(Stroke. 2007;38:2726.)
© 2007 American Heart Association, Inc.
Original Contributions |
From the Department of Cardiovascular Sciences (M.J.M., A.H.G., A.R.N., D.H.E.), University of Leicester; the Medical Physics Department (E.M.L.C., K.V.R., L.F., D.H.E.), University Hospitals of Leicester NHS Trust; and the Department of Engineering (S.V.H), University of Leicester, Leicester, England; and Bracco Research SA (A.D.M.), Plan-les-Ouates, Geneva, Switzerland.
Correspondence to Prof David H. Evans, Level 1, Sandringham Building, Leicester Royal Infirmary, Leicester, LE1 5WW, UK. E-mail dhe{at}le.ac.uk
| Abstract |
|---|
|
|
|---|
Methods— Binding of microbubbles to the surface of the thrombus was confirmed by scanning electron microscopy. Targeted and control bubbles were then introduced to thrombus and tissue-mimicking material circulated under pulsatile-flow conditions in an in vitro flow rig. Embolic signal intensities before and after introduction of the bubbles were measured by TCD.
Results— Targeted microbubbles enhanced TCD signal intensities from thrombus emboli by up to 13 dB. The bubbles were capable of binding to moving thrombus when injected into the flow circuit in low concentrations (
36 bubbles per 100 mL) and were retained on the thrombus under pulsatile-flow conditions. Signal intensities from similarly sized pieces of tissue-mimicking material were not enhanced by injection of targeted bubbles.
Conclusions— Injection of appropriately targeted microbubbles significantly enhances TCD detection of circulating thrombus emboli in vitro.
Key Words: atherosclerosis embolism microbubbles thrombus ultrasonics
| Introduction |
|---|
|
|
|---|
Embolization of thrombus contributes toward stroke and transient ischemic attack in patients with carotid disease and is implicated in the development of major neurologic deficits after carotid surgery.4–6 Endothelial damage associated with plaque rupture and cardiovascular surgery promotes formation of thrombus, which can be susceptible to embolization. The ability to distinguish whether emboli are principally composed of thrombus or other pathogenic materials could aid in patient diagnosis by providing an early indication of plaque stability and embolus composition.
Targeted microbubbles could also improve sensitivity in the detection of weakly scattering emboli. It is estimated that approximately one third of emboli detected after carotid surgery produce TCD embolic signal intensities that fall below the threshold for reliable detection.7 Histologic evidence suggests that TCD detects only a tiny fraction of emboli reaching the brain. In autopsy studies, it was found that some patients received several million microemboli during cardiac surgery.8 These produced large numbers of pathologically identifiable lesions, suggesting that microemboli smaller than the detectable threshold for TCD monitoring may be more clinically significant than currently assumed.
Previous studies have demonstrated attachment of targeted ultrasound contrast agents to static thrombus in vitro2,9 and in animal experiments.2,10,11 These studies indicate improved visualization of thrombus by vascular ultrasound imaging. However, possible applications of targeted microbubbles to TCD detection of moving thromboemboli were not investigated.
Here we propose a new method for enhancing TCD detection of emboli by "labeling" pieces of thrombus with antibody fragment–grafted microbubbles (immunobubbles). Injection of immunobubbles has potential to enhance detection of weakly scattering emboli owing to an increase in backscattered ultrasound power produced by the attached bubbles. Detection of an increase in embolic signal intensity has potential to provide a means of distinguishing one embolic material from another. Microbubbles also have a number of other promising potential applications, such as encapsulation of "clot-busting" pharmaceuticals12 and enhanced embolus destruction by sonothrombolysis.13
In the present study, we tested abciximab-grafted microbubbles designed to attach to activated platelets on the surface of thrombus. We confirmed binding of the bubbles by scanning electron microscopy (SEM) and TCD. Experiments simulating the flow of blood and emboli through an artery showed significant enhancement of Doppler embolic signal intensities after introduction of targeted microbubbles.
| Materials and Methods |
|---|
|
|
|---|
3 hours.
Embolus Preparation
Clinically realistic thrombi were generated by the Chandler loop method.15–17 Thrombi produced in the Chandler loop closely resemble arterial thrombi formed in vivo, with a platelet-rich "head" and a fibrin-rich "tail."17 Freshly prepared thrombi were cut into small pieces and measured with a traveling optical microscope equipped with a Vernier scale capable of ±0.01-mm accuracy. As a control, similarly sized "emboli" were also cut from pieces of an agar-based tissue-mimicking material.18
Scanning Electron Microscopy
To prepare the samples for SEM, fresh thrombi of approximately equal size were exposed for
3 seconds to a
3.6x108 bubbles/mL suspension of either the targeted or control microbubbles. The bubble-thrombus complex was gently washed in saline and transferred via pipette to a mixture of HEPES buffer and 25% glutaraldehyde solution. Samples were then dehydrated through a series of ethanol concentrations increasing from 50% to 100%19 and dried with LCO2 with a Bal-Tec CPD-030 critical-point drier (Bal-Tec, Balzers, Liechtenstein). The dried samples were sputter-coated with gold-palladium with a Polaron SC 7640 (VG Microtech). SEM images were obtained with a field-emission gun SEM (FEI Sirion 200, FEI Electron Optics, Eindhoven, Netherlands).20
Flow-Rig Experiments
Flow-rig experiments were conducted with a closed-loop flow phantom (Figure 1). The circuit was constructed of C-flex tubing with a 4.6-mm internal diameter and a 0.8-mm wall thickness (Cole-Parmer, London, UK). Pulsatile flow, with a peak "systolic" velocity of
30 cm · s–1, was produced by a peristaltic pump (HR flow-inducer type 2000, Watson Marlow, Cornwall, UK). The circulating fluid consisted of 100 mL of 0.9% saline mixed with
2 mL blood-mimicking fluid.21
|
The Doppler system used in this study was a noncommercial multigate TCD device with exceptionally high dynamic range.22 System settings were similar to those of a commercial TC22 TCD machine (SciMed, Bristol, UK), but with a lower thermal index and pulse-pressure amplitude and a similar signal-to-noise ratio.23 The artificial vessel was insonated with a nominal 10-mm sample length at a depth of 50 mm and an insonation angle of
30°. Recorded data were saved directly to hard disk for later analysis.
Because the circulating fluid was translucent, passage of an embolus through the sample volume could be verified by visual inspection. Embolic signals were accepted for analysis only when their times of appearance in the sonogram matched those visually recorded by an observer. Analyses of the recordings were performed with an in-house embolic signal analysis program to replay the sonogram and audio signal in real time. Embolic signal intensities were measured relative to the background signal with a measured embolus-to-background ratio (MEBR) in decibels.24 All of the analyzed emboli had properties that were consistent with international embolus detection criteria.25
Thrombi Immersed in Microbubbles
After prior circulation of the fluid to remove air from the system, single pieces of thrombus of known size were introduced to a microbubble-free circuit via the funnel reservoir (Figure 1). Each piece of thrombus circulated past the transducer 10 times, thus generating 10 MEBR values. Optical microscopy was performed before and after circulation to confirm that no fragmentation had occurred. If the thrombus was still intact, it underwent an
3-second exposure to a
3.6x108 bubble/mL suspension of either the targeted or the control bubbles. The bubble-thrombus complex was gently washed to remove any unbound bubbles and reintroduced to the flow phantom. The thrombus-bubble complex was then allowed to circulate another 10 times in the same circulating fluid, and the new average MEBR and SD were estimated. Finally, the embolus dimensions were measured again. Fragmentation was apparent in
1 in 4 emboli, which were not included in our analysis.
Microbubbles Injected Into the Flow Circuit
In the last set of experiments, thrombi were circulated 10 times, and then 0.1 mL of a dilute (
360 bubbles/mL) suspension was injected into the funnel reservoir (Figure 1). Once introduced to the flow circuit, the dilution of bubbles was sufficiently low that discrete signals from the passage of individual bubbles through the sample volume could be observed. Doppler signals from the circulating thrombus-bubble complex were distinguishable from those of free circulating microbubbles by their higher intensity and longer duration. The time taken to complete 1 circuit was
60 seconds, with a total duration for 10 circulations varying between 8 and 10 minutes. After the embolus had circulated another 10 times, average MEBR values before and after injection of the bubbles were compared. In calculating the intensity of the thrombus-bubble complex, we made every effort to ensure that there were no free microbubble signals in the background window used for calculation of MEBR. All experiments were performed within 20 minutes, which is considerably shorter than the
3-hour half-life of the bubbles.
Statistics
Mean MEBRs and SDs were estimated from an average of at least 10 measurements. The difference in MEBR before and after introduction of bubbles (
MEBR=MEBRafter–MEBRbefore) was assumed to have an SE of
|
|
The a priori hypothesis that microbubbles would produce an enhancement in MEBR was tested with a 1-tailed Mann–Whitney U test. Nominal significance levels (
) are stated in the text. The level of correlation of MEBR with time was characterized by a line-of-regression analysis, estimation of an R2 value, correlation coefficient (r), and significance level P. The symbol
MEBR denotes the mean increase in MEBR averaged for several samples.
| Results |
|---|
|
|
|---|
3-µm diameter, some of which resemble "squashed" or decomposing microbubbles (Figure 2c). In thrombi not exposed to any form of bubble, no bubblelike structures were observed.
|
|
Microscopy images for thrombi exposed to targeted bubbles clearly showed attachment of intact, targeted microbubbles to the thrombus (Figure 3). Bubbles appeared as
1- to 4-µm spherical structures, which were typically larger and more spherical than platelets. Microbubbles can be distinguished from white blood cells because of the smoothness of the bubble surface (Figure 3c).
Thrombi Immersed in Microbubbles
Immersion of thrombi (with mean diameters of 1.5, 2.1, and 2.5 mm) in a concentrated solution of targeted bubbles produced an average increase in mean MEBR of
13 dB (the Table). Representative data from 2 of these thrombi are shown in Figure 4a. Increases in MEBR were of high statistical significance, based on a 1-tailed Mann–Whitney U test (
0.0005). Variations in MEBR over time were investigated by determining the gradient of the line of regression for consecutive measurements of the same embolus and were not found to be statistically significant. For targeted bubbles, enhancement of embolic signal intensity persisted under pulsatile-flow conditions and did not diminish during the 8 to 10 minutes required for 10 circulations. This suggests that bound, targeted bubbles were robust to pulsatile-flow conditions.
|
|
As a control, 3 similar sizes of thrombi (1.5, 2.15, and 2.3 mm) underwent the same process but were instead bathed in a suspension containing control bubbles (the Table). Representative data from 2 thrombi are shown in Figure 4b. Initial exposure to control bubbles appeared to increase MEBRs by a similar amount as for targeted bubbles. However, the enhancement from control bubbles was short-lived (<8 minutes) and decreased with time in the circuit. All 3 thrombi exposed to control bubbles exhibited a statistically significant negative correlation between MEBR and the times of consecutive measurements (r
–0.9, P<0.001). This suggests that "binding" of control bubbles did not persist under pulsatile-flow conditions. Gradual detachment of control bubbles reduced the increase in MEBR, averaged for 10 cycles, to
8 dB.
With a null hypothesis stating that "there was no change in MEBR before and after introduction of the bubbles," the control microbubbles produced a statistically significant increase in MEBR in 2 of the 3 thrombi. For these thrombi, the null hypothesis was rejected with a level of significance of 
0.0005 (1.5-mm thrombus) and 
0.001 (2.3-mm thrombus; Figure 4b). For the third thrombus (2.15 mm), the null hypothesis was accepted, indicating that there was no statistically significant increase in MEBR.
Microbubbles Injected Into the Flow Circuit
To mimic injection of microbubbles under pulsatile-flow conditions, measurements were performed for 4 sizes of thrombi (with mean diameters of 0.7, 1, 1.7, and 1.9 mm) before and after injecting 0.1 mL of a dilute suspension containing
360 microbubbles/mL into the flow circuit. The targeted bubbles produced a clear enhancement in MEBRs (the Table). Representative data are shown in Figure 5a. Levels of significance were 
0.001 for the 0.7-mm thrombus and 
0.0005 for the other 3 sizes (1, 1.7, and 1.9 mm).
|
The experiment was repeated with control bubbles for 3 sizes of thrombus (1.1, 1.5, and 1.8 mm), but no statistically significant increase in MEBR for any of these thrombi was observed (the Table and Figure 5b). To confirm active binding to the thrombus, we also tested for attachment of targeted microbubbles to synthetic emboli composed of tissue-mimicking material. These tissue-mimicking material emboli do not contain glycoprotein IIb-IIIa receptors. No statistically significant increase in MEBR for tissue-mimicking material was observed.
| Discussion |
|---|
|
|
|---|
Our investigation differs from previous reports by using TCD to demonstrate binding of targeted microbubbles to circulating thrombus emboli. To test whether bubbles were capable of attaching to moving thrombus under pulsatile-flow conditions, we injected dilute solutions of targeted and control microbubbles into the flow circuit. This experiment was designed to verify whether a dilute solution of microbubbles injected upstream of the source of embolization would be capable of successfully adhering to moving thrombus emboli in flowing blood. We observed a clear difference between the injection of targeted and control bubbles: thrombi introduced to targeted bubbles were significantly enhanced (
MEBR=8.2±1.8 dB), whereas signals from thrombi receiving the control bubbles were not (
MEBR=1.4±2.1 dB; the Table). We concluded that only targeted bubbles were capable of attaching to and enhancing the MEBRs of moving thrombus under clinically realistic flow conditions.
The advantages of ultrasound contrast agents for enhancement of conventional ultrasound images are well known9,10,22 but have not yet been applied to TCD embolus detection. Thromboemboli have a density similar to that of blood, and are therefore, the most difficult composition of embolus to detect with TCD.7 A major benefit of targeted microbubbles would be enhancement of the detection of weak embolic signals.
One limitation of our study was that the sizes of thrombi used in these experiments were larger than expected in vivo. Doppler signal enhancement of smaller emboli may not be as marked as described in this study. Another limitation is that free bubbles in the circuit continued to circulate rather than be expelled. This prevented us from studying the dose dependence of the enhancement, as large numbers of free microbubbles in the fluid would have enhanced the background signal and reduced the MEBR. In clinical practice, any unbound bubbles would be flushed from the circulation, so the optimum concentration in vivo would be different from that determined in vitro.
Brief exposure of thrombi to a concentrated solution of targeted bubbles produced an average rise in MEBR of
13.1±5 dB (the Table). Despite the pressure cycles of the pump, ultrasound insonation, fluid flow forces, diffusion of gas from the bubbles, and bubbles becoming trapped in the reservoir, targeted bubbles remained bound to the thrombus for the duration of our measurements. No significant decrease in MEBR with time between consecutive circulations over 10 minutes was observed (Figure 4a). Increases in MEBRs of
8 dB were observed for doses of targeted microbubbles as low as
36 microbubbles in 100 mL, which indicates a high affinity between the targeted bubbles and thrombus. Concentrated control bubbles produced an increase in MEBR of
7.9±3.7 dB, but there was a significant decrease in MEBR with time (Figure 4b). This implied that control bubbles were "washed away" by flow forces in the circuit. When control bubbles were injected into the flow circuit in dilute concentrations, no MEBR enhancement was observed.
Administration of immunobubbles to patients with suspected "unstable" carotid or cardiac plaques could greatly enhance ultrasound images and lead to improved assessment of plaque stability and embolus composition.26,27 Although free bubbles would be rapidly expelled from the circulation, we have demonstrated that bound bubbles are robust to physiologic flow conditions and have the potential to survive in the body for longer durations, allowing ultrasound imaging or embolus detection to be performed. In prolonged TCD monitoring of patients with carotid stenosis, emboli are extremely infrequent, and either bubbles with longer lifetimes or a regular "top up" may be required. Bubbles of increased longevity are currently under development.28
During TCD embolus detection, an increase in average MEBR after injection of immunobubbles could indicate that a large proportion of emboli were of either thrombus or platelet origin. Microbubbles may therefore be potentially useful for distinguishing between different embolic materials, especially if various, more specific targeting moieties are used. To verify that immunobubbles examined in the current study were target-specific, experiments were performed with thrombus-targeted bubbles injected into a circuit containing synthetic tissue-mimicking material. No enhancement of signal from tissue-mimicking material emboli was observed (
MEBR=0.0±0.3 dB). However, because plaque surfaces and plaque debris can be thrombogenic and contain superimposed thrombus, immunobubbles may be unable to distinguish atheromatous from purely thrombotic embolic materials in vivo. The bubbles investigated herein adhered to activated platelets,2,11 but there is a range of potential targeting antibodies that could selectively bind to other pathogenic materials and might not suffer from the same attraction to superimposed thrombus as thrombus-targeted bubbles. The potential for distinguishing between pathogenic materials needs to be investigated in a more physiologically and hematologically realistic study.
The possibility that microbubbles could produce false-positive signals by binding to other material such as platelets or platelet microparticles was also considered. Single platelets or platelet microparticles would be tiny compared with the immunobubbles and therefore, result in an ultrasound signal that would be similar to that from a free circulating bubble.
Last, we considered the possibility that microbubbles could induce a pharmacologic antiplatelet effect. The c7E3 antibody (abciximab) is used therapeutically to prevent the binding of fibrinogen to platelets. However, to inhibit platelet aggregation, there needs to be >70% receptor occupancy of the drug (ie, >70% of
50 000 glycoprotein IIb-IIIa receptors on the platelet have to be occupied). The ratio of microbubbles to platelets would be
1:106, which would not significantly inhibit platelet aggregation.
In conclusion, we confirm that targeted microbubbles are capable of selectively binding to moving thromboemboli under pulsatile-flow conditions. Targeted microbubbles produce pronounced enhancements in Doppler signals from thrombus but not from tissue-mimicking material. We therefore propose that targeted microbubbles could form a new basis for enhanced target-specific detection of emboli by TCD.
| Acknowledgments |
|---|
This study was supported by the European Union Ultrasonic Monitoring and Early Diagnosis of Stroke (UMEDS) project, QLG1-CT-2002-01518. We also acknowledge the financial assistance of our employers, the University of Leicester (M.J.M, A.R.N, A.H.G, D.H.E., S.V.H.) and the University Hospitals of Leicester NHS Trust (E.M.L.C, L.F., K.V.R). A.D.M. was funded from the UMEDS grant but was based at Bracco Research SA while performing this research.
Disclosures
None.
| Footnotes |
|---|
Received March 28, 2007; accepted April 17, 2007.
| References |
|---|
|
|
|---|
2. Schumann PA, Christiansen JP, Quigley RM, McCreery TP, Sweitzer RH, Unger EC, Lindner JR, Matsunaga TO. Targeted-microbubble binding selectively to GPIIb IIIa receptors of platelet thrombi. Invest Radiol. 2002; 37: 587–593.[CrossRef][Medline] [Order article via Infotrieve]
3. Markus HS, Brown MM. Differentiation between different pathological cerebral embolic materials using transcranial Doppler in an in vitro model. Stroke. 1993; 24: 1–5.
4. Levi CR, OMalley HM, Fell G, Roberts AK, Hoare MC, Royle JP, Chan A, Beiles BC, Chambers BR, Bladin CF, Donnan GA. Transcranial Doppler detected cerebral embolisation following carotid endarterectomy: high microembolic signal loads predict post-operative cerebral ischaemia. Brain. 1997; 120: 621–629.
5. Stork JL, Levi CR, Chambers BR, Abbott AL, Donnan GA. Possible determinants of early microembolism after carotid endarterectomy. Stroke. 2002; 33: 2082–2085.
6. Payne DA, Jones CI, Hayes PD, Thompson MM, London NJ, Bell PR, Goodall AH, Naylor AR. Beneficial effects of clopidogrel combined with aspirin in reducing cerebral emboli in patients undergoing carotid endarterectomy. Circulation. 2004; 30: 109:1476–1481.
7. Chung EML, Fan L, Naylor AR, Evans DH. Impact of clinical environment on embolus detection: a comparison of automated and manual detection of embolic signals. Cerebrovasc Dis. 2007; 23: 169–174.[CrossRef][Medline] [Order article via Infotrieve]
8. Moody DM, Bell MA, Challa VR, Johnston WE, Prough WS. Brain microemboli during cardiac surgery or aortography. Ann Neurol. 1990; 28: 477–486.[CrossRef][Medline] [Order article via Infotrieve]
9. Unger EC, McCreery TP, Sweitzer RH, Shen D, Wu G. In vitro studies of a new thrombus-specific ultrasound contrast agent. Am J Cardiol. 1998; 81: 58G–61G.[CrossRef][Medline] [Order article via Infotrieve]
10. Takeuchi M, Ogunyankin K, Pandian NG, McCreery TP, Sweitzer RH, Caldwell VE, Unger EC, Avelar E, Sheahan M, Connolly R. Enhanced visualization of intravascular and left atrial appendage thrombus with the use of a thrombus-targeting ultrasonographic contrast agent (MRX-408A1): in vivo experimental echocardiographic studies. J Am Soc Echocardiogr. 1999; 12: 1015–1021.[CrossRef][Medline] [Order article via Infotrieve]
11. Unger E, Metzger P, Krupinski E, Baker M, Hulett R, Gabaeff D, Mills J, Ihnat D, McCreery T. The use of a thrombus-specific ultrasound contrast agent to detect thrombus in arteriovenous fistulae. Invest Radiol. 2000; 35: 86–89.[CrossRef][Medline] [Order article via Infotrieve]
12. Unger EC, Matsunaga TO, McCreery T, Schumann P, Sweitzer R, Quigley R. Therapeutic applications of microbubbles. Eur J Radiol. 2002; 42: 160–168.[CrossRef][Medline] [Order article via Infotrieve]
13. Wu Y, Unger EC, McCreery TP, Sweitzer RH, Shen D, Wu G, Vielhauer MD. Binding and lysing of blood clots using MRX-408. Invest Radiol. 1998; 33: 880–885.[CrossRef][Medline] [Order article via Infotrieve]
14. Della Martina A, Allémann E, Bettinger T, Busstat P, Lassus A, Pochon S, Schneider M. Grafting of abciximab to a microbubble-based ultrasound contrast agent for targeting to platelets expressing GPIIb/IIIa: characterization and in vitro testing. Eur J Pharm Biopharm. In press.
15. Chandler AB. In vitro thrombotic coagulation of the blood: a method for producing a thrombus. Lab Invest. 1958; 7: 110–114.[Medline] [Order article via Infotrieve]
16. Poole JCF. A study of artificial thrombi produced by a modification of Chandlers method. Q J Exp Physiol. 1959; 44: 377–384.
17. Robbie LA, Young SP, Bennett B, Booth NA. Thrombi formed in the Chandler loop mimic human arterial thrombi in structure and PAI-1 content and distribution. Thromb Haemost. 1997; 77: 510–515.[Medline] [Order article via Infotrieve]
18. Ramnarine KV, Anderson T, Hoskins PR. Construction and geometrical stability of physiological flow rate wall-less stenosis phantoms. Ultrasound Med Biol. 2001; 27: 245–250.[CrossRef][Medline] [Order article via Infotrieve]
19. Tsutsui JM, Xie F, Cano M, Chomas J, Phillips P, Radio SJ, Lof J, Porter TR. Detection of retained microbubbles in carotid arteries with real-time low mechanical index imaging in the setting of endothelial dysfunction. J Am Coll Cardiol. 2004; 44: 1036–1046.
20. Humphreys FJ, Brough I. High resolution electron backscatter diffraction with a field emission gun scanning electron microscope. J Microsc. 1999; 195 (pt 1): 6–9.[Medline] [Order article via Infotrieve]
21. Ramnarine KV, Nassiri DK, Hoskins PR, Lubbers J. Validation of a new blood-mimicking fluid for use in Doppler flow test objects. Ultrasound Med Biol. 1998; 24: 451–459.[CrossRef][Medline] [Order article via Infotrieve]
22. Fan L, Boni E, Tortoli P, Evans DH. Multigate transcranial Doppler ultrasound system with real-time embolic signal identification and archival. IEEE Trans Ultrasonics Ferroelec Freq Contr. 2006; 53: 1853–1861.[CrossRef]
23. Fan L, Ramnarine KV, Gittins J, Evans DH. Performance analyses of two transcranial Doppler systems using an electronic phantom and a flow phantom. Ultrasound. 2006; 14: 272. Abstract.
24. Evans DH. Detection of microemboli. In: Babikian VL, Wechsler LR, eds. Transcranial Doppler Ultrasonography, 2nd ed. Boston: Butterworth Heinemann; 1999: 141–155.
25. Ringelstein EB, Droste DW, Babikian VL, Evans DH, Grosset DG, Kaps M, Markus HS, Russel D, Siebler M. Consensus on microembolus detection by TCD. Stroke. 1998; 29: 725–729.
26. Tardy I, Pochon S, Theraulaz M, Nanjappan P, Schneider M. In vivo ultrasound imaging of thrombi using a target-specific contrast agent. Acad Radiol. 2002; 9 (suppl 2): S294–S296.[CrossRef][Medline] [Order article via Infotrieve]
27. Alonso A, Della Martina A, Stroick M, Fatar M, Griebe M, Pochon S, Schneider M, Hennerici M, Allemann E, Meairs S. Molecular imaging of human thrombus with novel abiciximab immunobubbles and ultrasounds. Stroke. 2007; 38: 1508–1514.
28. Gerber F, Krafft M, Waton G, Vandamme T. Microbubbles with exceptionally long life: synergy between shell and internal phase components. New J Chem. 2006; 30: 524–527.[CrossRef]
This article has been cited by other articles:
![]() |
T. R. Porter The utilization of ultrasound and microbubbles for therapy in acute coronary syndromes Cardiovasc Res, September 1, 2009; 83(4): 636 - 642. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Stroke Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2007 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |