To the Editor:
With much delight we have read in the issues 2, 5, and 7 of Stroke1,2,3 that ultrasonic perfusion imaging of the brain achieves more and more recognition in the community. In these very well conducted studies, the authors have proven the ability of the method of predicting localization and size of final infarction in the acute state. Even though of different technical background, there are now 5 studies with a considerable cohort of about 120 patients altogether suffering from acute stroke examined by ultrasonic perfusion imaging.1,2,3,4,5 With parametric imaging, the diagnosis becomes quick and easy,1 clinical outcome can be forecasted,2 and a microbubble destructive approach decreases time of examination considerably.3 However, two main drawbacks of the different methods remain: only 1 hemisphere is examined so that quantitative comparison with healthy referential tissue is only possible with a second (contralateral) examination; the evaluation of the microbubble kinetic is performed in a way that quantification is only possible in a "perfusion yes/no" manner.
Since visualization of perfusion deficits has obviously improved so much since the early days, we propose that future studies should gain for a semiquantitative bilateral examination with the aim of differentiating ischemia and hypoperfusion in the acute state. A bilateral examination (having both hemispheres in the field of view) using the bolus kinetic yields stable parameters throughout the field of view, where regions of one hemisphere can be compared with the same regions of the other hemisphere.6 Interindividual ranges of the resulting parameters are considerable due to various cardiovascular reasons. However, intraindividual ranges of eg, the time-to-peak intensity (TPI) are very small.7 In 20 healthy volunteers, the mean quartile deviation of TPI values in individual 14 regions throughout both hemispheres was 0.68 seconds, and there was no case in which any TPI exceeded the individual mean for more than 2 seconds.
Our conclusion is that TPI might be a reliable semiquantitative marker of cerebral perfusion, since its variation is so small. In accordance with other technical approaches,8 we hypothesize that a delay of TPI forecasts hypoperfused and potentially still salvageable tissue, whereas regions without detectable perfusion kinetic forecasts ischemia, ie, the core of the infarction. As a cut-off for a significant TPI delay we determined the double of the maximum variation in the healthy cohort, ie, 4 seconds. These considerations were supported by preliminary cases of acute stroke patients examined by the same protocol,9 in one of which we could demonstrate a region without a perfusion kinetic (where eventually infarction occurred) and a region with a delayed perfusion kinetic, where eventually tissue stayed viable after clinically successful thrombolytic therapy (Figure).
|
Thus, we propose that for future ultrasonic perfusion imaging protocols using the bolus-kinetic, the bilateral approach should be applied with settings optimized for a quantitative evaluation of the kinetic parameters in order to differentiate "penumbra" and core of infarction.
References
Department of Neurology, University of Schleswig-Holstein, Lübeck, Germany
Response:
Eyding et al state that there is a growing body of evidence that ultrasound perfusion imaging with the bolus or the diminution approach expands the diagnostic potential of ultrasound from the cerebral macrocirculation to the microcirculation. As discussed, there are several limitations of these new approaches like the limited insonation depth, insonation artifacts and, of course, the fact that brain perfusion (cerebral blood flow or cerebral blood volume) cannot be assessed quantitatively.
The statement that "quantitative comparison with healthy referential tissue" is not possible with the conventional technique is not entirely correct because at 10 cm insonation depth ipsilateral and contralateral thalamus can be reliably investigated and, being supplied by the posterior circulation, serve as reference regions in middle cerebral artery (MCA) ischemia. The other statement that "quantification is only possible in a perfusion yes/no manner" does not match the results of three studies cited,1,2,3 which showed very reliably that a delay (>3 seconds to the surrounding tissue) and a decrease (>50% of normalized signal amplitude) of the contrast signal predict the area of infarction beside a missing signal.
In case of a MCA ischemia, the "bilateral approach" with an insonation depth of 15 cm covers the arterial territories of both sides.4 The advantage of this approach would be a direct comparison of both MCA territories after one bolus injection. Yet, because of the known depth dependence of the ultrasound signal, a comparison of the signal amplitude (the strongest predictor of infarction) remains impossible. Only time dependent parameters of the time-intensity curve could be compared, provided that the extracranial vascular status is normal. This is likely the case in healthy volunteers, but not in elderly ischemic stroke patient. Another limitation of the "bilateral approach" is the increase of insonation artifacts which occur because of calcified structures in the midline of the brain like the pineal gland and the choroid plexus as shown in the figure given by Eyding et al (notice the small window of proper insonation and the shadowing artifacts at the left lower margin of the image). Because of these theoretical limitations and the limited number of published cases (n=4) the "bilateral approach"4 may be a promising new feature which should be further investigated but, by now, cannot be considered state of the art for ultrasound perfusion imaging. This approach should be evaluated with new ultrasound technologies like power modulation or contrast pulse sequencing which increase the sensitivity for contrast agent detection in deeper regions of the brain, compared with conventional harmonic imaging or phase inversion harmonic imaging.
The conclusion of the letter that, by using parameters of the time-intensity curve after ultrasound contrast agent bolus injection, the detection of ischemic core and penumbra could be predicted seems an over-interpretation of a study with four clinical cases authored by Eyding et al.4