(Stroke. 2003;34:1936.)
© 2003 American Heart Association, Inc.
Original Contributions |
From the Interventional Cardio-Angiology Unit, Villa Maria Cecilia Hospital, Cotignola (A.C., R.M., F.C.), and Department of General and Vascular Surgery, University of Siena, Siena (F.S., C.S.), Italy.
Correspondence to Alberto Cremonesi, MD, Interventional Cardio-Angiology Unit, Department of Medical and Surgical Cardiology, Villa Maria Cecilia Hospital, Via Corriera 1, 48010 Cotignola (RA), Italy.
| Abstract |
|---|
|
|
|---|
Methods From 1999 to June 2002, a total of 442 consecutive patients underwent percutaneous angioplasty and/or stenting of the extracranial carotid artery. The endovascular procedure was conducted under embolic protection devices.
Results The percutaneous procedure was successful in 440 of 442 patients (99.5%). No periprocedural death occurred with any embolic protection device. All in-hospital stroke/death and 30-day ipsilateral stroke/death rate was 1.1%. The overall complication rate was 3.4%. Major adverse events included 1 major stroke (0.2%), 4 intracranial hemorrhages (0.9%), 1 carotid artery wall fissuration (0.2%), and 1 diffuse cardioembolism (0.2%). Minor adverse events included 4 minor strokes (0.9%) and 4 transient ischemic attacks (0.9%). The cerebral protection devicerelated complications were 4 (0.9%): 1 case of abrupt closure of the internal carotid artery because of spiral dissection (0.2%), 1 case of trapped guide wire (0.2%), and 2 cases of intimal dissection (0.5%). Transient loss of consciousness, tremors, and fasciculations were present in 6 of 40 patients (15%) in whom occlusive protection devices were used.
Conclusions Our data suggest that percutaneous stenting of the carotid artery when a cerebral protection device is used is feasible and effective but not without potential complications. However, a long learning curve may exist for the proper use of some embolic protection devices.
Key Words: carotid arteries protective devices stents
| Introduction |
|---|
|
|
|---|
At present, distal cerebral microembolization is demonstrated by transcranial Doppler monitoring and diffusion-weighted MRI; both techniques have revealed significant and increased dislocation of microemboli during carotid angioplasty and stenting. The microemboli are greater in percutaneous treatment compared with a surgical approach.79
In the last 3 years, numerous cerebral protection systems have been proposed to limit the acute neurological complications related to distal embolization. The aim of this study was to investigate the feasibility, safety, and efficacy of different endovascular protection systems during elective stenting of the extracranial carotid arteries.
See Editorial Comment, page 1941
| Subjects and Methods |
|---|
|
|
|---|
|
|
Patient Inclusion Criteria
Inclusion criteria included the following: age >50 years; monolateral or bilateral carotid critical stenosis (>75% carotid lesion); symptomatic for the culprit carotid lesion; asymptomatic for the culprit carotid lesion but with a positive cerebral CT scan for silent cerebral focal ischemia ipsilateral to the culprit carotid lesion; asymptomatic for the culprit carotid lesion with a negative cerebral CT scan for cerebral focal ischemia but with carotid echo-Doppler findings demonstrating a severe complex stenosis with fast progression of carotid atherosclerotic disease; and refusal by surgeons as high-risk surgical subsets.
Patient Exclusion Criteria
Exclusion criteria included thrombocytopenia, leucopenia, neutropenia, or gastrointestinal bleeding in the previous 3 months; allergy to aspirin, clopidogrel, ticlopidine; angiographic appearance of fresh thrombus at the carotid lesion site; and angiographic appearance of carotid chronic total occlusion or long preocclusive lesion ("string sign" lesion).
Patient Assessment
Before treatment, all patients underwent careful neurological examination performed by an independent board-certified neurologist (including National Institutes of Health Stroke Scale [NIHSS]), echo/color-flow Doppler (lesion site and intracranial cerebral blood flow assessment), cerebral CT scan, and angiographic evaluation.
Within 24 hours after the procedure and at the 30-day follow-up, all patients underwent another neurological examination performed by the same independent board-certified neurologist and a complete echo/color-flow Doppler evaluation. A postprocedure cerebral CT scan was performed only in patients with documented neurological complications.
Preprocedure Echo Analysis of Carotid Plaques
Plaque echogenicity as assessed by B-mode ultrasound has been found to reliably predict the content of soft tissue and the amount of calcification in carotid plaques.1012 Plaque morphology in terms of echogenicity, defined as reflectance of the emitted ultrasound signal, was assessed in a modified version of the classification proposed by Gray-Weale et al10 and graded from 1 to 4 as echolucent, predominantly echolucent, predominantly echogenic, or echogenic.13,14 Echo plaque characteristics and complexity were checked in all treated patients and are presented in Table 3.
|
A total of 263 patients (59.5%) showed plaque echo analysis patterns compatible for high embolization risk (echolucent plaques, n=171; predominantly echolucent plaques, n=92); Additionally, 287 patients (62.9%) showed echo patterns of plaque complexity (heavy calcifications, n=92; plaque erosion/ulceration, n=186).
Definitions
Procedural success of protected carotid stent deployment was defined as quantitative coronary angiography showing <30% residual diameter stenosis of all treated lesions without alterations in the intracranial circulation at the postprocedural angiographic examination (residual diameter stenosis was assessed by averaging at least 2 matched views on quantitative angiography) and echo/color-flow Doppler showing the absence of significant residual stenosis and pathological acceleration in blood flow (<1.5 m/s). Transient ischemic attack (TIA) was defined as a focal retinal or hemispheric event from which the patient made a complete recovery within 24 hours. Minor stroke was defined as a new neurological deficit that either resolved completely within 30 days or increased the NIHSS by
3. Major stroke was defined as a new neurological deficit that persist for >30 days and increased the NIHSS by
4. Finally, fatal stroke was defined as death attributed to an ischemic stroke or intracerebral hemorrhagic stroke.
Medical Treatment
Before the Procedure
All patients were treated with acetylsalicylic acid at a mean dosage of 125 mg/d and ticlopidine at a mean dosage of 500 mg/d at least 4 to 5 days before admission.
During the Procedure
The mean dosage of sodium heparin used during the procedure was 100 U/kg. A mean of 1 mg atropine was given to patients before balloon inflation.
Postprocedure
Ticlopidine (500 mg/d) was continued for at least 30 days after the interventional procedure (hemochrome and white blood count were checked 7 to 10 days after the percutaneous intervention). Antiplatelet therapy with aspirin is being continued indefinitely.
Description of the Procedure
All procedures were carried out via puncture of the right and/or left femoral artery. The vascular approach did not change, depending on the use of different types of cerebral protection devices. The common carotid artery was selectively engaged directly by use of a primary guiding catheter. When use of a primary guiding catheter was not possible because of the particular anatomy of supra-aortic vessels, we placed a stiff wire into the external carotid artery for positioning of a long sheath or a guiding catheter into the common carotid artery.
Once the common carotid artery was engaged, all patients underwent an angiographic examination of the culprit carotid lesion in 2 different projections and an angiographic examination of the intracranial circulation in the anteroposterior and/or lateral projection. The same angiographic checkup was performed at the end of the procedure to determine whether there was any variation in the intracranial blood flow.
Cerebral Protection Devices
We used Johnson & Johnson-Cordis Angioguard Filter (167, 37.8%), Boston Scientific FilterWire EX (111, 25.1%), EV3 Microvena Trap Filter (83, 18.8%), and Mednova NeuroShield (41, 9.3%) distal filters. We also used Medtronic Percusurge (25, 5.7%) and Medicorp (1, 0.2%) distal occlusive balloons.
Proximal Endovascular Clamping Devices
For proximal endovascular clamping devices, we used the Arteria Parodi reversal flow system (7, 1.6%) and the Invatec MO.MA proximal occlusive system (7, 1.6%).
Carotid Stenting
Carotid stenting was carried out by use of self-expandable stents in 339 cases and balloon-expandable stents in 3 cases. The self-expandable stents were Boston Scientific Carotid Wallstent (378, 85.5%) and Easy Wallstent (4, 0.9%), Medicorp Expander (4, 0.9%), Johnson & Johnson-Cordis Smart (15, 3.4%), Guidant Acculink (35, 7.9%), and Medtronic AVE (9, 2.0%). The balloon-expandable stents were the Johnson & Johnson-Cordis Palmaz-Schatz (2, 0.5%) and Guidant Syncro (1, 0.2%) stents.
Predilation was performed with coronary balloons in tight or subocclusive carotid stenoses (167, 33.7%). Separate wires (buddy wire technique) were used in 91 patients (20.6%). The predilation balloons were routinely undersized (artery/balloon ratio: 1.8 to 1.5) to reduce vessel dissection and/or distal embolization.
Stent placement was optimized through multiple dilations by using suitably sized balloons based on quantitative analysis of the vessel. In 1 case, direct stenting with a nitinol self-expandable stent was performed without postdilation. During the poststenting dilation phase, atropine (mean dosage, 1 mg IV) was given to all patients before balloon inflation to reduce the bradycardia and hypotension potentially associated with carotid dilation.
At the end of the procedure, the arterial introducer was immediately removed from most patients, and hemostasis of the femoral artery was achieved by use of the St Jude Medical Angio-Seal hemostatic device.
| Results |
|---|
|
|
|---|
In-Hospital and 30-Day Adverse Events in the Study Group
All in-hospital stroke/death and 30-day ipsilateral stroke/death rate was 1.1%. The overall complication rate was 3.4% (Tables 4 and 5
). In summary, there were 7 major adverse events: 1 major stroke (0.2%), 4 intracranial hemorrhages (0.9%), 1 carotid artery wall fissuration (0.2%), and 1 diffuse cardioembolism (0.2%). Minor adverse events included 4 minor strokes (0.9%) and 4 TIAs (0.9%). No periprocedural death occurred in the study group.
|
|
The cases complicated by intracranial hemorrhage achieved complete resolution of symptoms (hemisensory loss, hemiplegia, and epileptic seizures) and radiological findings (CT and MR) within 30 days.
One case was complicated by carotid artery fissuration (very small and limited carotid artery wall rupture) with a periarterial extravasation immediately after stent postdilatation. This acute complication was successfully treated by sealing the wall rupture with long and repeated low-pressure balloon inflations inside the carotid stent. The patient left the laboratory asymptomatic, without angiographic patterns of residual significant periarterial extravasation.
One case was complicated by an unusual peripheral embolic event temporarily involving 3 different arterial sites (left common carotid artery, left brachial artery, right common femoral artery) occurring 12 hours after carotid stenting. This event, reported as diffuse cardioembolism, was related to the detachment of emboli from left atrium in an elderly patients with chronic atrial fibrillation determined by the full anticoagulation needed for the endovascular procedure. The patient was successfully treated via surgical multivascular approach.
In-Hospital and 30-Day Adverse Events in Symptomatic Patient Subset
All in-hospital stroke/death and 30-day ipsilateral stroke/death rate was 1.2%. The overall complication rate was 3.2% (Table 5). The major adverse events were as follows: 1 major stroke (0.4%), 3 intracranial hemorrhages (1.2%), and 1 diffuse cardioembolism (0.4%). There were 3 minor adverse events: 2 minor strokes (0.8%) and 1 TIA (0.4%).
In-Hospital and 30-Day Adverse Events in Asymptomatic Patient Subset
All in-hospital stroke/death and 30-day ipsilateral stroke/death rate was 1.1%. The overall complication rate was 3.7% (Table 5). The 2 major adverse events included 1 intracranial hemorrhage (0.5%) and 1 carotid wall fissuration (0.5%). Minor adverse events included 2 minor strokes (1.1%) and 3 TIAs (1.6%).
Temporal Distribution of Embolic Complications
Timing of any clinical appearance of neurological symptoms (major stroke, minor stroke, TIA) related to cerebral embolization was measured in all 442 study patients. Acute embolic complications were defined as neurological events occurring during the endovascular procedure or a few minutes after retrieval of the protection device. Late embolic complications were defined as neurological events occurring after the endovascular procedure completion, during the patients hospital stay and in the 30 following days. Temporal distribution analysis showed the following data.
Complications Related to the Use of EPDs
There were 4 (0.9%) EPD-related complications: dissection treated with an additional stent (2, 0.5%), vessel occlusion by spiral dissection (1, 0.2%), and "trapped" guide wire (1, 0.2%).
EPD Major Complications
Two major adverse events were reported, both asymptomatic. The first was a case of spiral dissection of the left internal carotid artery distal to the stenosis site as a result of Medtronic PercuSurge occlusive balloon vessel injury. This injury was attributed to diffuse artery disease also involving the distal portion of the internal carotid artery. Although we were unable to cross the spiral dissection and the intracranial left internal carotid artery was completely occluded, the patient did not complain of any symptoms. Neurological examination did not show any neurological deficit in either the Catheterization Laboratory or the ward. CT and MRI evaluation excluded the presence of any lesion. This behavior resulted because the hemisphere supplied by the occluded carotid artery received sufficient blood flow from the contralateral carotid circulation and intracranial posterior circulation.
The second event occurred during the endovascular treatment of a subocclusive ostial stenosis of the left common carotid artery. This procedure, because of the complete occlusion of the contralateral internal carotid artery, was protected by the use of a 7-mm Johnson & Johnson-Cordis AngioGuard system. At the end of stent delivery (Johnson & Johnson-Cordis Palmaz-Schatz P205 manually crimped on 7/20-mm balloon), the 0.014 wire of the AngioGuard system remained trapped in the proximal edge of the stent. This probably occurred because of the significant acute angle between the aortic arch and the takeoff of the internal carotid artery or the high-pressure stent dilation carried out with a balloon on a second stiff wire placed parallel to the AngioGuard wire. Because we were unable to recapture the nitinol basket and its potential embolic content, we referred the patient to the vascular surgical team to remove the system. The protection device was easily retrieved via surgical cutdown of the distal portion the common carotid artery. The patient was asymptomatic, and neurological examination did not show any neurological deficit.
EPD Minor Complications
Two cases of tear dissection of the right internal carotid artery distal to the stenosis site as a result of the Medtronic PercuSurge occlusive balloon vessel injury occurred. The localized wall dissection did not cause flow impairment and was easily treated by implanting an adjunctive self-expandable Boston Scientific Easy Wallstent above the first one to cover the intimal flap.
EPD Side Effects
Transient spasm of the intracranial segment of the carotid artery, presumably related to the protection device (distal occlusive balloon, filter), was observed in 35 cases (7.9%). In all cases, spasm was resolved after intracarotid nitroglycerin administration.
Carotid flow impairment resistant to nitroglycerin administration occurred in 58 cases (13.1%) protected by filter devices. All patients remained asymptomatic during the episode. Flow was completely restored after filter removal. A large amount of macroscopically visible debris was noticed inside the filters and considered to be responsible for the significant flow impairment.
Transient loss of consciousness, tremors, and fasciculations were present during 6 of 40 carotid procedures protected by either distal or proximal occlusive systems (PercuSurge, Parodi, MO.MA). In all cases, the neurological symptoms appeared after a mean cerebral occlusion time of 6±2.5 minutes. Immediately after deflation of the occlusive system, all patients returned to baseline neurological conditions.
Distal Protection Debris
In 304 patients (69.0%), macroscopically visible plaque debris was captured and retrieved. Plaque debris ranged from 75 to 6000 µm in diameter. In 17.7% of patients with visible particles (54 cases), the embolic debris was
2 mm in diameter.
Histopathological analysis was performed on all filters with macroscopically visible plaque debris with light microscopy after staining with hematoxylin-eosin and Azan-Mallory. The debris consisted of cholesterol crystals, fibrin material, atheromatous plaque, and macrophage foam cells.
| Discussion |
|---|
|
|
|---|
Comparing our reported data with recently published literature6,13,1517 shows that the use of cerebral protection systems can be considered positive in terms of high procedural success, low device-related complications, and low in-hospital/30-day complications.
No periprocedural death occurred in this neuroprotected group, with a composite all stroke/death rate of 1.1% confirmed at 30-day follow-up. The overall complication rate did not exceed 3.4%.
When we compare the overall complication rate in symptomatic and asymptomatic patients, no significant difference was demonstrated (symptomatic, 3.2%; asymptomatic, 3.7%). The all stroke/death analysis gave the same result (symptomatic, 1.1%; asymptomatic, 1.2%).
When comparing all major and minor complication analysis in symptomatic and asymptomatic patients, we found more frequent major events (major stroke and intracranial hemorrhage) in the symptomatic subgroup than in asymptomatic group (2.0% versus 1.1%); conversely, minor events (minor stroke, TIA) were more frequent in the asymptomatic subgroup than in the symptomatic subgroup (2.6% versus 1.2%).
Analysis of temporal distribution of embolic complications (timing of any clinical appearance of neurological symptoms: major stroke, minor stroke, TIA related to cerebral embolization) demonstrated that neuroprotection devices allowed us to protect the procedure but did not affect late embolic events. We reported in the present series 5 acute embolic events (1.2%) and 4 late ones (0.9%).
Intracranial hemorrhage was encountered in 4 patients (0.9%) and was the most dangerous nonembolic complication reported in this series. In 3 cases, the event happened in very symptomatic and hypertensive patients with subocclusive carotid lesions and was attributed to cerebral reperfusion syndrome. In 1 case, the intracranial hemorrhage complicated a quite simple procedure in an asymptomatic patient and was not attributed to any specific clinical variable but to the full heparinization during the procedure and the prior acetylsalicylic acid and ticlopidine regimen.
The use of protection devices was not free of complications; at least 2 potentially dangerous situations (0.5%) related to the use of 2 very different protection devices occurred. Although both complications were managed without negative results to the patientsthey did not complain of any symptoms and did not show any neurological deficitwe are aware of the objective danger these situations might involve.
The present study was not designed either to compare various protection systems or to detect differences between balloon occlusive devices, filter devices, and reversal flow devices. The primary end point was to evaluate the clinical impact of protection devices during carotid endovascular interventions. From our experience, we can state that occlusive balloon devices and reversal flow devices have a favorable crossing profile and do not require lesion crossing. However, balloon inflation may cause transient cerebral ischemia, which is not always predictable or well tolerated by all patients.
Transient loss of consciousness, tremors, and fasciculations were present in 6 cases (1.3% of the entire protected group, 15% of the occlusive system subset) of carotid procedures protected by either distal or proximal occlusive systems. In contrast, filter devices, although they have higher crossing profiles and more stiffness, maintain antegrade cerebral flow and allow visualization, which may result in more precise stent placement. In the present series, filter devices were well tolerated by all patients.
Study Limitations
The patients did not undergo transcranial Doppler monitoring during the endovascular carotid procedure and did not receive a postprocedure CT scan or MRI examination. For this reason, we are not able to provide any objective information about the degree of embolization occurring while crossing the lesion with wires, balloons, stents, and protection devices; the comparison between the 2 groups in terms of degree of embolization; and the efficiency of the protection systems in capturing all the particles produced during carotid angioplasty and stenting.
Conclusions
Our experience demonstrates that the extensive use of cerebral protection devices is feasible and effective in preventing distal embolization during carotid endovascular procedures. Despite the fact that the immediate periprocedural and in-hospital results are encouraging, we are aware that EPDs allowed operators to protect the procedure but unfortunately did not affect late embolic events. Moreover, our experience suggests that the correct use of all protection systems requires a proper learning curve and that the use of protection devices is not currently complication free. Cerebral protection will most likely decrease the complication rate of carotid stent procedures further, possibly turning this intervention into the therapy of choice for patients in whom carotid artery angioplasty and stenting is indicated.
Received February 4, 2003; revision received March 25, 2003; accepted April 4, 2003.
| References |
|---|
|
|
|---|
2. Roubin GS, New G, Iyer SS, Vitek JJ, Al-Mubarak N, Liu MW, Yadav J, Gomez C, Kuntz RE. Immediate and late clinical outcomes of carotid artery stenting in patients with symptomatic and asymptomatic carotid artery stenosis: a 5-year prospective analysis. Circulation. 2001; 103: 532537.
3. Cremonesi A, Castriota F, Manetti R, Balestra G, Liso A. Endovascular treatment of carotid atherosclerotic disease: early and late outcome in a non-selected population. Ital Heart J. 2000; 1: 801809.[Medline] [Order article via Infotrieve]
4. Mathur A, Roubin GS, Iyer SS, Piamsonboon C, Liu MW, Gomez CR, Yadav JS, Chastain HD, Fox LM, Dean LS, Vitek JJ. Predictors of stroke complicating carotid artery stenting. Circulation. 1998; 97: 12391245.
5. Topol EJ, Yadav JS. Recognition of the importance of embolization in atherosclerotic vascular disease. Circulation. 2000; 101: 570580.
6. Reimers B, Corvaja N, Moshiri S, Saccà S, et al. Cerebral protection with filter devices during carotid artery stenting. Circulation. 2001; 104: 1215.
7. Jordan WD Jr, Voellinger DC, Doblar DD, Plyushcheva NP, Fisher WS, McDowell HA. Microemboli detected by transcranial Doppler monitoring in patients during carotid angioplasty versus carotid endarterectomy. Cardiovasc Surg. 1999; 7: 3338.[CrossRef][Medline] [Order article via Infotrieve]
8. Ackerstaff RG, Moons KG, van de Vlasakker CJ, Moll FL, Vermeulen FE, Algra A, Spencer MP. Association of intraoperative transcranial Doppler monitoring variables with stroke from carotid endarterectomy. Stroke. 2000; 31: 18171823.
9. Jaeger HJ, Mathias KD, Hauth E, Dresher R, Gissler HM, Hennings S, Christman A. Cerebral ischemia detected with diffusion-weighted MR imaging after stent implantation in the carotid artery. AJNR Am J Neuroradiol. 2002; 23: 200207.
10. Gray-Weale AC, Graham JC, Burnett JR, et al. Carotid artery atheroma: comparison of preoperative B-mode ultrasound appearance with carotid endarterectomy specimen pathology. J Cardiovasc Surg. 1988; 29: 676681.[Medline] [Order article via Infotrieve]
11. Mathiesen EB, Bonaa KH, Joakimsen O. Low levels of high-density lipoprotein cholesterol are associated with echolucent carotid artery plaques: the Tromsø Study. Stroke. 2001; 32: 19601965.
12. Joakimsen O, Bønaa KH, Stensland-Bugge E. Reproducibility of ultrasound assessment of carotid plaque occurrence, thickness, and morphology: the Tromsø Study. Stroke. 1997; 28: 22012207.
13. Castriota F, Cremonesi A, Manetti R, Liso A, Oshoala K, Ricci E, Balestra G. Impact of cerebral protection devices on early outcome of carotid stenting. J Endovasc Ther. 2002; 9: 786792.[CrossRef][Medline] [Order article via Infotrieve]
14. North American Symptomatic Carotid Endarterectomy Trial Collaborators. Beneficial effect of carotid endarterectomy in symptomatic patients with high-grade carotid stenosis. N Engl J Med. 1991; 325: 445453.[Abstract]
15. Albuquerque FC, Teitelbaum GP, Lavine SD, Larsen DW, Giannotta SL. Balloon-protected carotid angioplasty. Neurosurgery. 2000; 46: 918921.[CrossRef][Medline] [Order article via Infotrieve]
16. Ohki T, Parodi J, Veith FJ, Bates M, Bade M, Chang D, Mehta M, Rabin J, Goldstein K, Harvey J, Lipsitz E. Efficacy of a proximal occlusion catheter with reversal of flow in the prevention of embolic events during carotid artery stenting: an experimental analysis. J Vasc Surg. 2001; 33: 504509.[CrossRef][Medline] [Order article via Infotrieve]
17. Ohki T, Veith FJ. Carotid artery stenting: utility of cerebral protection devices. J Invasive Cardiol. 2001; 13: 4755.[Medline] [Order article via Infotrieve]
18. European Carotid Surgery Trialists Collaborative Group. MRC European Carotid Surgery Trial: interim results for symptomatic patients with severe (7099%) or with mild (029%) carotid stenosis. Lancet. 1991; 337: 12351243.[CrossRef][Medline] [Order article via Infotrieve]
19. Asymptomatic Carotid Atherosclerosis Study. Clinical advisory: carotid endarterectomy for patients with asymptomatic internal carotid artery stenosis. Stroke. 1994; 25: 25232524.[Abstract]
This article has been cited by other articles:
![]() |
M. Roffi, D. Mukherjee, and D. G. Clair Carotid artery stenting vs. endarterectomy Eur. Heart J., November 2, 2009; 30(22): 2693 - 2704. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. Jansen, J. Fiehler, M. Hartmann, and H. Bruckmann Protection or Nonprotection in Carotid Stent Angioplasty: The Influence of Interventional Techniques on Outcome Data From the SPACE Trial Stroke, March 1, 2009; 40(3): 841 - 846. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Perona, G. Castellazzi, L. Valvassori, E. Boccardi, L. de Girolamo, G. P. Cornalba, and K. Kandarpa Safety of Unprotected Carotid Artery Stent Placement in Symptomatic and Asymptomatic Patients: A Retrospective Analysis of 30-day Combined Adverse Outcomes Radiology, January 1, 2009; 250(1): 178 - 183. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Verzini, P. De Rango, G. Parlani, G. Panuccio, and P. Cao Carotid Artery Stenting: Technical Issues and Role of Operators' Experience Perspectives in Vascular Surgery and Endovascular Therapy, September 1, 2008; 20(3): 247 - 257. [Abstract] [PDF] |
||||
![]() |
M. Bosiers, K. Deloose, J. Verbist, and P. Peeters The Impact of Embolic Protection Device and Stent Design on the Outcome of CAS Perspectives in Vascular Surgery and Endovascular Therapy, September 1, 2008; 20(3): 272 - 279. [Abstract] [PDF] |
||||
![]() |
M. K. Eskandari Preventable Complications of Carotid Stenting Perspectives in Vascular Surgery and Endovascular Therapy, March 1, 2008; 20(1): 17 - 25. [Abstract] [PDF] |
||||
![]() |
M. Schillinger, M. Gschwendtner, B. Reimers, J. Trenkler, L. Stockx, J. Mair, S. Macdonald, F. Karnel, K. Huber, and E. Minar Does Carotid Stent Cell Design Matter? Stroke, March 1, 2008; 39(3): 905 - 909. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Maynar, S. Baldi, R. Rostagno, T. Zander, M. Rabellino, R. Llorens, J. Alvarez, and F. Barajas Carotid Stenting without Use of Balloon Angioplasty and Distal Protection Devices: Preliminary Experience in 100 Cases AJNR Am. J. Neuroradiol., August 1, 2007; 28(7): 1378 - 1383. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Zahn, T. Ischinger, M. Hochadel, U. Zeymer, W. Schmalz, N. Treese, K. E. Hauptmann, H. Seggewiss, I. Janicke, H. Haase, et al. Carotid artery stenting in octogenarians: results from the ALKK Carotid Artery Stent (CAS) Registry Eur. Heart J., February 1, 2007; 28(3): 370 - 375. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. F. Bonvini, M. Righini, M. Hamon, J. W. Riddell, A. O. Maree, K. A. Rosenfield, J.-L. Mas, G. Chatellier, and the EVA-3S Investigators Endarterectomy versus Stenting for Carotid Stenosis N. Engl. J. Med., January 18, 2007; 356(3): 305 - 307. [Full Text] [PDF] |
||||
![]() |
G. Maleux, P. Demaerel, E. Verbeken, K. Daenens, S. Heye, F. Van Sonhoven, A. Nevelsteen, and G. Wilms Cerebral ischemia after filter-protected carotid artery stenting is common and cannot be predicted by the presence of substantial amount of debris captured by the filter device. AJNR Am. J. Neuroradiol., October 1, 2006; 27(9): 1830 - 1833. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Hofmann, A. Niessner, A. Kypta, C. Steinwender, J. Kammler, K. Kerschner, M. Grund, F. Leisch, and K. Huber Risk Score for Peri-Interventional Complications of Carotid Artery Stenting Stroke, October 1, 2006; 37(10): 2557 - 2561. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Cremonesi, C. Setacci, A. Bignamini, L. Bolognese, F. Briganti, G. Di Sciascio, D. Inzitari, G. Lanza, L. Lupattelli, S. Mangiafico, et al. Carotid Artery Stenting: First Consensus Document of the ICCS-SPREAD Joint Committee Stroke, September 1, 2006; 37(9): 2400 - 2409. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. du Mesnil de Rochemont, S. Schneider, B. Yan, A. Lehr, M. Sitzer, and J. Berkefeld Diffusion-Weighted MR Imaging Lesions after Filter-Protected Stenting of High-Grade Symptomatic Carotid Artery Stenoses AJNR Am. J. Neuroradiol., June 1, 2006; 27(6): 1321 - 1325. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Shilling, B. F. Uretsky, and G. C. Hunter Entrapment of a Cerebral Embolic Protection Device: A Case Report Vascular and Endovascular Surgery, May 1, 2006; 40(3): 229 - 233. [Abstract] [PDF] |
||||
![]() |
G. S. Roubin, S. Iyer, A. Halkin, J. Vitek, and C. Brennan Realizing the Potential of Carotid Artery Stenting: Proposed Paradigms for Patient Selection and Procedural Technique Circulation, April 25, 2006; 113(16): 2021 - 2030. [Abstract] [Full Text] [PDF] |
||||
![]() |
M.D. Hill, W. Morrish, G. Soulez, A. Nevelsteen, G. Maleux, C. Rogers, K.E. Hauptmann, A. Bonafe, R. Beyar, L. Gruberg, et al. Multicenter evaluation of a self-expanding carotid stent system with distal protection in the treatment of carotid stenosis. AJNR Am. J. Neuroradiol., April 1, 2006; 27(4): 759 - 765. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. C. Lin, R. R. Kolvenbach, and L. Pinter Protected Carotid Artery Stenting and Angioplasty via Transfemoral versus Transcervical Approaches Vascular and Endovascular Surgery, November 1, 2005; 39(6): 499 - 503. [Abstract] [PDF] |
||||
![]() |
Y. P. Liew and J. R Bartholomew Atheromatous embolization Vascular Medicine, November 1, 2005; 10(4): 309 - 326. [Abstract] [PDF] |
||||
![]() |
E. I. Bluth, J. A. Vos, and R. G. A. Ackerstaff Value of US in Selecting Patients for Carotid Angioplasty and Stent Placement * Drs Vos and Ackerstaff respond: Radiology, October 1, 2005; 237(1): 374 - 375. [Full Text] [PDF] |
||||
![]() |
K. I. Paraskevas, S. S. Daskalopoulou, M. E. Daskalopoulos, and C. D. Liapis Secondary Prevention of Ischemic Cerebrovascular Disease. What Is the Evidence? Angiology, September 1, 2005; 56(5): 539 - 552. [Abstract] [PDF] |
||||
![]() |
G. Orlandi, S. Fanucchi, S. Gallerini, C. Sonnoli, M. Cosottini, M. Puglioli, F. Sartucci, and L. Murri Impaired Clearance of Microemboli and Cerebrovascular Symptoms During Carotid Stenting Procedures Arch Neurol, August 1, 2005; 62(8): 1208 - 1211. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Zahn, T. Ischinger, B. Mark, S. Gass, U. Zeymer, W. Schmalz, K. Haerten, K. E. Hauptmann, E.-R. von Leitner, W. Kasper, et al. Embolic Protection Devices for Carotid Artery Stenting: Is There a Difference Between Filter and Distal Occlusive Devices? J. Am. Coll. Cardiol., June 7, 2005; 45(11): 1769 - 1774. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. P. Lownie, D. M. Pelz, D. H. Lee, S. Men, I. Gulka, and P. Kalapos Efficacy of Treatment of Severe Carotid Bifurcation Stenosis By Using Self-Expanding Stents without Deliberate Use of Angioplasty Balloons AJNR Am. J. Neuroradiol., May 1, 2005; 26(5): 1241 - 1248. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. J. Coward, R. L. Featherstone, and M. M. Brown Safety and Efficacy of Endovascular Treatment of Carotid Artery Stenosis Compared With Carotid Endarterectomy: A Cochrane Systematic Review of the Randomized Evidence Stroke, April 1, 2005; 36(4): 905 - 911. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. A. Vos, J. C. van den Berg, S. M. P. G. Ernst, M. J. Suttorp, T. T. C. Overtoom, H. W. Mauser, O. J. M. Vogels, H. P. M. van Heesewijk, F. L. Moll, Y. van der Graaf, et al. Carotid Angioplasty and Stent Placement: Comparison of Transcranial Doppler US Data and Clinical Outcome with and without Filtering Cerebral Protection Devices in 509 Patients Radiology, February 1, 2005; 234(2): 493 - 499. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. K. Sztriha, E. Voros, K. Sas, R. Szentgyorgyi, A. Pocsik, P. Barzo, P. Szikra, A. Makai, A. Szolics, P. Elek, et al. Favorable Early Outcome of Carotid Artery Stenting Without Protection Devices Stroke, December 1, 2004; 35(12): 2862 - 2866. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Schmidt, K.-W. Diederich, S. Scheinert, S. Braunlich, T. Olenburger, G. Biamino, G. Schuler, and D. Scheinert Effect of two different neuroprotection systems on microembolization during carotid artery stenting J. Am. Coll. Cardiol., November 16, 2004; 44(10): 1966 - 1969. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Zahn, B. Mark, N. Niedermaier, U. Zeymer, P. Limbourg, T. Ischinger, K. Haerten, K. E. Hauptmann, E.-R. v. Leitner, W. Kasper, et al. Embolic protection devices for carotid artery stenting: better results than stenting without protection? Eur. Heart J., September 1, 2004; 25(17): 1550 - 1558. [Abstract] [Full Text] |
||||
![]() |
M. A. Bettmann, M. D. Dake, L. N. Hopkins, B. T. Katzen, C. J. White, A. C. Eisenhauer, W. H. Pearce, K. A. Rosenfield, R. W. Smalling, T. A. Sos, et al. Atherosclerotic Vascular Disease Conference: Writing Group VI: Revascularization Circulation, June 1, 2004; 109(21): 2643 - 2650. [Full Text] [PDF] |
||||
![]() |
D. Pelz, P. Lylyk, and M. Negoro Interventional Neuroradiology Stroke, February 1, 2004; 35(2): 381 - 382. [Full Text] [PDF] |
||||
![]() |
M. Forsting Editorial Comment--With or Without Protection? The Second Important Question in Carotid Artery Stenting Stroke, January 1, 2004; 35 (1): e20 - e21. [Full Text] [PDF] |
||||
![]() |
J. D. Barr, J. J. Connors III, D. Sacks, J. C. Wojak, G. J. Becker, J. F. Cardella, B. Chopko, J. E. Dion, A. J. Fox, R. T. Higashida, et al. Quality Improvement Guidelines for the Performance of Cervical Carotid Angioplasty and Stent Placement: Developed by a Collaborative Panel of the American Society of Interventional and Therapeutic Neuroradiology, the American Society of Neuroradiology, and the Society of Interventional Radiology AJNR Am. J. Neuroradiol., November 1, 2003; 24(10): 2020 - 2034. [Full Text] [PDF] |
||||
![]() |