(Stroke. 2009;40:952.)
© 2009 American Heart Association, Inc.
Original Contributions |
From the Department of Biomedical Engineering (C.S., B.B.L., A.R., Q.H.), the Department of Radiology (B.B.L.), and the Endovascular Research Center, Vascular Biology Institute (L.C.), University of Miami, Coral Gables, Fla; the Department of Engineering and Physics (J.S.), University of Central Oklahoma, Edmond; the Department of Radiology (A.K.W.), University of Massachusetts Medical School, Worcester; and the Department of Medicine (F.O.T.), University of Texas Health Science Center, San Antonio.
Correspondence to Baruch B. Lieber, PhD, Professor of Biomedical Engineering and of Radiology, 1251 Memorial Drive, Coral Gables, FL 33146. E-mail blieber{at}miami.edu
| Abstract |
|---|
|
|
|---|
Methods— Three different configurations of an original flow diversion device were implanted across thirty elastase-induced aneurysm models in rabbits. Ten animals per device configuration were followed-up for 3 weeks (n=3), 3 months (n=3), or 6 months (n=4), and tissue explanted postsacrifice was sent for histology. The temporal variation in angiographic contrast intensity within each aneurysm was fitted with a mathematical model to quantify the alteration in local hemodynamics caused by the implanted device. A predictive index, called the washout coefficient, was constructed to estimate long-term aneurysm occlusion probabilities immediately after treatment with any flow diversion device.
Results— The device with a porosity of 70% and pore density of 18 pores/mm2 performed better at occluding aneurysms than devices with 70% porosity, 12 pores/mm2 and 65% porosity, 14 pores/mm2. A value of the washout coefficient less than 30 predicted greater than 97% angiographic aneurysm occlusion over a period of 6 months with a sensitivity of 73% and specificity of 82%.
Conclusions— The flow diversion devices effected successful and stable aneurysm occlusion. Pore density, rather than porosity, may be the critical factor modulating efficacy of such devices.
Key Words: stents histology scanning electron microscopy washout coefficient side-branch patency
| Introduction |
|---|
|
|
|---|
Seminal studies in the early to mid-90s introduced the notion that intracranial aneurysms could be successfully treated solely by the deployment of stents within the parent artery.8–11 Their hypothesis was that stents would divert flow away from the aneurysm and into the parent artery, and the resultant sluggish flow within the aneurysm would promote thrombus formation that would eventually organize into scar tissue and thus exclude the aneurysm from the circulation. Moreover, the stent implanted in the parent artery would, over time, be incorporated into the vessel wall thereby precluding regrowth of the aneurysm by hemodynamic mechanisms. Their results showed complete and stable occlusion of sidewall venous pouch canine carotid aneurysms treated with stents only over follow-up periods ranging from 1 to 9 months.9–11 These initial experimental results have been corroborated by subsequent clinical evidence showing progressive thrombosis or complete occlusion of aneurysms (mostly sidewall) at follow-up when treated with stents alone.12–14 On the other hand, other clinical reports have shown aneurysms treated with stents alone to have persistent filling at follow-up.14,15 It is important to note that none of these devices was designed to act as a flow diversion device, but instead as a scaffold for coils or stenotic arteries. Moreover, the follow-up times in the clinical studies noting incomplete aneurysm occlusion is less (average of about 1 month) than those that report complete aneurysm occlusion (average around 4 months).
The primary intent of a flow diversion device (as opposed to a stent) is to optimally alter the flow exchange between the parent artery and the aneurysm so as to promote complete thrombosis of the sac as rapidly as possible while eliciting minimal neointimal hyperplasia. There are no commercially available flow diversion devices as yet, although recent reports16–18 suggest that the development of such devices is being actively pursued. Based on our previous experience with flow diversion for treating intracranial aneurysms,10,19–22 we have constructed a flow diversion device of original design. Six potentially optimal device configurations with varying porosities and pore densities were constructed with the intent of homing in on the optimal configuration through experimentation. Detailed particle image velocimetry21 and preliminary in vivo22 studies were subsequently performed on the rabbit elastase-induced aneurysm model, which indicated 3 device configurations as having the best performance of the 6. We proceeded to implant these 3 devices in a large cohort of experimental aneurysms, the results of which are presented here.
| Methods |
|---|
|
|
|---|
|
Animal Experiments
All experimental protocols were approved by the Animal Care and Use Committee at our institution. Initiation and maintenance of anesthesia were identical for all procedures. New Zealand White rabbits were preanesthetized with an intramuscular injection of Glycopyrrolate (0.1 mg/kg) and sedated with intramuscular injections of Xylazine (5 mg/kg) and Ketamine (35 mg/kg); sedation was maintained with 1% to 2% Isoflurane throughout the procedure. After completion of survival procedures, the animals were given intramuscular Buprenorphine (0.04 mg/kg) for pain relief and subcutaneous Ketoprofen (2 mg/kg) as antiinflammatory and antipyretic immediately after surgery and twice a day for 3 days postsurgery. The procedure for creating elastase-induced aneurysms in rabbits is described elsewhere23; model aneurysms were created in 40 rabbits.
After allowing sufficient time for the aneurysm to mature (nominally 3 weeks), the animals were brought back to the angiography suite for flow diversion device implantation. Ten animals were implanted with each of the 3 device configurations, and the remainder 10 animals served as controls. Treated animals were orally given 20 mg of aspirin once a day beginning 2 days before the procedure and continuing until 10 days after the procedure to mitigate acute thrombotic events. Through a right transfemoral approach, the devices were implanted in the innominate-to-subclavian artery with the intent of maintaining equal lengths on either side of the aneurysm while also covering the vertebral artery ostium. High-speed angiographic sequences were acquired at 30 frames/s over 20 seconds while injecting 5 cc of angiographic contrast (Visipaque, 320 mgI/mL, GE Healthcare) at a rate of 2 cc/s both before, and immediately after, device implantation. Identical imaging (image to source distance, C-arm orientation, acquisition mode) and injection (injection volume, injection rate, preset delay) parameters were maintained during both acquisitions. The choice of the device to be implanted in any given animal was solely dependent on the study schedule and was otherwise arbitrary.
Animals were followed-up at 21 days (n=12, 3 animals per treatment group), 90 days (n=12, 3 animals per treatment group), or 180 days (n=16, 4 animals per treatment group) after the implantation phase of the study. High-speed follow-up angiograms were acquired via left transfemoral access. The animal was euthanized with an overdose of sodium pentobarbital (1 mL intravenous injection of Euthasol, Delmarva Laboratories), flushed with saline and 10% formalin, after which the aneurysm-parent vessel complex including the aorta, vertebral artery, superficial cervical and costocervical trunk, and distal subclavian artery was explanted. Explanted tissue was pressure-fixed with 10% formalin under approximately 50 mmHg for about 20 minutes.
Tissue Processing
Tissue samples evaluated by histology were processed through a graded series of ethanol, xylene, and were embedded in methyl methacrylate. Sections were taken at approximately 600-µm intervals, polished down to 6 µm, and stained with a metachromatic (toluidine blue-basic fuchsin) stain. Neointimal thicknesses at any section were calculated as the difference between device and lumen radii based on measurements of internal elastic membrane area (device area) and lumen area. Tissue samples obtained at 180 days follow-up from 1 animal per treatment group were evaluated by scanning electron microscopy (SEM). Before tissue harvesting for SEM evaluation, the vasculature was sequentially perfused with 5% dextrose in water, 0.25% silver nitrate, 5% dextrose in water, and formalin. The implanted parent artery was cut longitudinally to obtain 2 half-cylinder shape segments, and the silver nitrate was developed by exposing the endothelial cells to Kodak T-Max solution. The tissue was then placed in 1% Zetterquists Osmium, rinsed in Zetterquists Buffer, washed in water, dehydrated in a graded series of ethanol and acetone solutions, and dried in hexamethyldisilazane (HMDS). Samples were mounted and sputter-coated with gold or palladium for SEM observation.
Angiographic Analysis
The angiographic image with maximal opacification of the aneurysm was selected in each of the acquired sequences, and the aneurysm was manually delineated as a region of interest (ROI). The percentage angiographic aneurysm occlusion at follow-up was calculated as AFU%=(A1–A2)/A1x100, where A1 was the aneurysm area (in pixels) before device implantation and A2 was the aneurysm area at follow-up. The flow exchange between the parent artery and the aneurysm was measured by calculating the temporal variation in average grayscale intensity (ratio of the sum of grayscale intensities of all pixels to total number of pixels) within the ROI. These aneurysmal washout curves were then normalized and fit to a mathematical model described previously24 in the least squares sense. The model demarcates the aneurysm-parent vessel flow exchange into convective and diffusive components as characterized by amplitudes and time constants. The alteration in local hemodynamics produced by a flow diversion device could then be quantified by the extent to which the device increased the diffusive mode of flow exchange as seen by an increase in the amplitude of the diffusive component (
diff) and the diffusive time constant (
diff).
In addition to increasing the proportion and washout time of the diffusive region within the aneurysm, a flow diversion device would also reduce the net amount of dye entering the aneurysm, as reflected by changes in the amplitude of the aneurysmal washout curve. To account for these different mechanisms in a single quantity, an index of the effectiveness of the device could be constructed. This index, called the washout coefficient here, was given as
|
|
where
% was the washout curve amplitude and
diff% and
diff% were the amplitude and time constant of the diffusive component of the model obtained immediately after device implantation expressed as percentages of the corresponding values before device implantation. Such a parameter obtained immediately after device implantation could be used to assess whether or not any given device will occlude the treated aneurysm at follow-up. A prognostic test for aneurysm occlusion could then be developed based on a suitable threshold value for W. A threshold was chosen so as to maximize the sensitivity and specificity of the test based on the receiver operating characteristic (ROC) curve for the data, and statistical significance of the test was assessed by Fisher exact test for the resulting contingency table. Statistical significance of the differences in angiographic aneurysm occlusion and neointimal thickness between devices was compared by analysis of variance or Student t tests. All statistical evaluations were made with the InStat software (version 5.1, GraphPad Software).
| Results |
|---|
|
|
|---|
|
|
Histological sections of the aneurysms showed a progression in maturity of intraaneurysmal thrombus from acute, unorganized clot at 21 days through to complete organization with neovascularization at 6 months (Figure 4a through 4c). A thin smooth neointima was observed in all samples at all time points in histological evaluations. The average neointimal thicknesses formed within each device at different follow-up time points are graphed in Figure 3b. Device III consistently had significantly lower neointimal thicknesses as compared to the other 2 devices. For a given device, there were no significant differences over various follow-up times (P>0.59). Endothelial cells were seen to cover the entire luminal surface of all the devices in scanning electron microscope (SEM) sections of samples explanted at 6 months (Figure 5a). The vertebral artery and the costocervical trunk jailed by the flow diversion devices remained angiographically patent in all cases at all follow-up time points (Figure 2c). SEM sections at the vertebral artery corroborated side branch patency in all processed samples (Figure 5b).
|
|
Aneurysmal washout curves from one case obtained before device implantation and immediately after implantation of device III are shown in Figure 6a. In this case, the amount of dye entering the aneurysm (amplitude of the washout curve) after device implantation was reduced to approximately 40% of the corresponding value before device implantation. The mathematical model faithfully captured the trend of all the aneurysmal washout curves, and the model-fits to the 2 curves in Figure 6a are shown in Figure 6b and 6c as examples. The convective and diffusive components of the model are superposed on the plots. The proportion of the aneurysm-parent vessel flow exchange dominated by diffusion was increased by about 45% due to the flow diversion device in this case; the time constant of this diffusive flow exchange was concomitantly increased to 70 times the corresponding value before device implantation.
|
There was a statistically significant difference in the washout coefficient (W) values when correlated with the angiographic aneurysm occlusion at follow-up. The average W value for cases with
97% angiographic occlusion of the aneurysm at follow-up (21±15) was significantly lower than the average value for cases with <97% angiographic aneurysm occlusion at follow-up (51±34); Welch-corrected t test P=0.02. The threshold value of W which gave the maximum sensitivity and specificity for the prognostic test was 30. Figure 6d shows the ROC curve for the washout coefficient data with the inset showing the contingency table based on a threshold value of 30. The sensitivity and the specificity of this test were 73% and 82%, respectively. The Fisher exact test 2-sided probability value was 0.03, suggesting that the row-column association was statistically significant.
| Discussion |
|---|
|
|
|---|
In contrast to an ideal stent, an ideal flow diversion device has low porosity and high pore density values optimized to promote intraaneurysmal thrombosis while maintaining patency of the parent vessel and side branches. Moreover, lower radial forces are required of this device as compared to a stent, which facilitates the optimization of other device characteristics such as longitudinal flexibility, trackability, and conformability. Recognition of the potential for aneurysm treatment by flow diversion is evidenced by the recent development of these devices by various groups. One such device that is furthest in development (Pipeline Neuroendovascular Device, Chestnut Medical Technologies) consists of 16 platinum and 16 stainless steel filaments that are braided to a 70% porosity. The treatment of 17 rabbit elastase-induced aneurysms with this device showed complete to near-complete angiographic occlusion in 5 of 6, 5 of 5, and 5 of 6 aneurysms followed-up at 1 month, 3 months, and 6 months, respectively.17 Jailed vertebral arteries were patent in all cases. Histological sections of the aneurysm showed acute unorganized thrombus at the dome in all animals at 1 month, poorly organized thrombus in 3/4 animals and organized connective tissue in 1/4 animals at 3 months, and poorly organized thrombus in 4/6 and organized connective tissue in 2/6 animals at 6 months. By comparison, the flow diversion devices used in this report resulted in complete organization of aneurysm contents in 2/4, 2/5, and 3/6 of animals at 3 weeks, 3 months, and 6 months, respectively. The flow diversion devices seem to perform better than the Pipeline device based on these data. Neointimal thicknesses within both devices are in the same range. Further studies specifically aimed at device comparisons are required to judge the equivalency of our flow diversion devices and these other devices.
The jailing of perforators, which are side branches of intracranial arteries ranging from 100 µm to 1 mm, by device struts was simulated by implanting the devices across the vertebral artery ostium originating adjacent to the aneurysm. This entrance diameter is probably on the higher side of actual perforator ostia, but none of the vertebral arteries were occluded angiographically or histologically through 180 days after device implantation. An in vitro study with flow diversion devices having the same nominal porosities as the ones used here showed that the maximal reduction in mean flow rate through the vertebral artery was about 15%.21 Although it has been suggested that perforators remain patent if less than 50% of their ostial diameter is covered by the device mesh,26 methodological (possibly bench-top) studies on the effect of varying device porosities covering 100 to 200 µm diameter side branches are required to identify the critical ranges of device parameters vis-à-vis perforator filling or occlusion. More than three-quarters of the luminal surface of all devices was noted to be lined with endothelial cells on all histological sections at all follow-up time points, which is evidence of minimal injury to the arterial wall produced by these self-expanding devices. Further, the neointimal thickness within a given device did not change significantly over the follow-up period of 3 weeks to 6 months. These results suggest stable incorporation of the flow diversion device within the arterial lumen within a few weeks.
Our results suggest that the device with the highest pore density (device III) performed the best, even in comparison with devices with the same (device I) or lower (device II) nominal porosities. Presumably, the finer meshes implied by higher pore densities further stratify the flow exchange (flow movement perpendicular to the mesh) between the parent vessel and the aneurysm, which leads to delayed washout of blood and increased probability of intraaneurysmal thrombosis. Further refinement of the best device by decreasing its filament diameter and concomitantly maximizing its pore density to an optimum value may lead to yet superior success rates. Patency of cerebral perforators must, however, be considered while increasing device pore density because pore size reductions below orders of cellular lengths may allow neointimal cells to bridge the gap between device struts and completely cover perforator ostia.
The 97% aneurysm occlusion rate used to stratify the prognostic test for treatment success based on the washout coefficient was a posthoc measure chosen to group enough number of animals with complete to near-complete occlusion and reach a significant difference between groups. Because occlusion percentages were grouped over the follow-up times, the test only predicts treatment success at 180 days (longest follow-up time used in this study). Also, the washout coefficient value for a control aneurysm (no flow diverter) would be 1, so the test cannot be used to evaluate cases for which the value comes to be
1 after flow diverter implantation. The sensitivity of the test (73%) is probably not high enough to guide clinical course of management. The true negative rate (82%) is less crucial because if the aneurysm actually occludes after the test predicts that it will not (W>30), there is negligible loss in patient care. The physician may choose to take additional treatment steps (multiple devices or coiling) if the test is negative. In either case, the test can at least supplement the physicians experience in evaluating the next course of action. As mentioned previously, optimization of the washout curves and the prognostic test based on the washout coefficient can be performed immediately after device deployment, while the patient is on the angiographic table. This test is based on a data set of experimental aneurysms and its validity will have to be evaluated on large sets of clinical data.
Summary
Three novel flow diversion device configurations developed and refined based on past experience were tested in a large cohort of experimental aneurysms. Successful and stable aneurysm occlusion was effected by the devices over the study period of 6 months. Results suggest that the device with a porosity of 70% and pore density of 18 pores/mm2 performed better than devices of 65% porosity, 14 pores/mm2 and 70% porosity, 12 pores/mm2. The pore density of flow diversion devices may thus be the critical factor modulating treatment success. Angiographic quantification suggested a parameter, called the washout coefficient here, which could be used to estimate aneurysm occlusion probabilities immediately after the treatment and thus preclude unnecessary additional treatment manipulations. Based on the aneurysms treated and the devices used, a value of this parameter less than 30 predicts greater than 97% angiographic occlusion of the aneurysm over a period of 6 months. Generalization of this threshold value to the treatment of any given aneurysm by flow diversion requires further testing.
| Acknowledgments |
|---|
The study was supported by a National Institutes of Health grant (NS045753-01A1) and Siemens Medical Solutions.
Disclosures
B.B.L. and A.K.W. have ownership interest in Surpass Medical LTD.
Received August 4, 2008; accepted August 26, 2008.
| References |
|---|
|
|
|---|
2. Hop JW, Rinkel GJ, Algra A, van Gijn J. Case-fatality rates and functional outcome after subarachnoid hemorrhage: a systematic review. Stroke. 1997; 28: 660–664.
3. Molyneux AJ, Kerr RS, Yu LM, Clarke M, Sneade M, Yarnold JA, Sandercock P; International Subarachnoid Aneurysm Trial (ISAT) Collaborative Group. International subarachnoid aneurysm trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised comparison of effects on survival, dependency, seizures, rebleeding, subgroups, and aneurysm occlusion. Lancet. 2005; 366: 809–817.[CrossRef][Medline] [Order article via Infotrieve]
4. Pelz DM, Levy EI, Hopkins LN. Advances in interventional neuroradiology 2007. Stroke. 2008; 39: 268–272.
5. Higashida RT, Halbach VV, Dowd CF, Juravsky L, Meagher S. Initial clinical experience with a new self-expanding nitinol stent for the treatment of intracranial cerebral aneurysms: the Cordis Enterprise stent. AJNR Am J Neuroradiol. 2005; 26: 1751–1756.
6. Fiorella D, Albuquerque FC, Deshmukh VR, McDougall CG. Usefulness of the Neuroform stent for the treatment of cerebral aneurysms: results at initial (3–6-mo) follow-up. Neurosurgery. 2005; 56: 1191–1201.[CrossRef][Medline] [Order article via Infotrieve]
7. Wakhloo AK, Mandell J, Gounis MJ, Brooks C, Linfante I, Winer J, Weaver JP. Stent-assisted reconstructive endovascular repair of cranial fusiform atherosclerotic and dissecting aneurysms - Long-term clinical and angiographic follow-up. Stroke. 2008; 39: 3288–3296.
8. Wakhloo AK, Shellhammer F, de Vries J, et al. Coated and non-coated stents for vessel reconstruction and treatment of aneurysms and av-fistulas: An experimental study. ESNR 92; XVIIIth Congress of the European Society of Neuroradiology. 1992;34:S24. Abstract
9. Turjman F, Acevedo G, Moll T, Duquesnel J, Eloy R, Sindou M. Treatment of experimental carotid aneurysms by endoprosthesis implantation: preliminary report. Neurol Res. 1993; 15: 181–184.[Medline] [Order article via Infotrieve]
10. Wakhloo AK, Schellhammer F, de Vries J, Haberstroh J, Schumacher M. Self-expanding and balloon-expandable stents in the treatment of carotid aneurysms: an experimental study in a canine model. AJNR Am J Neuroradiol. 1994; 15: 493–502.[Abstract]
11. Geremia G, Haklin M, Brennecke L. Embolization of experimentally created aneurysms with intravascular stent devices. AJNR Am J Neuroradiol. 1994; 15: 1223–1231.[Abstract]
12. Doerfler A, Wanke I, Egelhof T, Stolke D, Forsting M. Double-stent method: therapeutic alternative for small wide-necked aneurysms. J Neurosurg. 2004; 100: 150–154.[Medline] [Order article via Infotrieve]
13. Zenteno MA, Murillo-Bonilla LM, Guinto G, Gomez CR, Martinez SR, Higuera-Calleja J, Lee A, Gomez-Llata S. Sole stenting bypass for the treatment of vertebral artery aneurysms: technical case report. Neurosurgery. 2005; 57: E208.[CrossRef][Medline] [Order article via Infotrieve]
14. Szikora I, Berentei Z, Kulcsar Z, Barath K, Berez A, Bose A, Nyary I. Endovascular treatment of intracranial aneurysms with parent vessel reconstruction using balloon and self expandable stents. Acta Neurochir (Wien). 2006; 148: 711–723.[CrossRef][Medline] [Order article via Infotrieve]
15. Lanzino G, Wakhloo AK, Fessler RD, Hartney ML, Guterman LR, Hopkins LN. Efficacy and current limitations of intravascular stents for intracranial internal carotid, vertebral, and basilar artery aneurysms. J Neurosurg. 1999; 91: 538–546.[Medline] [Order article via Infotrieve]
16. Ionita CN, Paciorek AM, Hoffmann KR, Bednarek DR, Yamamoto J, Kolega J, Levy EI, Hopkins LN, Rudin S, Mocco J. Asymmetric Vascular Stent. Feasibility Study of a New Low-Porosity Patch-Containing Stent. Stroke. 2008; 39: 2105–2113.
17. Kallmes DF, Ding YH, Dai D, Kadirvel R, Lewis DA, Cloft HJ. A new endoluminal, flow-disrupting device for treatment of saccular aneurysms. Stroke. 2007; 38: 2346–2352.
18. Ahlhelm F, Roth C, Kaufmann R, Schulte-Altedorneburg G, Romeike BF, Reith W. Treatment of wide-necked intracranial aneurysms with a novel self-expanding two-zonal endovascular stent device. Neuroradiology. 2007; 49: 1023–1028.[CrossRef][Medline] [Order article via Infotrieve]
19. Lieber BB, Stancampiano AP, Wakhloo AK. Alteration of hemodynamics in aneurysm models by stenting: influence of stent porosity. Ann Biomed Eng. 1997; 25: 460–469.[Medline] [Order article via Infotrieve]
20. Lieber BB, Livescu V, Hopkins LN, Wakhloo AK. Particle image velocimetry assessment of stent design influence on intra-aneurysmal flow. Ann Biomed Eng. 2002; 30: 768–777.[CrossRef][Medline] [Order article via Infotrieve]
21. Seong J, Wakhloo AK, Lieber BB. In vitro evaluation of flow divertors in an elastase-induced saccular aneurysm model in rabbit. J Biomech Eng. 2007; 129: 863–872.[CrossRef][Medline] [Order article via Infotrieve]
22. Lieber BB, Sadasivan C, Miskolczi L, et al. Flow divertors to treat cerebral aneurysms: preliminary results in the rabbit elastase-induced aneurysm model. Proceedings of the 2006 Summer Bioengineering Conference. 2006.
23. Miskolczi L, Gounis MJ, Onizuka M, et al. Elastase-induced saccular aneurysms in rabbits: Instructions for the rest of us. Proceedings of the 42nd Annual Meeting Am Society of Neuroradiology. 2004;352. Abstract.
24. Sadasivan C, Lieber BB, Gounis MJ, Lopes DK, Hopkins LN. Angiographic quantification of contrast medium washout from cerebral aneurysms after stent placement. AJNR Am J Neuroradiol. 2002; 23: 1214–1221.
25. Seong J, Sadasivan C, Onizuka M, Gounis MJ, Christian F, Miskolczi L, Wakhloo AK, Lieber BB. Morphology of elastase-induced cerebral aneurysm model in rabbit and rapid prototyping of elastomeric transparent replicas. Biorheology. 2005; 42: 345–361.[Medline] [Order article via Infotrieve]
26. Lopes DK, Ringer AJ, Boulos AS, Qureshi AI, Lieber BB, Guterman LR, Hopkins LN. Fate of branch arteries after intracranial stenting. Neurosurgery. 2003; 52: 1275–1278.[Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
D Fiorella, P Lylyk, I Szikora, M E Kelly, F C Albuquerque, C G McDougall, and P K Nelson Curative cerebrovascular reconstruction with the Pipeline embolization device: the emergence of definitive endovascular therapy for intracranial aneurysms JNIS, July 1, 2009; 1(1): 56 - 65. [Abstract] [Full Text] [PDF] |
||||
![]() |
D.F. Kallmes, Y.H. Ding, D. Dai, R. Kadirvel, D.A. Lewis, and H.J. Cloft A Second-Generation, Endoluminal, Flow-Disrupting Device for Treatment of Saccular Aneurysms AJNR Am. J. Neuroradiol., June 1, 2009; 30(6): 1153 - 1158. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Stroke Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2009 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |