(Stroke. 2003;34:1645.)
© 2003 American Heart Association, Inc.
Original Contributions |
From the Departments of Neurological Surgery (R.A., S.R.L., D.W.N.), Medicine (Cardiology) (G.H.B.), and Anesthesiology (W.H.G.), University of Washington School of Medicine, University of Washington Medical Center, Harborview Medical Center, Seattle, Wash.
Correspondence to David W. Newell, MD, Department of Neurological Surgery, Harborview Medical Center, 325 9th Ave, Seattle, WA 98104. E-mail dwn{at}u.washington.edu
| Abstract |
|---|
|
|
|---|
Methods TCD velocities and ABP in the radial artery were recorded before, during, and after 28 episodes of VF in 13 patients. CCPs were calculated by 3 different methods: (1) linear extrapolation from data during VF (gold standard); (2) linear extrapolation from normal heartbeat data; and (3) first harmonic Fourier filtering of normal heartbeat data.
Results The CCP during VF calculated from long diastoles was 32.9±11 mm Hg (mean±SD). The regular heartbeat estimate was 6.0±4.3 mm Hg lower (P<0.05). The CCP estimate with the use of a Fourier filter was 1.4±3.9 mm Hg (P=NS) lower than during VF. During hyperemia after defibrillation, the CCP decreased by 13.3 mm Hg, while velocity increased by 63%. The decrease in CCP could explain half of the increase in flow velocity during hyperemia.
Conclusions CCP can be accurately estimated from regular heartbeat data and is an important factor in regulation of the cerebral circulation.
Key Words: cerebrovascular circulation ultrasonography, Doppler, transcranial vasoreactivity
| Introduction |
|---|
|
|
|---|
|
With the use of ultrasound, the velocity of flow in the basal cerebral arteries can be assessed noninvasively from the Doppler shift. Because the calibers of the insonated vessels are not known precisely, this technique cannot be used to determine absolute values of flow. Therefore, the slope of the relationship between pressure and flow is less meaningful when Doppler techniques are used. Nonetheless, quantitative determinations of the CCP can be made if, as shown in some studies,5,6 the vessel walls of the basal cerebral arteries are stiff and the caliber changes are small over the normal range of pressures. Comparisons of transcranial Doppler ultrasound (TCD) recordings in the middle cerebral artery (MCA) with electromagnetic flowmetry demonstrated good proportionality between Doppler estimates of velocity and volume flow.7,8 Moreover, it is clear that at the zero-flow pressure, flow velocity must also be zero. In principle, the possibility exists that the CCP determined by ultrasound Doppler techniques may emerge as a physiologically relevant index of cerebrovascular tone.917
The dynamic pressure-flow relationships have been studied extensively in the coronary circulation.4,1821 The zero-flow pressure is significantly higher than the outflow pressure in the right atrium, and it reflects the level of vascular tone. Measurements in the coronary arteries of conscious humans showed a decrease in Pf=0 from 51.7 to 37.9 mm Hg after vasodilation with angiographic contrast media.19 In the cerebral circulation, a 25.9% decrease in CCP and a 20.7% increase in flow velocity were reported after the subject breathed 5% CO2 in air.15 Vasoconstriction during moderate hyperventilation increased the CCP from 13 to 23 mm Hg.17 These early reports suggest that the "back pressure" effect of the CCP is an important factor in the regulation of CBF.
In the heart, because of the complicated intramural distribution of pressure in systole, calculation of the zero-flow pressure requires either long diastoles induced experimentally18 or slow heartbeats occurring naturally.19 In the cerebral circulation, this complication does not exist, and normal heartbeats have been used for calculation of CCP.1,2,917 However, this technique has not been validated against the prolonged diastole method, which would be expected to give more accurate estimates because data closer to the zero-flow intercept are used.
The aim of the present study was to validate estimates of CCP based on regular heartbeat data against long diastole (CCPd) determinations after induced ventricular fibrillation (VF) during defibrillator testing. Two different regular beat methods were tested: (1) We used the conventional method917 of determining the CCP by linear regression (CCPr), using only mildly filtered recordings of pressure and flow velocity waveforms. (2) To minimize potential errors due to waveform distortion in the arm arterial system,22 we also investigated whether using maximally filtered waveforms would improve the estimates. For this purpose we determined the first harmonic Fourier filtered CCP (CCPf). Furthermore, the study protocol allowed an assessment of the relative role of the decrease in back pressure in mediating the flow increase during reactive hyperemia.
| Subjects and Methods |
|---|
|
|
|---|
2 of the following: isoflurane, nitrous oxide, enflurane, midazolam, sodium thiopental, sufentanil citrate, vecuronium bromide, alfentanil hydrochloride, metocurine, and propofol. Dobutamine, calcium carbonate, phenylephrine hydrochloride, or epinephrine was administered when necessary for control of ABP. The patients were not receiving other medication that would have influenced the results of the study significantly. No medication changes were made within 3 minutes before or after induction of VF by either alternating current or rapid ventricular pacing. Mechanical ventilation was used, and a constant hypocapnic end-tidal PCO2 measured by a Datex 223 monitor was maintained. The left MCA velocity signals were obtained with the use of a 2-MHz probe held in position by a headband. The probe was connected to a TC2-64B TCD instrument (EME GmbH) with analog output of the CBF velocity (CBFV) spectral outline waveform. The ABP was recorded from a radial artery catheter. Care was taken to ensure that the ABP waveform was undamped by blockage or air bubbles. After sampling at 50 Hz and analog-to-digital conversion, CBFV, ABP, end-tidal PCO2, and ECG signals were stored on the hard disk of a PC-compatible laptop computer. The different estimates of CCP were determined offline from the stored data with the use of a custom program written in Visual C++ for Windows 2000. Microsoft Excel 97 was used for statistical calculations.
Processing of Long Diastole Data After Induced VF
Analysis of the long diastolic fall in pressure and flow after cessation of cardiac function is the "classic" approach to estimating the extrapolated zero-flow pressure intercept.18 This procedure is illustrated by a representative case in Figure 1A and 1B. The flow velocity drops rapidly during the first 2.5 seconds after the induction of VF. Since Doppler instruments have high-pass filters to eliminate the artifacts from arterial wall movements, the quality of the spectral outline is poor at lower velocities. In the unit used, the filter cutoff was 150 Hz, corresponding to 6 cm/s. When the velocity falls below this level, noise will influence the output of the spectral outline estimator, and the waveform data are not reliable. As illustrated by the bar on the time axis in Figure 1A, the interval from the end of the dicrotic notch of the ABP wave to the point where CBFV fell below 6 cm/s was used for calculation of CCPd. The 6 cm/s velocity end point was reached in 11 patients before a ventricular contraction occurred. In 2 subjects, systolic activity was seen in all tests before the 6 cm/s end point was reached. However, in 3 instances the diastolic period was >2.5 seconds, and these data were considered to qualify as long diastoles.
A total of 60 episodes of VF were recorded. Thirty-two of these were discarded from analysis because the VF was not complete (long diastoles interrupted by early systolic waves in the ABP) or because of spikes or noise in the TCD recording. The remaining 28 episodes were analyzed as the example shown in Figure 1B. Corresponding samples of ABP (x axis) and CBFV (y axis) were plotted throughout the long diastolic interval. Extrapolation to the zero flow velocity pressure was done by conventional linear regression analysis.
Processing of Regular Heartbeat Waveforms
The 8 heartbeats immediately before induction of VF were used for estimation of CCPr. After low-pass filtering of both ABP and CBFV at 5 Hz (to eliminate high-frequency noise in the velocity), the computer plotted the data pairs of pressure versus velocity during the cardiac cycle for each of the 8 heartbeats immediately preceding VF. The plot from 1 such beat is shown in Figure 1C, and the linear extrapolation was done by the same procedure as for the prolonged diastoles. To compensate for time delays in pulse propagation and measurement systems, linear regression was performed with the velocity data delayed by up to 10 samples (=0.2 seconds). The delay that gave the highest r value was used for determination of the CCPr. The results from the 4 beats with the highest r value of the linear regression were averaged.
For the determination of pressureflow velocity relationships, both measurements should preferably be made in the same anatomic location. In the clinical environment this is difficult to achieve, and for ethical reasons the radial artery was the only available location from which a blood pressure recording could be made. The pressure waveform in this location will be somewhat distorted compared with the waveform in the cerebral vessels. This distortion is caused by the transmission of the pulse as well as by reflections of the pulse wave.22 The higher-frequency components of the waveform are more likely to be distorted than the fundamental (first harmonic) component. Therefore, removing the higher frequency content before processing may possibly yield a more accurate estimate of CCP. A similar approach has been described earlier and used to estimate cerebral perfusion pressure.23
Fourier analysis is the preferred method to remove the higher harmonics from the data. The conventional Fourier transform was used to determine the amplitudes of the first harmonics (ABP1 and CBFV1) of pressure and velocity data from each of the 8 regular heartbeats. The corresponding time means of the waveforms in each beat were also calculated (ABP0 and CBFV0). The inverse Fourier transform can be used to reconstruct the filtered waveform, as shown in Figure 1D. Since the higher harmonics are eliminated, the regression analysis will be done on near-perfect linear data. Mathematically, the intercept on the pressure axis will therefore be given by the following equation23: CCPf=ABP0-CBFV0xABP1/CBFV1.
The 2 highest and 2 lowest values of the 8 beats were discarded, and the remaining 4 intermediate values were averaged to determine the estimate of CCPf used for comparison with CCPd.
Processing of Data During Hyperemic Response After VF
A clear hyperemic response in CBFV was seen after ventricular defibrillation and heart restart in most of the recordings. However, in some recordings there was either no or only a weak response. Moreover, the response was sometimes masked by large changes, mostly increases, in the ABP level. The more detailed study of the hyperemic response was made with the use of the following selection criteria for recordings used: (1) an ABP difference of <10 mm Hg between the pre-VF and the hyperemic interval and (2) an increase in CBFV of >30%. Twelve episodes of VF from 6 patients met these criteria and were selected for further analysis. An interval of 8 heartbeats at the peak of the CBFV response was analyzed by the CCPf calculation method, as described above.
| Results |
|---|
|
|
|---|
The linear regression analysis of the regular beat data also showed a high degree of linear correlation between ABP and CBVF (r=0.978±0.010). The average time delay giving the highest r values was 0.041 second, and the maximum delay was 0.08 second. The CCPr was 26.9±11 mm Hg (range, 10 to 41 mm Hg). When compared with the CCPd (from long diastoles), the CCPr was 6.0±4.3 mm Hg lower. This difference was significant (P<0.05). The correlation with the gold standard CCPd was good (r=0.924); the results are plotted in Figure 2A.
|
The first harmonic Fourier analysisbased method resulted in a CCPf estimate of 31.5±10.9 mm Hg (range, 15 to 52 mm Hg). This was only 1.4±3.9 mm Hg lower than the gold standard CCPd. This difference was not statistically significant at the P<0.05 level. The correlation between CCPd and CCPf was good (r=0.935); the individual data are shown in Figure 2B.
A comparison between the pressureflow velocity relationships before VF (control state) and during peak hyperemia is shown in Figure 3. Flow velocity was normalized as 100% in the control period. During hyperemia, the pressure axis intercept (CCP) was shifted to the left by 13.3 mm Hg. Alone this shift would have resulted in an increase in CBFV to 131%, as illustrated by the dotted line and the circle in Figure 3. In addition, the slope of the relationship also increased so that the flow velocity at peak hyperemia was 163% of its pre-VF value. The mean ABP in hyperemia (80.6 mm Hg) was almost equal to the ABP in the control state (80.4 mm Hg).
|
| Discussion |
|---|
|
|
|---|
Vascular bed compliance has frequently been mentioned in the coronary hemodynamic literature as a factor that may cause errors in the estimates of the extrapolated zero-flow pressure.18,20 In the cerebral circulation, this factor is of less concern because the cerebral arterial bed is stiff, with the total compliancealso referred to as "windkessel"component of flow being 2% to 3% of mean flow during a heartbeat.25 This estimate was based on the rapid inflow filling of the windkessel after release of common carotid artery compression maneuvers. In the formula used for calculating CCPf, the second term contains the ratio of the first harmonic amplitude of ABP divided by the equivalent amplitude of flow velocity. If we assume that this ratio decreases by 2.5% (because of increase of the flow component into the windkessel), the CCPf estimate would increase by approximately 1.5 mm Hg in our series. An error of this magnitude seems acceptable. Moreover, this is probably a conservative estimate of the error since the excursions in pressure during normal heartbeats are smaller than those provoked by the compression maneuver used to assess the compliance component of flow.25
While the aforementioned windkessel effect refers to the entire cerebral (precapillary) vascular bed, we also need to consider elastic properties of the arterial wall and their effects on the caliber of the conducting artery that is insonated. The cross-sectional area would increase slightly with increased pressure. Velocity is given as flow divided by area. Therefore, any (relative) velocity increase would be slightly less than the (relative) flow increase. The ratio ABP1/CBFV1 would increase because of this effect, and the CCPf estimate would be lower. For a cross-sectional area variation of 5%, the underestimation of the CCP would be approximately 3 mm Hg. Again, this level of error should be tolerable. We shall also mention an interesting possibility: if the peripheral windkessel effect had the same numerical value as that from the elasticity of the insonated conductance artery, the errors due to these 2 phenomena would cancel out.
The concept of the CCP broadens our understanding of how flow in the cerebrovascular bed is regulated. The results summarized in Figure 3 show that, during the hyperemic phase, almost half (31%) of the increase in flow velocity can be explained by a reduction in the CCP. The remaining 32% increase can be attributed to the increase in slope. The slope can be defined as the inverse cerebrovascular resistance or the inverse of the resistance area product.11 Similar to the coronary circulation,19 flow in the cerebral vascular bed seems to be regulated by 2 factors: CCP (back pressure) and slope ("resistance proper"). We should, however, caution that in a normal physiological state the relative contribution of each factor could be different than for the present data.
The CCP levels reported in this study were much higher than the cerebral venous sinus pressures and the intracranial pressure. For obvious reasons, we had no direct access to the intracranial pressure, but there were no indications that it was abnormal. Moreover, the CCP fell markedly during hyperemia. The vasodilation in the post-VF phase would have caused a substantial increase in intracranial pressure if it had been elevated. Weyland et al17 had access to intracranial pressure recordings, and they showed that the CCP increased during vasoconstriction despite a decrease in intracranial pressure.
The CCP in the present series was also higher than previously reported.13,1517 This may possibly be explained by an increased cerebral vascular tone during the procedures. The low flow velocity (less than half the normal value, while the ABP was only slightly lower than normal) indicated vasoconstriction. Probably the vasoactive effects were caused mainly by the induced hypocapnia. Any effects of medication and/or anesthetic agents would have been overridden by the strong vasoconstrictive influence of an end-tidal PCO2 of 24 mm Hg. With all the interventions/medications necessary for this type of validation study, the values found should not be considered "normal physiological values" of CCP.
While the results of this report are encouraging with respect to methodological aspects, further studies are necessary to establish normal values. In such studies, if radial artery ABP recordings are used, filtering of the data by Fourier transform techniques should produce estimates of CCP that are closely correlated to the gold standard method of CCP determination with the use of prolonged diastoles. It would also be of interest to study whether CCP can be assessed with sufficient accuracy by entirely noninvasive means, without having to use invasive ABP measurements.
| Acknowledgments |
|---|
Received November 27, 2002; revision received January 21, 2003; accepted February 19, 2003.
| References |
|---|
|
|
|---|
2. Dewey RC, Pieper HP, Hunt WE. Experimental cerebral hemodynamics: vasomotor tone, critical closing pressure, and vascular bed resistance. J Neurosurg. 1974; 41: 597606.[Medline] [Order article via Infotrieve]
3. Burton AC. On the physical equilibrium of the small blood vessels. Am J Physiol. 1951; 164: 319329.
4. Bellamy RF. Diastolic coronary artery pressure-flow relations in the dog. Circ Res. 1978; 43: 92101.
5. Giller CA, Bowman G, Dyer H, Mootz L, Krippner W. Cerebral arterial diameters during changes in blood pressure and carbon dioxide during craniotomy. Neurosurgery. 1993; 32: 737741.[Medline] [Order article via Infotrieve]
6. Giller CA, Aaslid R. Estimates of pulse wave velocity and measurement of pulse transit time in the human cerebral circulation. Ultrasound Med Biol. 1994; 20: 101105.[CrossRef][Medline] [Order article via Infotrieve]
7. Lindegaard K-F, Lundar T, Wiberg J, Sjøberg D, Aaslid R, Nornes H. Variations in middle cerebral artery blood flow investigated with noninvasive transcranial blood velocity measurements. Stroke. 1987; 18: 10241030.
8. Newell DW, Aaslid R, Lam A, Mayberg TS, Winn HR. Comparison of flow and velocity during dynamic autoregulation testing in humans. Stroke. 1994; 25: 793797.[Abstract]
9. Aaslid R. Cerebral hemodynamics. In: Newell DW, Aaslid R, eds. Transcranial Doppler. New York, NY: Raven Press; 1992: 4955.
10. Panerai RB, Coughtrey H, Rennie JM, Evans DH. A model of the instantaneous pressure-velocity relationships of the neonatal cerebral circulation. Physiol Meas. 1993; 14: 411418.[CrossRef][Medline] [Order article via Infotrieve]
11. Panerai RB, Kelsall AW, Rennie JM, Evans DH. Estimation of critical closing pressure in the cerebral circulation of newborns. Neuropediatrics. 1995; 26: 168173.[Medline] [Order article via Infotrieve]
12. Michel E, Zernikow B, von Twickel J, Hillebrand S, Jorch G. Critical closing pressure in preterm neonates: towards a comprehensive model of cerebral autoregulation. Neurol Res. 1995; 17: 149155.[Medline] [Order article via Infotrieve]
13. Richards HK, Czosnyka M, Pickard JD. Assessment of critical closing pressure in the cerebral circulation as a measure of cerebrovascular tone. Acta Neurochir (Wien). 1999; 141: 12211227.[CrossRef]
14. Dawson SL, Panerai RB, Potter JF. Critical closing pressure explains cerebral hemodynamics during the Valsalva maneuver. J Appl Physiol. 1999; 86: 675680.
15. Panerai RB, Deverson ST, Mahony P, Hayes P, Evans DH. Effects of CO2 on dynamic cerebral autoregulation measurement. Physiol Meas. 1999; 20: 265275.[CrossRef][Medline] [Order article via Infotrieve]
16. Czosnyka M, Smielewski P, Piechnik S, Al-Rawi PG, Kirkpatrick PJ, Matta BF, Pickard JD. Critical closing pressure in cerebrovascular circulation. J Neurol Neurosurg Psychiatry. 1999; 66: 606611.
17. Weyland A, Buhre W, Grund S, Ludwig H, Kazmaier S, Weyland W, Sonntag H. Cerebrovascular tone rather than intracranial pressure determines the effective downstream pressure of the cerebral circulation in the absence of intracranial hypertension. J Neurosurg Anesthesiol. 2000; 12: 210216.[CrossRef][Medline] [Order article via Infotrieve]
18. Dole WP, Bishop VS. Influence of autoregulation and capacitance on diastolic coronary artery pressure-flow relationships in the dog. Circ Res. 1982; 51: 261270.
19. Dole WP, Richards KL, Hartley CJ, Alexander GM, Campbell AB, Bishop VS. Diastolic coronary artery pressure-flow velocity relationships in conscious man. Cardiovasc Res. 1984; 18: 548554.[Medline] [Order article via Infotrieve]
20. Hoffman JIE, Spaan JAE. Pressure-flow relations in the coronary circulation. Physiol Rev. 1990; 70: 331389.
21. Jin XY, Gibson DG, Pepper JR. The effects of cardioplegia on coronary pressure-flow velocity relationships during aortic valve replacement. Eur J Cardiothorac Surg. 1999; 16: 324330.
22. ORourke MF, Nichols WW. McDonalds Blood Flow in Arteries: Theoretical, Experimental and Clinical Principles. London, UK: Arnold Publishers; 1998: 187190.
23. Aaslid R, Lundar T, Lindegaard K-F, Nornes H. Estimation of cerebral perfusion pressure from arterial blood pressure and transcranial Doppler recordings. In: Miller JD, Teasdale GM, Rowan JO, Galbraith SL, Mendelow AD, eds. Intracranial Pressure VI. Berlin, Germany: Springer-Verlag; 1986: 226229.
24. Aaslid R, Brubakk AO. Accuracy of an ultrasound Doppler servo method for noninvasive determination of instantaneous and mean arterial blood pressure. Circulation. 1981; 64: 753759.
25. Aaslid R, Newell DW, Stooss R, Sorteberg W, Lindegaard K-F. Assessment of cerebral autoregulation dynamics from simultaneous arterial and venous transcranial Doppler recordings in humans. Stroke. 1991; 22: 11481154.
This article has been cited by other articles:
![]() |
M. Bundgaard-Nielsen, C. C. Jorgensen, T. B. Jorgensen, B. Ruhnau, N. H. Secher, and H. Kehlet Orthostatic intolerance and the cardiovascular response to early postoperative mobilization Br. J. Anaesth., June 1, 2009; 102(6): 756 - 762. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. J. McCulloch and M. J. Turner The Effects of Hypocapnia and the Cerebral Autoregulatory Response on Cerebrovascular Resistance and Apparent Zero Flow Pressure During Isoflurane Anesthesia Anesth. Analg., April 1, 2009; 108(4): 1284 - 1290. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. S. Darwish, E. Ahn, and N. S. Amiridze Role of Transcranial Doppler in Optimizing Treatment of Cerebral Vasospasm in Subarachnoid Hemorrhage J Intensive Care Med, July 1, 2008; 23(4): 263 - 267. [Abstract] [PDF] |
||||
![]() |
R. Aaslid, M. Blaha, G. Sviri, C. M. Douville, and D. W. Newell Asymmetric Dynamic Cerebral Autoregulatory Response to Cyclic Stimuli Stroke, May 1, 2007; 38(5): 1465 - 1469. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. A. H. R. Claassen, R. Zhang, Q. Fu, S. Witkowski, and B. D. Levine Transcranial Doppler estimation of cerebral blood flow and cerebrovascular conductance during modified rebreathing J Appl Physiol, March 1, 2007; 102(3): 870 - 877. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. B. Panerai, M. Moody, P. J. Eames, and J. F. Potter Cerebral blood flow velocity during mental activation: interpretation with different models of the passive pressure-velocity relationship J Appl Physiol, December 1, 2005; 99(6): 2352 - 2362. [Abstract] [Full Text] [PDF] |
||||
![]() |
A Kumar, E A Schmidt, M Hiler, P Smielewski, J D Pickard, and M Czosnyka Asymmetry of critical closing pressure following head injury J. Neurol. Neurosurg. Psychiatry, November 1, 2005; 76(11): 1570 - 1573. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Moody, R. B. Panerai, P. J. Eames, and J. F. Potter Cerebral and systemic hemodynamic changes during cognitive and motor activation paradigms Am J Physiol Regulatory Integrative Comp Physiol, June 1, 2005; 288(6): R1581 - R1588. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. D. Marval, M. E. Perrin, S. M. Hancock, and R. P. Mahajan The Effects of Propofol or Sevoflurane on the Estimated Cerebral Perfusion Pressure and Zero Flow Pressure Anesth. Analg., March 1, 2005; 100(3): 835 - 840. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Soehle, M. Czosnyka, J. D. Pickard, and P. J. Kirkpatrick Critical Closing Pressure in Subarachnoid Hemorrhage: Effect of Cerebral Vasospasm and Limitations of a Transcranial Doppler-Derived Estimation Stroke, June 1, 2004; 35(6): 1393 - 1398. [Abstract] [Full Text] [PDF] |
||||
![]() |
H.-Y. Hsu, Y.-T. Chen, H.-H. Hu, R. Aaslid, and D. W. Newell Limitations in Estimating Critical Closing Pressure by Noninvasive Blood Pressure Measurements * Response: Limitations in Estimating Critical Closing Pressure by Noninvasive Blood Pressure Measurements Stroke, May 1, 2004; 35(5): e91 - e92. [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Stroke Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2003 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |