Background and Purpose The ability to detect asymptomatic circulating cerebral emboli may contribute to the management of patients with stroke, but its clinical usefulness will depend on effective systems for automatically detecting embolic signals (ES) and differentiating them from artifact. A new method involves the use of a multidepth probe that allows recording from both distal and proximal sample volumes along the same vessel. Theoretically, an embolus should appear sequentially, with a time delay, between the two channels, whereas an artifact should appear simultaneously in the two channels.
Methods We evaluated this method in an in vitro flow model and in patients. In an in vitro model, with a flow pattern mimicking intracerebral flow, 181 air bubbles and 193 thrombus emboli were compared with the signals resulting from 368 episodes of artifact; a sample volume of 5 mm and a channel separation of 10 mm were used. ES from two groups of patientsthose with carotid artery stenosis (141 ES) and those with mechanical prosthetic cardiac valves (125 ES)were studied and compared with 222 episodes of artifact produced in the same patients.
Results In the model the mean (SD) time delay was 17.32 (9.94) ms for air emboli and 17.78 (10.66) ms for thrombus emboli compared with -0.01 (0.39) ms for artifact (air and thrombus emboli versus artifact, P<.0001). A sensitivity of 100% and specificity of 100% were obtained when a cutoff of >2 ms was used for an embolus. The method allowed equally good detection of those air emboli that resulted in receiver overload and aliasing. In patients the mean (SD) time delay was 29.6 (28.2) ms for valve ES and 14.9 (15.42) for carotid ES compared with 0.00 (0.46) for artifact (carotid and valve ES versus artifact, P<.0001). Considering only those signals that were visible in both Doppler time domains resulted in a sensitivity for valve ES of 98.9% and for carotid ES of 94.0%, with a specificity of 99.0%. However, in one patient in the valve group some ES were visible only in the proximal channel, possibly because of passage of emboli down branch vessels between the two sample volumes. In addition, for the less intense carotid ES some signals were unclear or absent in one or both of the time domain signals at either depth, although visible in the postfast Fourier transform spectra. Including those ES visible in only one channel reduced the sensitivity to 75.2% for valve ES and 92.6% for carotid ES.
Conclusions The multigated technique offers a new method of detecting ES and differentiating them from artifact and is the first reliable method for differentiating intense ES resulting in receiver overload from artifact. Occasionally its sensitivity is reduced when ES do not appear in the distal channel, probably because they pass down a side branch; this may be reduced by reducing gate separation. Some less intense carotid ES can be difficult to detect if the amplitude increase is small compared with the amplitude of the background Doppler signal.
Key Words: carotid artery diseases cerebral embolism heart valve prosthesis ultrasonics
This article has been cited by other articles:
![]() |
Y. Kalairajah, A. J. Cossey, G. M. Verrall, G. Ludbrook, and A. J. Spriggins Are systemic emboli reduced in computer-assisted knee surgery?: A PROSPECTIVE, RANDOMISED, CLINICAL TRIAL J Bone Joint Surg Br, February 1, 2006; 88-B(2): 198 - 202. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. K. Moppett and R. P. Mahajan Transcranial Doppler ultrasonography in anaesthesia and intensive care Br. J. Anaesth., November 1, 2004; 93(5): 710 - 724. [Full Text] [PDF] |
||||
![]() |
G. Devuyst, G.A. Darbellay, J.-M. Vesin, V. Kemeny, M. Ritter, D.W. Droste, C. Molina, J. Serena, R. Sztajzel, P. Ruchat, et al. Automatic Classification of HITS Into Artifacts or Solid or Gaseous Emboli by a Wavelet Representation Combined With Dual-Gate TCD Stroke, December 1, 2001; 32(12): 2803 - 2809. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Cullinane, G. Reid, R. Dittrich, Z. Kaposzta, R. Ackerstaff, V. Babikian, D. W. Droste, D. Grossett, M. Siebler, L. Valton, et al. Evaluation of New Online Automated Embolic Signal Detection Algorithm, Including Comparison With Panel of International Experts Stroke, June 1, 2000; 31(6): 1335 - 1341. [Abstract] [Full Text] [PDF] |
||||
![]() |
Z. Kaposzta, E. Young, P. M. W. Bath, and H. S. Markus Clinical Application of Asymptomatic Embolic Signal Detection in Acute Stroke : A Prospective Study Stroke, September 1, 1999; 30(9): 1814 - 1818. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. L. Taylor, M. A. Borger, R. D. Weisel, L. Fedorko, and C. M. Feindel Cerebral microemboli during cardiopulmonary bypass: increased emboli during perfusionist interventions Ann. Thorac. Surg., July 1, 1999; 68(1): 89 - 93. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Kemeny, D. W. Droste, S. Hermes, D. G. Nabavi, G. Schulte-Altedorneburg, M. Siebler, and E. B. Ringelstein Automatic Embolus Detection by a Neural Network Stroke, April 1, 1999; 30(4): 807 - 810. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Bernd Ringelstein, D. W. Droste, V. L. Babikian, D. H. Evans, D. G. Grosset, M. Kaps, H. S. Markus, D. Russell, and M. Siebler Consensus on Microembolus Detection by TCD Stroke, March 1, 1998; 29(3): 725 - 729. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Georgiadis, A. Wenzel, H. R. Zerkowski, S. Zierz, and A. Lindner Automated Intraoperative Detection of Doppler Microembolic Signals Using the Bigate Approach Stroke, January 1, 1998; 29(1): 137 - 139. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. S. Markus and J. Molloy Use of a Decibel Threshold in Detecting Doppler Embolic Signals Stroke, April 1, 1997; 28(4): 692 - 695. [Abstract] [Full Text] |
||||
![]() |
D. W. Droste, G. Hagedorn, A. Notzold, H.-J. Siemens, H. H. Sievers, and M. Kaps Bigated Transcranial Doppler for the Detection of Clinically Silent Circulating Emboli in Normal Persons and Patients With Prosthetic Cardiac Valves Stroke, March 1, 1997; 28(3): 588 - 592. [Abstract] [Full Text] |
||||
|
Stroke Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1996 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |