(Stroke. 2003;34:2576.)
© 2003 American Heart Association, Inc.
Original Contributions |
From the Departments of Neurology (F.M.M., A.S., G.S.V.) and Cardiology (K.S.C.), Royal Hallamshire Hospital, and Sheffield Vascular Institute, Northern General Hospital (T.J.C., P.A.G., J.D.B.), Sheffield Teaching Hospitals National Health Service Trust, Sheffield, UK.
Reprint requests to Dr Fiona McKevitt, Department of Neurology, Royal Hallamshire Hospital, Glossop Rd, Sheffield S10 2JF, UK. E-mail fionamckevitt{at}hotmail.com
| Abstract |
|---|
|
|
|---|
Methods Patients were randomized to CEA (n=49) or endovascular treatment (n=55) that comprised percutaneous transluminal angioplasty alone (n=31), balloon-expandable stent (n=13), or self-expandable stent (n=11). A baseline 24-hour ambulatory blood pressure recording was made before carotid intervention and repeated at 24 hours, 1 month, and 6 months after the procedure.
Results In the first 24 hours after the procedure, episodes of hypotension occurred in 75% of the CEA group and 76% of the endovascular group; hypertension occurred in 11% and 13%, respectively. There was a significant fall in blood pressure at 1 hour after the procedure in both groups (24 and 16 mm Hg fall in CEA and endovascular groups, respectively), but this was only sustained in the endovascular group. The pattern of blood pressure response in the first 24 hours was significantly different (P<0.0001, ANCOVA). Systolic blood pressure was significantly lower at 1 and 6 months only in the surgical group (6 and 5 mm Hg fall, respectively).
Conclusions Both CEA and endovascular treatment have an effect on blood pressure stability, particularly within the first 24 hours after the procedure.
Key Words: angioplasty, balloon blood pressure carotid endarterectomy carotid stenosis
| Introduction |
|---|
|
|
|---|
See Editorial Comment, page 2581
This is the first study to compare the effect of both CEA and endovascular treatment on blood pressure within the first 6 months after treatment and the first to assess the longer-term effects of endovascular treatment on blood pressure.
| Subjects and Methods |
|---|
|
|
|---|
Carotid Endarterectomy
All patients underwent a standardized general anesthetic for the duration of the study period. Diazepam 10 mg was used as a premedication, with midazolam (2 to 3 mg), etomidate (10 to 20 mg), and alfentanil (2.5 mg) used as induction agents. Vecuronium (10 mg) was used for neuromuscular blockade. Nitric oxide, oxygen, and isoflurane were used during maintenance of general anesthesia. The surgical technique was also standardized. The carotid sinus nerve was preserved and injected with lidocaine before carotid artery clamping. Shunting and Dacron patch angioplasty were performed routinely. Patients were then transferred to a high-dependency unit for 24 hours after the procedure. Blood pressure was maintained at ±30 mm Hg of the preprocedure systolic blood pressure, with the use of esmolol as a hypotensive agent (to reduce the risk of hyperperfusion injury9) or phenylephrine as a hypertensive agent if required.
Carotid Angioplasty and Stent
After selective angiography of the femoral artery was performed, a 7F long sheath or 9F guiding catheter was passed into the common carotid artery. Intra-arterial atropine (1.2 mg) was given, and the lesion was crossed with a V18 control wire (Boston Scientific). If balloon angioplasty was used, the balloon size was selected after measurement of the internal carotid artery. If the result was considered insufficient, a stent was placed (secondary stenting). During the study period, routine use of stents with prior 3 mm predilatation became the strategy (primary stenting). Postdilatation of the stent was routine, with the use of a balloon compatible with the internal carotid artery above the stenosis. Both balloon-expandable stents (Palmaz-Schatz, Cordis) and self-expanding stents (Wallstent, Schneider) were used.
An intravenous heparin bolus of 5000 U was given during the procedure and then continued as an intravenous infusion for 24 hours to achieve an activated partial thromboplastin time ratio of 1.5 to 2.5. Patients were then transferred back to the neurology ward for observation. No active management of blood pressure by esmolol or phenylephrine was used.
In addition to the CAVATAS protocol,4 a previously validated 24-hour ambulatory blood pressure recording (Space Labs SL-90207) was taken, before carotid intervention, to obtain a mean baseline systolic blood pressure.10 This was then repeated during the first 24 hours after the procedure. The number of patients experiencing episodes of hypotension or hypertension after CEA or endovascular treatment was documented, and the difference between the 2 procedures was assessed. An episode of hypotension was defined as a fall in systolic blood pressure of >30 mm Hg, and an episode of hypertension was defined as a rise in systolic blood pressure of >30 mm Hg from the baseline blood pressure reading. If the fall in systolic blood pressure was sustained for >1 hour, it was then defined as persistent hypotension. Twenty-fourhour blood pressure recordings were also taken at 1 and 6 months.
Statistical Analysis
Baseline systolic blood pressure was derived from the average reading of the preprocedure 24-hour ambulatory blood pressure recordings. The 2 groups were compared with an independent t test. The number of patients who experienced episodes of hypotension or hypertension was recorded, and the time that the first episode of hemodynamic disturbance occurred was noted. The difference between the 2 groups was compared with the
2 or Fisher exact test.
The average systolic and diastolic blood pressure at each hourly interval for both groups was calculated for the first 24 hours. Comparisons with preprocedure values were made with the paired t test. An ANCOVA with a general linear model, with group and time point as fixed factors, was used to compare both groups in the first 24 hours after the procedures.
The average systolic and diastolic blood pressure was also calculated for 1 and 6 months for both groups. Comparisons with preprocedure values were made with the paired t test.
We also performed a subgroup analysis of the endovascular group, taking into account the change of technique from PTA alone to angioplasty with carotid artery stenting and from balloon-expandable stents to self-expanding stents.
The threshold for statistical significance was taken as P<0.05.
| Results |
|---|
|
|
|---|
|
In the immediate postprocedural period, 5 patients (all surgical) were excluded from analysis because postprocedural 24-hour blood pressure data were not available as a result of technical failure of the blood pressure monitoring equipment.
Episodes of hypotension were seen frequently in both groups (75.0% in the surgical group and 75.9% in the endovascular group; P=0.92). The number of patients in each group who experienced a hypertensive episode was also similar: 11.4% in the surgical group versus 13.0% in the endovascular group (P=0.81). No patients in the endovascular group received the phenylephrine/esmolol regimen, whereas 33 patients in the surgical group received phenylephrine and 5 patients received esmolol. Persistent hypotension occurred in 50.0% of the surgical group and 44.4% of the endovascular group (P=0.58).
Immediate Changes
Accurate timing of hemodynamic disturbance after carotid intervention did not occur in 13 procedures (7 surgical, 6 endovascular), and therefore these were not included in this analysis. The first significant hemodynamic change in blood pressure occurred at or before 6 hours in 89.7% of the surgical group and in 73.2% of the endovascular group (P=0.09). This occurred at or before 10 hours in 93.1% of the surgical group and in 85.4% of the endovascular group (P=0.46).
At 1 hour there was a significant fall in blood pressure for both groups from baseline (surgical group mean systolic difference, 24.3 mm Hg; 95% CI, 17.1 to 31.5; P<0.001; endovascular group mean systolic difference, 16.2 mm Hg; 95% CI, 9.5 to 22.9; P<0.001) (Table 2).
|
Because of a marked reduction in data obtained at 24 hours (<50%), the average fall in blood pressure was compared at 20 hours. The surgical groups blood pressure returned to baseline, but there was a sustained fall in blood pressure in the endovascular group (surgical group mean systolic difference, -0.8 mm Hg; 95% CI, -12.7 to 11.1; P=0.89; endovascular group mean systolic difference, -22.6 mm Hg; 95% CI, -15.3 to -30.1; P<0.001).
Overall, when the 2 groups were assessed over the 24-hour time period, the difference in blood pressure response was highly significant (95% CI, 15.2 to 33.8; P<0.0001, ANCOVA) (Figure).
|
Changes at 1 Month
At 1 month, data were available for 92 procedures (41 in surgery group, 51 in endovascular group). Medication changes occurred in 8 patients (6 in surgery group, 2 in endovascular group), and these were omitted from analysis in addition to the patient who had no preprocedure 24-hour recording for comparison. In the surgical group there was a significant reduction in blood pressure (95% CI, 0.9 to 10.8; P<0.05). There was no difference in systolic blood pressure from preprocedure values in the endovascular group (95% CI, -0.91 to 6.5; P=0.14) (Table 2).
Changes at 6 Months
At 6 months, data were available for 77 procedures (37 in surgery group, 40 in endovascular group); a total of 7 patients were excluded from analysis because of changes in medication (6 in surgery group, 1 in endovascular group). There was a significant reduction in blood pressure in the surgical group (95% CI, 0.31 to 9.5; P<0.05); however, there was no difference in systolic blood pressure in the endovascular group (95% CI, -0.8 to 2.6; P=0.29) (Table 2).
Endovascular Group
PTA alone was performed on 31 patients (1 was excluded from analysis because of lack of preprocedure data; see above). Twenty-four patients underwent carotid artery stenting; of these, 7 were secondary and 17 were primary stenting procedures. The use of balloon-expandable stents occurred in 13 patients, and self-expanding stents were used in 11 patients.
There was no difference in frequency of hypotensive events between PTA and carotid artery stenting groups (76.7% and 75.0%, respectively; P=1). Persistent hypotension occurred in 43.4% of the PTA group and 45.8% of the carotid artery stenting group (P=1). The baseline average blood pressures for both groups were not significantly different. However, when the 2 groups were assessed over the 24-hour time period, the difference in blood pressure response was highly significant (95% CI, 30.1 to 51.2; P<0.0001, ANCOVA), with a lower blood pressure sustained over the first 24 hours in the carotid artery stenting group. No long-term blood pressure differences were seen between the PTA and carotid artery stenting groups (Table 2).
There was no difference in frequency of hypotensive events between balloon-expandable and self-expandable stent groups (72.7% and 76.9%, respectively; P=1). However, persistent hypotension occurred in 61.5% of the balloon-expandable stent group and 27.3% of the self-expandable stent group (P=0.12). The baseline average blood pressures for both stent types were not significantly different. However, when the 2 stent types were assessed over the 24-hour time period, the difference in blood pressure response was highly significant (95% CI, 37.9 to 57.0; P<0.0001, ANCOVA), with the balloon-expandable stent resulting in lower systolic blood pressures.
Neurological Complications
Within the perioperative period (ie, first 30 days), the all stroke/death rates were 10.2% in the surgical group and 5.5% in the endovascular group.
Of those who experienced a hemodynamic disturbance, 16.0% had a neurological complication in the perioperative period, including all transient ischemic attack, amaurosis fugax, stroke, and stroke-related death. Only 5.9% of those who were hemodynamically stable did so. This was not statistically significant (P=0.46, Fisher exact test) (Table 3).
|
Excluded Data
The baseline characteristics of the patients excluded in the immediate analysis did not vary from the baseline characteristics of those retained. The 30-day outcome for this group included 2 nondisabling major strokes only. Therefore, the neurological complication rate and the all stroke/death rate were both 10.5%. For the remaining procedures, the 30-day neurological complication rate and the all stroke/death rate were 14.1% and 7.1%, respectively.
The reasons for loss of data at 1 and 6 months included death (n=4), stroke (n=1), move from the country (n=1), inaccurate timing of data (n=5), refusals (n=19), and surgical procedure (n=1).
| Discussion |
|---|
|
|
|---|
This study has several important strengths. It is the first to compare the effects on blood pressure of both CEA and endovascular treatment of carotid artery stenosis, and the blood pressure measurements were obtained prospectively. To help eliminate factors that reduce the reliability of single blood pressure values, such as
white coat effect,
time of measurement, and interrater reliability, 24-hour blood pressure readings were used both before and after the procedure.11
Although the baseline systolic blood pressure is slightly lower in the endovascular group, it is the change in blood pressure from each groups baseline that is highly significant. Interpretation of results must take into account, however, that the postoperative care differed between the 2 groups. The surgical patients were monitored on a high-dependency unit and consequently had more active management of blood pressure. This would not have altered the initial documentation of an occurrence of hemodynamic disturbance because pharmacological intervention was only triggered if the change in systolic blood pressure was >30 mm Hg. Nevertheless, the treatment received by the surgical group could potentially result in further hemodynamic disturbance or limit the degree of disturbance; this may explain the difference seen between the 2 groups. However, despite this intervention, persistent hypotension was still observed in 50% of the surgical group. Furthermore, the surgical procedure was standardized throughout the study period, but the endovascular technique varied, with 44% of patients having stent insertion. Separate analysis of these patients was therefore performed. Medication changes that potentially had effects on blood pressure occurred in 9 patients in the 6-month follow-up period. In 4 patients changes were made for reasons not related to blood pressure management, in 1 patient antihypertensive drugs were stopped because of prolonged hypotension, and 4 patients were treated for hypertension. Since these changes were a potential source of bias, these patients were omitted from analysis. However, the inclusion of these patients does not change the results except that the reduction in 6-month diastolic blood pressure in the surgical group would reach statistical significance (mean diastolic difference, 3.4 mm Hg; 95% CI, 0.7 to 6.1; P<0.05). A limitation of this study is the reduction in data seen within the first 24 hours but particularly at 1 and 6 months. This is unfortunate in that this represents a potential source of bias and limits any definite conclusion being drawn.
Hemodynamic disturbances immediately after carotid intervention are well recognized. Hypertension has been noted with a frequency of 9% to 38% after CEA2,12 and 39% after carotid stent insertion.7 A wide variation in the frequency of hypotension has been observed (12% to 50% after endarterectomy1,2 and 18.8% to 56.1% after endovascular intervention6,8). In this study a low frequency of episodes of hypertension was noted (11.4% after surgery, 13.0% after endovascular treatment), but a much higher frequency of episodes of hypotension was observed (75.0% after surgery and 75.9% after endovascular treatment). The wide variation in frequency of blood pressure fluctuations seen in other studies and the high incidence of hypotensive episodes observed in this study may be explained in part by the definitions used. A strict definition of a change of >30 or <30 mm Hg was chosen for our study because this was the degree of change previously allowed in postsurgical patients before active treatment was given. Persistent hypotension lasting >1 hour was also assessed because this has previously been shown to relate to adverse neurological outcome,6 although the magnitude of the change in systolic blood pressure (>50 mm Hg) has recently also been shown to relate to adverse neurological events after carotid stenting.13 Comparison between the 2 groups in this regard was subject to bias because the surgical group received more active treatment. Furthermore, the surgical technique can influence the type of blood pressure disturbances seen, so that dissection methods that preserve the carotid nerve result in fewer episodes of hypertension.1 The methods used in this study would favor hypotensive rather than hypertensive changes.
The initial difference in blood pressure response between the 2 groups is interesting. During CEA, removal of the atheromatous plaque reduces pressure wave dampening, thus increasing baroreceptor stimulation, resulting in hypotension.14 The more sustained fall seen in the endovascular group may be explained partly by increased distension of the carotid sinus from compression of the atherosclerotic plaque during balloon angioplasty. Further stimulation may also result from additional pressure exerted by a stent. In this study both angioplasty alone and carotid artery stenting result in a sustained reduction in systolic blood pressure in the immediate postprocedural period, which was more pronounced when a stent was used. Persistent hypotension has been found to be more frequent after the use of balloon-expandable rather than self-expanding stents.6 This is due to the fact that balloon-expandable stents are less flexible and have a stronger radial force. Although the number of episodes of hypotension did not differ between the 2 types of stent in this study, a greater proportion of patients in the balloon-expandable group experienced persistent hypotension. There was a highly significant difference seen in blood pressure response in the balloon-expandable group, with a more marked lowering of systolic blood pressure over the first 24 hours. Current endovascular practice is based on the use of self-expandable stents; the effect on blood pressure with the use of these stents in the immediate perioperative period is shown in the Figure.
Although the increase in baroreceptor activity is thought to be a short-term phenomenon, the baroreceptors can be reset to function at higher levels secondary to age, atherosclerosis, or chronic hypertension.15 The change in vessel wall and removal of the atherosclerotic plaque after surgery may cause further resetting of the baroreceptors and hence a long-term fall in blood pressure. Previous studies assessing blood pressure changes at 6 months and 2 years after endarterectomy did not find a long-term effect; however, 24-hour blood pressure monitoring was not used.16,17 A study that used 24-hour ambulatory monitoring also found a reduction in blood pressure 2 weeks after endarterectomy.18
Blood pressure management plays a crucial role in the prevention of stroke. Any decrease in blood pressure is associated with a decrease in the relative risk of stroke.19 A study of antihypertensive treatments in patients with a previous history of cerebrovascular events found that a reduction of 9/4 mm Hg was associated with a reduced risk of stroke of 28%, and therefore even small decreases in blood pressure are highly effective.20 Blood pressure instability in the immediate postprocedural period is a common occurrence after both CEA and the endovascular treatment of carotid artery stenosis. The long-term effect, however, requires further confirmation to assess whether endarterectomy rather than endovascular treatment is associated with a sustained reduction. Despite this, regular monitoring of blood pressure should continue to play a key role in ongoing management for both groups since the importance of blood pressure control for all patients at risk of cerebrovascular disease cannot be overemphasized.
Received December 3, 2002; revision received July 16, 2003; accepted July 16, 2003.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
E. Martinez-Fernandez, F. Boza Garcia, J. R. Gonzalez-Marcos, A. Gil-Peralta, A. Gonzalez Garcia, and A. Mayol Deya Risk Factors and Neurological Consequences of Syncopes Induced by Internal Carotid Artery Angioplasty Stroke, April 1, 2008; 39(4): 1336 - 1339. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Martinez-Fernandez, F. B. Garcia, J.R. Gonzalez-Marcos, A. G. Peralta, A. G. Garcia, and A. M. Deya Clinical and Electroencephalographic Features of Carotid Sinus Syncope Induced by Internal Carotid Artery Angioplasty AJNR Am. J. Neuroradiol., February 1, 2008; 29(2): 269 - 272. [Abstract] [Full Text] [PDF] |
||||
![]() |
M Paciaroni, V Caso, M Acciarresi, R W Baumgartner, and G Agnelli Management of asymptomatic carotid stenosis in patients undergoing general and vascular surgical procedures J. Neurol. Neurosurg. Psychiatry, October 1, 2005; 76(10): 1332 - 1336. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Reinhard, M. Roth, T. Muller, B. Guschlbauer, J. Timmer, M. Czosnyka, and A. Hetzel Effect of Carotid Endarterectomy or Stenting on Impairment of Dynamic Cerebral Autoregulation Stroke, June 1, 2004; 35(6): 1381 - 1387. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Dangas Editorial Comment--Hypotension After Carotid Revascularization Stroke, November 1, 2003; 34(11): 2581 - 2582. [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Stroke Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2003 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |