(Stroke. 2006;37:3019.)
© 2006 American Heart Association, Inc.
Original Contributions |
From the Departments of Interventional and Diagnostic Neuroradiology (J.G., G.S., L.R., J.S., C.B.), Interventional and Diagnostic Radiology (J.G.), and Neurology (K.N.), University of Bern, Switzerland.
Correspondence to Caspar Brekenfeld, Department of Neuroradiology, Freiburgstrasse 4, CH-3010 Bern, Switzerland. E-mail caspar.brekenfeld{at}insel.ch
| Abstract |
|---|
|
|
|---|
Methods Angiography and embolization with a radioopaque whole blood thrombus was performed in 10 swine. Mechanical thrombectomy was performed in 20 cranial vessels using a proximal aspiration device (Vasco35) and a distal basket-like device (Catch) with and without proximal balloon occlusion. Fifty-six retrieval attempts were made.
Results The proximal device allowed fast repeated application with a low risk of thromboembolic events (3%) and vasospasm, but it had a significantly lower success rate (39.4%) in retrieving thrombotic material than the distal device (DD) (82.6%; odds ratio, 7.3; 95% CI, 2.0 to 26.4). The compaction of the thrombus during retrieval with DD increased the risk of vessel wall irritation significantly (P<0.01) and complicated retrieval into the guiding catheter. The number of embolic events was significantly higher with DD (26%; odds ratio, 11.3; 95% CI, 1.35 to 101.6) unless proximal balloon occlusion was used.
Conclusions The proximal and the distal approaches to mechanical thrombectomy proved to be effective at achieving recanalization of cranial vessels. The proximal device is faster in application and allowed repeated attempts with a low complication rate. The DD is more successful at removing thrombotic material, but its method of application and attendant thrombus compaction increase the risk of thromboembolic events and vasospasms.
Key Words: animal models embolic stroke interventional neuroradiology stroke
| Introduction |
|---|
|
|
|---|
All mechanical thrombectomy devices are delivered by endovascular access proximal to the occlusion site. The various devices can be divided into 2 major groups according to where they apply force on the thrombus: proximal devices (PDs) apply force to the proximal base of the thrombus (this group includes various aspiration catheters); distal devices (DDs) approach the thrombus proximally but then are advanced by guide wire and microcatheter past the thrombus to be unsheathed behind it, where force is applied to the distal base of the thrombus (this group includes snare-like, basket-like, or coil-like devices).6,9,12 Two commercially available DDs are the Catch device (Balt) and the Merci Retrieval System (Concentric Medical, Inc). No in vivo studies have yet systematically compared these 2 major principles of mechanical thrombectomy in the treatment of acute cerebrovascular stroke.
The present study uses an animal model using a radio-opaque whole blood thrombus13 to compare the in vivo effectiveness of the proximal approach (PD Vasco 35 catheter) to the distal approach (DD Catch device) with regard to thrombusdevice interaction, retrieval success, and complication rates.
| Methods |
|---|
|
|
|---|
Thrombus Preparation
In acute cerebrovascular stroke, whole blood thrombus or "red thrombus" predominates in the occluded vessel.14 To reproduce the situation in human stroke, thrombi of 25x3 mm were generated. We therefore prepared the blood clot for selective thromboembolization by mixing 10 mL autologous blood with 25 IU bovine thrombin (Dade Behring Inc). To increase x-ray absorption, 1 g barium sulfate was added. After mixing for 10 s, the blood was injected into a silicone tube (3 mm; ClinicoMedical) and incubated for 60 minutes at room temperature. The radio-opacity allows visualization of thrombus movement in real time as well as detection of the dislocation of fragments to peripheral vessels.
The thrombus was then washed in physiological saline solution and incubated for an additional 20 minutes before application. The radioopaque barium sulfate does not cause local or systemic reactions during the study interval.13
Study Protocol
Surgery was performed on each animal through a groin approach for preparation of the common femoral artery and vein under general anesthesia. A 9 French catheter sheath (Arrow International Inc.) was introduced into the common femoral artery and flushed continuously with heparinized physiological saline (10 U/mL). A central venous catheter was placed in the common femoral vein for constant venous access.
Selective intra-arterial digital subtraction angiography was performed on a biplane high-resolution angiography system (Toshiba CAS500). Selective angiography was done on the lingual artery (LA) and internal carotid artery (ICA). Two vessels per animal were occluded. The mean vessel size was 2.7 mm for both proximal and distal clot retrieval (range 2.4 to 3.2 mm). After thrombus application, the guiding catheter was removed for 5 minutes to restore arterial flow and allow embedding of the thrombus (Figure 1).
|
The 4.2 French in diameter Vasco +35 catheter (Balt) for thrombus aspiration and the Catch device (Balt) were used for mechanical thrombectomy (Figure 2). Aspiration was performed by vacuum suction with a 20-mL syringe. The Catch device was delivered using a microwire (SilverSpeed 14; MTI) and Vasco 21 microcatheter (Balt). To assess the influence of proximal balloon occlusion on the success rate and complications such as thrombus dislocation and vasospasm, the devices were tested with (half of the occlusions) and without proximal balloon occlusion using a 9 French Merci balloon guider (Concentric Medical, Inc) in the common carotid artery.
|
The number of attempts for a device to retrieve the thrombus was limited to 5 attempts per thromboembolic occlusion. After the fifth attempt without success, the clot retrieval was rated as failed. During the study, a total of 20 vessels (10 LA and 10 ICA) were occluded, and 56 retrieval attempts were performed. A follow-up angiography was performed to evaluate the recanalization rate and detect potential complications.
The following data were recorded: (1) thrombus movement and distal dislocation, fragmentation, and thromboembolization; (2) recanalization rate using the thrombolysis in myocardial infarction scale (TIMI; 0 to 3);15 (3) number of vasospasms in case of recanalization using a grading scale:16 0, no narrowing; 1, slight narrowing (<25% reduction in lumen diameter); 2, moderate narrowing (25% to 50% stenosis or 50% to 75% stenosis affecting only a short segment of vessel); and 3, severe narrowing (50% to 75% stenosis affecting a long segment of vessel or any stenosis >75%); (4) vessel dissection or perforation; (5) success rate and time per attempt; and (6) site where thrombotic material was lost during retrieval.
Student t test was used for between-group analysis and MannWhitney test for comparison of vasospasm grading. The odds ratio (OR) and 95% CI were applied to compare the probability of successful thrombus retrieval and embolization rate with the 2 devices.
| Results |
|---|
|
|
|---|
The mean time between introduction of the device into the sheath and the first retrieval attempt was 9 minutes. The mean time per attempt for subsequent attempts was slightly shorter (6.3 minutes). The mean time in case of successful recanalization with proximal aspiration was 20.7 minutes.
ThrombusDevice Interaction
Evaluation of the thrombusdevice interaction revealed that the PD typically led to an initial elongation of the thrombus. In the 20 unsuccessful attempts, aspiration produced initial movement of the thrombus, but the applied force was not sufficient for retrieval or fragmentation of the thrombus. In 9 attempts, the proximal base of the thrombus entered the microcatheter tip during aspiration. This phenomenon led to a significantly higher retrieval rate compared with the attempts without thrombus entering the catheter tip (OR, 20.3; 95% CI, 3.1 to 132.3). Once the applied force exceeded the adhesion force of the thrombus in the vessel, the clot returned to its original shape and could be pulled into the guiding catheter (Figure 3). No thrombotic material was lost at the tip of the guiding catheter.
|
Distal Device
Efficiency and Complications
Overall, 10 vessel occlusions were treated with the DD. The distal approach allowed total recanalization (TIMI 3) in 8 of 10 occlusions (80%). In 2 occlusions, the thrombotic material was partially removed, but no recanalization was achieved (TIMI 0). A total of 23 attempts (mean 2.3; range 1 to 3) were made with the DD. In 2 occlusions, the very first attempt led to complete thrombus removal and recanalization. In 19 of 23 attempts (82.6%), the device was able to remove thrombotic material, and in only 4 attempts (17.4%), it failed to retrieve any clot. This retrieval success was significantly higher than the results in the PD trial (OR, 7.3; 95% CI, 2.0 to 26.4). In 2 attempts (2 of 23; 8.7%), the initial passage of the microcatheter led to distal dislocation of the thrombus in the carrying vessel (unrelated to the proximal balloon occlusion). In 6 of 23 attempts (26%), thrombotic material was lost during retrieval in the carrying vessel and led to thromboembolic occlusion of a different vessel (Figure 4). This rate is significantly higher than with PD (OR, 11.3; 95% CI, 1.3 to 101.7). The risk of embolic events using DD was considerably lower with proximal balloon occlusion (1 of 11; 9.1%) than without (5 of 12, 41.6%; OR, 7.1; 95% CI, 0.7 to 75.2). In 1 occlusion, a dissection occurred, but no perforation of a vessel was found. The mean vasospasm rating of 2.6 (range 1 to 3) for the DD was relevantly higher than that for the PD (1.6). In 1 case, severe vasospasm hindered any additional retrieval attempts. The occurrence and grade of vasospasm were independent of the use of proximal balloon occlusion.
|
The time of the initial attempt per occlusion was mean of 14.1 minutes. In contrast to PD, the mean time per attempt increased with each successive attempt (mean 18.3 minutes). This increase in time was mainly attributable to the procedure for repositioning the device into the microcatheter and navigation of the microwire and microcatheter distally to the thrombus. Therefore, the mean time in case of successful recanalization was significantly (P<0.01) prolonged to 42.1 minutes compared with PD (20.8 minutes; Table).
|
ThrombusDevice Interaction
Device navigation showed that the guide wire and microcatheter could be advanced easily past the thrombus and the device unsheathed distally to the thrombus. In none of the attempts, thrombus penetration was observed, the devices always passing between the thrombus and vessel wall. Restriction of the distal space because of branching of the vessel or distal obstruction made deployment of the device more difficult and time consuming. Typically, retrieval of the DD began with compaction of the thrombus. In successful thrombus retrieval, the thrombus was partially caught in the proximal third of the basket-like device and pushed sideways into the guiding catheter (Figure 5). Because of the compaction, retrieval into the tip of the guiding catheter was complicated. In 6 of 23 attempts (26%), small fragments of the thrombus were sheared off at the tip of the guiding catheter. In another attempt, the device compressed and lost the thrombus in the carrying vessel; the resulting obstruction could not be passed with either the guide wire or the microcatheter. Furthermore, in 2 attempts (2 of 23; 8.7%), thrombotic material was pressed into small side branches and cut off by the device. In none of the 23 retrieval attempts did the device fracture.
|
| Discussion |
|---|
|
|
|---|
|
The present study also revealed that DDs have a relevant higher risk and severity of vasospasm, hindering reperfusion in case of recanalization. This may be as well attributable to the irritation caused by the procedure for advancing past the thrombus, the force trajectory of the device, and resulting compaction of the thrombus. If the thrombus has extended into small side branches, a phenomenon sometimes seen after previous unsuccessful retrieval attempts with thrombus compaction, the DD cuts off this thrombotic material in the main carrying vessel.
The distal approach demands a mechanically more complex device than the proximal approach. The repeated passing and the need to reposition the device in the microcatheter after unsuccessful retrieval increase recanalization time. Although fewer attempts are necessary than with PD, the time to achieve total recanalization is much longer. In addition, the design of the device requires space distal to the occlusion site, limiting access to the thrombus if there is distal branching or bifurcation.
Compared with the DD, the approach of the PD is uncomplicated and its design mechanically simple. Without the need for time-consuming repositioning and passing procedures, the PD allowed repeated fast retrieval attempts. Because it does not require passing procedures, no distal dislocation occurred. Analysis of the thrombusdevice interaction shows that the thrombus was initially elongated during PD retrieval. Both of these characteristics appear to have reduced irritation to the vessel wall and thus lowered the risk and severity of vasospasm compared with DD. Furthermore, the in-line position of the thrombus behind the PD decreased the risk of thromboembolic events attributable to flow. Accordingly, we found proximal balloon occlusion to have only a minor influence on the embolization rate. However, per attempt, the success rate for retrieving thrombotic material was relatively low. The limiting factor for successful aspiration was the amount of suction the tip of the microcatheter could apply to the proximal base of the thrombus. In more than half of all attempts, the force created at the tip of the PD by suction with the 20-mL syringe was insufficient, thus preventing removal of the thrombus. In case the proximal end of the thrombus entered the microcatheter during aspiration, the chance of thrombus retrieval increased significantly. These results indicate that microcatheter diameter, tip shape, and the syringe size influence the success rate and point out room for device improvement.
| Acknowledgments |
|---|
None.
Received January 17, 2006; revision received April 13, 2006; accepted April 28, 2006.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
C. Brekenfeld, G. Schroth, M. El-Koussy, K. Nedeltchev, M. Reinert, J. Slotboom, and J. Gralla Mechanical Thromboembolectomy for Acute Ischemic Stroke: Comparison of the Catch Thromboectomy Device and the Merci Retriever In Vivo Stroke, April 1, 2008; 39(4): 1213 - 1219. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Gralla, M. Burkhardt, G. Schroth, M. El-Koussy, M. Reinert, K. Nedeltchev, J. Slotboom, and C. Brekenfeld Occlusion Length Is a Crucial Determinant of Efficiency and Complication Rate in Thrombectomy for Acute Ischemic Stroke AJNR Am. J. Neuroradiol., February 1, 2008; 29(2): 247 - 252. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Slotboom Localized therapeutic hypothermia in the brain for the treatment of ischemic stroke J Appl Physiol, April 1, 2007; 102(4): 1303 - 1304. [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Stroke Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2006 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |