(Stroke. 2007;38:881.)
© 2007 American Heart Association, Inc.
Original Contributions |
From the Departments of Neurosurgery and Neuroradiology (D.F., H.W., P.A.R., T.J.M.), Cleveland Clinic Foundation, Cleveland, Ohio; the Departments of Neurosurgery and Radiology and the Toshiba Stroke Research Center (E.I.L., R.A.H., L.N.H.), School of Medicine and Biomedical Sciences, State University of New York, and Millard Fillmore Gates Hospital, Kaleida Health, Buffalo, NY; the Departments of Neurosurgery and Neuroradiology (A.S.T., D.B.N., B.A.-K.), University of Wisconsin, Madison; and the Department of Neurosurgery (F.C.A., C.G.M.), Barrow Neurological Institute, Phoenix, Ariz.
Correspondence to David Fiorella, MD, PhD, Cleveland Clinic Foundation, S-80, Departments of Endovascular and Cerebrovascular Neurosurgery, 9500 Euclid Ave, Cleveland, OH 44195. E-mail fioreld{at}ccf.org
| Abstract |
|---|
|
|
|---|
Methods All patients undergoing angioplasty and stenting with the Gateway balloonWingspan stent system were prospectively tracked.
Results During a 9-month period, treatment with the stent system was attempted in 78 patients (average age, 63.6 years; 33 women) with 82 intracranial atheromatous lesions, of which 54 were
70% stenotic. Eighty-one of 82 lesions were successfully stented (98.8%) during the first treatment session. In 1 case, the stent could not be delivered across the lesion; the patient was treated solely with angioplasty and stented at a later date. Lesions treated involved the internal carotid (n=32; 8 petrous, 10 cavernous, 11 supraclinoid segment, 3 terminus), vertebral (n=14; V4 segment), basilar (n=14), and middle cerebral (n=22) arteries. Mean±SD pretreatment stenosis was 74.6±13.9%, improving to 43.5±18.1% after balloon angioplasty and to 27.2±16.7% after stent placement. Of the 82 lesions treated, there were 5 (6.1%) major periprocedural neurological complications, 4 of which ultimately led to patient death within 30 days of the procedure.
Conclusions Angioplasty and stenting for symptomatic intracranial atheromatous disease can be performed with the Gateway balloonWingspan stent system with a high rate of technical success and acceptable periprocedural morbidity. Our initial experience indicates that this procedure represents a viable treatment option for this patient population.
Key Words: angioplasty intracranial atheromatous disease stenting Wingspan
| Introduction |
|---|
|
|
|---|
Strategies for revascularization have included surgical bypass, percutaneous transluminal angioplasty (PTA) alone, and percutaneous transluminal angioplasty and stenting (PTAS) using balloon-mounted coronary stents (BMCSs) off-label in the cerebrovasculature. Although several retrospective case series suggest that PTA alone is a safe and effective treatment strategy,3 surgical bypass was shown to be inferior to medical therapy for middle cerebral artery (MCA) stenosis,4 and PTAS with BMCSs has been associated with high rates of periprocedural morbidity and mortality.57
The Wingspan, a flexible, self-expanding, microcatheter-delivered, nitinol microstent (Boston Scientific/Target) is the first self-expanding stent designed specifically for the treatment of symptomatic ICAD. The Gateway PTA balloon catheter (Boston Scientific/Target) is used to predilate the target lesion before Wingspan stenting. The Gateway balloonWingspan stent system was designed to combine the safety advantages of the angioplasty-alone strategy with the superior immediate luminal gain achieved with a stenting strategy. We report the immediate and periprocedural results of our prospective, multicenter experience of the Wingspan system for the treatment of ICAD.
| Patients and Methods |
|---|
|
|
|---|
50% stenosis) were treated with the Gateway balloonWingspan stent system at the Barrow Neurological Institute (n=16 lesions), Cleveland Clinic (n=21), State University of New York at Buffalo (n=21), and University of Wisconsin (n=24) between November 2005 and July 2006 and were prospectively tracked in a collaborative endovascular database. Patient characteristics, lesion description, procedural details, and periprocedural complications were recorded. The institutional review board at each institution approved the collection and review of patient data for the study period. Composite data were collected for evaluation. Procedural success was defined as completion of Gateway balloon angioplasty and Wingspan stent placement across the target lesion, despite the degree of residual stenosis or any complications related to the procedure. The primary outcome end point was periprocedural stroke and death. Complications were adjudicated by the primary operators involved in the procedures.
Interventional Procedure
In brief, the presented interventions were performed via common femoral artery access with the exception of 1 procedure that was performed through direct puncture of the right common carotid artery (Figure 1a through 1d) and 2 that were performed through a brachial approach. All interventions were performed through a 6F access system. The targeted parent vessel was accessed with either a 6F guiding catheter (Envoy or Envoy XB; Cordis) or a 6F KSAW Shuttle-Select sheath system (Cook). Heparinization was instituted to a targeted activated coagulation time of 250 to 300 seconds. In most cases, after conventional angiography, an SL-10 (Boston Scientific), Prowler-10 (Cordis), or Eschelon-10 (Microtherapeutics, Irvine, Calif) microcatheter was manipulated across the target lesion using a 0.014-inch Synchro (Boston Scientific) or Transcend EX Soft Tip (Boston Scientific) microwire. The microcatheter was then exchanged over a 0.014-inch Transcend Floppy (Boston Scientific), Luge (Boston Scientific), or PVS (Boston Scientific) exchange microwire for a Gateway angioplasty balloon. The remaining lesions were primarily crossed with the Gateway angioplasty balloon and a Transcend 300-cm exchange-length 0.014-inch microwire (Boston Scientific). In each case, the balloon diameter was sized to 80% of the "normal" parent vessel diameter. The balloon length was selected to match the length of the lesion. Angioplasty was typically performed with a slow, graded inflation of the balloon to a pressure of between 6 and 12 atm for at least 120 seconds. After angioplasty, the balloon was removed and conventional angiography was repeated.
|
Next, the Wingspan delivery system was prepared and advanced over the exchange wire across the target lesion. The stent diameter was sized to exceed the diameter of the normal parent vessel by 0.5 to 1.0 mm. The stent length was selected to equal or exceed the length of the angioplasty balloon. In addition, the stent length was selected both to completely cover the entire diseased segment and to allow positioning of the proximal end of the stent so as not to preclude future endovascular access into the stented segment. The diameter of the stenotic lesion was measured by biplane angiography and compared with a reference diameter of the normal vessel (usually proximal to the lesion) as per the technique used in the WASID study.1 Pretreatment and postprocedural angiographic images were available for all 78 patients successfully treated with the stent. Postangioplasty images were available for 69 of 82 lesions treated. In 13 cases, a postangioplasty angiogram either was not obtained by the operator before stent placement (in these cases, a roadmap was performed and not permanently captured) or was not interpretable due to patient motion, image quality, or distortion of the regional anatomy by the exchange wire or balloon catheter. Poststenting imaging was available for 81 lesions; in 1 case, the stent could not be delivered across the target lesion (see Results). The sequence of steps involved in stent delivery and deployment are illustrated in Figure 2a through 2d.
|
All patients were pretreated with antiplatelet agents (typically, both aspirin and clopidogrel) and were typically discharged on both aspirin (325 mg daily) and clopidogrel (75 mg daily). The dual-antiplatelet regimen was maintained for a minimum of 4 weeks after the procedure, after which time patients remained on aspirin therapy (325 mg daily).
| Results |
|---|
|
|
|---|
Initial Treatment Results
Of the 82 lesions approached, 81 were treated with both the Gateway balloon and Wingspan stent (98.8% technical success rate for stent deployment across the stenotic lesion). In 1 patient, successful Gateway balloon angioplasty was performed with a 32% residual stenosis at the conclusion of the procedure; however, a tortuous carotid anatomy precluded delivery of the Wingspan stent. This patient presented later with recurrent stenosis and successfully underwent PTAS with Wingspan at that time. The 82 lesions were distributed within the following arterial locations: internal carotid (n=32; 8 petrous, 10 cavernous, 11 supraclinoid segment, 3 terminus), vertebral (n=14; V4 segment), basilar (n=14), and middle cerebral (n=22). The mean±SD pretreatment stenosis was 74.6±13.9%, with 54 of the 82 lesions having a
70% stenosis (Figure 3a through 3c). After Gateway angioplasty, the average stenosis was 43.5±18.1%. After Wingspan stent placement, the average stenosis was 27.2±16.7% (Figure 4).
|
|
Imaging and Clinical Follow-Up
Five major procedural complications have been encountered thus far during the course of the registry (the Table and Figure 5a through 5f). The remaining lesions were treated without any permanent neurological sequelae (6.1% major periprocedural morbidity and mortality). Two complications (lesions 44 and 58, Table), 1 vessel rupture after angioplasty and 1 microwire perforation during stent delivery, were clearly device related. One patient experienced a reperfusion hemorrhage (lesion 52) immediately after an otherwise technically successful procedure. One patient experienced a nondevice-related procedural complication, a large hemispheric infarction contralateral to the treated left MCA stenosis (lesion 71). The final complication was encountered in a patient with a symptomatic basilar stenosis who was treated emergently in the context of an actively evolving stroke syndrome (lesion 55). This patient continued to deteriorate clinically during and after the procedure. Magnetic resonance (MR) imaging performed after the procedure demonstrated extensive posterior circulation infarctions.
|
|
One patient experienced transient visual symptoms that completely resolved within 36 hours of the procedure. Five extracranial parent-vessel dissections related to guide catheter manipulation were encountered, 2 of which were flow limiting and required stenting. One flow-limiting intracranial dissection was encountered after Gateway angioplasty, which completely resolved after Wingspan stent placement and did not result in any neurological morbidity (Figure 6a through 6c).
|
Postprocedural MR diffusion-weighted imaging was performed on 38 patients within 72 hours of the procedure, 13 (34.2%) of whom had new ischemic lesions after the procedure. Ten of the 13 lesions were asymptomatic; the other 3 were in patients with major periprocedural complications (Table).
| Discussion |
|---|
|
|
|---|
Wingspan Stent System Design and Treatment Strategy: Effect on the Periprocedural Complication Rate Associated With Intracranial PTAS
Most series describing the outcomes of patients undergoing treatment with BMCSs report periprocedural complication rates in the range of 15% to 30%.5,811 In 1 of the largest available series of patients with symptomatic vertebrobasilar ICAD undergoing treatment with BMCSs, 9 periprocedural neurological complications (stroke or death) were encountered during treatment of 39 patients (23.1%).5 The present prospective, all-inclusive series indicates a significant reduction in overall event rates during PTAS in patients treated with the Wingspan system. Further evidence of procedural safety was provided by postprocedural MR diffusion-weighted imaging, which demonstrated new lesions after PTAS with the Wingspan system in 34% of patients, a rate that compares favorably with the 70% rate of diffusion-positive lesions reported after PTAS with BMCSs.12
The substantial reduction in periprocedural complications with the Wingspan can be attributed to both the device design and the recommended treatment strategy. Before delivery of the Wingspan stent, an angioplasty is performed with the Gateway balloon. The balloon is undersized to 80% of the normal parent-vessel diameter. This conservative predilation reduces the amount of vascular trauma induced and likely minimizes both the risk of target-vessel perforation as well as the likelihood of downstream embolization of atheromatous debris caused by plaque disruption.
After the angioplasty, the Wingspan stent is navigated across the lesion over a floppy exchange-length microwire and deployed (Figure 2a through 2d). The Wingspan microstent is composed of nitinol and housed in a low-profile, hydrophilic microcatheter delivery system. These properties make the stent delivery system considerably easier to navigate to and across an intracranial target lesion than even the newest generation of BMCSs (which remain significantly more rigid than the Wingspan). The enhanced navigability of the Wingspan system results in several important practical advantages during the procedure that seem to translate into lower complication rates: (1) stent delivery can be achieved with less aggressive guiding catheter positions within the parent vessel, resulting in less iatrogenic parent-vessel spasm and dissection; (2) stent delivery can be performed over flexible exchange microwires with much less distortion of the cerebrovascular tree, decreasing the severity of spasm and iatrogenic intracranial vascular injury; and (3) less aggressive distal microwire access is required to support the delivery of the stent system, decreasing the risk of distal microwire perforation.
The self-expanding nature of the Wingspan stent also provides several unique advantages in comparison with the available BMCSs. The Wingspan stent is slightly oversized, chosen to measure
0.5 to 1.0 mm larger than the diameter of the normal parent vessel in an unconstrained, fully expanded configuration. As such, once deployed, the stent exerts a continuous outward radial force on the vessel wall. This outward radial force acts to prevent early vessel recoil, thus consolidating the gains achieved with the initial angioplasty. As the stent gradually expands, it functions not only to maintain the luminal diameter but also to secure any foci of dissection or intimal injury (Figure 6a through 6c). The self-expanding nature of the stent facilitates optimal stentparent-vessel wall apposition over varying luminal diameters (which are frequently encountered within the cerebrovasculature). Finally, although the Wingspan is a new device, most neurointerventionists have had extensive experience with the Neuroform stent (Boston Scientific), another self-expanding nitinol microstent used for the treatment of intracranial aneurysms. As such, the "learning curve" for the application of this device is limited, as it represents an iteration of existing technology that is routinely applied to support the embolization of intracranial aneurysms.
Viable Endovascular Strategies for the Management of Symptomatic ICAD: Angioplasty Alone and Gateway PTAWingspan Stenting
Initial case series reported relatively high rates of periprocedural morbidity (28% periprocedural stroke or death) with the angioplasty-alone strategy.13 The introduction of routine adjunctive periprocedural platelet inhibition, along with a more conservative approach to balloon sizing and a slow balloon inflation technique, made intracranial angioplasty considerably safer.14 Several single-institution retrospective series and 1 multicenter retrospective series have suggested that with these modern techniques, intracranial angioplasty without stenting can be performed with a relatively low periprocedural risk profile.3,15,16 The major potential limitations of angioplasty without stenting include vessel recoil with acute recurrent stenosis, acute vessel occlusion secondary to either procedural dissection or recoil with regional platelet aggregation, and a lack of long-term durability, with some patients requiring multiple procedures because of recurrent stenosis.16
Although PTA alone has merit as a primary strategy for the treatment of symptomatic ICAD, there are several theoretical benefits to augmenting the procedure with the added step of stenting with the Wingspan. First, the present data indicate that the added risk of placing the self-expanding stent after successful angioplasty is low. Second, the self-expanding nature of the Wingspan augments the results of the initial angioplasty and prevents acute vessel recoil, providing better and more reliable initial results with greater immediate luminal gain. Not infrequently after PTA alone, a considerable residual stenosis remains. In the multicenter experience reported by Marks et al,3 40.7% of patients had residual stenoses of >50% at the conclusion of the procedure. In an additional 12.9% of patients treated, stents were required because the initial angioplasty result was "unchanged or worse" than the pretreatment stenosis. In the current series, only 4 of 81 (4.9%) lesions treated with both the Gateway balloon and Wingspan stent were left with a stenosis of 50% or more at the conclusion of the initial procedure. Third, post-PTA vessel dissections, which occurred 20.2% of the time in the series of Marks et al,3 can be immediately repaired with stent placement (Figure 6a through 6c).
Role of the Wingspan System in the Treatment of Symptomatic ICAD
The current data indicate that Gateway-Wingspan PTAS can be performed by multiple operators at different institutions with a level of periprocedural safety that makes it a potentially viable option for the treatment of symptomatic ICAD. Provided that the ongoing multicenter registry continues to demonstrate reasonable periprocedural safety and, as these patients are followed clinically, low rates of postprocedural neurological events, this technique may supplant medical therapy as the treatment of choice in some patients with symptomatic ICAD. This determination would be best made through a direct comparison with "best medical therapy" in the context of a multicenter, randomized, controlled trial.
Limitations
Complications were tracked by the primary operators involved in the procedures. As such, it is possible that neurological morbidity in these patients could have been underestimated in the absence of independent neurological adjudication. However, it is unlikely that any major procedural morbidity (periprocedural stroke or death) would have gone undetected in the periprocedural period. The current series represents an initial experience in a relatively small population of patients. Until larger numbers of patients are treated, the actual complication profile of the procedure remains an approximation. Equally important, no longitudinal follow-up of these patients is available at this time. Correspondingly, the efficacy of Wingspan for the prevention of recurrent ischemic events cannot be assessed at this point. Until these data are better understood, the precise role of the Wingspan system for the treatment of symptomatic ICAD remains poorly defined.
| Acknowledgments |
|---|
Sources of Funding
This US Wingspan Registry is supported by a research grant from Boston Scientific.
Disclosures
Drs Aagaard-Kienitz and Masaryk have no potential conflicts of interest to disclose. Dr Albuquerque is a consultant/advisory board member of Boston Scientific, Cordis, and Micrus. Dr Fiorella has received grant support from Boston Scientific, and is on the speakers bureau of and has received honoraria from Boston Scientific. Dr Hanel has received grant support from Boston Scientific, Cordis, and Microvention and is a scientific advisor to Neurovasx. Dr Hopkins has received industry grant support and consultant fees from Boston Scientific, Cordis, EndoTex, and Micrus; is a stock or shareholder of EndoTex and Micrus; has received honoraria from Bard, Boston Scientific, Cordis, and Medsn; and has a board, trustee, or officer position with Access Closure, marketRx, and Micrus. Dr Levy has received industry grant support from Boston Scientific and Cordis and honoraria from Boston Scientific and Cordis, and receives patent royalties from Zimmer Spine. Dr McDougall has received honoraria from Boston Scientific. Dr Niemann is a consultant/advisory board member of Boston Scientific, Cordis, ev3, and Micrus; is on the speakers bureau of and has received honoraria from Cordis, ev3, and Micrus; and has received research grants from Cordis, ev3, and Micrus. Dr Rasmussen is on the speakers bureau and has received honoraria from Boston Scientific. Dr Turk has received research support from Boston Scientific and GE Healthcare and is a consultant to Boston Scientific. Dr Woo has ownership interest in ev3 for consulting services and is a consultant/advisory board member of ev3.
Received September 15, 2006; accepted October 4, 2006.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. E. Kelly, A. J. Furlan, and D. Fiorella Recanalization of an Acute Middle Cerebral Artery Occlusion Using a Self-Expanding, Reconstrainable, Intracranial Microstent as a Temporary Endovascular Bypass Stroke, June 1, 2008; 39(6): 1770 - 1773. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Gupta Symptomatic Intracranial Atherosclerotic Disease: What Is the Best Treatment Option? Stroke, June 1, 2008; 39(6): 1661 - 1662. [Full Text] [PDF] |
||||
![]() |
O. O. Zaidat, R. Klucznik, M. J. Alexander, J. Chaloupka, H. Lutsep, S. Barnwell, M. Mawad, B. Lane, M. J. Lynn, M. Chimowitz, et al. The NIH registry on use of the Wingspan stent for symptomatic70-99% intracranial arterial stenosis Neurology, April 22, 2008; 70(17): 1518 - 1524. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. C. Haley Jr Registries: They're not just for weddings anymore Neurology, April 22, 2008; 70(17): 1508 - 1509. [Full Text] [PDF] |
||||
![]() |
S. Muller-Schunk, J. Linn, N. Peters, M. Spannagl, M. Deisenberg, H. Bruckmann, and T.E. Mayer Monitoring of Clopidogrel-Related Platelet Inhibition: Correlation of Nonresponse with Clinical Outcome in Supra-Aortic Stenting AJNR Am. J. Neuroradiol., April 1, 2008; 29(4): 786 - 791. [Abstract] [Full Text] [PDF] |
||||
![]() |
D.C. Suh, J.K. Kim, J.W. Choi, B.S. Choi, H.W. Pyun, Y.J. Choi, M.-H. Kim, H.R. Yang, H.I. Ha, S.J. Kim, et al. Intracranial Stenting of Severe Symptomatic Intracranial Stenosis: Results of 100 Consecutive Patients AJNR Am. J. Neuroradiol., April 1, 2008; 29(4): 781 - 785. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Prabhakaran, K.R Wells, V.H. Lee, C.A. Flaherty, and D.K. Lopes Prevalence and Risk Factors for Aspirin and Clopidogrel Resistance in Cerebrovascular Stenting AJNR Am. J. Neuroradiol., February 1, 2008; 29(2): 281 - 285. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Puetz, G. Gahn, U. Becker, D. Mucha, A. Mueller, N.U. Weir, B. Wiedemann, and R. von Kummer Endovascular Therapy of Symptomatic Intracranial Stenosis in Patients With Impaired Regional Cerebral Blood Flow or Failure of Medical Therapy AJNR Am. J. Neuroradiol., February 1, 2008; 29(2): 273 - 280. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. J. Quinn and K. R. Lees Advances in Emerging Therapies 2007 Stroke, February 1, 2008; 39(2): 255 - 257. [Full Text] [PDF] |
||||
![]() |
A.S. Turk, E.I. Levy, F.C. Albuquerque, G.L. Pride Jr, H. Woo, B.G. Welch, D.B. Niemann, P.D. Purdy, B. Aagaard-Kienitz, P.A. Rasmussen, et al. Influence of Patient Age and Stenosis Location on Wingspan In-Stent Restenosis AJNR Am. J. Neuroradiol., January 1, 2008; 29(1): 23 - 27. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Gupta, O. Zaidat, and A. Majid US Multicenter Experience With the Wingspan Stent System for the Treatment of Intracranial Atheromatous Disease: Periprocedural Results Stroke, September 1, 2007; 38(9): e80 - e80. [Full Text] [PDF] |
||||
![]() |
D. Fiorella and H. H. Woo Emerging Endovascular Therapies for Symptomatic Intracranial Atherosclerotic Disease Stroke, August 1, 2007; 38(8): 2391 - 2396. [Full Text] [PDF] |
||||
![]() |
D. F. Kallmes and H. M. Do Wherefore Wingspan? AJNR Am. J. Neuroradiol., June 1, 2007; 28(6): 997 - 998. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|