| Introduction |
|---|
|
|
|---|
According to the National Institute of Neurological Disorders and Stroke (NINDS), about 700 000 people have a stroke each year, a leading cause of long-term disability and the third leading cause of death for Americans after heart disease and cancer.1 The regulation of medical devices by CDRH includes devices indicated for stroke. This article will review research and development of such devices, pre- and postmarket evaluation, and provide an FDA perspective on future development.
| Medical Device Regulation |
|---|
|
|
|---|
Device Classification
Medical devices are categorized into 3 classes with regulatory control increasing from Class I to III. The device classification regulations define the regulatory requirements for a general device type. Most Class I devices are exempt from FDA notification; Class II devices require Premarket Notification 510(k); and most Class III devices require Premarket Approval (PMA).3 Humanitarian use devices (HUDs) are marketed for a limited population in an entirely separate process termed a Humanitarian Device Exemption (HDE). The majority of medical devices reviewed by CDRH are evaluated under premarket notification (510(k)).
Premarket Notification 510(k) Process
The 510(k) clearance process requires that the Sponsor demonstrate their device to be substantially equivalent to a legally marketed predicate device. It is the responsibility of the sponsor to identify an appropriate predicate device or devices to which the sponsor demonstrates that their new device is substantially equivalent in design, function, and indication for use. Given the incremental changes that occur with device development, the majority of device applications cleared under the 510(k) program are based on preclinical testing that demonstrates this equivalence to a predicate. Before marketing clearance is obtained, the manufacturer must also assure that the device is properly labeled in accordance with FDAs labeling regulations. In some cases, however, when there are concerns regarding safety and effectiveness, the FDA can require clinical data for 510(k) clearance.
PMA Process
A second regulatory pathway for marketing of medical devices is the PMA process. This market approval process is required for most Class III devices. These devices are typically referred to as significant risk devices that support or sustain human life, are important in preventing health impairment, or present risk of serious injury or death in case of failure. The PMA approval process requires that the manufacturer has demonstrated, through preclinical and clinical data, a reasonable assurance of safety and effectiveness when used for the label indication in the intended population. In FY 2005, FDA authorized marketing of 3148 devices with 510(k) clearance and 32 devices through the PMA program. Neurological devices have comprised
5% of all medical devices authorized for marketing.4
Approval of a device through the PMA process represents a rigorous type of device marketing application required by FDA and is based on a determination by FDA that the PMA application contains valid scientific evidence to assure that the device is safe and effective for its intended use(s) (21CFR814 [Subpart C]). An approved PMA is, in effect, a license granting the applicant (or owner) permission to market the device. The PMA applicant is usually the person (ie, company, also called sponsor) who owns the rights, or has authorized access, to the data and other information submitted in support of safety and effectiveness. In some cases, before approving or denying a PMA, an FDA advisory committee composed of nonagency experts (special government employees), may review the data submitted in support of the PMA. A public meeting provides a venue for making a recommendation whether to approve the submission. After carefully weighing the evidence contained in the submission and recommendations from the advisory panel, the FDA will notify the applicant of its decision. In those instances where a PMA is approved and after FDA notifies the applicant that the PMA has been approved, a notice is published on the Internet announcing the data on which the decision is based, providing an opportunity for the public to petition FDA within 30 days for reconsideration of the decision.5 To assist potential users in understanding the basis for PMA approval, FDA provides a Summary of Safety and Effectiveness that identifies the type of safety and effectiveness data as the basis for each approval decision. The clinical community will often continue to study the medical device after marketing (postmarket surveillance), to develop a more refined understanding of the patient population most likely to benefit.
Medical device regulation requires FDA to tailor the data requested from a manufacturer to address the specific safety and effectiveness questions that need to be addressed before a marketing authorization can be granted. The Center for Devices and Radiological Health, by law, considers valid scientific evidence to include all of the following: well-controlled investigations, partially controlled studies, studies and objective trials without matched controls, well-documented case histories conducted by qualified experts, and reports of significant human experience with a marketed device, to determine whether there is reasonable assurance that the device is safe and effective (21CFR 860.7). The agency has considered whether these guidelines are in conflict with adequate scientific integrity and believes there can be several approaches to addressing regulatory requirements.
HDE
In 1996 the FDA finalized rules (21 CFR part 814 [Subpart H]) regarding HUDs. In contrast to the PMA, a HUD designation requires that <4000 individuals in the United States per year are eligible for use of the device under the proposed indications. After approval of an HUD designation, the device must receive a Humanitarian Device Exemption to allow the intended use. Unlike a PMA approval, the sponsor of an HDE is not required to provide clinical data to demonstrate reasonable safety and effectiveness but must provide sufficient information demonstrating that the device does not pose unreasonable risks and that there is probable benefit of use that outweighs the risks. These data must be interpreted in light of currently available treatment options. Lastly, the sponsor must demonstrate that no comparable devices are available to treat or diagnose the disease or condition, and that the device could not be brought to market without HUD status. A device approved under an HDE for marketing, however, remains restricted as an investigational device limited to the specific indications stated in the product labeling. Investigators are required to obtain approval from their Institutional Review Boards for use of the device in each case.
There are currently no PMAs cleared for treatment of acute ischemic stroke. However, the FDA has recently approved two HDEs for intracranial stent systems for the treatment of symptomatic and medically refractory intracranial atherosclerotic disease, a very select patient population with an extremely high stroke risk: the Wingspan Stent System (Boston Scientific, 2005) and NeuroLink System (Guidant, 2002).
| Research and Development of Medical Devices |
|---|
|
|
|---|
During the research and development of a medical device, sponsors may approach FDA for guidance on the suitable design of a clinical or nonclinical study. Investigational devices for the treatment of stroke are usually evaluated by the Office of Device Evaluations (ODE), Division of General, Restorative, and Neurological Devices (DGRND), one of five divisions in CDRH, responsible for regulating medical devices. Clearance of a Class II medical device is typically based on both bench and animal performance testing. In some cases, the FDA may require a clinical study to demonstrate adequate performance characteristics, or substantial equivalence to a legally marketed predicate device. Safety and effectiveness are important considerations in FDAs review of premarket applications for medical devices.
Regardless of the medical device or regulatory pathway necessary to market the medical device, sponsors and principal investigators usually collect clinical data for each of the applications in a clinical study permitted by the FDA with approval of a protocol providing an Investigational Device Exemption (IDE). In most cases, if there is reasonable assurance that a device is safe (ie, the probable benefits to health from use of the device for its intended uses and conditions of use outweigh the probable risks), the FDA will allow a clinical investigation to proceed.
In general, devices indicated for the treatment of acute ischemic stroke have been categorized as significant risk devices and therefore warrant study under an IDE application. In addition to the requirement for having an FDA-approved IDE, sponsors of such trials must comply with the regulations governing institutional review boards (21 CFR Part 56) and informed consent (21 CFR Part 50).
Acute Ischemic Stroke
FDAs role is to regulate medical devices, drugs, and biologics, but not regulate the practice of medicine. There have been several clinical studies conducted to evaluate medical products for acute ischemic stroke. Prolyse in Acute Cerebral Thromboembolism (PROACT II),6 the Mechanical Embolus Removal in Cerebral Ischemia (MERCI) Retrieval Study,7 and other trials8,9,10 support investigation of acute stroke, generally within the first few hours of stroke onset. Based on this experience, initial treatment in IDE stroke trials usually occurs within hours of symptom onset. Defining the exact time of symptom onset is an important component of all study protocols to ensure identification of appropriate subjects. For those studies or devices with a therapeutic time window beyond intervals studied in prior studies, evidence is encouraged from either preclinical or clinical testing to support treating subjects for FDA approval of an IDE.
An issue in the investigation of acute ischemic stroke is the difficulty in recruiting subjects, particularly when there is a need to provide a cohort of control subjects treated with the current standard of care for ischemic stroke especially if a time restriction for use of the device is set. An unbiased way to interpret the outcomes of a medical device study can be obtained with a randomized controlled and double-blinded study to reliably infer the use of a device. Simple comparison of pretreatment versus posttreatment assessments, without a control group, may not provide adequate proof of a reasonable assurance of safety and effectiveness. Therefore, in some cases, prospective randomized controlled studies are encouraged. Although we are aware of the difficulty in the design and recruitment of subjects for acute ischemic stroke, alternate methods may be considered only when valid methods of inference are used and there are valid reasons for not using a randomized control group (eg, evaluation of medical device to restore blood flow). In some cases, study designs using single-arm nonrandomized clinical protocols with comparison of study results to historical or concurrent matched controls may be acceptable if they are scientifically sound and address the relevant safety and effectiveness concerns important to the proper use of the device by the community. In those cases where alternative study designs are presented, early collaboration with the FDA and discussion of new statistical issues with FDA, before study initiation, are strongly recommended.
Other important aspects of study design that are important to device manufacturers and will be evaluated by FDA include selection of an adequate and representative patient population, clinical outcome assessments by an appropriate, validated neurological impairment scale, disability measure, or handicap scale; device- and study design–dependent selection of appropriate clinical end points and statistical approaches; and the reproducibility of any technique.
| Postapproval Requirements |
|---|
|
|
|---|
What Clearance for Marketing Means to the User
Activase (alteplase), a genetically engineered version of tissue plasminogen activator, is a medical product approved for use during the acute onset of ischemic stroke. Other advances in the management of acute ischemic stroke include recent FDA clearance of the MERCI Retriever for US marketing.11,12,13 In comparison to tissue plasminogen activator, the MERCI retriever is intended to remove thrombus obstructing blood flow in the neurovasculature as cause of acute ischemic stroke (the MERCI Retriever had been previously cleared by the FDA for use in the retrieval of foreign bodies misplaced during interventional radiological procedures in the neuro, peripheral and coronary vasculature). Patients who are ineligible for treatment with IV tissue plasminogen activator or who fail IV tissue plasminogen activator therapy are candidates for intervention with the MERCI retriever.
The MERCI retriever was cleared along the 510(k) regulatory pathway rather than the PMA process because there already existed a predicate device with a similar use. The 510(k) device regulatory process is different from the regulatory process for drugs. In some cases, medical devices can have local effects whereas drugs can have systemic effects. Moreover, unlike the 510(k) regulatory path where one device may be similar to another and potentially appropriate for marketing clearance based on an incremental change in technology, approval of one drug based on an incremental change in molecular identity may not be appropriate because slight differences in molecular identity can have significant differences in safety and effectiveness. These characteristics highlight the potential differences between medical device and drug clearances.
Because of concerns regarding safety and potential effectiveness for the new use of the MERCI retriever, a clinical study was still required for device clearance. This study was designed as a nonrandomized study using historical data from prior clinical studies, such as PROACT II, as the control. The FDA also involved its advisory panel in the review of data for the MERCI retriever, and FDA considered comments made by advisory panel members very carefully when evaluating the evidence submitted by the manufacturer.11 Because the primary end point of this study was restoration of cerebral circulation and not improved clinical outcome after stroke, and because of the nonrandomized design, the MERCI retriever was cleared with an indication for use regarding cerebral revascularization after stroke and not an indication for use of acute stroke treatment. Although this may appear clinically to be an arbitrary distinction, this distinction was necessary because the data supporting market clearance was determined by radiographic evidence for accomplishing this end point and one of several considerations that the agency used in clearance of the device.
Once a device reaches the market, there is postmarket surveillance regulations with which a manufacturer must comply. These requirements include the Quality Systems (also known as Good Manufacturing Practices) and Medical Device Reporting regulations. The Quality Systems regulation is a quality assurance requirement that covers the design, packaging, labeling and manufacturing of a medical device. The Medical Device Reporting regulation is an adverse event reporting program. In the case of acute ischemic stroke, FDA continues to monitor Medical Device Reporting and other postmarket studies related to devices that are indicated for use in ischemic stroke to determine whether additional labeling or other modifications are needed to improve the safety and effectiveness profile of such devices.
Acute stroke management is a dynamic field in which medical devices will continue to play an important role in its development. Further studies are needed to evaluate outcomes for patients undergoing mechanical thrombectomy as a treatment modality.
| Future Perspectives |
|---|
|
|
|---|
Equally important in this process is the cooperation that is necessary in the development of medical technologies between academic, industry, advocacy, and government. Through ongoing interaction of the various parties involved, validation of safety and effectiveness of new treatments is important. The regulatory controls the FDA imposes provides important assurance that this will be accomplished without unacceptable restriction of the availability of these tools to the healthcare environment. Indeed, the Food Drug Modernization Act of 1997 requires that the agency cooperates with other stakeholders to assure a "least burdensome approach" to device regulation. The agency is also committed to monitoring devices through the products total life cycle, cooperating with the diverse parties primarily involved at each phase of the cycle, including its research, development, and clinical investigation, performance in the real world of clinical practice after market release, and any subsequent technological improvements or changes. Despite the regulatory pathway, an evaluation of device safety and effectiveness will be needed for any device. These principles are especially important for the field of acute stroke because of the tremendous public health impact of this disease and the recent technological advances in the treatment of cerebrovascular occlusive disease.
With the proper steps and dialogue among various stakeholders, ongoing research and development of medical products will hopefully lead to more safe and effective treatments for the treatment of this devastating disease.
| Acknowledgments |
|---|
This article represents the professional opinion of the authors and is not an official document, guidance or policy of the US Government, the Department of Health and Human Services, or the Food and Drug Administration, nor should any official endorsement be inferred.
Received September 26, 2006; revision received November 6, 2006; accepted November 14, 2006.