(Stroke. 2003;34:e222.)
© 2003 American Heart Association, Inc.
Research Reports |
From the Laboratory of Cardiovascular Physiology, Montpellier I University, Faculty of Medicine, Nimes, France (H.V., J.P.S., M.D.), and Departments of Neuroradiology (B.J., S.L., A.B., I.Q.) and Vascular Medicine (J.P.L.), Montpellier University Hospital, Montpellier Cedex, France.
Correspondence to Michel Dauzat, Laboratory of Cardiovascular Physiology, Montpellier I University, Faculty of Medicine, Avenue Kennedy, Nimes F 30907 France. E-mail dauzat{at}univ-montp1.fr
| Abstract |
|---|
|
|
|---|
Methods Carotid Wallstents (Schneider) were placed in the extracranial carotid arteries of 15 patients. B-mode ultrasound examination was performed with a 7.5-MHz probe on the carotid artery upstream; at the proximal, mid, and distal stent levels; downstream from the stent; and on the contralateral internal and common carotid arteries. Carotid diameter (d) and systolic diameter changes (
d) were measured with a dedicated image processing system (IÔ version 3.1, IÔDP), while pulse blood pressure (
P) was measured. Diameter compliance (Cd) and distensibility coefficient (DC) were calculated as Cd=2
d/
P and DC=2
d/
P/d and compared between measurement sites.
Results The evaluation could be completed in 8 of 15 patients. Compliance was significantly lower at the proximal, mid, and distal stent levels (27.77±1.11, 27.38±1.08, 27.38±1.09x10-3 mm · kPa-1) than upstream (103.3±36.7x10-3 mm · kPa-1), downstream (91.5±41.3x10-3 mm · kPa-1), or on the contralateral internal (87.6±28x10-3 mm · kPa-1) and common (149.3±47.6x10-3 mm · kPa-1) carotid arteries.
Conclusions Stenting of the extracranial carotid artery induces a compliance mismatch between the native and the stented artery.
Key Words: carotid arteries compliance stents
| Introduction |
|---|
|
|
|---|
| Materials and Methods |
|---|
|
|
|---|
Stent Placement
The stents were placed according to the ANAES consensus recommendations.4 A Carotid Wallstent (Schneider) was used in 7 patients (nominal size, 7x30 mm in 5 patients, 7x40 mm in 2 patients), whereas a Wallstent (Schneider) of nominal size 7x55 mm was placed in the patient with the dissected carotid artery. All stents were placed across the carotid bifurcation with their proximal part located in the common carotid artery and their distal part in the internal carotid artery.
Diameter and Flow Measurements
B-mode ultrasound images were obtained with a real-time ultrasound scanner and a 7.5-MHz convex probe (Sequoia, Acuson). B-mode, duplex, and color Doppler sonography was used in search of atheromatous plaques, homogeneous thickening of the carotid wall, and hemodynamically significant stenoses. Diameter measurement was performed 10 mm upstream (common carotid artery), at the stent level, and 10 mm downstream from the stent (internal carotid artery) on the stented artery and on the contralateral common and internal carotid arteries.
Real-time B-mode sequences were transferred to the image processing system (IÔ version 3.1, IÔDP) through the Y/C video output and then digitized, archived, and processed. The ECG signal was simultaneously acquired and used for cardiac cycle recognition. A strict parallel scanning of the artery was required so that the angle of approach was strictly perpendicular to the vessel axis, thus offering the largest carotid diameter and ensuring optimal axial resolution. At least 3 acquisitions were performed at the same anatomic site, and all values obtained during at least 20 cardiac cycles at each site were averaged.
Pulse pressure (systolic minus end-diastolic pressure) was obtained for each subject by monitoring the right brachial artery with an automated blood pressure measurement device.
Processing of B-mode Images and Sequences
On the stored B-mode images, the observer drew a region of interest encompassing the proximal and distal vessel walls on a length of at least 10 mm. The intima-to-lumen interface of the proximal wall and the lumen-to-intima interface of the distal wall were detected automatically by analyzing the mean gray-level profiles of the frame. The relative displacement of these interfaces on consecutive frames was measured (Figure 1). The waveform of vessel diameter and diameter changes over time was automatically generated, and the average systolic and diastolic diameters and diameter changes were calculated (Figure 2).5 Diameter compliance (Cd) was calculated as Cd=2
d/
P and distensibility coefficient (DC) as DC=Cd/d, where d is diastolic diameter,
P is average systolic-diastolic blood pressure change, and
d is average systolic-diastolic diameter change.
|
|
Statistical Analysis
Measurement sites were compared for diameter, compliance, and distensibility coefficient by use of a paired t test.
| Results |
|---|
|
|
|---|
The diameter was smaller in both the stented (P<0.05) and nonstented (P<0.005) internal carotid artery than in the ipsilateral common carotid artery. Compliance was significantly lower at the 3 measurement sites within the stent than downstream (P<0.0005), upstream (P<0.005), and on the contralateral common (P<0.0001) and internal (P<0.001) carotid arteries. The distensibility coefficient was significantly lower at the 3 measurement sites within the stent than downstream (P<0.005) and upstream from the stent (P<0.005) and in the contralateral common (P<0.0005) and internal (P<0.0005) carotid arteries. The distensibility coefficient was significantly lower on the common carotid artery upstream from the stent than on the contralateral common carotid artery (P<0.05) and at the proximal stent level than at the distal stent level (P<0.05) (the Table).
|
| Discussion |
|---|
|
|
|---|
The main technical limitation of our study is, despite the stent-related stiffness of the carotid artery, the difficulty in accurately analyzing arterial segments that were not perpendicular to the plane section, especially the internal carotid artery downstream from the stent, so 50% of the patients were excluded. Our study demonstrated a marked alteration of wall mechanics in the stented segment, supporting the compliance mismatch hypothesis. Further investigation of the correlation between distensibility and intima thickening is needed to completely demonstrate this physiopathological link.
| Acknowledgments |
|---|
We thank Margaret Manson for her meticulous and helpful reading of the manuscript.
Received May 28, 2003; accepted July 8, 2003.
| References |
|---|
|
|
|---|
2. Baird RN, Abbott WM. Pulsatile blood flow in arterial grafts. Lancet. 1976; 2: 948949.[Medline] [Order article via Infotrieve]
3. Vernhet H, Demaria R, Juan JM, Oliva-Lauraire MC, Marty-Double C, Sénac JP, Dauzat M. Wall mechanics of the stented rabbit aorta: long-term study and correlation with histologic findings. J Endovasc Ther. 2003; 10: 577584.[Medline] [Order article via Infotrieve]
4. Long A, Bousser MG. Medical recommendations and references 1997: stenosis of the origin of the cervical internal carotid and the carotid bifurcation: surgery, angioplasty: ANAES: National Agency for Accreditation and Evaluation in Health [in French]. J Mal Vasc. 1998; 23: 125143.[Medline] [Order article via Infotrieve]
5. Vernhet H, Demaria R, Juan JM, Oliva-Lauraire MC, Quere I, Gariepy J, Senac JP, Dauzat M. Validation of non invasive B-mode image processing versus wall tracking ultrasound for the study of wall mechanics in small caliber arteries. Clin Physiol Funct Imaging. 2002; 22: 180186.[Medline] [Order article via Infotrieve]
6. Giannattasio C, Failla M, Emanuelli G, Grappiolo A, Boffi L, Corsi D, Mancia G. Local effects of atherosclerotic plaque on arterial distensibility. Hypertension. 2001; 38: 11771180.
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Stroke Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2003 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |