the corresponding IRF may be calculated by standard methods of system analysis.25,26 In our model, the ABP
ICP IRF was represented by 26 coefficients. The computational rules for nICP were given byequation
|
|
where n=25, wi was a 26-coefficient ABP
ICP IRF, and the consecutive samples ABPi, i=0...25, were equally spread over a time period of 3 cardiac cycles. For TCD characteristics (tcdj), we used a 13-coefficient ABP
FV IRF (tj), 13 ratios of tj/ABP, pulsatility index (systolic FV minus end-diastolic FV divided by mean FV), and pulse length. The benefit for nICP assessment of this certain combination was validated by extending the preceding tests of different sets of TCD characteristics.
ABP
ICP IRF (wi) was calculated from tcdj by multiplying a 26x28 matrix A and adding a 26-dimensional vector B, expressed by the formula (wi)=Ax(tcdj)+B.
The compound of A and B, the 26x29 matrix AB, determined the properties of the nICP assessment procedure and was called the nICP matrix. AB was computed by a multiple regression analysis of data recordings from a patient reference group consisting of FV, ABP, and (measured) ICP curves. The multiple regression process related the TCD characteristics to the time-corresponding ABP
ICP IRF coefficients.
Adaptation of the nICP Matrix to SCA
The improved nICP assessment procedure in this study (Figure 2a and 2b) consisted of a dynamically modified nICP matrix ABCUR, which was a linear combination of 2 specific nICP matrixes, ABAUT and ABNAUT. Both were previously calculated from reference groups of patients with preserved CA (ABAUT) and disturbed CA (ABNAUT). The nICP procedure started with a "neutral" nICP matrix, ABSTART (ABSTART=0.4xABAUT+0.6xABNAUT), where 0.4 was found to be best for our purposes. The autoregulation indexes nMx and nPRx were computed every 50 seconds from FV, ABP, and nICP signals, and the control matrix ABCUR was adapted to the current SCA by ABCUR=(0.5-MPR)xABAUT+(0.5+MPR)xABNAUT,where MPR=(nMx+nPRx)/2.
It is easily verified that ABCUR=ABAUT, if MPR=-0.5, indicating preserved CA; ABCUR=ABNAUT, if MPR=0.5, indicating impaired CA; and ABCUR lies between ABCUR and ABAUT, if MPR lies between -0.5 and 0.5.
Coefficients of ABCUR and ABAUT might also take on negative values. In this case, the properties of the opposite nICP matrix were boosted.
Received April 11, 2002; revision received August 9, 2002; accepted August 19, 2002.
| References |
|---|
|
|
|---|
2. Enevoldsen EM, Jensen FT. Autoregulation and CO2 responses of cerebral blood flow in patients with severe head injury. J Neurosurg. 1978; 48: 689703.[Medline] [Order article via Infotrieve]
3. Czosnyka M, Smielewski P, Piechnik S, Steiner LA, Pickard JD. Cerebral autoregulation following head injury. J Neurosurg. 2001; 95: 756763.[CrossRef][Medline] [Order article via Infotrieve]
4. Aaslid R, Lindegaard KF, Sorteberg W, Nornes H. Cerebral autoregulation dynamics in humans. Stroke. 1989; 20: 4552.
5. Giller CA. A bedside test for cerebral autoregulation using transcranial Doppler ultrasound. Acta Neurochir (Wien). 1991; 108: 714.
6. Smielewski P, Czosnyka M, Kirkpatrick P, Pickard JD. Evaluation of transient hyperemic response test in head injured patients. J Neurosurg. 1997; 86: 773778.[CrossRef][Medline] [Order article via Infotrieve]
7. Diehl RR, Linden D, Lücke D, Berlit P. Phase relationship between cerebral blood flow velocity and blood pressure: a clinical test of autoregulation. Stroke. 1989; 20: 13.
8. Piechnik SK, Yang X, Czosnyka M, Smielewski P, Fletcher SH, Jones AL, Pickard JD. The continuous assessment of cerebrovascular reactivity: a validation of the method in healthy volunteers. Anesth Analg. 1999; 89: 944949.
9. Czosnyka M, Smielewski P, Kirkpatrick P, Menon DK, Pickard JD. Monitoring of cerebral autoregulation in head-injured patients. Stroke. 1996; 27: 18291834.
10. Czosnyka M, Smielewski P, Kirkpatrick P, Laing RJ, Menon D, Pickard JD. Continuous assessment of the cerebral vasomotor reactivity in head injury. Neurosurgery. 1997; 41: 1119.[CrossRef][Medline] [Order article via Infotrieve]
11. Steinmeier R, Bauhuf C, Hübner U, Bauer RD, Fahlbusch R, Laumer R, Bondar I. Slow rhythmic oscillations of blood pressure, intracranial pressure, microcirculation, and cerebral oxygenation. Stroke. 1996; 27: 22362243.
12. Panerai RB, White RP, Markus HS, Evans DH. Grading of cerebral dynamic autoregulation from spontaneous fluctuations in arterial blood pressure. Stroke. 1998; 29: 23412346.
13. Schmidt B, Klingelhöfer J, Schwarze JJ, Sander D, Wittich I. Noninvasive prediction of intracranial pressure curves using transcranial Doppler ultrasonography and blood pressure curves. Stroke. 1997; 28: 24652472.
14. Schmidt B, Schwarze JJ, Czosnyka M, Sander D, Wittich I, Klingelhöfer J. A method for a simulation of continuous intracranial pressure curves. Comp Biomed Res. 1998; 31: 231243.[CrossRef][Medline] [Order article via Infotrieve]
15. Schmidt B, Czosnyka M, Schwarze JJ, Sander D, Gerstner W, Lumenta CB, Pickard JD, Klingelhöfer J. Cerebral vasodilatation causing acute intracranial hypertension: a method for non-invasive assessment. J Cereb Blood Flow Metab. 1999; 19: 990996.[CrossRef][Medline] [Order article via Infotrieve]
16. Schmidt B, Czosnyka M, Schwarze JJ, Sander D, Gerstner W, Lumenta CB, Klingelhöfer J. Evaluation of a method for noninvasive intracranial pressure assessment during infusion studies in patients with hydrocephalus. J Neurosurg. 2000; 92: 793800.[Medline] [Order article via Infotrieve]
17. Zabolotny W, Czosnyka M, Smielewski P. Portable software for intracranial pressure recording and waveform analysis. In: Nagai H, Kamiya K, Ishii S, eds. Intracranial Pressure IX. Berlin, Germany: Springer Verlag; 1994: 439440.
18. Rosner MJ, Becker DP. Origins and evaluation of plateau waves: experimental observations and theoretical model. J Neurosurg. 1984; 60: 312324.[Medline] [Order article via Infotrieve]
19. Muizelaar JP, Ward JD, Marmarou A, Newlon PG, Wachi A. Cerebral blood flow and metabolism in severely head-injured children. Autoregulation J Neurosurg. 1989; 71: 7276.
20. Lundberg N. Continuous recording and control of ventricular fluid pressure in neurosurgical practice. Acta Psych Neurol Scand. (Suppl). 1960; 149: 1193.
21. Newell DW, Aaslid R, Stooss R, Reulen HJ. The relationship of blood flow velocity fluctuations to intracranial pressure B waves. J Neurosurg. 1992; 76: 415421.[Medline] [Order article via Infotrieve]
22. Enevoldsen EM. CBF in head injury. Acta Neurochir Suppl (Wien). 1986; 36: 133136.[Medline] [Order article via Infotrieve]
23. Shigemori M, Nakashima H, Moriyama T, Tokutomi T, Nishio N, Harada K, Kuramoto S. Noninvasive study of critical thresholds of intracranial pressure and cerebral perfusion pressure for cerebral circulation and brain function. Neurol Res. 1989; 11: 165168.[Medline] [Order article via Infotrieve]
24. Paulson OB, Strandgaard S, Edvinsson L. Cerebral autoregulation. Cerebrovasc Brain Metab Rev. 1990; 2: 161192.[Medline] [Order article via Infotrieve]
25. Marmarelis P, Marmarelis V. Analysis of Physiological Systems. New York, NY: Plenum Press; 1978: 1221.
26. Kasuga Y, Nagai H, Hasegawa Y, Nitta M. Transmission characteristics of pulse waves in the intracranial cavity of dogs. J Neurosurg. 1987; 66: 907914.[Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
K. H. Polderman, S. A. Mayer, D. Menon, P. M. Kochanek, M. J. Bell, P. D. Adelson, A. Kornecki, G. Morrison, D. Karakitsos, A. Karabinis, et al. Hypothermia Therapy after Traumatic Brain Injury in Children N. Engl. J. Med., September 11, 2008; 359(11): 1178 - 1180. [Full Text] [PDF] |
||||
![]() |
T. Nottidge, I. Edemekong, J. Emejulu, and T. Shokunbi Transcranial Sonogram Through a Sutural Fracture for the Assessment of Posttraumatic Intracranial Mass Lesion in a Child Journal of Diagnostic Medical Sonography, March 1, 2008; 24(2): 107 - 109. [Abstract] [PDF] |
||||
![]() |
M. Smith Monitoring Intracranial Pressure in Traumatic Brain Injury Anesth. Analg., January 1, 2008; 106(1): 240 - 248. [Abstract] [Full Text] [PDF] |
||||
![]() |