Stroke. 2006;37:e15-e17
Published online before print December 29, 2005,
doi: 10.1161/01.STR.0000198880.28827.84
(Stroke. 2006;37:e15.)
© 2006 American Heart Association, Inc.
Magnetic Resonance Angiography of Cerebral Arteries After Neonatal Venoarterial and Venovenous Extracorporeal Membrane Oxygenation
Jeroen Hendrikse, MD, PhD;
Linda S. de Vries, MD, PhD
Floris Groenendaal, MD, PhD
From the Department of Radiology (J.H.), University Medical Center Utrecht, The Netherlands; and Department of Neonatology (L.S.d.V., F.G.), University Medical Center Utrecht/Wilhelmina Childrens Hospital, The Netherlands.
Correspondence to J. Hendrikse, MD, PhD, University Medical Center Utrecht, Department of Radiology (Hp E 01.132), PO Box 85500, 3508 GA Utrecht, The Netherlands. E-mail j.hendrikse{at}azu.nl
 |
Abstract
|
|---|
Background and Purpose In newborns with severe respiratory
failure, extracorporeal membrane oxygenation (ECMO) has been
proven to improve survival. With venoarterial ECMO, the right
common carotid artery is ligated, whereas with venovenous ECMO,
this carotid artery remains intact. The authors present the
magnetic resonance angiography (MRA) evaluation of cerebral
hemodynamics in a case of venoarterial and a case of venovenous
ECMO.
Methods With 2D phase-contrast MRA flow volume (mL/min), measurements are obtained of the brain feeding arteries and draining veins. Time-of-flight and flow-directional MRA are used to demonstrate the anatomy and flow patterns at the level of the circle of Willis.
Results A total volume flow of 63 mL/min at the skull base and collateral flow via the anterior circle of Willis were measured in the infant after venoarterial ECMO, and a total volume flow of 86 mL/min and a fetal type configuration of the circle of Willis were observed in the infant after venovenous ECMO.
Conclusion MRA is a useful technique to quantify cerebral blood flow in neonates after venoarterial or venovenous ECMO and could be used to select neonates with insufficient collateral compensation after venoarterial ECMO, requiring revascularization surgery.
Key Words: carotid arteries extracorporeal membrane oxygenation magnetic resonance angiography occlusion
 |
Introduction
|
|---|
In newborns with severe respiratory failure, extracorporeal
membrane oxygenation (ECMO) has been proven to improve survival.
1 The combination of the pre-existent severe hypoxia and the ECMO
procedure itself with ligation of the right common carotid artery
may result in brain lesions in 20% to 50% of the neonates.
1 To prevent ischemic complications, revascularization surgery
has been performed after venoarterial ECMO.
2 Alternatively,
ligation of the common carotid artery can be prevented by venovenous
ECMO, with a purely venous access in the circulation. A flow
velocity comparison with ultrasound between venoarterial and
venovenous ECMO demonstrated decreased intracranial flow velocities
during venoarterial and preserved flow velocities during venovenous
ECMO.
3
In our institution, a routine MRI examination is performed after an ECMO procedure to evaluate the presence of ischemic brain damage. Advantages of MRI examination after an ECMO procedure are the potential of a combined evaluation brain anatomy and the cerebral hemodynamics with magnetic resonance angiography (MRA) techniques. We report the results of MRA measurements of flow volume (mL/min) collateral flow and vascular anatomy in 2 full-term neonates, 1 after venoarterial ECMO and 1 after venovenous ECMO.
 |
Case Reports
|
|---|
Patient A
In patient A (birth weight 3460 g), venoarterial ECMO with occlusion
of the right common carotid artery and right jugular vein was
performed because of persistent pulmonary hypertension of the
neonate. Anatomical MRI performed at 3 weeks and 4 days postnatal
age demonstrated no hemorrhagic or ischemic lesions. A fast
phase-contrast (PC) sagittal survey was obtained to plan the
2D PC MRA scans for volume flow measurements (
Figure 1A). Two-dimensional
PC MRA scan time was 40 s, with a velocity sensitivity of 30
cm/s.
4 Quantitative flow values were calculated in each vessel
separately by integrating over manually drawn regions of interest
that enclosed the vessel lumen of the arteries and veins of
interest (
Figure 1B and 1C). The intraobserver variation of
2D PC MRA volume flow measurements, as expressed by the SD of
the difference, was 2.2 mL/min corresponding to 5.6%. The interobserver
variation (J.H., F.G.) of volume flow measurements was 2.1 mL/min
corresponding to 5.5%. At the skull base, a total arterial flow
was measured of 63 mL/min, with a volume flow of 44 mL/min for
the left internal carotid artery (ICA) and 19 mL/min for the
basilar artery. Below the carotid bifurcation, a total arterial
flow was measured of 95 mL/min, with a volume flow of 61 mL/min
for the left common carotid artery, 24 mL/min for the right
vertebral artery, and 10 mL/min for the right vertebral artery.
The total venous volume flow draining via the left jugular vein
below the level of the carotid bifurcation was 97 mL/min. The
time-of-flight MRA, planned parallel to the circle of Willis,
demonstrated occlusion of the right ICA (
Figure 2A). Flow-directionalsensitive
MRA images demonstrated collateral filling of the right middle
cerebral artery via the anterior part of the circle of Willis,
with reversed flow in the precommunicating segment (A1) of the
right anterior cerebral artery (
Figure 2B and 2C). No collateral
flow was detected via the right posterior communicating artery.

View larger version (113K):
[in this window]
[in a new window]
|
Figure 1. Planning (A) 2D PC MRA volume flow measurement at the skull base (B) and below the carotid bifurcation (C). 1 represents left ICA; 2, basilar artery; 3, sinus ethmoidales inferior; 4, left common carotid artery; 5, vertebral arteries; 6, left jugular vein.
|
|

View larger version (67K):
[in this window]
[in a new window]
|
Figure 2. Patient A, Venoarterial ECMO. The time of flight MRA scan (A) demonstrates an occluded right carotid artery and 2D PC MRA scans (B and C) demonstrate collateral flow via the anterior circle of Willis toward the right middle cerebral artery with retrograde flow in the right A1 segment of the anterior cerebral artery (arrowheads).
|
|
Patient B
In patient B (birth weight 3500 g), venovenous ECMO without ligation of the right common carotid artery was performed because of meconium aspiration syndrome. Anatomical MRI performed at 3 weeks and 3 days postnatal age demonstrated a small subcortical parietooccipital ischemic lesion on the right side. Using 2D PC MRA, the total arterial volume flow measured at the skull base was 86 mL/min, with a 33 mL/min contribution of the right ICA, 37 of the left ICA, and 16 mL/min of the basilar artery. Below the carotid bifurcation a total arterial volume flow was measured of 115 mL/min, with a volume flow of 46 mL/min for the right common carotid artery, 39 mL/min for the left common carotid artery, 16 mL/min for the right vertebral artery, and 14 mL/min for the left vertebral artery. The total venous volume flow just below the carotid bifurcation was 121 mL/min, with a 28 mL/min contribution of the right jugular vein and 93 mL/min contribution of the left jugular vein. The time-of-flight MRA demonstrated a fetal type circle of Willis, with filling of the right anterior, middle, and posterior cerebral artery from the right ICA, with a hypoplastic precommunicating segment (P1) of the right posterior cerebral artery (Figure 3A). Flow-directionalsensitive MRA images demonstrate anterior to posterior flow via the right posterior communicating artery in accordance with the dominant feeding of the right posterior cerebral artery from the right ICA (Figure 3B and 3C).

View larger version (63K):
[in this window]
[in a new window]
|
Figure 3. Patient B, Venoarterial ECMO. The time of flight MRA scan (A) and 2D PC MRA scans (B and C) demonstrate a right fetal type posterior cerebral artery, with blood flowing in the posterior direction (arrowheads).
|
|
 |
Discussion
|
|---|
In the current study, 2D PC MRA was used for volume flow (mL/min)
measurements in the brain feeding arteries. With 2D PC MRA,
the blood flow velocityinduced changes in magnetic resonance
(MR) signal (phase shift) are exploited to measure cerebral
blood volume quantitatively.
4 As demonstrated, MRA volume flow
measurements can be added to existing MR protocols, with an
additional scan time of

40 s for combined volume flow measurements
of the ICAs and basilar artery. Alternatively, arterial volume
flow measurements in the extracranial arteries in preterm and
term neonates can be performed with color duplex sonography.
5 When MRI is routinely performed for detailed anatomical evaluation
in the follow-up of an ECMO procedure, adding the described
MRA scans provides quantitative information of flow through
arteries and veins (
Figure 1) and detailed information of the
intracranial vasculature. The mean total volume flow of 95 mL/min
for neonate A and 115 mL/min for neonate B with both neonates
of 45 weeks postmenstrual age are in accordance with previous
ultrasound measurements in preterm and term neonates, with a
measured volume flow of 85 mL/min at 42 weeks postmenstrual
age and an observed flow increase in these ultrasound studies
of 6 mL/min per week increase in postmenstrual age.
5 Most likely,
physiological variation in volume flow, as has been observed
in previous ultrasound studies, may explain the difference in
flow between the 2 neonates of similar age in the present study.
5,6 The degree of respiratory failure severity, positioning in the
dedicated neonate head coil, and MRI acquisition protocol were
similar for both neonates.
A previous study demonstrated the use of time-of-flight MRA for anatomical evaluation of the circle of Willis after ECMO.7 In addition to flow quantification, we combined anatomical MRA images with flow-directionalsensitive MRA at the level of the circle of Willis. With anatomical and functional information, we demonstrate collateral flow from the contralateral ICA via the anterior communicating artery and flow direction in a fetal variant type circle of Willis. Visualization of the intracranial vasculature is important because 50% of the neonates has a variant type of the circle of Willis.8 Potentially, neonates with a variant type circle of Willis, such as a missing or hypoplastic precommunicating segment (A1) of the anterior cerebral artery or a fetal type feeding of the posterior cerebral artery, will be more vulnerable for ischemia after venoarterial ECMO with less collateral capacity.
In the present study a velocity encoding of 30 cm/s was used to benefit from the full dynamic range of the 2D PC MRA flow volume measurements with relatively low flow velocities. The pixel size of the 2D PC MRA scans of 1 mm was small compared with the mean vessel size of 3.4±1.0 mm (mean±SD) to avoid potential underestimation of flow measurements. Previously, validation studies have demonstrated the quantitative accuracy of 2D PC MRA for volume flow measurements in the brain feeding arteries,4,9 and comparisons have been made between MR PC flow quantification and ultrasound flow measurements.10
In conclusion, intracranial and extracranial volume flow measurements are reported in 2 neonates after an ECMO procedure. Collateral flow via the anterior circle of Willis is demonstrated in patient A after venoarterial ECMO, and a fetal configuration of the circle of Willis is demonstrated in patient B with a venovenous ECMO. MRA may be used in future studies to select neonates with insufficient collateral compensation after venoarterial ECMO requiring revascularization surgery.
Received October 17, 2005;
accepted November 2, 2005.
 |
References
|
|---|
- Bulas D, Glass P. Neonatal ECMO: neuroimaging and neurodevelopmental outcome. Semin Perinatol. 2005; 29: 5865.[CrossRef][Medline]
[Order article via Infotrieve]
- Fukuda S, Aoyama M, Yamada Y, Saitoh N, Honjoh T, Hasegawa T, Futamura M, Katoh J, Niimi N, Tanaka S, Nagaya M. Comparison of venoarterial versus venovenous access in the cerebral circulation of newborns undergoing extracorporeal membrane oxygenation. Pediatr Surg Int. 1999; 15: 7884.[CrossRef][Medline]
[Order article via Infotrieve]
- Mitchell DG, Merton D, Desai H, Needleman L, Kurtz AB, Goldberg BB, Graziani L, Wolfson P Neonatal brain: color Doppler imaging. Part II. Altered flow patterns from extracorporeal membrane oxygenation. Radiology. 1988; 167: 307310.[Abstract/Free Full Text]
- Bakker CJG, Hartkamp MJ, Mali WPTM. Measuring blood flow by nontriggered 2D phase contrast MR angiography. Magn Reson Imaging. 1996; 14: 609614.[CrossRef][Medline]
[Order article via Infotrieve]
- Kehrer M, Goelz R, Krageloh-Mann I, Schoning M. Measurements of volume of cerebral blood flow in healthy preterm and term neonates with ultrasound. Lancet. 2002; 360: 17491750.[CrossRef][Medline]
[Order article via Infotrieve]
- Kehrer M, Goelz R, Schoning M. The development of hemodynamics in the extracranial cerebral arteries of healthy preterm and term neonates. Ultrasound Med Biol. 2004; 30: 283287.[CrossRef][Medline]
[Order article via Infotrieve]
- Maas K, Barkovich AJ, Dong L, Edwards MS, Piecuch RE, Charlton V. Selected indications for and applications of magnetic resonance angiography in children. Pediatric Neurosurg. 1994; 20: 113125.
- Van Overbeeke JJ, Hillen B, Tulleken CA. A comparative study of the circle of Willis in fetal and adult life. The configuration of the posterior bifurcation and the posterior communicating artery. J Anat. 1991; 176: 4554.[Medline]
[Order article via Infotrieve]
- Bakker CJG, Kouwenhoven M, Hartkamp MJ, Hoogeveen RM, Mali WPTM. Accuracy and precision of time-averaged flow as measured by non-triggered 2D phase-contrast MR angiography: a phantom evaluation. Magn Reson Imaging. 1995; 13: 959965.[CrossRef][Medline]
[Order article via Infotrieve]
- Ho SS, Chan YI, Yeung DK, Metreweli C. Blood flow volume quantification of cerebral ischemia: comparison of three noninvasive imaging techniques of carotid and vertebral arteries. AJR Am J Roentgenol. 2002; 178: 551556.[Abstract/Free Full Text]